| DICOM PS3.3 2024c - Information Object Definitions |
|---|
Table C.36.3-1 specifies the Attributes of the Enhanced RT Series Module, which identify and describe general information about the Enhanced RT Series.
The RT Second Generation IODs use the General Series Module, specialized by this Module.
Table C.36.3-1. Enhanced RT Series Module Attributes
|
Type of device, process or method that originally acquired the data used to create the Instances in this Series. See Section C.36.3.1.1. |
|||
|
Uniquely identifies the Performed Procedure Step SOP Instance that resulted in creation of the Series (e.g. a Modality or Unified Procedure Step SOP Instance). Only a single Item shall be included in this Sequence. Required if the series has been created as a result of a single procedure step request and the Instance-Level Referenced Performed Procedure Step Sequence (3010,0044) is not present. If different Instances in the series are created as a result of a procedure step, the Instance-Level Referenced Performed Procedure Step Sequence (3010,0044) in the Radiotherapy Common Instance Module shall be used. |
|||
|
>Include Table 10-11 “SOP Instance Reference Macro Attributes” |
|||
|
A unique identifier of the RT Treatment Session to which Instances in this Series belong. |
|||
| DICOM PS3.3 2024c - Information Object Definitions |
|---|