| DICOM PS3.3 2024c - Information Object Definitions |
|---|
Table A.35.17-1 specifies the Modules of the Simplified Adult Echo SR IOD.
The document shall be constructed from TID 5300 “Simplified Echo Procedure Report” invoked at the root node.
Value Type (0040,A040) in Content Sequence (0040,A730) of the SR Document Content Module is constrained to the following Enumerated Values (see Table C.17.3-7 for Value Type definitions):
Relationships between Content Items in the content of this IOD shall be conveyed in the by-value mode. See Table C.17.3-8 for Relationship Type definitions.
Relationships by-reference are forbidden. Therefore, Referenced Content Item Identifier (0040,DB73) is not present in any of the Content Items within the SR Document Content Module.
Table A.35.17-2 specifies the relationship constraints of this IOD.
Table A.35.17-2. Relationship Content Constraints for Simplified Adult Echo SR IOD
| DICOM PS3.3 2024c - Information Object Definitions |
|---|