DICOM PS3.3 2024c - Information Object Definitions
A.34.3.4 12-Lead ECG IOD Content Constraints
(Current)
Prev
A.34.3 12-Lead ECG IOD
Next
PS3.3
>
Composite Information Object Definitions (Normative)
>
Waveform IODs
>
12-Lead ECG IOD
>
12-Lead ECG IOD Content Constraints
A.34.3.4 12-Lead ECG IOD Content Constraints
A.34.3.4.1 Modality
The value of Modality (0008,0060) shall be ECG.
DICOM PS3.3 2024c - Information Object Definitions