| DICOM PS3.2 2024c - Conformance |
|---|
[Note that the appendices defined in the following subsections are a mandatory part of the DICOM Conformance Statement and must be filled in by any product, that creates DICOM SR objects.]
[For each SR TID (including Private TIDs) that is created by the system (See Overview Section N.1.1.1) provide an Annex B.x.]
[If you are extending a TID by adding additional concepts indicate this extension by adding an asterisk to the TID number in the TID column (e.g.,4000*).]
[If your product creates SR Instances of a TID which includes long lists of measurements, they can also be documented in an external file. For details refer to the instructions right before Section N.10.2.]
This section provides the detailed content encoding for all TIDs supported by <product>.
Throughout the tables listed in Section N.10 the following codes are used for the "Source" and "Presence of Content Item" columns.
In the "Source" column, the following Values can be used:
CONFIGURATION: The Value is copied from the system configuration.
QUERY: The Value is determined by performing a query of any of the supported Query/Retrieve Services.
SCANNED: The Value is read from a barcode scanner or similar device.
SRC_INSTANCE: The Value is copied from previously created/received SOP Instances.
In the "Presence of Content Item" the following Values can be used:
ALWAYS: the module, functional group macro, Attributes or Value is always present.
CONDITIONAL: the presence of the module, functional group macro, Attributes or Value is dependent on a condition. The condition must be listed in the "Comments" column.
SRC_COPY: The presence of the Attributes and Values depends on the availability of these in the source instances, which are used for copying this information.
EMPTY: The Attribute is present but without a Value (zero length).
Table N.10-1 shows the encoding of content of a DICOM Mammography CAD SR (TID 4000).
[The following table shows how to document TID content usage, with TID 4000 as an example. Modify to match your product implementation, e.g., select supported concepts and Values and add additional templates as needed. In the "Value" column you can either list the coded Values directly, reference a CID from DICOM PS3.16 if used unmodified or provide a table in Section N.10.1.1, if you are using more than two codes (otherwise codes can be added directly to the table). For more complex TIDs it is possible to split the table below into multiple tables following the Template Structure defined in DICOM PS 3.16.]
| DICOM PS3.2 2024c - Conformance |
|---|