DICOM PS3.3 2024c - Information Object Definitions
A.34.12.4 Routine Scalp Electroencephalogram IOD Constraints
(Current)
Prev
A.34.12 Routine Scalp Electroencephalogram IOD
Next
PS3.3
>
Composite Information Object Definitions (Normative)
>
Waveform IODs
>
Routine Scalp Electroencephalogram IOD
>
Routine Scalp Electroencephalogram IOD Constraints
A.34.12.4 Routine Scalp Electroencephalogram IOD Constraints
A.34.12.4.1 Modality
The value of Modality (0008,0060) shall be EEG.
DICOM PS3.3 2024c - Information Object Definitions