This Module specifies the Attributes of the Display System Module, and describes the Display System.
Table C.32.1-1. Display System Module Attributes
|
Mailing address of institution where the Display System is located. |
||
|
Department name in the institution where the Display System is located. |
||
|
A coded description of the type of Department or Service within the healthcare facility. |
||
|
The date the Display System was originally manufactured or re-manufactured (as opposed to refurbished). |
||
|
The date the Display System was installed in its current location. The printer may or may not have been used prior to installation in its current location. |
||
|
Name of the person or role that is administrating the equipment. |
||
|
>Include Table 10-1 “Person Identification Macro Attributes” |
||
|
Number that identifies a specific display subsystem within a Display System. The value shall be unique within the Display Subsystem Sequence (0028,7023). |
||
|
Manufacturer of the Display Device in the Display Subsystem. If the Display Subsystem is provided integrated into the Display System (e.g., for a tablet) and the Manufacturer is not otherwise known, the Manufacturer of the Display System may be recorded here. |
||
|
Manufacturer's serial number of the Display Device (i.e., monitor device) in the Display Subsystem. If the Display Subsystem is provided integrated into the Display System (e.g., for a tablet) and the Device Serial Number is not otherwise known, the Device Serial Number of the Display System may be recorded here. |
||
|
Manufacturer's model name of the Display Device in the Display Subsystem. If the Display Subsystem is provided integrated into the Display System (e.g., for a tablet) and the Model Name is not otherwise known, the Model Name of the Display System may be recorded here. |
||
|
The date the Display Device in the Display Subsystem was originally manufactured or re-manufactured (as opposed to refurbished). |
||
|
The date the Display Device in the Display Subsystem was installed in its current location. The printer may or may not have been used prior to installation in its current location. |
||
|
Current overall status of the Display Subsystem. See Section C.32.1.1 for Defined Terms. |
||
|
A free text comment about the status of the Display Subsystem. |
||
|
Number that identifies a specific display subsystem configuration within the Display Subsystem Configuration Sequence (0028,700A). The value shall be unique within the Display Subsystem Configuration Sequence (0028,700A). |
||
|
A free text description of the Display Subsystem Configuration. |
||
|
References the intended luminance characteristics for this Configuration of this Display Subsystem. The value is the Luminance Characteristics ID (0028,7009) within the Target Luminance Characteristics Sequence (0028,7008) of the Target Luminance Characteristics Module. |
||
|
References the current configuration of the Display Subsystem. The value is the Configuration ID (0028,700B) of the Item in the Display Subsystem Configuration Sequence (0028,700A). |
||
|
Describes measurement equipment (if any) built into the Display Subsystem. |
||
|
Measurement functions of the measurement equipment. Enumerated Values: This multi-valued Attribute shall contain at most one of each enumerated value. When multiple values are present, the equipment is capable of performing multiple functions. |
||
|
Characteristic(s) which are measured by the measurement equipment. Enumerated Values: This multi-valued Attribute shall contain at most one of each enumerated value. When multiple values are present, the equipment is capable of measuring multiple characteristics. |
||
|
Type of measurement equipment. See Section C.32.1.2 for Enumerated Values. |
||
|
The date the measurement equipment was originally manufactured or re-manufactured (as opposed to refurbished). |
||
|
The date the measurement equipment was installed in its current location. The measurement equipment may or may not have been used prior to installation in its current location. |
||
|
Date and time of the last calibration of the measurement equipment. |
||
Defined Terms for System Status (0028,7006):
It is up to the implementation how to set the System Status (0028,7006) based on the contents of the status code sequences.
Enumerated Values for Measurement Equipment Type (0028,7014):
See [IEC 62563-1], Annex B, section B.2.1 to B.2.4 for examples of these types.