| DICOM PS3.3 2024c - Information Object Definitions |
|---|
Table C.8.19.3-1 specifies the Attributes of the XA/XRF Acquisition Module.
Table C.8.19.3-1. XA/XRF Acquisition Module Attributes
|
Average of the peak kilo voltage outputs of the X-Ray generator used for all frames. |
|||
|
Average of the nominal X-Ray tube currents in milliamperes for all frames. Required if Exposure in mAs (0018,9332) is not present. May be present otherwise. |
|||
|
Duration of X-Ray exposure in milliseconds. See Section C.8.7.2.1.1. Required if Exposure in mAs (0018,9332) is not present. May be present otherwise. |
|||
|
The exposure expressed in milliampereseconds, for example calculated from Exposure Time and X-Ray Tube Current. Required if either Exposure Time in ms (0018,9328) or X-Ray Tube Current in mA (0018,9330) are not present. May be present otherwise. |
|||
|
The time in seconds needed for the complete acquisition. See Section C.7.6.16.2.2.1 for further explanation. |
|||
|
Distance in mm between the receptor plane and the detector housing. The direction of the distance is positive from receptor plane to X-Ray source. Note
|
|||
|
Describes for C-arm positioner type systems if positioner and tabletop has the same geometrical reference system. |
|||
|
X-Ray dose, measured in dGy*cm*cm, to which the patient was exposed for the acquisition of this image only. |
| DICOM PS3.3 2024c - Information Object Definitions |
|---|