| DICOM PS3.3 2024c - Information Object Definitions |
|---|
Table C.36.2.2.5-1 specifies the Attributes of the External Beam Control Point General Macro, which specify the RT Control Point Attributes used to model external beam radiation.
Table C.36.2.2.6-1. External Beam Control Point General Macro Attributes
|
Include Table C.36.2.2.5-1 “RT Control Point General Macro Attributes”. |
|||
|
The value of the intended nominal rate of delivery of the specified Cumulative Meterset (300A,063C). The units are defined in the Delivery Rate Unit Sequence (300A,063E). Required if the conditions in Section C.36.2.2.5.1.1 are satisfied. See Section C.36.2.2.5.1. |
|||
|
The unit of the delivery rate value. Required if Delivery Rate (300A,063D) is present and has a value. See Section C.36.2.2.5.1. |
|||
|
Area within which the treatment beam must be contained, for example when using MLC tracking for a moving target. Only a single Item shall be included in this Sequence. Required if beam shall be limited. See Section C.36.2.2.5.1. |
|||
|
>Include Table 10.38-1 “Outline Definition Macro Attributes”. |
The Outline is defined on the Beam Modifier Definition Plane. |
||
| DICOM PS3.3 2024c - Information Object Definitions |
|---|