| DICOM PS3.3 2024c - Information Object Definitions |
|---|
Table C.11.16-1 specifies the Attributes of the Structured Display Module.
Table C.11.16-1. Structured Display Module Attributes
|
Include Table 10-12 “Content Identification Macro Attributes” |
|||
|
The number of screens for this Structured Display. If SOP Class UID (0008,0016) equals 1.2.840.10008.5.1.4.1.1.131(Basic Structured Display), the value shall be 1. |
|||
|
Sequence of Items that describes the set of screens for this Structured Display. The number of Items shall equal the value of Number of Screens (0072,0100). One or more Items shall be included in this Sequence. NoteThe Basic Structured Display IOD limits this Sequence to one Item. |
|||
|
>Include Table C.23.2-2 “Screen Specifications Macro Attributes” |
|||
|
This icon image is representative of the structured display. |
|||
|
See Section C.7.6.1.1.6 for further explanation. |
|||
|
A value in which it is recommended that Structured Display background (i.e., the area outside of Image Boxes) be rendered on a color display. The units are specified in PCS-Values, and the value is encoded as CIELab. See Section C.10.7.1.1. |
|||
|
A value in which it is recommended that empty Image Boxes be rendered on a color display. The units are specified in PCS-Values, and the value is encoded as CIELab. See Section C.10.7.1.1. |
|||
|
Identifies the hanging protocol that was used in the creation of this structured display. See Table OO.1.1-1 “Hanging Protocol Names for Dental Image Layout based on JSOMR classification” in PS3.17 for standard hanging protocols for dentistry. |
|||
|
Identifies the creator of the Hanging Protocol. See Table OO.1.1-1 “Hanging Protocol Names for Dental Image Layout based on JSOMR classification” in PS3.17 for standard hanging protocols for dentistry. |
|||
| DICOM PS3.3 2024c - Information Object Definitions |
|---|