DICOM PS3.2 2024c - Conformance
N.5.2.8.7 SCP of the Basic Grayscale Print Management Meta SOP Class

The Basic Grayscale Print Management Meta SOP Class is composed of the mandatory SOP Classes listed in Table N.5-40:

Table N.5-40. Basic Grayscale Print Management Meta SOP Classes - SCP

SOP Class Name

SOP Class UID

Basic Film Session

1.2.840.10008.5.1.1.1

Basic Film Box

1.2.840.10008.5.1.1.2

Basic Grayscale Image Box

1.2.840.10008.5.1.1.4

Printer

1.2.840.10008.5.1.1.16


N.5.2.8.7.1 Basic Film Session SOP Class

Table N.5-41 lists the supported DIMSE Services for the Basic Film Session SOP Class:

[List the supported DIMSE Service Elements. Remove the non-supported one.]

Table N.5-41. Services for the Basic Film Session SOP Class - SCP

DIMSE Service Element

Purpose

N-CREATE

Create the Film Session

N-SET

Update the Film Session

N-DELETE

Delete the Film Session

N-ACTION

Print all Film Boxes in the Film Session


Table N.5-42 lists the supported N-CREATE and N-SET Attributes for Basic Film Session:

[List the supported Attributes and their possible values/ranges. Indicate the default Value when relevant. See example below.]

Table N.5-42. - Supported N-CREATE and N-SET Attributes for Basic Film Session - SCP

Attribute Name

Tag

Values

Default

Number of Copies

(2000,0010)

<range or fixed Value>

1

Print Priority

(2000,0020)

<<HIGH

LOW

MED>>

LOW

Medium Type

(2000,0030)

<<BLUE FILM

CLEAR FILM

MAMMO BLUE FILM

MAMMO CLEAR FILM

PAPER

…>>

Film Destination

(2000,0040)

<<MAGAZINE

PROCESSOR

BIN_I

…>>

PROCESSOR

Film Session Label

(2000,0050)

Memory Allocation

(2000,0060)

Owner ID

(2100,0160)


[If the SCP supports N-ACTION for the Film Session SOP Class, then the SCP must specify the maximum number of collated films.]

N.5.2.8.7.2 Basic Film Box SOP Class

Table N.5-43 lists the supported DIMSE Services for the Basic Film Box SOP Class:

[List the supported DIMSE Service Elements. Remove the non-supported one.]

Table N.5-43. Services Supported for the Basic Film Box SOP Class - SCP

DIMSE Service Element

Purpose

N-CREATE

Create the Film Box in a previously created Film Session

N-ACTION

Print the Film Box

N-DELETE

Delete the Film Box

N-SET

Update the Film Box


Table N.5-44 lists the supported N-CREATE and N-SET Attributes for Basic Film Box:

[List the supported Attributes and their possible Values. Indicate the default Value when relevant. See example below.]

Table N.5-44. Supported N-CREATE and N-SET Attributes for Basic Film Box - SCP

Attribute Name

Tag

Values

Default

Image Display Format

(2010,0010)

<<STANDARD\C,R

ROW\R1,R2,R3, etc.

COL\C1,C2,C3, etc.

SLIDE

SUPERSLIDE

CUSTOM\i>>

STANDARD\1,1

Annotation Display Format ID

(2010,0030)

<Possible Values>

Film Orientation

(2010,0040)

<<PORTRAIT

LANDSCAPE>>

PORTRAIT

Film Size ID

(2010,0050)

<<8INX10IN

8_5INX11IN

10INX12IN

11INX14IN

11INX17IN

14INX14IN

14INX17IN

24CMX24CM

24CMX30CM

A4

A3

…>>

Magnification Type

(2010,0060)

<<REPLICATE

BILINEAR

CUBIC

NONE

…>>

CUBIC

Smoothing Type

(2010,0080)

<possible Values or range>

143

Border Density

(2010,0100)

<<BLACK

WHITE

i, where i represents the desired density in hundredths of OD

…>>

BLACK

Empty Image Density

(2010,0110)

<<BLACK

WHITE

i, where i represents the desired density in hundredths of OD

…>>

BLACK

Minimum Density

(2010,0120)

<possible Values or range in hundredths of OD>

20

Maximum Density

(2010,0130)

<possible Values or range in hundredths of OD>

320

Trim

(2010,0140)

<<YES

NO>>

NO

Configuration Information

(2010,0150)

Illumination

(2010,015E)

<possible Values or range>

2000

Reflective Ambient Light

(2010,0160)

<possible Values or range>

10

Referenced Film Session Sequence.

(2010,0500)

> Referenced SOP Class UID

(0008,1150)

1.2.840.10008.5.1.1.1

> Referenced SOP Instance UID

(0008,1155)

Referenced Image Box Sequence

(2010,0510)

Provided in the N-CREATE-RSP

> Referenced SOP Class UID

(0008,1150)

1.2.840.10008.5.1.1.4

> Referenced SOP Instance UID

(0008,1155)

Referenced Annotation Box Sequence

(2010,0520)

> Referenced SOP Class UID

(0008,1150)

1.2.840.10008.5.1.1.15

> Referenced SOP Instance UID

(0008,1155)

Referenced Presentation LUT Sequence

(2050,0500)

> Referenced SOP Class UID

(0008,1150)

1.2.840.10008.5.1.1.23

> Referenced SOP Instance UID

(0008,1155)


[Describe each supported custom Image Display Format (2010,0010) and provide details such as position and dimensions of each composing Image Box, including the numbering scheme of the image positions.]

[Describe each supported Annotation Display Format ID (2010,0030) (e.g., position and dimensions of annotation box, font, number of characters).]

[Describe supported configuration information (e.g., identification, content).]

N.5.2.8.7.3 Basic Grayscale Image Box SOP Class

Table N.5-45 lists the supported DIMSE Service for the Basic Grayscale Image Box SOP Class:

Table N.5-45. Services for the Basic Grayscale Image Box SOP Class- SCP

DIMSE Service Element

Purpose

N-SET

Set each Image Attributes for a previously created Film Box


Table N.5-46 lists the supported N-SET Attributes for Basic Grayscale Image Box:

[List the supported Attributes and their possible Values. Indicate the default Value when relevant. See example below.]

Table N.5-46. Supported N-SET Attributes for Basic Grayscale Image Box - SCP

Attribute name

Tag

Values

Default

Magnification Type

(2010,0060)

<<REPLICATE

BILINEAR

CUBIC

NONE

…>>

CUBIC

Smoothing Type

(2010,0080)

<possible Values or range>

143

Minimum Density

(2010,0120)

<possible Values or range in hundredths of OD>

Maximum Density

(2010,0130)

<possible Values or range in hundredths of OD>

320

Configuration Information

(2010,0150)

Image Box Position

(2020,0010)

<possible Position Values or range>

Polarity

(2020,0020)

<<NORMAL

REVERSE>>

NORMAL

Requested Image Size

(2020,0030)

width, x-dimension, in mm

Requested Decimate/Crop Behavior

(2020,0040)

<<DECIMATE

CROP

FAIL>>

Basic Grayscale Image Sequence

(2020,0110)

>Samples per Pixel

(0028,0002)

1

>Photometric Interpretation

(0028,0004)

<<MONOCHROME1

MONOCHROME2>>

>Rows

(0028,0010)

>Columns

(0028,0011)

>Pixel Aspect Ratio

(0028,0034)

1\1

>Bits Allocated

(0028,0100)

<<8

16>>

>Bits Stored

(0028,0101)

<<8

12>>

>High Bit

(0028,0102)

<<7

11>>

>Pixel Representation

(0028,0103)

0

0

>Pixel Data

(7FE0,0010)

Referenced Presentation LUT Sequence

(2050,0500)

>Referenced SOP Class UID

(0008,1150)

1.2.840.10008.5.1.1.23

>Referenced SOP Instance UID

(0008,1155)


[If cropping or decimating of images is supported, describe the algorithm for removing rows and columns from the image.]

N.5.2.8.7.4 Printer SOP Class

Table N.5-47 lists the supported DIMSE Services for the Printer SOP Class:

Table N.5-47. Services for the Printer SOP Class - SCP

DIMSE Service Element

Purpose

N-EVENT-REPORT

Report the printer status in an asynchronous way

N-GET

Retrieve printer information and status.


Table N.5-48 lists the Printer SOP Class N-EVENT-REPORT Behavior:

Table N.5-48. Printer SOP Class N-EVENT-REPORT Behavior

Event Type Name

Event Type ID

Attribute Name

Tag

Values

Normal

1

N/A

Warning

2

Printer Status info

(2110,0020)

[Indicate the possible Values supported by the printer out of the defined terms table. See PS 3.3 Section C.13.9.1 for Defined Terms when the Printer Status is equal to WARNING or FAILURE]

Film Destination

(2000,0040)

Printer Name

(2110,0030)

Failure

3

Printer Status info

(2110,0020)

[Indicate the possible Values supported by the printer out of the defined terms table. See PS 3.3 Section C.13.9.1 for Defined Terms when the Printer Status is equal to WARNING or FAILURE]

Film Destination

(2000,0040)

Printer Name

(2110,0030)


Table N.5-49 lists the supported N-GET Attributes for Printer SOP Class:

[List the supported Attributes. Remove the non-supported Attributes from the Table]

Table N.5-49. Supported N-GET Attributes for the Printer SOP Class - SCP

Attribute Name

Tag

Values

Printer Status

(2110,0010)

<<NORMAL

WARNING

FAILURE>>

Printer Status Info

(2110,0020)

[Indicate the possible Values supported by the printer out of the defined terms table. See PS 3.3 Section C.13.9.1 for Defined Terms when the Printer Status is equal to WARNING or FAILURE]

Printer Name

(2110,0030)

Manufacturer

(0008,0070)

Manufacturer Model Name

(0008,1090)

Device Serial Number

(0018,1000)

Software Versions

(0018,1020)

Date Last Calibration

(0018,1200)

Time Last Calibration

(0018,1201)


DICOM PS3.2 2024c - Conformance