| DICOM PS3.3 2024c - Information Object Definitions |
|---|
Table C.36.13-1 specifies the Attributes of the RT Radiation Common Module, which are shared by all RT Radiation IODs used for radiation treatment delivery.
Table C.36.13-1. RT Radiation Common Module Attributes
|
Include Table 10.9.1-1 “Enhanced Content Identification Macro Attributes”. |
|||
|
Include Table C.36.2.1.6-1 “RT Radiation Common Base Macro Attributes”. |
|||
|
The maximum delivery time in seconds. See Section C.36.13.1.3. |
|||
The Treatment Time Limit (300A,062E) is the maximum time span allowed to deliver a single fraction of this RT Radiation SOP Instance. The behavior of the treatment delivery device on exceeding the Treatment Time Limit is up to the implementation. Treatment is expected to terminate upon reaching the Treatment Time Limit (300A,062E) independent of the Meterset and so this Attribute should have a value that accommodates normal variations in delivery.
| DICOM PS3.3 2024c - Information Object Definitions |
|---|