| DICOM PS3.3 2024c - Information Object Definitions |
|---|
Table A.66.3-1 specifies the Modules of the Intravascular Optical Coherence Tomography Image IOD.
Table A.66.3-1. Intravascular Optical Coherence Tomography Image IOD Modules
|
C - Required if Pixel Presentation (0008,9205) in the Intravascular OCT Image Module equals COLOR. |
|||
|
C - Required if Presentation Intent Type (0008,0068) is FOR PROCESSING. |
|||
|
C - Required if the SOP Instance was created in response to a Frame-Level retrieve request |
The following constraints on Image Attributes take precedence over the descriptions given in the Module Attribute Tables.
For Contrast/Bolus Agent Sequence (0018,0012) DCID 3850 “Contrast Bolus Substance” shall be used.
The Overlay Plane Module and VOI LUT Module shall not be used in a Standard Extended SOP Class of the Intravascular Optical Coherence Tomography Image.
In order to annotate images, whether during acquisition or subsequently, SOP Instances of the Grayscale Softcopy Presentation State Storage or the Structured Report Storage SOP Classes that reference the image SOP Instance may be used.
Pseudo-color presentation information may be applied through the use of separate Pseudo-color Softcopy Presentation State SOP Instances.
No standard mechanism is provided for inclusion of annotations within the image SOP Instance itself and implementers are discouraged from using private extensions to circumvent this restriction.
| DICOM PS3.3 2024c - Information Object Definitions |
|---|