| DICOM PS3.3 2024c - Information Object Definitions |
|---|
Table C.2-2 specifies the Attributes of the Patient Identification Module, which define the Attributes relevant to identifying a Patient.
Table C.2-2. Patient Identification Module Attributes
|
Primary identifier for the Patient. NoteIn the case of imaging a group of small animals simultaneously, the single value of this identifier corresponds to the identification of the entire group. See also Section C.7.1.4.1.1. |
||
|
The type of identifier in the Patient ID (0010,0020). Note
|
||
|
A Sequence of identification numbers or codes used to identify the Patient, which may or may not be human readable, and may or may not have been obtained from an implanted or attached device such as an RFID or barcode. |
||
|
>Include Table 10-18 “Issuer of Patient ID Macro Attributes” |
||
|
The type of identifier in the Patient ID (0010,0020) in this Item. Note
|
||
|
A photo to confirm the identity of a Patient. Only a single Item is permitted in this Sequence. See C.2.2.1.1. |
||
|
>Include Table 10-3b “Referenced Instances and Access Macro Attributes” |
||
Previously, Other Patient IDs (0010,1000) and Medical Record Locator (0010,1090) were included in this table. These Attributes have been retired. See PS3.3-2017a.
A Patient Photo may be provided as an additional means to confirm the identity of the Patient.
When the Referenced SOP Sequence (0008,1199) is present, the Referenced SOP Class UID (0008,1150) shall be either VL Photographic Image Storage (1.2.840.10008.5.1.4.1.1.77.1.4) or Secondary Capture Image Storage (1.2.840.10008.5.1.4.1.1.7).
When the WADO Retrieval Sequence (0040,E023) or WADO-RS Retrieval Sequence (0040,E025) is present, the Retrieve URI (0040,E010) shall refer to a JPEG image if the referenced server is not WADO-compliant.
| DICOM PS3.3 2024c - Information Object Definitions |
|---|