| DICOM PS3.3 2024c - Information Object Definitions |
|---|
Table C.8-118 specifies the Attributes of the CT Acquisition Type Macro.
Table C.8-118. CT Acquisition Type Macro Attributes
|
Description of the method used during acquisition of this frame. See Section C.8.15.3.2.1 for Defined Terms. Required if Frame Type (0008,9007) Value 1 of this frame is ORIGINAL. May be present otherwise. |
|||
|
The constant angle at which the X-Ray source is located during acquisition. 0 degrees means that the source is located at the highest point of the gantry orbit. Degrees increase from 0 to positive 360 in a clockwise direction as viewed when facing the gantry where the table enters the gantry. Required if Frame Type (0008,9007) Value 1 of this frame is ORIGINAL and Acquisition Type (0018,9302) is CONSTANT_ANGLE. May be present otherwise if Frame Type (0008,9007) Value 1 of this frame is DERIVED and Acquisition Type (0018,9302) is CONSTANT_ANGLE. |
|||
|
Identifies that the acquisition was performed by repetitively acquiring the same volume set over a period of time. NoteAcquisition Type (0018,9302) may be SEQUENCED, SPIRAL or STATIONARY depending on whether table movement is necessary to cover the volume. Required if Frame Type (0008,9007) Value 1 of this frame is ORIGINAL. May be present otherwise. |
|||
|
Identifies that near real-time display of a block of continuously acquired data was performed, which may result in a lower than usual image quality. Required if Frame Type (0008,9007) Value 1 of this frame is ORIGINAL. May be present otherwise. |
Defined Terms:
identifies that the acquisition was performed by acquiring single or multi detector data while rotating the source about the gantry while the table is not moving; additional slices may be acquired by incrementing the table position and again rotating the source about the gantry while the table is not moving
identifies that the acquisition was performed by acquiring data while rotating the source about the gantry while continuously moving the table
identifies that the acquisition was performed by holding the source at a constant angle and moving the table to obtain a projection image (e.g., a localizer image)
identifies that the acquisition was performed by holding the table at a constant position and performing one or more acquisitions over time at the same location
identifies that the acquisition was performed while rotating the source about the gantry while the table movement is under direct control of a human operator or under the control of an analysis application (e.g., fluoroscopic image)
| DICOM PS3.3 2024c - Information Object Definitions |
|---|