DICOM PS3.2 2024c - Conformance
N.5.2.8.9 SCP of the Basic Annotation Box SOP Class

Table N.5-53 lists the supported DIMSE Service for the Basic Annotation Box SOP Class:

Table N.5-53. Services for the Basic Annotation Box SOP Class - SCP

DIMSE Service Element

Purpose

N-SET

Set each Image Attributes for a previously created film box


Table N.5-54 lists the supported N-SET Attributes for Basic Annotation Box SOP Class:

[List the supported Attributes and their possible Values. Indicate the default Value when relevant. See example below.]

Table N.5-54. Supported N-SET Attributes for Basic Annotation Box SOP Class - SCP

Attribute Name

Tag

Values

Default

Annotation Position

(2030,0010)

Text String

(2030,0020)

Free text


N.5.2.8.10 SCP of the Print Job SOP Class

Table N.5-55 lists the supported DIMSE Services for the Print Job SOP Class:

Table N.5-55. Services for the Print Job SOP Class - SCP

DIMSE Service Element

Purpose

N-EVENT-REPORT

Report the printer status in an asynchronous way

N-GET

Retrieve printer information and status.


An N-EVENT-REPORT request can be sent by the SCP at any time during an Association if the print Job SOP Class has been negotiated by the SCU.

Table N.5-56 lists the supported Event Types and Attributes within the N-EVENT-REPORT.

Table N.5-56. Print Job SOP Class N-EVENT-REPORT- SCP

Event Type name

Event Type ID

Attribute Name

Tag

Values

Pending

1

Execution Status Info

(2100,0030)

[Indicate the possible Values supported by the printer out of the Defined Terms Table. See PS 3.3 Section C.13.9.1 for Defined Terms when the Execution Status info is PENDING or FAILURE]

Film Session Label

(2000,0050)

Printer Name

(2110,0030)

Printing

2

Execution Status Info

(2100,0030)

NORMAL

Film Session Label

(2000,0050)

Printer Name

(2110,0030)

Done

3

Execution Status Info

(2100,0030)

NORMAL

Film Session Label

(2000,0050)

Printer Name

(2110,0030)

Failure

4

Execution Status Info

(2100,0030)

[Indicate the possible Values supported by the printer out of the Defined Terms Table See PS 3.3 Section C.13.9.1 for Defined Terms when the Execution Status info is PENDING or FAILURE]

Film Session Label

(2000,0050)

Printer Name

(2110,0030)


[Remove the complete table if N-GET is not supported.]

Table N.5-57 lists the supported N-GET Attributes for Print Job SOP Class:

[List the supported Attributes and the supported Values when relevant. Remove the non-supported Attributes from the table.]

Table N.5-57. Supported N-GET Attributes for the Print Job SOP Class - SCP

Attribute Name

Tag

Values

Print Priority

(2000,0020)

<<HIGH

MEDIUM

LOW>>

Execution Status

(2100,0020)

<<PENDING

PRINTING

DONE

FAILURE>>

Execution Status Info

(2100,0030)

[Indicate the possible Values supported by the printer out of the Defined Terms Table. See PS3.3 Section C.13.9.1 for Defined Terms when the Execution Status info is PENDING or FAILURE]

Creation Date

(2100,0040)

Creation Time

(2100,0050)

Originator

(2100,0070)

Printer Name

(2110,0030)


N.5.2.8.11 SCP of the Presentation LUT SOP Class

Table N.5-58 lists the supported DIMSE Services for the Presentation LUT SOP Class:

Table N.5-58. Services for the Presentation LUT SOP Class SCP

DIMSE Service Element

Purpose

N-CREATE

Create the Presentation LUT Instance

N-DELETE

Delete the Presentation LUT Instance


Table N.5-59 lists the supported N-CREATE Attributes for Presentation LUT:

[List the supported Attributes in Table N.5-59.]

Table N.5-59. Supported N-CREATE Attributes for Presentation LUT - SCP

Attribute Name

Tag

Values

Default

Presentation LUT Sequence

(2050,0010)

>LUT Descriptor

(0028,3002)

>LUT Explanation

(0028,3003)

>LUT Data

(0028,3006)

Presentation LUT Shape

(2050,0020)

<<IDENTITY

LIN OD>>


N.5.2.8.12 SCP of the Printer Configuration Retrieval SOP Class

Table N.5-60 lists the supported DIMSE Services for the Printer Configuration Retrieval SOP Class:

Table N.5-60. Services for the Printer Configuration Retrieval SOP Class

DIMSE Service Element

Purpose

N-GET

Retrieve printer configuration.


DICOM PS3.2 2024c - Conformance