  <chapter label="" status="1" xml:id="chapter_Notice">
    <title>Notice and Disclaimer</title>
    <para xml:id="para_c2e206ef-20e4-4509-93a5-e03025aa0868">The information in this publication was considered technically sound by the consensus of
      persons engaged in the development and approval of the document at the time it was developed.
      Consensus does not necessarily mean that there is unanimous agreement among every person
      participating in the development of this document.</para>
    <para xml:id="para_ac4a3c48-e3a6-452e-baa1-f0f76d183113">NEMA standards and guideline publications, of which the document contained herein is one,
      are developed through a voluntary consensus standards development process. This process brings
      together volunteers and/or seeks out the views of persons who have an interest in the topic
      covered by this publication. While NEMA administers the process and establishes rules to
      promote fairness in the development of consensus, it does not write the document and it does
      not independently test, evaluate, or verify the accuracy or completeness of any information or
      the soundness of any judgments contained in its standards and guideline publications.</para>
    <para xml:id="para_025b5f35-5167-42ae-8161-a222e1c32613">NEMA disclaims liability for any personal injury, property, or other damages of any nature
      whatsoever, whether special, indirect, consequential, or compensatory, directly or indirectly
      resulting from the publication, use of, application, or reliance on this document. NEMA
      disclaims and makes no guaranty or warranty, expressed or implied, as to the accuracy or
      completeness of any information published herein, and disclaims and makes no warranty that the
      information in this document will fulfill any of your particular purposes or needs. NEMA does
      not undertake to guarantee the performance of any individual manufacturer or seller's products
      or services by virtue of this standard or guide.</para>
    <para xml:id="para_0c73fab9-dc62-4205-abc9-f205c3446a9a">In publishing and making this document available, NEMA is not undertaking to render
      professional or other services for or on behalf of any person or entity, nor is NEMA
      undertaking to perform any duty owed by any person or entity to someone else. Anyone using
      this document should rely on his or her own independent judgment or, as appropriate, seek the
      advice of a competent professional in determining the exercise of reasonable care in any given
      circumstances. Information and other standards on the topic covered by this publication may be
      available from other sources, which the user may wish to consult for additional views or
      information not covered by this publication.</para>
    <para xml:id="para_e085cb9a-9ce0-474e-bc3f-26f202d9760e">NEMA has no power, nor does it undertake to police or enforce compliance with the contents
      of this document. NEMA does not certify, test, or inspect products, designs, or installations
      for safety or health purposes. Any certification or other statement of compliance with any
      health or safety-related information in this document shall not be attributable to NEMA and is
      solely the responsibility of the certifier or maker of the statement.</para>
  </chapter>
  <chapter label="" status="1" xml:id="chapter_Foreword">
    <title>Foreword</title>
    <para xml:id="para_27d3534f-c737-4874-b246-099bb90d6f88">This DICOM Standard was developed according to the procedures of the DICOM Standards Committee.</para>
    <para xml:id="para_c86dc638-fa1f-48e9-b595-9f7aa79d3a33">The DICOM Standard is structured as a multi-part document using the guidelines established in <xref linkend="biblio_ISODirectives2"/>.</para>
  </chapter>
  <chapter label="1" status="1" xml:id="chapter_1">
    <title>Scope and Field of Application</title>
    <para xml:id="para_6ebcd549-9e81-445f-a340-798e5d95435c">This Part of the DICOM Standard specifies the transformations between DICOM and other representations of the same information.</para>
    <para xml:id="para_5c4dd9aa-12f4-47a5-9ab6-e611f2e3d27d">Currently, this Part addresses:</para>
    <itemizedlist>
        <listitem>
          <para xml:id="para_221823d2-2e00-43df-97ff-5f5993476370">compatible components of the NCI Annotation and Image Markup (AIM) to and from DICOM SR Measurement Templates, applicable to quantitative and categorical description of regions of interest in images</para>
        </listitem>
      </itemizedlist>
    <para xml:id="para_bd1eecec-8bb7-4440-9388-d9078437459c">Transformation of diagnostic imaging reports between DICOM SR and HL7 CDA representations is addressed in PS3.20.</para>
  </chapter>
  <chapter label="2" status="1" xml:id="chapter_2">
    <title>Normative and Informative References</title>
    <para xml:id="para_c511c4d9-1dec-40d0-a969-fcf00f640570">The following standards contain provisions that, through reference in this text, constitute provisions of this Standard. At the time of publication, the editions indicated were valid. All standards are subject to revision, and parties to agreements based on this Standard are encouraged to investigate the possibilities of applying the most recent editions of the standards indicated below.</para>
    <bibliography>
      <biblioentry xml:id="biblio_ISODirectives2">
        <abbrev>ISO/IEC Directives, Part 2</abbrev>
        <author>
          <orgname>ISO/IEC</orgname>
        </author>
        <date>2016/05</date>
        <edition>7.0</edition>
        <title>Rules for the structure and drafting of International Standards</title>
        <bibliosource>
          <link xl:href="http://www.iec.ch/members_experts/refdocs/iec/isoiecdir-2%7Bed7.0%7Den.pdf"/>
        </bibliosource>
      </biblioentry>

      <biblioentry xml:id="biblio_IEEE754">
        <abbrev>IEEE 754</abbrev>
        <author>
          <orgname>IEEE</orgname>
        </author>
        <date>1985</date>
        <title>32-bit and 64-bit Floating Point Number Representations</title>
      </biblioentry>
      
      <biblioentry xml:id="biblio_ISO_21090">
        <abbrev>ISO 21090</abbrev>
        <author>
          <orgname>ISO</orgname>
        </author>
        <date>2011</date>
        <title>Health informatics -- Harmonized data types for information interchange</title>
        <bibliosource>
          <link xl:href="http://www.iso.org/iso/iso_catalogue/catalogue_tc/catalogue_detail.htm?csnumber=35646"/>
        </bibliosource>
      </biblioentry>
      
      <biblioentry xml:id="biblio_HL7V3DTR1">
        <abbrev>V3 DT R1</abbrev>
        <author>
          <orgname>HL7</orgname>
        </author>
        <date>2004</date>
        <title>HL7 Version 3 Standard: XML Implementation Technology Specification - Wire Format Compatible Release 1 Data Types, Release 1</title>
        <bibliosource>
          <link xl:href="http://www.hl7.org/implement/standards/product_brief.cfm?product_id=357"/>
        </bibliosource>
      </biblioentry>
      
      <biblioentry xml:id="biblio_XML">
        <abbrev>XML</abbrev>
        <author>
          <orgname>W3C</orgname>
        </author>
        <date>2006/09/29</date>
        <title>Extensible Markup Language (XML) 1.1</title>
        <bibliosource>
          <link xl:href="http://www.w3.org/TR/2006/REC-xml11-20060816/"/>
        </bibliosource>
      </biblioentry>
      
      <biblioentry xml:id="biblio_XSD_DataTypes">
        <abbrev>XSD Datatypes</abbrev>
        <author>
          <orgname>W3C</orgname>
        </author>
        <date>2004/10/28</date>
        <title>XML Schema Part 2: Datatypes</title>
        <bibliosource>
          <link xl:href="http://www.w3.org/TR/xmlschema-2/"/>
        </bibliosource>
      </biblioentry>
       
      <biblioentry xml:id="biblio_AIMV3V4Changes">
        <abbrev>AIM v3 v4 changes</abbrev>
        <author>
          <orgname>NCI</orgname>
        </author>
        <date>2014/04/10</date>
        <title>Changes from AIM 3.0, Revision 11 to AIM Foundation - Summary of Changes Between AIM 3.0 and 4.0 Models</title>
        <bibliosource>
          <link xl:href="https://wiki.nci.nih.gov/display/AIM/Changes+from+AIM+3.0%2C+Revision+11+to+AIM+Foundation#ChangesfromAIM3.0,Revision11toAIMFoundation-ISO21090DataTypes">https://​wiki.nci.nih.gov/​display/​AIM/​Changes+from+AIM+3.0%2C​+Revision+11+to+AIM+​Foundation#​ChangesfromAIM3.0,Revision11toAIMFoundation​-ISO21090DataTypes</link>
        </bibliosource>
      </biblioentry>
      
      
      <biblioentry xml:id="biblio_AIM_Model_v4_2">
          <abbrev>AIM Model v4.2</abbrev>
        <author>
          <orgname>NCI</orgname>
        </author>
          <date>2018/02</date>
        <title>Annotation and Image Markup model version 4.2</title>
        <bibliosource>
          <link xl:href="https://wiki.nci.nih.gov/display/AIM/Annotation+and+Image+Markup+-+AIM#AnnotationandImageMarkup-AIM-AIMModel">https://​wiki.nci.nih.gov/​display/​AIM/​Annotation+​and+​Image+​Markup​+-+AIM#​Annotationand​ImageMarkup​-AIM-​AIMModel</link>
        </bibliosource>
      </biblioentry>
      
      <biblioentry xml:id="biblio_AIM_Model_Extending">
        <abbrev>Extending AIM</abbrev>
        <author>
          <orgname>NCI</orgname>
        </author>
        <date>2014/12/03</date>
        <title>Extending the AIM Model</title>
        <bibliosource>
          <link xl:href="https://wiki.nci.nih.gov/display/AIM/Extending+the+AIM+Model"/>
        </bibliosource>
      </biblioentry>
      
      <biblioentry xml:id="biblio_AIM_2010">
        <abbrev>caBIG™ AIM</abbrev>
        <biblioset relation="journal">
          <title>J Digit Imaging</title>
        </biblioset>
        <biblioset relation="article">
          <authorgroup>
            <author>
              <personname>Channin, David S</personname>
            </author>
            <author>
              <personname>Mongkolwat, Pattanasak</personname>
            </author>
            <author>
              <personname>Kleper, Vladimir</personname>
            </author>
            <author>
              <personname>Sepukar, Kastubh</personname>
            </author>
            <author>
              <personname>Rubin, Daniel L</personname>
            </author>
          </authorgroup>
          <date>2010/04</date>
          <volumenum>23</volumenum>
          <issuenum>2</issuenum>
          <pagenums>217–225</pagenums>
          <title>The caBIG™ Annotation and Image Markup Project</title>
          <biblioid class="doi">10.1007/s10278-009-9193-9</biblioid>
          <bibliosource>
            <link xl:href="http://www.ncbi.nlm.nih.gov/pmc/articles/PMC2837161/"/>
          </bibliosource>
        </biblioset>
      </biblioentry>
      
      
      <biblioentry xml:id="biblio_AIM_2012">
        <abbrev>AIM Template Builder</abbrev>
        <biblioset relation="journal">
          <title>Radiographics</title>
        </biblioset>
        <biblioset relation="article">
          <authorgroup>
            <author>
              <personname>Mongkolwat, Pattanasak</personname>
            </author>
            <author>
              <personname>Channin, David S</personname>
            </author>
            <author>
              <personname>Kleper, Vladimir</personname>
            </author>
            <author>
              <personname>Rubin, Daniel</personname>
            </author>
          </authorgroup>
          <date>2012/07</date>
          <volumenum>32</volumenum>
          <issuenum>4</issuenum>
          <pagenums>1223–1232</pagenums>
          <title>An Open-Source and Open-Access Cancer Biomedical Informatics Grid Annotation and Image Markup Template Builder</title>
          <biblioid class="doi">10.1148/rg.324115080</biblioid>
          <bibliosource>
            <link xl:href="http://www.ncbi.nlm.nih.gov/pmc/articles/PMC3393884/"/>
          </bibliosource>
        </biblioset>
      </biblioentry>
      
      <biblioentry xml:id="biblio_AIM_2014">
        <abbrev>AIM Foundation Model</abbrev>
        <biblioset relation="journal">
          <title>J Digit Imaging</title>
        </biblioset>
        <biblioset relation="article">
          <authorgroup>
            <author>
              <personname>Mongkolwat, Pattanasak</personname>
            </author>
            <author>
              <personname>Kleper, Vladimir</personname>
            </author>
            <author>
              <personname>Talbot, Skip</personname>
            </author>
            <author>
              <personname>Rubin, Daniel</personname>
            </author>
          </authorgroup>
          <date>2014/12</date>
          <volumenum>27</volumenum>
          <issuenum>6</issuenum>
          <pagenums>692–701</pagenums>
          <title>The National Cancer Informatics Program (NCIP) Annotation and Image Markup (AIM) Foundation Model</title>
          <biblioid class="doi">10.1007/s10278-014-9710-3</biblioid>
          <bibliosource>
            <link xl:href="http://www.ncbi.nlm.nih.gov/pmc/articles/PMC4391072/"/>
          </bibliosource>
        </biblioset>
      </biblioentry>
      

    </bibliography>
  </chapter>
  <chapter label="3" status="1" xml:id="chapter_3">
    <title>Definitions</title>
    <para xml:id="para_3f0db0be-357e-4065-9bb3-66eed0426b44">For the purposes of this Standard the following definitions apply.</para>
      <glossary>
    <glossdiv label="3.1" status="2" xml:id="sect_3.1">
      <title>Codes and Controlled Terminology Definitions:</title>
        <para xml:id="para_ee8c43ae-dcc1-4117-913d-ee6f0bd3ea07">This Part of the Standard makes use of the following terms defined in <olink targetdoc="PS3.16" targetptr="PS3.16" xrefstyle="select: labelnumber"/>:</para>
        <glossentry xml:id="glossentry_CodingScheme" xreflabel="Coding Scheme">
            <glossterm>Coding Scheme</glossterm>
            <glossdef>
                <para xml:id="para_64f34980-8d9f-4bbb-9b95-7f542c6ab9df">
<olink targetdoc="PS3.16" targetptr="glossentry_CodingScheme" xrefstyle="template:See %t in %o"/>.</para>
            </glossdef>
        </glossentry>
            <glossentry xml:id="glossentry_ContextGroup" xreflabel="ContextGroup">
                <glossterm>Context Group</glossterm>
                <glossdef>
                    <para xml:id="para_dad30c69-6f47-4bdd-a783-9d74385dffce">
<olink targetdoc="PS3.16" targetptr="glossentry_ContextGroup" xrefstyle="template:See %t in %o"/>.</para>
                </glossdef>
            </glossentry>
        <glossentry xml:id="glossentry_ContextID" xreflabel="Context ID (CID)">
            <glossterm>Context ID</glossterm>
            <acronym>CID</acronym>
            <glossdef>
                <para xml:id="para_82b77c8e-590a-45c5-8515-43e9633218d2">
<olink targetdoc="PS3.16" targetptr="glossentry_ContextID" xrefstyle="template:See %t in %o"/>.</para>
            </glossdef>
        </glossentry>
        <glossentry xml:id="glossentry_Template" xreflabel="Template">
            <glossterm>Template</glossterm>
            <glossdef>
                <para xml:id="para_85f11614-951d-4f68-8937-fb1fc8017251">
<olink targetdoc="PS3.16" targetptr="glossentry_Template" xrefstyle="template:See %t in %o"/>.</para>
            </glossdef>
        </glossentry>
        <glossentry xml:id="glossentry_TemplateID" xreflabel="Template ID (TID)">
            <glossterm>Template ID</glossterm>
            <acronym>TID</acronym>
            <glossdef>
                <para xml:id="para_bdb46102-4663-47e3-b241-7f35ccfa04da">
<olink targetdoc="PS3.16" targetptr="glossentry_TemplateID" xrefstyle="template:See %t in %o"/>.</para>
            </glossdef>
        </glossentry>
    </glossdiv>
          <glossdiv label="3.2" status="2" xml:id="sect_3.2">
          <title>DICOM Introduction and Overview Definitions:</title>
          <para xml:id="para_ff80733f-b221-4532-b4e0-bcdeb3ecb776">This Part of the Standard makes use of the following terms defined in <olink targetdoc="PS3.1" targetptr="PS3.1" xrefstyle="select: labelnumber"/>:</para>
              <glossentry xml:id="glossentry_Attribute" xreflabel="Attribute">
                  <glossterm>Attribute</glossterm>
                  <glossdef>
                      <para xml:id="para_e03e3037-9ba1-4e67-a0e2-e91f9bb6a052">
<olink targetdoc="PS3.1" targetptr="glossentry_Attribute" xrefstyle="template:See %t in %o"/>.</para>
                  </glossdef>
              </glossentry>
              <glossentry xml:id="glossentry_ServiceObjectPairClass" xreflabel="Service-Object Pair Class (SOP Class)">
                  <glossterm>Service-Object Pair Class</glossterm>
                  <acronym>SOP Class</acronym>
                  <glossdef>
                      <para xml:id="para_8582d8e3-08f9-4410-b1cb-97c7f1a87dae">
<olink targetdoc="PS3.1" targetptr="glossentry_ServiceObjectPairClass" xrefstyle="template:See %t in %o"/>.</para>
                  </glossdef>
              </glossentry>
      </glossdiv>
          <glossdiv label="3.3" status="2" xml:id="sect_3.3">
          <title>DICOM Service Class Definitions:</title>
          <para xml:id="para_470dfb67-4a30-4e93-8aad-ba657aecc655">This Part of the Standard makes use of the following terms defined in <olink targetdoc="PS3.4" targetptr="PS3.4" xrefstyle="select: labelnumber"/>:</para>
              <glossentry xml:id="glossentry_ServiceObjectPairInstance" xreflabel="Service-Object Pair Instance (SOP Instance)">
                  <glossterm>Service-Object Pair Instance</glossterm>
                  <acronym>SOP Instance</acronym>
                  <glossdef>
                      <para xml:id="para_d644ed99-4b7e-47fe-9ed0-cd5de6ee0bbf">
<olink targetdoc="PS3.4" targetptr="glossentry_ServiceObjectPairInstance" xrefstyle="template:See %t in %o"/>.</para>
                  </glossdef>
              </glossentry>
      </glossdiv>
          <glossdiv label="3.4" status="2" xml:id="sect_3.4">
              <title>DICOM Information Object Definitions:</title>
              <para xml:id="para_aee0766f-e413-40e0-956e-ddc17d7b8759">This Part of the Standard makes use of the following terms defined in <olink targetdoc="PS3.3" targetptr="PS3.3" xrefstyle="select: labelnumber"/>:</para>
              <glossentry xml:id="glossentry_AttributeMacro" xreflabel="Attribute Macro">
                  <glossterm>Attribute Macro</glossterm>
                  <glossdef>
                      <para xml:id="para_553b34dd-0d87-410e-909d-d6a21894a808">
<olink targetdoc="PS3.3" targetptr="glossentry_AttributeMacro" xrefstyle="template:See %t in %o"/>.</para>
                  </glossdef>
              </glossentry>
              <glossentry xml:id="glossentry_ContentItem" xreflabel="Content Item">
                  <glossterm>Content Item</glossterm>
                  <glossdef>
                      <para xml:id="para_17149a46-ef3e-4472-9c0c-31fb42fb491c">
<olink targetdoc="PS3.3" targetptr="glossentry_ContentItem" xrefstyle="template:See %t in %o"/>.</para>
                  </glossdef>
              </glossentry>
              <glossentry xml:id="glossentry_ContentTree" xreflabel="Content Tree">
                  <glossterm>Content Tree</glossterm>
                  <glossdef>
                      <para xml:id="para_c9eb6cfd-74ab-44dc-bde3-dcac6a7c1b1d">
<olink targetdoc="PS3.3" targetptr="glossentry_ContentTree" xrefstyle="template:See %t in %o"/>.</para>
                  </glossdef>
              </glossentry>
              <glossentry xml:id="glossentry_Module" xreflabel="Module">
                  <glossterm>Module</glossterm>
                  <glossdef>
                      <para xml:id="para_5a0948b3-6f98-41d9-b06a-e52d8945791d">
<olink targetdoc="PS3.3" targetptr="glossentry_Module" xrefstyle="template:See %t in %o"/>.</para>
                  </glossdef>
              </glossentry>              
          </glossdiv>
          <glossdiv label="3.5" status="2" xml:id="sect_3.5">
              <title>DICOM Data Structures and Encoding</title>
              <para xml:id="para_27b8bb93-f6d3-4e7d-89d2-c0cb855a8531">This Part of the Standard makes use of the following terms defined in <olink targetdoc="PS3.5" targetptr="PS3.5" xrefstyle="select: labelnumber"/>:</para>
              <glossentry xml:id="glossentry_DataSet" xreflabel="Data Set">
                  <glossterm>Data Set</glossterm>
                  <glossdef>
                      <para xml:id="para_9184976e-d2c8-488e-a9ea-61b269b11a91">
<olink targetdoc="PS3.5" targetptr="glossentry_DataSet" xrefstyle="template:See %t in %o"/>.</para>
                  </glossdef>
              </glossentry>
              <glossentry xml:id="glossentry_ValueMultiplicity" xreflabel="Value Multiplicity (VM)">
                  <glossterm>Value Multiplicity</glossterm>
                  <acronym>VM</acronym>
                  <glossdef>
                      <para xml:id="para_990dc00f-d6ab-409b-84bb-2f09abbf0b22">
<olink targetdoc="PS3.5" targetptr="glossentry_ValueMultiplicity" xrefstyle="template:See %t in %o"/>.</para>
                  </glossdef>
              </glossentry>
              <glossentry xml:id="glossentry_ValueRepresentation" xreflabel="Value Representation (VR)">
                  <glossterm>Value Representation</glossterm>
                  <acronym>VR</acronym>
                  <glossdef>
                      <para xml:id="para_2a4c0f7a-b641-4d01-ae19-10be3d93de76">
<olink targetdoc="PS3.5" targetptr="glossentry_ValueRepresentation" xrefstyle="template:See %t in %o"/>.</para>
                  </glossdef>
              </glossentry>
          </glossdiv>
      </glossary>
  </chapter>
  <chapter label="4" status="1" xml:id="chapter_4">
    <title>Symbols and Abbreviations</title>
    <para xml:id="para_3dbfe831-434b-4415-b98d-994263194f57">The following symbols and abbreviations are used in this Part of the Standard.</para>
    <variablelist>
      <varlistentry>
        <term>AIM</term>
        <listitem>
          <para xml:id="para_1e42a4f2-7939-4566-9054-5de2916bf66e"> Annotation and Image Markup</para>
        </listitem>
      </varlistentry>
      <varlistentry>
        <term>DICOM</term>
        <listitem>
          <para xml:id="para_10698848-2ce5-4388-8da5-e40ea1b28953">Digital Imaging and Communications in Medicine</para>
        </listitem>
      </varlistentry>
      <varlistentry>
        <term>IOD</term>
        <listitem>
          <para xml:id="para_dd91013c-5d0d-4185-8b57-23082f65a986">Information Object Definition</para>
        </listitem>
      </varlistentry>
      <varlistentry>
        <term>ISO</term>
        <listitem>
          <para xml:id="para_b5a181a9-5afe-4811-858f-5d1f0b2f0f9c">International Standards Organization</para>
        </listitem>
      </varlistentry>
      <varlistentry>
        <term>NCI</term>
        <listitem>
          <para xml:id="para_b9ab0bd2-f2e5-4e30-916b-18dd3e5c2fde">National Cancer Institute</para>
        </listitem>
      </varlistentry>
      <varlistentry>
        <term>NEMA</term>
        <listitem>
          <para xml:id="para_c51aac7b-e05e-419a-8f88-33c0e50c0c84">National Electrical Manufacturers Association</para>
        </listitem>
      </varlistentry>
      <varlistentry>
        <term>OID</term>
        <listitem>
          <para xml:id="para_90280ea8-6671-4507-97df-485e4e41c5c7">Object Identifier (ISO 8824)</para>
        </listitem>
      </varlistentry>
      <varlistentry>
        <term>SR</term>
        <listitem>
          <para xml:id="para_e682208d-dccc-4c6d-8f0c-233e971e99c7">Structured Reporting</para>
        </listitem>
      </varlistentry>
      <varlistentry>
        <term>UCUM</term>
        <listitem>
          <para xml:id="para_9405bb6a-fac0-4ba5-aeed-2391630050f8">Unified Code for Units of Measure</para>
        </listitem>
      </varlistentry>
      <varlistentry>
        <term>UID</term>
        <listitem>
          <para xml:id="para_b863a9e9-4bc3-47e3-90d6-5dea965f2285">Unique Identifier</para>
        </listitem>
      </varlistentry>
      <varlistentry>
        <term>XML</term>
        <listitem>
          <para xml:id="para_8c3c28b3-8e89-4cdf-893e-9f9d6be4a4ef">Extensible Markup Language</para>
        </listitem>
      </varlistentry>
      <varlistentry>
        <term>XSLT</term>
        <listitem>
          <para xml:id="para_7ff4076d-8d2f-45b1-a587-f0243a791d44">Extensible Stylesheet Language Transformations</para>
        </listitem>
      </varlistentry>
    </variablelist>
  </chapter>
  <chapter label="5" status="1" xml:id="chapter_5">
    <title>Conventions</title>
    <para xml:id="para_6b1753b6-bff7-4f4c-b88c-382e467c95d1">Terms listed in <xref linkend="chapter_3" xrefstyle="template:Section %n"/> Definitions are capitalized throughout the document.</para>
  </chapter>
  <chapter label="A" status="1" xml:id="chapter_A">
    <title>Transformation Between AIM and DICOM SR</title>
    <section label="A.1" status="2" xml:id="sect_A.1">
      <title>Scope and Field of Application</title>
      <para xml:id="para_56b88c55-f7d4-4e3c-aa00-eb78b4c3e6ac">NCI AIM objects that are constrained to specific use cases can be transformed to DICOM SR documents that are based on <olink targetdoc="PS3.16" targetptr="sect_TID_1500" xrefstyle="select: labelnumber"/>.</para>
      <para xml:id="para_7c63a006-cee2-4ad8-9e3c-467db67c402b">NCI AIM specifies a generic model for encoding structured information about medical images. The AIM model and its XML encoding are version specific. The version transformed in this document is Version 4 <xref linkend="biblio_AIM_Model_Extending"/>. Though AIM instances can be generated according to application-specific templates [ref. AIM template builder], such templates are too use-case specific to be detailed in the transformation described in this Part. Rather, common patterns of use implemented by well known implementations (ref. CC, ref ePAD) have been abstracted and are mapped as described in this document.</para>
      <para xml:id="para_033c3dd0-53e9-4113-8ade-94946482106a">DICOM SR specifies a generic model for encoding structured information about DICOM instances. DICOM <olink targetdoc="PS3.16" targetptr="sect_TID_1500" xrefstyle="select: labelnumber title"/> specifies a basic DICOM SR report template for quantitative measurements and categorical statements for single identifiable patient subjects including regions of interest defined by spatial coordinates or segmentations.</para>
    </section>
    <section label="A.2" status="2" xml:id="sect_A.2">
      <title>Use Cases</title>
      <para xml:id="para_3b073ffc-428b-4b23-a641-c7bb2fa0feb7">The basic use case for the transformation from AIM to DICOM SR is exchange of quantitative and categorical information about DICOM images and regions of interest in DICOM images.</para>
      <para xml:id="para_c2aea29d-7828-4e14-81ae-1ae42869086e">AIM and DICOM SR instances both contain references to DICOM images, and both make use of references to other types of DICOM instances such as segmentations. The flow of information between systems that might make use of a conversion between AIM and DICOM SR instances is considered in <xref linkend="figure_A.2-1" xrefstyle="select: label"/>.</para>
      <para xml:id="para_96a98c59-a772-4156-8341-2a24af54d0d5">
        <figure label="A.2-1" pgwide="1" xml:id="figure_A.2-1">
          <title>Annotation Creation and Transformation</title>
          <mediaobject>
            <imageobject>
              <imagedata fileref="figures/PS3.21_A.2-1.svg"/>
            </imageobject>
          </mediaobject>
        </figure>
      </para>
      <para xml:id="para_94fc1b22-2d3f-4ceb-b16a-9df19103d419">Various different transformation scenarios should be considered:</para>
      <orderedlist numeration="loweralpha">
        <listitem>
          <para xml:id="para_e2d88808-b2e0-4fc1-a68b-ac3338e5c132">Transformation of a complete AIM instance to a complete DICOM SR Measurement Report. The receiver optionally selects relevant parts of the transformed document for further processing or static or interactive visualization by the user.</para>
        </listitem>
        <listitem>
          <para xml:id="para_c43212d9-1d95-4fe5-8d2c-965202d2cfc5">Transformation of a subset of a AIM instance to a DICOM SR Measurement Report. This subset comprises the relevant information for a specific use-case.</para>
        </listitem>
        <listitem>
          <para xml:id="para_3ea9d820-2141-4148-9e67-a03e73542eba">Transformation of multiple related AIM instances to a single DICOM SR Measurement Report. The selected related AIM instances comprise the relevant information for a specific use-case.</para>
        </listitem>
      </orderedlist>
      <para xml:id="para_eff50d7c-e351-40bd-b9a3-c6064e13119a">This Part of the Standard does not mandate any particular transformation scenario. Transformations of various compatible components of AIM and DICOM SR <olink targetdoc="PS3.16" targetptr="sect_TID_1500" xrefstyle="select: labelnumber"/> are described, allowing each of these scenarios to be implemented as appropriate. This Part of the Standard enables a deterministic transformation of the first scenario (complete mapping of an AIM instance to DICOM).</para>
      <para xml:id="para_3e852895-49a3-451b-9d3b-7b4f535b7b4d">The primary use-case for the transformation is that of:</para>
      <para xml:id="para_491b352b-dd75-4698-9e54-32796cef2cba">
        <itemizedlist>
          <listitem>
            <para xml:id="para_06aa15e9-dd9a-4722-9bb2-2978fafe699c">taking an existing AIM instance containing one or more annotations, converting it into DICOM SR, storing it in a DICOM storage system, possibly displaying it with a DICOM SR aware application, transforming it back into an AIM instance after retrieval from the DICOM storage system, and reusing it in an AIM-aware application</para>
          </listitem>
        </itemizedlist>
      </para>
      <para xml:id="para_6342c366-fe00-4a3f-a420-aaa15b0e6eb9">An important secondary use-case is transformation of annotation information in different formats into a single format for data aggregation (e.g., "analytics", "data mining", "big data", "machine learning", and "deep learning"). The conversion described here is intended to allow preservation of semantics sufficient for such purposes, regardless of the source format. </para>
      <para xml:id="para_8ed13aec-0fc1-419b-9056-8c31d001c463">It should be understood that DICOM SR created by transforming an AIM object will not necessarily be identical in structure and content to a DICOM SR that might have been created <foreignphrase xml:lang="la">de novo</foreignphrase> by a similar application. For instance, there are various encoding choices that an application implementer may make, especially with respect to the degree of post-coordination of ROI and measurement descriptions, which might result in different structures. It may not be possible to transform a DICOM SR TID 1500 instance into AIM and retain all of its content. Post-coordinated concepts, such as measurement and derivation methods, may be preserved by using multiple CalculationEntity/typeCode entries or by pre-coordinating into a single concept during the transformation.</para>
      <para xml:id="para_2d598bc6-dd3d-4878-89c8-e98d68e15078">Multiple regions of interest, or multiple measurements and categorical statements about a single region of interest, or about the same real world entity (e.g., lesion) identified with different regions of interest (e.g., at different time points or with different modalities), may be encoded in single or multiple AIM or DICOM SR instances. Whether or not a source AIM or DICOM SR implementation encodes more than one region of interest (and their accompanying measurements and categorical statements) in a single instance or in separate instances, and whether the conversion from one form to the other "bundles" multiple instances into a single instance, or "unbundles" a single instance into multiple instances, is not prescribed.
          The AIM 4.2 model allows for the encoding of multiple marked up regions and multiple measurements in a single annotation instance,
          and unlike earlier AIM model versions provides mechanisms for identifying which markup is associated with which measurement using AIM statements;
          otherwise it is necessary to assume that all markup applies to all measurements and vice versa.</para>
        <para xml:id="para_a9c51aa3-df28-410d-a7fe-559f9078f9e8">For example, for RECIST measurements that involve the long and short axis of a lesion,
            it is not only possible to encode in AIM (and map to TID 1500)
            the boundaries of and the area measurements derived from a planar volumetric ROI,
            it is also possible to encode the endpoints of the measured axes of the pair of linear measurements
            and their derived values.</para>
    </section>
    <section label="A.3" status="2" xml:id="sect_A.3">
      <title>Structure of DICOM SR Documents</title>
      <para xml:id="para_0b54e54b-69a8-4d5c-b41c-a6e39d454d7b">DICOM SR documents can be thought of as consisting of a document header and a document body. The header metadata Attribute values are grouped into Modules such as "Patient", "General Study" in <olink targetdoc="PS3.3" targetptr="PS3.3" xrefstyle="select: labelnumber"/>.</para>
      <para xml:id="para_96bafad4-f8af-42e2-82f1-7e19a866837e">The SR Document Content Module (<olink targetdoc="PS3.3" targetptr="sect_C.17.3" xrefstyle="select: docname label"/>) contains the Attributes for the root Content Item, which includes the coded report title. The Content Tree (structured content) of the document body is contained in the nested Content Sequence Items of that Module. "Container" Content Items are part of the Content Sequence. They are structural elements of the SR document body structure. Content items are DICOM SR document nodes within the Content Tree that are connected through "by-value" relationships (for Enhanced SR IODs). The transformations defined in this Part do not support the use of "by-reference" relationships between Content Items.</para>
      <para xml:id="para_1878e5b7-f633-4d87-a6a0-266a9751b98c">
        <figure label="A.3-1" pgwide="1" xml:id="figure_A.3-1">
          <title>SR Document Structure</title>
          <mediaobject>
            <imageobject>
              <imagedata fileref="figures/PS3.21_A.3-1.svg"/>
             </imageobject>
          </mediaobject>
        </figure>
      </para>
      <section label="A.3.1" status="3" xml:id="sect_A.3.1">
        <title>Header</title>
        <para xml:id="para_d647dfce-6b37-4e1d-a7e2-bce1c40bb5eb">The Modules used in a DICOM SR are defined by the Information Object Definition (IOD). A particular DICOM SR template may be encoded using a variety of DICOM SR IODs, depending on the features supported by the template and used by a particular instance. Each SR IOD constrains the Value Types and Relationship Types that are permitted. The Enhanced SR IOD is sufficient to encode TID 1500 instances unless 3D patient-relative coordinates (rather than 2D image-relative coordinates or segmentations) are used to define regions of interest on images, in which case use of the Comprehensive 3D SR IOD or Extensible SR IOD would be required.</para>
        <para xml:id="para_b616016e-a08a-47a1-8b49-aa007a3eabb5">
<xref linkend="table_A.3.1-1" xrefstyle="select: label"/> summarizes the Modules common to the SR IODs that can encode the <olink targetdoc="PS3.16" targetptr="sect_TID_1500" xrefstyle="select: labelnumber"/> as specified in <olink targetdoc="PS3.3" targetptr="PS3.3" xrefstyle="select: labelnumber"/>.</para>
        <table frame="box" label="A.3.1-1" rules="all" xml:id="table_A.3.1-1">
          <caption>Transformation of DICOM SR IOD Modules</caption>
          <thead>
            <tr valign="top">
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_3e777db3-e94e-4866-b4de-5f8d806087d3">IE</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_2be9472c-e183-43b8-ad77-57fbfcd2dc67">Module</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_5bc0652f-be8a-4c1e-b538-61397ec9fe05">Reference</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_5bfa355b-977e-41ed-8fb0-a9b0f98e2ea8">Usage</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_134f2893-4f59-47ed-b797-0e7a5348584f">Transformation</para>
              </th>
            </tr>
          </thead>
          <tbody>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="2">
                <para xml:id="para_9bf5e87d-2307-440b-9686-b2079774e792">Patient</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_856f86eb-dd51-4fa9-b1e8-5316c9651965">Patient</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <olink targetdoc="PS3.3" targetptr="sect_C.7.1.1" xrefstyle="select: labelnumber"/>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para xml:id="para_93e2acb5-4a07-494c-ab5d-58b433164a27">M</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_2c479045-4e55-4fe9-a2c0-e0dd92a89adf">
<xref linkend="sect_A.6.1.1.1" xrefstyle="select: labelnum"/>
</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_fbaf5335-8bfd-4fa3-a14e-38d99d85bc80">Clinical Trial Subject</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <olink targetdoc="PS3.3" targetptr="sect_C.7.1.3" xrefstyle="select: labelnumber"/>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para xml:id="para_9820108c-d97a-4e92-bf88-7e2352f891c9">U</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_465be742-b413-4cbb-afb9-4c2a22efa542">
<xref linkend="sect_A.6.1.1.2" xrefstyle="select: labelnum"/>
</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="3">
                <para xml:id="para_c1b8565a-4334-47ba-b736-f0fe67d8c8dd">Study</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d5e329f7-cfba-4641-8d69-aa3080e28813">General Study</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <olink targetdoc="PS3.3" targetptr="sect_C.7.2.1" xrefstyle="select: labelnumber"/>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para xml:id="para_bd4d28cd-02e4-4ed1-9a44-2a5668414768">M</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b4822b26-faf6-4ebe-baef-9bd0d40cdebf">
<xref linkend="sect_A.6.1.1.3" xrefstyle="select: labelnum"/>
</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_4ee5d7b2-4f2b-45ae-b65e-864c8196a482">Patient Study</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <olink targetdoc="PS3.3" targetptr="sect_C.7.2.2" xrefstyle="select: labelnumber"/>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para xml:id="para_3982e2dd-257a-4bbd-98c5-74f0d60964ea">U</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a2c30ba1-1653-43ba-8526-8581fa2bc2cb">
<xref linkend="sect_A.6.1.1.4" xrefstyle="select: labelnum"/>
</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_8d0146bb-117c-404c-9f79-352962651a99">Clinical Trial Study</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <olink targetdoc="PS3.3" targetptr="sect_C.7.2.3" xrefstyle="select: labelnumber"/>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para xml:id="para_cf265905-e7d9-4204-9dee-3eea60815ffa">U</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_45f30666-dbbf-43f0-be63-b2757c0db400">
<xref linkend="sect_A.6.1.1.5" xrefstyle="select: labelnum"/>
</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="2">
                <para xml:id="para_c8ce99b1-2ef8-48b7-8352-1d151af886c9">Series</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c8626815-e749-4565-8437-b7259692fad6">SR Document Series</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <olink targetdoc="PS3.3" targetptr="sect_C.17.1" xrefstyle="select: labelnumber"/>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para xml:id="para_902fea94-0fe3-499e-ba7f-878610a96415">M</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a21876e7-da14-4932-91e8-3a3bf275fcd0">
<xref linkend="sect_A.6.1.1.6" xrefstyle="select: labelnum"/>
</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_354b8e6e-c804-4a0b-95bf-e7da6971cd6f">Clinical Trial Series</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <olink targetdoc="PS3.3" targetptr="sect_C.7.3.2" xrefstyle="select: labelnumber"/>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para xml:id="para_34c2b825-65b0-4754-b4c2-1c6ebab546b3">U</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_3406ca12-78fa-4d2c-8f81-f35e94fec2b4">
<xref linkend="sect_A.6.1.1.7" xrefstyle="select: labelnum"/>
</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_4fde8886-1c47-4e70-a332-0d1c984ac25a">Equipment</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_431916db-f68a-4d30-ad8e-9d8ed777201b">General Equipment</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <olink targetdoc="PS3.3" targetptr="sect_C.7.5.1" xrefstyle="select: labelnumber"/>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para xml:id="para_f198c49d-ce2d-41fe-9016-5bf44b9e6784">M</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_940e4855-2eca-4836-bd6f-0b8ec4ca528e">
<xref linkend="sect_A.6.1.1.8" xrefstyle="select: labelnum"/>
</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="3">
                <para xml:id="para_ac6dfaef-3236-4777-9197-f597512e86c5">Document</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d32f26ff-9c7c-438d-8a09-9c94c0efeb6c">SR Document General</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <olink targetdoc="PS3.3" targetptr="sect_C.17.2" xrefstyle="select: labelnumber"/>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para xml:id="para_5dd4094b-8bcb-4d09-a9c6-39c28f234723">M</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_de2beac5-6fd0-446e-80f3-d52914db3117">
<xref linkend="sect_A.6.1.1.9" xrefstyle="select: labelnum"/>
</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f054174b-72f4-4d36-b539-a76426c112ca">SR Document Content</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <olink targetdoc="PS3.3" targetptr="sect_C.17.3" xrefstyle="select: labelnumber"/>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para xml:id="para_40f0dd9a-384d-4de8-a0c9-9eee4c2f20df">M</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_8f6cc8d5-9e16-4a3e-b499-2490adc62d6c">
<xref linkend="sect_A.6.1.1.10" xrefstyle="select: labelnum"/>
</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_922d2e4f-dd5c-4644-afbb-ba80bbcb2e58">SOP Common</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <olink targetdoc="PS3.3" targetptr="sect_C.12.1" xrefstyle="select: labelnumber"/>
              </td>
              <td align="center" colspan="1" rowspan="1">
                <para xml:id="para_6ae3a085-effc-4bf2-92c9-4babd21be29e">M</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_05147fe1-b923-479f-8cb6-6a0005483766">
<xref linkend="sect_A.6.1.1.11" xrefstyle="select: labelnum"/>
</para>
              </td>
            </tr>
          </tbody>
        </table>
        <para xml:id="para_a4684a4f-4225-44a9-ade2-0f05331fd2b9">
          <emphasis role="bold">Patient Module</emphasis>
        </para>
        <para xml:id="para_4a4503cf-d5ab-408c-a9ba-08026cfa06d9">The Patient Module specifies the Attributes of the Patient that describe and identify the Patient who is the subject of a Study. This Module contains Attributes of the patient that are needed for interpretation of the Image and are common for all studies performed on the patient.</para>
        <para xml:id="para_cc3dd948-33df-44ec-a9d4-a0fa309ee42e">
          <emphasis role="bold">Clinical Trial Subject Module</emphasis>
        </para>
        <para xml:id="para_bd881b8d-e865-4359-9b74-1ec1625fd056">The Clinical Trial Subject Module contains Attributes that identify a Patient as a clinical trial Subject.</para>
        <para xml:id="para_bac05a58-5239-4b02-b2f5-7cb6b9dcf459">
          <emphasis role="bold">General Study Module</emphasis>
        </para>
        <para xml:id="para_18a847b8-1d2f-42d6-a6df-7b4b63504267">The General Study Module specifies the Attributes that describe and identify the Study performed upon the Patient.</para>
        <para xml:id="para_f4841a00-128d-4376-9cce-a0576bc0c72a">
          <emphasis role="bold">Patient Study Module</emphasis>
        </para>
        <para xml:id="para_abae20db-738b-4a1b-b9b8-712e4a03d279">The Patient Study Module defines the Attributes that provide information about the Patient at the time the Study was performed.</para>
        <para xml:id="para_a4e8449d-c441-4904-9369-d6187cae8d39">
          <emphasis role="bold">Clinical Trial Study Module</emphasis>
        </para>
        <para xml:id="para_b704835e-e852-475b-bbec-6272e3b5aa23">The Clinical Trial Study Module contains Attributes that identify a Study in the context of a clinical trial.</para>
        <para xml:id="para_1920305a-e653-4359-9dc7-f2fcbe8c8bc7">
          <emphasis role="bold">SR Document Series Module</emphasis>
        </para>
        <para xml:id="para_ae9cc063-9439-4ff2-8a81-e8e2fd9161dd">The SR Document Series Module defines the Attributes of the SR Document Series. A Series of SR Documents may contain any number of SR Documents.</para>
        <para xml:id="para_0930b332-4d5d-457a-b841-e2b3fbf0f81a">
          <emphasis role="bold">Clinical Trial Series Module</emphasis>
        </para>
        <para xml:id="para_1c6f9b06-0dc0-4118-b374-d08161df8339">The Clinical Trial Series Module contains Attributes that identify a Series in the context of a clinical trial.</para>
        <para xml:id="para_7c33d933-de82-4a11-ba4f-966795b0b07b">
          <emphasis role="bold">General Equipment Module</emphasis>
        </para>
        <para xml:id="para_4195cd83-3f6a-4b8f-af10-3dfc0d625b70">The General Equipment Module specifies the Attributes that identify and describe the piece of equipment that produced a Series of Composite Instances.</para>
        <para xml:id="para_696ea8fe-20f4-41a0-9e54-a6c9f122b5be">
          <emphasis role="bold">SR Document General Module</emphasis>
        </para>
        <para xml:id="para_7b69a286-f8c7-4ab5-a4e9-1c00986d33ec">The SR Document General Module defines the general Attributes of an SR Document Instance. These Attributes identify the SR Document and provide context for the entire document.</para>
        <para xml:id="para_f60f4bc4-3e82-479b-bec0-fb9ea944502f">
          <emphasis role="bold">SOP Common Module</emphasis>
        </para>
        <para xml:id="para_7c5a3294-3a87-4f2f-93f9-ac9972c9447d">The SOP Common Module defines the Attributes that are required for proper functioning and identification of the associated SOP Instances.</para>
        <para xml:id="para_e92ef29a-6122-49b4-96d9-6ac7a84504e1">
          <emphasis role="bold">SR Document Content Module</emphasis>
        </para>
        <para xml:id="para_d434bc98-526a-4ba7-a9b6-ffe0bfc260a1">The Attributes in this Module convey the content of an SR Document. It specifies the root Content Item and the Content Tree (refer to <xref linkend="figure_A.3-1" xrefstyle="select: label"/>).</para>
      </section>
      <section label="A.3.2" status="3" xml:id="sect_A.3.2">
        <title>Document Body</title>
        <para xml:id="para_d1cdd0a0-347f-4cb9-8597-887958c2365d">The document body is the information that is stored in the DICOM SR Content Tree. The Content Tree is encoded in the SR Document Content Module.</para>
        <section label="A.3.2.1" status="4" xml:id="sect_A.3.2.1">
          <title>DICOM SR "Measurement Report" Template Structure</title>
          <para xml:id="para_616b02ad-f923-4f53-bc6b-5d1353f57730">
<olink targetdoc="PS3.16" targetptr="sect_TID_1500" xrefstyle="select: labelnumber"/> is the top-level template of the DICOM SR Measurement Report (<olink targetdoc="PS3.16" targetptr="PS3.16" xrefstyle="select: labelnumber"/>). It includes sub-templates as shown in <xref linkend="figure_A.3.2.1-1" xrefstyle="select: label"/>.</para>
          <para xml:id="para_c6b40b62-0db1-4fc0-90fa-7bfa90190f64">
            <figure label="A.3.2.1-1" pgwide="1" xml:id="figure_A.3.2.1-1">
              <title>Template Structure Summarized from PS 3.16</title>
              <mediaobject>
                <imageobject>
                  <imagedata fileref="figures/PS3.21_A.3.2.1-1.svg"/>
                </imageobject>
              </mediaobject>
            </figure>
          </para>
          <note>
            <orderedlist>
              <listitem>
                <para xml:id="para_49c25e8f-1739-4ecc-a7f0-1d50559b1bf5">The use of TID 1001 Observation Context within TID 1000 Quotation is not shown because it is not relevant to the mapping use cases.</para>
              </listitem>
              <listitem>
                <para xml:id="para_3827144e-3f15-4340-af6f-8a2b8ce86e86">The use of TID 311 Measurement Statistical Properties and TID 312 Normal Range Properties within TID 310 Measurement Properties is not shown because it is not relevant to the mapping use cases.</para>
              </listitem>
              <listitem>
                <para xml:id="para_5a2501a4-2653-4be4-84c7-7033a7668ba3">The use of TID 1410 Planar ROI Measurements, TID 1411 Volumetric ROI Measurements, and TID 310 Measurement Properties within TID 1420 is not shown because it is not relevant to the mapping use cases.</para>
              </listitem>
            </orderedlist>
          </note>
        </section>
        <section label="A.3.2.2" status="4" xml:id="sect_A.3.2.2">
          <title>Mapping Considerations</title>
          <para xml:id="para_778524ba-3099-4f16-8450-f5a284b054a0">The goal of this document is to specify a mapping between constrained AIM v4 instances and <olink targetdoc="PS3.16" targetptr="sect_TID_1500" xrefstyle="select: labelnumber title"/> DICOM SR documents. The following limitations apply to AIM instances that are mapped to DICOM SR Measurement Reports:</para>
          <itemizedlist>
            <listitem>
              <para xml:id="para_97a2adf8-e1f1-4300-8aad-d40b76f342df">Subject Context: The DICOM SR TID 1500 Measurement Report is restricted to cover exactly one patient subject; the mapping of subject context of fetuses, specimens or devices as subjects is out of scope. Small or large animal identifiers and descriptors (beyond reuse of the normal patient identifiers) are not specifically addressed since no such identifiers are present in the AIM model (e.g., multiple animals imaged as one, and strain descriptions are out of scope). </para>
            </listitem>
            <listitem>
              <para xml:id="para_6bb4f1cc-107c-480b-9ed7-0ea21ac25c4c">The mapping of DICOM SR clinical trial header data (Clinical Trial Subject Module, Clinical Trial Study Module, Clinical Trial Series Module) is not described since no such identifiers are present in the AIM model.</para>
            </listitem>
            <listitem>
              <para xml:id="para_694f40a7-2482-4cd6-a909-2b7d0c1c13f7">The transformation of de-identified objects is not specifically addressed in this mapping, since AIM does not address encoding of the history of de-identification explicitly. I.e., identifiers will be converted unchanged and whether they have been de-identified will not be explicitly signaled.</para>
            </listitem>
            <listitem>
              <para xml:id="para_3bffab92-be11-40da-bb20-8261e87c8263">A subset of spatial coordinate types are mapped, to the extent that both AIM and DICOM SR support the same graphic concepts.</para>
            </listitem>
          </itemizedlist>
        </section>
        <section label="A.3.2.3" status="4" xml:id="sect_A.3.2.3">
          <title>DICOM Composite Object References</title>
          <para xml:id="para_f23fef25-1b8c-423f-913f-fc22c6046a87">The hierarchical Attributes describing DICOM composite object references are used whenever DICOM composite objects are referenced in the Content Tree and are also described in the Image Library templates and the Current Requested Procedure Evidence Sequence (0040,A375) or Pertinent Other Evidence Sequence (0040,A385). Information on relevant DICOM objects referenced within the AIM instance are included in the AIM DicomImageReferenceEntity class.</para>
        </section>
      </section>
    </section>
    <section label="A.4" status="2" xml:id="sect_A.4">
      <title>Structure of AIM Version 4 Instances</title>
      <para xml:id="para_a11c7932-2f72-48e2-a283-2a56f673a71f">AIM instances are encoded in XML according to a schema generated from the AIM Model <xref linkend="biblio_AIM_Model_v4_2"/>, which is defined in UML.
      <xref linkend="figure_A.4-1" xrefstyle="select: label"/> is a simplified view rather than the entire model, showing only those
      UML classes and attributes relevant to the transformations described in this Part.</para>
      <para xml:id="para_bb5d08cd-703d-4e8b-8051-4fa046c2f2aa">
        <figure label="A.4-1" pgwide="1" xml:id="figure_A.4-1">
          <title>AIM v4 Instance Structure</title>
          <mediaobject>
            <imageobject>
              <!-- contentwidth="15cm" parameter was necessary to prevent error during PDF rendering from EA export, now to prevent page overflow; see also "http://www.sagehill.net/docbookxsl/ImageSizing.html" -->
              <imagedata contentwidth="18cm" fileref="figures/PS3.21_A.4-1.svg"/>
             </imageobject>
          </mediaobject>
        </figure>
      </para>
    </section>
    <section label="A.5" status="2" xml:id="sect_A.5">
      <title>AIM v4 Structure</title>
      <para xml:id="para_a413ca3b-92de-408a-928b-924c7a3673c4">Version 4 of AIM makes use of <xref linkend="biblio_ISO_21090"/> data types. See <xref linkend="sect_A.8" xrefstyle="select: label quotedtitle"/>.</para>
    </section>
    <section label="A.6" status="2" xml:id="sect_A.6">
      <title>AIM v4 to DICOM TID 1500 Mapping</title>
      <para xml:id="para_7c62f9db-e8cd-43f6-8c1d-70d57d44feab">The transformation is described in one direction, by enumerating the structures in the target (DICOM SR TID 1500) and describing where in the source (AIM v4) the corresponding information may be obtained. The information is tabulated in a manner that can be implemented as an automated "pull" conversion of an AIM instance into a DICOM SR instance, such as might be described using a transformation language such as XSLT. The transformation is intended to be reversible, i.e., by inverting the target and the source, even though round-trip full fidelity will not be achieved in some cases. Gaps that exist in the information required in the target to create a compliant object, which need to be filled by information from an out of band source or generated <foreignphrase xml:lang="la">de novo</foreignphrase>, are highlighted. Information in the source that is not "pulled" into the DICOM encoding will be lost; these omissions are deemed to be harmless from the perspective of the
        relevant use cases.</para>
      <para xml:id="para_8e4ad639-1bb2-42da-9a1e-0c7279c84e71">The tabular representations make use of the following conventions in order to simplify the automatic extraction and use as a formal syntax to drive implementations:</para>
      <itemizedlist>
        <listitem>
            <para xml:id="para_7ca2f14a-0dc2-469f-a7b8-882e56c591fe">DICOM Attributes are represented by keywords defined in <olink targetdoc="PS3.6" targetptr="PS3.6" xrefstyle="select: labelnumber"/>, rather than specifying the data element group and element tags.</para>
        </listitem>
        <listitem>
          <para xml:id="para_10ceadd8-3655-4d73-a374-9f1a7d1d4d45">DICOM Attributes that are nested within sequences are shown as a path from the top level Data Set separated by a "&gt;" symbol.</para>
        </listitem>
        <listitem>
          <para xml:id="para_73edceaa-5f96-4897-ade2-0ab81d7d983c">AIM classes, attributes and associations are represented using their XPath representation as encoded in XML instances.</para>
        </listitem>
        <listitem>
          <para xml:id="para_95b59578-677d-4468-8c98-d4dde5b90a21">DICOM Code Sequence Attributes are mapped from AIM CD data type attributes without fully enumerating the corresponding subordinate DICOM Attributes and XML elements and attributes.</para>
        </listitem>
        <listitem>
          <para xml:id="para_615fab3c-652b-4a5e-9efc-8f0bfc42ec39">Other DICOM Sequences are listed, without a mapping on the same row, but with the following rows describing the individual DICOM Attributes nested within that Sequence.</para>
        </listitem>
        <listitem>
          <para xml:id="para_57e85bd7-aaf4-41ae-a917-177c6c0fe982">All source and target paths are fully qualified relative to the root of the instance in order to make the transformation reversible. I.e., it would be possible to describe some transformations using the descendant-or-self axis XSLT operator ("//") if the source were unambiguous but that would not specify the location reversibly as a target. Accordingly, some of the explicit paths are quite long.</para>
        </listitem>
        <listitem>
          <para xml:id="para_fefb0d39-5d2e-404c-872e-ac10ef5a0fc9">The requirement type for DICOM Attributes is as defined in PS3.3 for Attributes in Modules and Attribute Macros, except that if the containing Module in the IOD is not required, e.g., is U rather than M, then a mandatory (Type 1 or 2) Attribute in a user optional (type U) Module is shown as optional (Type 3).</para>
        </listitem>
        <listitem>
          <para xml:id="para_fbce7c7a-f4e5-4dc3-903a-e2911c9bbfa0">The data type and cardinality are specified for both the source and the target, to highlight potential mismatches that may occur during transformation. For nested elements and attributes, the multiplicity is expressed as the combination of the multiplicity along the entire path. For example, the aim:name attribute has a multiplicity of 1:1 in an aim:Person class but is associated with the aim:Image​AnnotationClass with a multiplicity of 1 -&gt; 0:1, so the multiplicity is indicated as 0:1, not 1:1, since that is the net effect. The DICOM multiplicity is either the VM for the data element or the number of Sequence Items if the data element is a Sequence.</para>
          <para xml:id="para_cd5451c6-6952-4a2f-82a0-b8bfa935157f">The XML representation of the AIM UML collapses some associations and classes such that they are encoded as a single element, and this is reflected in the mapping paths. For example, the aim:Person class has an aim:person association from the aim:Image​AnnotationCollection class, so the path to the aim:name attribute is expressed as /Image​AnnotationCollection/​person/​name, as it appears in the XML instance.</para>
        </listitem>
        <listitem>
          <para xml:id="para_96d51e74-f678-461c-b9f4-61455bedbe43">Data type transformations are assumed and are not described further unless there is a specific requirement. For example, conversion from DICOM DA, TM and DT VR Attributes to AIM TS Data Type values is implicit, including extraction/​population of the appropriate sub-fields (i.e., only the date portion of a TS is used when creating a DICOM DA value).</para>
        </listitem>
        <listitem>
          <para xml:id="para_4bc89462-03fa-48fb-85c9-64ccd98f5826">If value sets are defined for both AIM and DICOM, then value mappings are described. In some cases, explicit value sets are not defined. For example, aim:Person/​sex has no explicitly defined value set but maps to DICOM PatientSex, which does; so if AIM implementations use the DICOM values, and the values are copied, then the transformation without value mapping will be successful, but not otherwise.</para>
        </listitem>
        <listitem>
          <para xml:id="para_5a18e689-25ca-49b5-b916-cdfc968bef31">When a mapping is defined but no transformation source is available but a value is required, a Generated Value is indicated, which may be a fixed constant (e.g., a Modality value of "SR"), an indication that an empty (zero length) value or sequence is required, or an indication that a new value of the appropriate VR needs to be generated (e.g., a new UID for a UI VR, a new integer for an IS VR, etc.). The need to generate new values will not produce a deterministic result without a memory of previous conversions.</para>
        </listitem>
        <listitem>
          <para xml:id="para_10250e0a-3c93-46e1-b682-fec55177a61c">Optional content in the target that has no defined source is not described (e.g., DICOM SeriesDescription in the General Series Module has no correlate in AIM, though it would be useful to populate with a generated value).</para>
        </listitem>
        <listitem>
          <para xml:id="para_cddacd91-5281-487f-b859-5eb1784ddbb3">When the same DICOM Attribute is described in two different Modules, the more specialized (restrictive) use is described in the mapping tables. E.g., InstanceNumber is Type 3 in the SOP Common Module but Type 1 in the SR Document General Module, so it is only described in the latter.</para>
        </listitem>
        <listitem>
          <para xml:id="para_e9946041-baa2-4069-97dc-76559d574c93">Capitalization and punctuation of DICOM keywords and AIM class, attribute and association names are significant.</para>
        </listitem>
      </itemizedlist>
      <section label="A.6.1" status="3" xml:id="sect_A.6.1">
        <title>Mapping of Constrained AIM v4</title>
        <section label="A.6.1.1" status="4" xml:id="sect_A.6.1.1">
          <title>Header</title>
          <para xml:id="para_697269e9-1ffc-43ad-a70e-148f1fde2d4d">General Remarks on the mapping of DICOM header Module Attributes:</para>
          <para xml:id="para_57c4866b-9183-4b0a-9b79-7c5e412eee5a">
            <emphasis role="bold">SR Document General Module</emphasis>
          </para>
          <itemizedlist>
            <listitem>
              <para xml:id="para_9d7252f8-e199-4274-a1c9-58c9c989f493">Mapped AIM objects are considered "unverified", so there is no requirement to record the identity of the Verifying Observer; if the "recording" observer identity is required (aim:user class attributes) it may be mapped to Author Observer Sequence (0040,A078) in the SR Document General Module (and entries in Observation Context in the Content Tree are not needed).</para>
            </listitem>
            <listitem>
              <para xml:id="para_9802d433-5523-4dbb-bfd7-d7771d0f977b">Attributes of the Predecessor Documents Sequence (0040,A360) and Identical Documents Sequence (0040,A525) are not described in this transformation since they are relevant only in the context of a managed DICOM SR document environment and have no correlate in AIM.</para>
            </listitem>
            <listitem>
              <para xml:id="para_26ce0627-6e1a-4162-bf4c-f6fcec3db40a">Attributes of the Current Requested Procedure Evidence Sequence (0040,A375) and Pertinent Other Evidence Sequence (0040,A385) are described in the transformation and provide the information described in the Hierarchical SOP Instance Reference Macro used to match composite instance references with their Study and Series context; the AIM DicomImageReferenceEntity class performs a similar function.</para>
            </listitem>
            <listitem>
              <para xml:id="para_8290e899-3113-43e8-a5a9-47d938f34c51">Attributes of the Equivalent Document Sequence (0040,A090) are not described in the transformation since they are relevant only in the context of the original DICOM SR document.</para>
              <para xml:id="para_973530f0-c914-492d-9015-d5dd993fe8cd">The name space of the AIM elements is elided, and is implied to be "gme://caCORE.caCORE/4.4/edu.northwestern.radiology.AIM".</para>
            </listitem>
          </itemizedlist>
          <para xml:id="para_0e119d37-2b20-4ba9-bfd3-2d6d0d635b4c">
            <emphasis role="bold">SOP Common Module</emphasis>
          </para>
          <itemizedlist>
            <listitem>
              <para xml:id="para_e371df25-3bef-461d-87ed-33804a88497a">Timezone Offset From UTC (0008,0201) shall be considered for Attributes of the DICOM SR document that are based on the DA or TM data type (<olink targetdoc="PS3.5" targetptr="PS3.5" xrefstyle="select: labelnumber"/>). AIM date and time attributes may or may not contain explicit timezone information that may be extracted to populate Timezone Offset From UTC (0008,0201). </para>
            </listitem>
            <listitem>
              <para xml:id="para_eb6a2ebc-f688-4531-9566-b2ab4d7b3d77">The Specific Character Set (0008,0005) is required (Type 1C), if the Basic Graphic Set is expanded or replaced. This is the basis for mapping DICOM character sets to AIM XML Unicode (&lt;?xml version="1.0" encoding="UTF-8"?&gt;).</para>
              <note>
                <para xml:id="para_1e12d58f-1181-4035-b22d-b0b61bcc6322">Ambiguities exist for mapping individual characters to Unicode (e.g., for Japanese characters). Resolution of those issues is beyond the scope of this document. Please refer to <xref linkend="sect_A.8" xrefstyle="select: label"/> for further details on data types and character sets.</para>
              </note>
            </listitem>
          </itemizedlist>
          <section label="A.6.1.1.1" status="5" xml:id="sect_A.6.1.1.1">
            <title>Mapping of DICOM Patient Module</title>
          <table frame="box" label="A.6.1.1.1-1" rules="all" xml:id="table_A.6.1.1.1-1">
            <caption>Mapping of DICOM Patient Module</caption>
            <thead>
                <tr valign="top">
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_fa5cd929-555f-462e-81dc-1563f431397f">DICOM Attribute</para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_864f3635-6437-4313-9c34-f517a128a4ac">DICOM VR</para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_1b7bb000-6a0e-4a54-8ce0-a0963b579dc0">DICOM VM</para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_87e2c17f-8c39-4aef-8dc5-f73d5f6a4319">DICOM Usage Type</para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_e2f9b985-0782-40cb-b625-5ebd7300f7d6">Generated Value</para>
                  </th>
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_67bd8da0-4db7-4a64-bc3b-7dbac28bc7d4">AIM Element or Attribute</para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_ade73865-eeb4-4ae9-8c30-adaca094edcd">AIM Data Type</para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_fb380799-2ad1-4451-8b94-16e62e540910">AIM Multiplicity</para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_db34b024-3e98-4f98-9803-5355ab3d3f03">Comment</para>
                </th>
              </tr>
            </thead>
                            <tbody>
                                <tr valign="top">
                                    <td align="left" colspan="1" rowspan="1">
                                        <para xml:id="para_4478d8af-0dfd-471e-9289-76f6158fb159">PatientName</para>
                                    </td>
                                    <td align="center" colspan="1" rowspan="1">
                                        <para xml:id="para_0f9fae41-3724-4861-ba64-9c0029a503e1">PN</para>
                                    </td>
                                    <td align="center" colspan="1" rowspan="1">
                                        <para xml:id="para_2b36e1ea-5b35-416f-ae43-469033570f07">1</para>
                                    </td>
                                    <td align="center" colspan="1" rowspan="1">
                                        <para xml:id="para_483c7bc7-683f-45af-8a04-61e57ae717f8">2</para>
                                    </td>
                                    <td align="left" colspan="1" rowspan="1">
                                        <para xml:id="para_cd35203b-3b3a-43f1-92d0-89901a3c5f19"/>
                                    </td>
                                    <td align="left" colspan="1" rowspan="1">
                                        <para xml:id="para_a5036d1a-48d8-4d89-8fd0-e413ee4d8a4e">Image​Annotation​​Collection/​person/​name/​@value</para>
                                    </td>
                                    <td align="center" colspan="1" rowspan="1">
                                        <para xml:id="para_6412cc08-b598-4888-af8c-205f8456d63b">ST</para>
                                    </td>
                                    <td align="center" colspan="1" rowspan="1">
                                        <para xml:id="para_89ddbcc7-0c56-491e-8bf8-0a5743cdbd2a">0..1</para>
                                    </td>
                                    <td align="left" colspan="1" rowspan="1">
                                        <para xml:id="para_c017a59c-0701-4213-a4f4-7c7c7ce9a2db"/>
                                    </td>
                                </tr>
                                <tr valign="top">
                                    <td align="left" colspan="1" rowspan="1">
                                        <para xml:id="para_89687a94-a830-48f2-859d-6cd70ae9415d">PatientID</para>
                                    </td>
                                    <td align="center" colspan="1" rowspan="1">
                                        <para xml:id="para_0a57157c-02a4-4d4c-9979-73444afbb1a8">LO</para>
                                    </td>
                                    <td align="center" colspan="1" rowspan="1">
                                        <para xml:id="para_1b5ec162-0009-4d9e-9cfc-f40bd4b7bc4d">1</para>
                                    </td>
                                    <td align="center" colspan="1" rowspan="1">
                                        <para xml:id="para_f42dbac3-2439-4e96-9918-c654bb94435b">2</para>
                                    </td>
                                    <td align="left" colspan="1" rowspan="1">
                                        <para xml:id="para_730cb899-65f2-4e34-a8d4-070dbf0ffff8"/>
                                    </td>
                                    <td align="left" colspan="1" rowspan="1">
                                        <para xml:id="para_99703b30-7afa-4e93-831f-5f650d231624">Image​Annotation​​Collection/​person/​id/​@value</para>
                                    </td>
                                    <td align="center" colspan="1" rowspan="1">
                                        <para xml:id="para_9b4e0f68-5207-4635-9815-db0179190245">ST</para>
                                    </td>
                                    <td align="center" colspan="1" rowspan="1">
                                        <para xml:id="para_33c53618-9084-49e6-a6c5-4aa35983ebad">0..1</para>
                                    </td>
                                    <td align="left" colspan="1" rowspan="1">
                                        <para xml:id="para_758f9969-6b1b-4d0e-bb57-cee1c4d26899"/>
                                    </td>
                                </tr>
                                <tr valign="top">
                                    <td align="left" colspan="1" rowspan="1">
                                        <para xml:id="para_2c2a4696-1b4d-41b0-bb0c-4d480ca69ee9">PatientBirthDate</para>
                                    </td>
                                    <td align="center" colspan="1" rowspan="1">
                                        <para xml:id="para_7f224fbe-c0d5-48dc-b796-9aa5c40b7602">DA</para>
                                    </td>
                                    <td align="center" colspan="1" rowspan="1">
                                        <para xml:id="para_88d5b0ba-c98c-407e-9d85-81be390aaa46">1</para>
                                    </td>
                                    <td align="center" colspan="1" rowspan="1">
                                        <para xml:id="para_17c1dea1-819f-4341-a245-09ccbb755e20">2</para>
                                    </td>
                                    <td align="left" colspan="1" rowspan="1">
                                        <para xml:id="para_377dd417-3e12-4850-941f-5dc69d7c095f"/>
                                    </td>
                                    <td align="left" colspan="1" rowspan="1">
                                        <para xml:id="para_3ea735bf-62ca-43fb-b90a-0ad48782c3e3">Image​Annotation​​Collection/​person/​birthDate/​@value</para>
                                    </td>
                                    <td align="center" colspan="1" rowspan="1">
                                        <para xml:id="para_ec9e73fc-5b8b-47e2-95c9-2c3762400313">TS</para>
                                    </td>
                                    <td align="center" colspan="1" rowspan="1">
                                        <para xml:id="para_f1c13123-367f-41a4-a98e-50979b669852">0..1</para>
                                    </td>
                                    <td align="left" colspan="1" rowspan="1">
                                        <para xml:id="para_e12967a0-d3f2-4095-bbb8-e988a306ccce"/>
                                    </td>
                                </tr>
                                <tr valign="top">
                                    <td align="left" colspan="1" rowspan="1">
                                        <para xml:id="para_0536c2d0-112a-45b4-9fa0-f4a4c699ad4d">PatientSex</para>
                                    </td>
                                    <td align="center" colspan="1" rowspan="1">
                                        <para xml:id="para_f26aa15f-3018-41c9-a877-12578e1b2f23">CS</para>
                                    </td>
                                    <td align="center" colspan="1" rowspan="1">
                                        <para xml:id="para_9b87e902-9af6-4080-adac-47c8942f8ae3">1</para>
                                    </td>
                                    <td align="center" colspan="1" rowspan="1">
                                        <para xml:id="para_7674eff9-7c00-4b39-915e-f7ed584fd8bc">2</para>
                                    </td>
                                    <td align="left" colspan="1" rowspan="1">
                                        <para xml:id="para_c8cff641-3c49-4731-a053-ccfbea03c398"/>
                                    </td>
                                    <td align="left" colspan="1" rowspan="1">
                                        <para xml:id="para_96152cc9-8600-4af6-8925-9d1e8eb8670c">Image​Annotation​​Collection/​person/​sex/​@value</para>
                                    </td>
                                    <td align="center" colspan="1" rowspan="1">
                                        <para xml:id="para_f0f93dd8-ade3-4097-b971-56db5cf3442c">ST</para>
                                    </td>
                                    <td align="center" colspan="1" rowspan="1">
                                        <para xml:id="para_a5369f64-09d1-4ccf-999e-cdd3d238cacd">0..1</para>
                                    </td>
                                    <td align="left" colspan="1" rowspan="1">
                                        <para xml:id="para_7f7960c6-a0c2-4b99-9b37-4b2c27a0f4ab"/>
                                    </td>
                                </tr>
                                <tr valign="top">
                                    <td align="left" colspan="1" rowspan="1">
                                        <para xml:id="para_8a665d3f-0b20-4c6c-831f-6f845d66d0de">EthnicGroup</para>
                                    </td>
                                    <td align="center" colspan="1" rowspan="1">
                                        <para xml:id="para_d6af8431-f684-450e-b9be-7ccf669d3e4f">SH</para>
                                    </td>
                                    <td align="center" colspan="1" rowspan="1">
                                        <para xml:id="para_a91bd22d-6ed1-401d-8c69-b9f2b15d1d85">1</para>
                                    </td>
                                    <td align="center" colspan="1" rowspan="1">
                                        <para xml:id="para_739eb05c-bb89-4c84-8d55-678995c2db2b">3</para>
                                    </td>
                                    <td align="left" colspan="1" rowspan="1">
                                        <para xml:id="para_513c29ce-c7cd-42df-b58d-cf088dc8fc83"/>
                                    </td>
                                    <td align="left" colspan="1" rowspan="1">
                                        <para xml:id="para_7c2eb898-0d6c-4511-ab71-0c1029f8bf4f">Image​Annotation​​Collection/​person/​ethnicGroup/​@value</para>
                                    </td>
                                    <td align="center" colspan="1" rowspan="1">
                                        <para xml:id="para_24027ee5-d905-4368-9dfc-d86f60b234ec">ST</para>
                                    </td>
                                    <td align="center" colspan="1" rowspan="1">
                                        <para xml:id="para_d9eded27-47e4-41ef-b3f7-5b440568be85">0..1</para>
                                    </td>
                                    <td align="left" colspan="1" rowspan="1">
                                        <para xml:id="para_09755853-064e-4700-9fb0-91cb13f8701a"/>
                                    </td>
                                </tr>
                                <tr valign="top">
                                    <td align="left" colspan="1" rowspan="1">
                                        <para xml:id="para_931a9290-cf7f-45e6-b448-2ba7ec9fd7d9">Source​Patient​Group​Identification​Sequence&gt;Patient​ID</para>
                                    </td>
                                    <td align="center" colspan="1" rowspan="1">
                                        <para xml:id="para_4d41bc09-7faf-438d-8e24-e0ad04e5b320">SQ</para>
                                    </td>
                                    <td align="center" colspan="1" rowspan="1">
                                        <para xml:id="para_e2c9c59a-c211-49fc-8498-daa181436afc">1</para>
                                    </td>
                                    <td align="center" colspan="1" rowspan="1">
                                        <para xml:id="para_c310a0b5-116d-4561-8d41-10299ee5b6c6">3</para>
                                    </td>
                                    <td align="left" colspan="1" rowspan="1">
                                        <para xml:id="para_a81a3b7e-420f-43b0-baa9-092c0ca75a15"/>
                                    </td>
                                    <td align="left" colspan="1" rowspan="1">
                                        <para xml:id="para_fdd3e8b9-e8da-4744-bb19-57a5206b7779">Image​Annotation​​Collection/​person/​source​Patient​Group​Id</para>
                                    </td>
                                    <td align="center" colspan="1" rowspan="1">
                                        <para xml:id="para_dacfe284-68fe-4e97-bf99-cda885004d81">ST</para>
                                    </td>
                                    <td align="center" colspan="1" rowspan="1">
                                        <para xml:id="para_22e3b8e1-312b-4e6f-bd5c-366f15ea9728">0..1</para>
                                    </td>
                                    <td align="left" colspan="1" rowspan="1">
                                        <para xml:id="para_7802cca6-79e6-474f-b248-998e6cfbcb45"/>
                                    </td>
                                </tr>
                            </tbody>
          </table>
            <note>
              <para xml:id="para_1d9e0236-4c8c-498f-b1ee-34bdb06c597e">If the annotation concerns a small animal that has been imaged as part of a group of animals in the same image,
                  then the PatientID and other Attributes of this Module will be those of that individual animal, not the group.
                  The AIM 4.2 model can optionally identify the group of animals imaged at the same time that corresponds to the DICOM SourcePatientGroupIdentificationSequence. </para>
            </note>
          </section>
          <section label="A.6.1.1.2" status="5" xml:id="sect_A.6.1.1.2">
          <title>Mapping of DICOM Clinical Trial Subject Module</title>
            <para xml:id="para_d830fee6-29c2-4a54-9e69-d2e89cadb706">No mapping of the DICOM Clinical Trial Subject Module is described since no corresponding content is present in the AIM model.</para>
        </section>
          <section label="A.6.1.1.3" status="5" xml:id="sect_A.6.1.1.3">
            <title>Mapping of DICOM General Study Module</title>
              <para xml:id="para_c5497901-1c37-4527-812a-c943a86047c6">The AIM 4.2 model provides optional Study identification information.
                  If available, this information shall be used during transformation from AIM to SR, otherwise either a new Study may be generated,
                  or the SR instance derived from the AIM object could be placed in (one of) the Study(ies) referenced by the AIM instance, assuming there are any,
                  which produces a predictable transformation.</para>
            <para xml:id="para_c3c76295-3907-47ea-b205-9cf64ccf11df">If there is more than one DICOM Study referenced by the AIM object
                and explicit Study identification information (Image​Annotation​​Collection/​study​Instance​Uid) is absent, 
                duplicates of the converted AIM SR instance may be placed in each of the referenced studies (with different SOP Instance UIDs),
                in which case the IdenticalDocumentsSequence is required in the SR Document General Module;
                see <olink targetdoc="PS3.3" targetptr="sect_C.17.2.2" xrefstyle="select: docname label quotedtitle"/>.</para>
            <table frame="box" label="A.6.1.1.3-1" rules="all" xml:id="table_A.6.1.1.3-1">
            <caption>Mapping of DICOM General Study Module</caption>
            <thead>
                <tr valign="top">
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_b7c86ad3-4380-4a10-b020-2ec2af29efc4">DICOM Attribute</para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_a9eb9423-634e-4ac6-956b-5b93e6896113">DICOM VR</para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_ed35abe7-d2f6-442a-9ad5-190b9c74e440">DICOM VM</para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_2fb56b37-0be2-4379-b670-32d9ab1c5521">DICOM Usage Type</para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_82df6db6-109f-484a-a000-438285a89510">Generated Value</para>
                  </th>
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_1febe810-9fb2-4347-ab6d-dfda5f2018f3">AIM Element or Attribute</para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_e0912eb3-2bc3-4b94-9ec8-af337c8565e6">Data Type</para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_5e7bbf12-1e4e-4948-8300-d569b39ddbc3">Multiplicity</para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_1a15484d-5c5c-4afd-98a5-d566255a1da9">Comment</para>
                </th>
              </tr>
            </thead>
            <tbody>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_9a5e37d2-c486-4b18-9639-f16ec445cf60">StudyInstanceUID</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_6b953d00-17c5-46b1-816c-d8299f6b9783">UI</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_3a433e03-d95f-4079-b57e-6e8417c62cde">1</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_5dddc1ff-44f6-4bb8-ae15-b58a1d18f163">1</para>
                </td>
                  <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_4d7c063d-5d51-48a1-ba4f-c1b30f382695">New or derived from referenced image if not in source</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_f9009c42-e382-4ed7-ab21-6cbc27a26aea">Image​Annotation​​Collection/​study​Instance​Uid/​@root</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_7dd0153e-1a00-4f76-97aa-21c2a6564344">II</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_944413a2-724f-4bb9-9d3e-a2f1d5e2ee8f">0..1</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_9d40b899-b24d-4c6e-a28a-022dd48f2552"/>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_7e15f087-f14f-4cae-96dc-a41436882c14">StudyDate</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_9386d3ba-66b8-4f66-b57c-95a008c406f8">UI</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_f147c2c3-617a-4ba1-a823-f6d3145bfa4c">1</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_53d3761f-00fd-4bf0-b2d0-47e40d11bbf9">2</para>
                </td>
                  <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_fb82d481-6a1e-4060-8bad-a04a7d7be70e">Empty if not in source</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_a6e002fd-0f47-457d-82d7-247aab74fcd7">Image​Annotation​​Collection/​image​Annotations/​Image​Annotation[1]/​image​Reference​Entity​​Collection/​ImageReferenceEntity[1]/​imageStudy[1]/​startDate/​@value</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_02ea279c-caf5-42f9-b83d-4e53ecade0a7">TS</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_04ad0bf4-0522-437f-909b-97723e783832">0..1</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_5fad9e3e-fc11-4f82-a6f5-72aa736424c7"/>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_fca1f7b7-a29d-4d9a-b3e7-65febcba1cb1">StudyTime</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_f306a44d-4171-48a2-bdff-68caa5b6a7b7">UI</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_ccd5d570-719f-4541-8017-ff8f5fa031e1">1</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_3664d0fb-2d88-4f58-9353-36fa9a43f0b2">2</para>
                </td>
                  <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_9ac7f1de-72d8-4e34-9f3b-2038862ec385">Empty if not in source</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_fb172213-a7c0-4095-92d7-638a5b54aff1">Image​Annotation​​Collection/​image​Annotations/​Image​Annotation[1]/​image​Reference​Entity​​Collection/​ImageReferenceEntity[1]/​imageStudy[1]/​startTime/​@value</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_14ed6951-42bd-4b69-be6a-52aa7f3731a5">TS</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_e07a36bd-828b-42eb-b502-17f902bdeebd">0..1</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_f3fc5260-931f-436e-a781-737688253643"/>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_b346a0ee-cddc-490e-8897-2621dcc25a7b">Referring​Physician​Name</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_0056a83f-6dc9-4abe-8249-e4324aed6695">PN</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_28812b49-c59e-4cbb-894f-04e48fdc1f1a">1</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_ed6b51a8-ca57-40df-8166-cbf828894797">2</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_38756d78-c3d5-4f4f-96a4-b810e1734d7f">Empty</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_18acb746-4fc3-434c-bce5-da04cd77c696"/>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_7a81fc52-8708-4da6-b687-02799f9288ed"/>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_1eed7cf8-e902-4351-8e01-c3c433650681"/>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_7bc73696-2113-4722-a3d5-50f4fd474c8f">Not in AIM.</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_e6f674b4-bffe-482d-9cc8-f321202575f6">Study​ID</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_c8b572c2-371c-4cdc-ad0b-73b887383630">SH</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_eaa1f059-261c-400f-a771-4d1cec2628a5">1</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_447a1137-e26a-47d8-889a-6d73a5d79648">2</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_fabe799d-ac7e-4d1a-95e7-7996ead706f7">Empty</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_b1459e50-a67c-4559-b153-d329784cbfdd"/>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_672862c4-27e5-4765-a80c-c151f4b0e3ff"/>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_77c72f2b-6ac1-4ed0-990a-85f6233545d8"/>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_98ea5bb7-93e9-4ac1-b63f-703f916c77e9">Not in AIM.</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_344b090f-b705-474f-ac15-77c642eb154c">Accession​Number</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_774367c4-fc04-4810-85d1-b78b6b37d850">SH</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_055353e3-bdf1-4c77-9830-8e6751e384de">1</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_b40b799f-1fd5-446c-8bc8-b14fc9845548">2</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_43a6b7d5-13c9-426f-8130-a7d1217b0079">Empty if not in source</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_6b295222-67fb-4b2e-96db-d845fc08dd4e"/>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_89fcac0d-2b21-4b45-aefc-03baa04ddc08"/>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_cfbe35c1-2cf2-44e2-bcbe-d65f8f6be2da"/>
                </td>
                <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_53bcf210-7da0-49aa-bb6f-7d6995791519">Image​Annotation​​Collection/accession​Number/​@value.</para>
                </td>
              </tr>
            </tbody>
          </table>
          </section>
          <section label="A.6.1.1.4" status="5" xml:id="sect_A.6.1.1.4">
            <title>Mapping of DICOM Patient Study Module</title>
            <para xml:id="para_2b9c2e11-d9b8-4c55-a445-0ca2c0e022d0">No mapping of the DICOM Patient Study Module is described since no corresponding content is present in the AIM model.</para>
          </section>
          <section label="A.6.1.1.5" status="5" xml:id="sect_A.6.1.1.5">
            <title>Mapping of DICOM Clinical Trial Study Module</title>
            <para xml:id="para_353b69e4-5e1e-4262-967d-4bd66164f00e">No mapping of the DICOM Clinical Trial Study Module is described since no corresponding content is present in the AIM model.</para>
          </section>
          <section label="A.6.1.1.6" status="5" xml:id="sect_A.6.1.1.6">
            <title>Mapping of DICOM SR Document Series Module</title>
            <para xml:id="para_cb04f55f-69f7-4699-8389-1a9c0c258a98">The AIM 4.2 model optionally supports the concept that an annotation itself is part of a Series.
                If available, this information shall be used,
                otherwise a new Series shall be generated since a converted instance cannot be made part of a referenced image Series, if any, because of the rule that all instances of a Series are generated on the same Equipment.</para>
            <table frame="box" label="A.6.1.1.6-1" rules="all" xml:id="table_A.6.1.1.6-1">
              <caption>Mapping of DICOM SR Document Series Module</caption>
              <thead>
                <tr valign="top">
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_e9df23eb-59c9-4424-adac-29cddfb96465">DICOM Attribute</para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_5bb9d24a-b0ea-437e-a438-718587227229">DICOM VR</para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_9355961c-1438-4b5c-a67d-6bd27a1bc41f">DICOM VM</para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_7cdc5752-0fd6-4099-ad85-2558ddacb7bf">DICOM Usage Type</para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_f3d14283-a4a9-4a51-b253-76b1f68de25c">Generated Value</para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_5aaa1e13-0ed1-4d47-b78e-227cf0cfc175">AIM Element or Attribute</para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_fc1e5787-a382-4519-81da-425f8adf7f61">AIM Data Type</para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_fd1d130e-2381-413e-b6b4-363de6badba7">AIM Multiplicity</para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_0843d3d5-4e28-4195-bc19-5eff06aff29d">Comment</para>
                  </th>
                </tr>
              </thead>
              <tbody>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_2c15ee6d-0491-4c3b-9389-5598b71048d7">Modality</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_992b11df-52bb-47a4-aa22-b218f68a4c14">CS</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_4932d5d3-129a-4326-b292-8abed98f0b9b">1</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_f200dfec-7890-4e2e-8f52-8437a8474dd4">1</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_59a7cc82-185d-45ce-91e2-aa1d2b754edf">"SR"</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_679cc6c7-03c8-428b-b5d4-4b0739908f4a"/>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_e29dc6b4-1515-4abe-b1bd-d0db11cf37e7"/>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_eda9a082-91a0-4caf-81fc-f0411c138f8f"/>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_b4c24303-1ebc-46f3-ba66-90a3c09c4835"/>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_1b94c967-adc2-4b13-b26b-1ff71d60eced">Series​Instance​UID</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_b84d79d5-aa5d-4c91-83d0-8b20ac41f249">UI</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_c5a2cbbb-9d66-4f4d-a68c-829a3c1d0806">1</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_a93a29f8-9bf0-4312-9e0b-6111d967da8a">1</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_392a4439-6b7b-4e50-bff6-d22265388bac">New if not in source.</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_afd9ed39-a965-4daa-af3f-1e2ac460fb48">Image​Annotation​​Collection/​series​Instance​Uid/​@root</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_146d6b98-b9ed-4ac6-8df1-f2401139108e">II</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_7a9e594a-08bb-494b-b0c6-5bbaf4723aa6">0..1</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_4f8a484a-a17b-4062-84f6-1dada540d17b">Using a generated value means that multiple round-trips may produce different values.</para>
                    <para xml:id="para_458cdd0d-b829-4efe-910f-817a40c3e3f3">When mapping multiple AIM instances related to the same study, if the same SeriesInstanceUID is used they will appear in the same Series.</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_c0233716-1e22-4c62-94c0-f60a49d0f180">Series​Number</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_6c891780-0b2b-47fe-b0c3-d120578ed63b">IS</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_a43f6236-6cba-4c50-b107-b0c114e31f80">1</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_e58800b3-eb43-49cc-837c-dedd699a8926">1</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_6e64b541-be92-4bea-8b6b-26cc97d83183">7291</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_8bb0271c-4633-4b9e-b132-7ef7e67fe511"/>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_323a01a0-03ff-4945-bd41-75b4e1a4c496"/>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_5571317d-99a1-483d-a9e0-6c90635abe5e"/>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_26d0b564-58f0-4b35-9fa6-dbad58be60ad">A well-known value means that multiple round-trips will use the same value.</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_c9b4d989-b10d-4ff4-893c-6cb4c487a567">Referenced​Performed​Procedure​Step​Sequence</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_0739f350-ee8c-4744-be89-09d9265799b5">SQ</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_f22b7840-4dba-470b-9ce5-c3891223fddd">1</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_2591f08c-777d-45dd-955f-b87c411e4862">2</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_e2d6d7be-2669-4795-a930-b43d75fed4d8">Empty</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_7ff84c2d-4e61-4c88-9ea8-3ade58f42c4b"/>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_1a0b7f18-3257-4138-af50-7571fced7153"/>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_f12df359-da82-463e-8bd6-3efcc2cf554e"/>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_28ceb761-37a2-4d9f-bbfa-4ccbb2dc11be"/>
                  </td>
                </tr>
                
              </tbody>
            </table>
          </section>
          <section label="A.6.1.1.7" status="5" xml:id="sect_A.6.1.1.7">
            <title>Mapping of DICOM Clinical Trial Series Module</title>
            <para xml:id="para_0eeb3824-341f-4336-8d0e-431a40d06389">No mapping of the DICOM Clinical Trial Series Module is described since no corresponding content is present in the AIM model.</para>
          </section>
          <section label="A.6.1.1.8" status="5" xml:id="sect_A.6.1.1.8">
            <title>Mapping of DICOM General Equipment Module</title>
            <table frame="box" label="A.6.1.1.8-1" rules="all" xml:id="table_A.6.1.1.8-1">
              <caption>Mapping of DICOM General Equipment Module</caption>
              <thead>
                <tr valign="top">
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_0f5bd900-ab28-49c1-b2d1-60492dfca9e7">DICOM Attribute</para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_cadf1626-4be6-4c87-8b68-c67acc952cda">DICOM VR</para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_d3608208-07f9-4f39-8fa6-e22a2a5c6b0d">DICOM VM</para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_b0aedb89-974d-430f-99b7-345203db58f1">DICOM Usage Type</para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_6bb34685-cff4-466b-af85-49d8b40550f1">Generated Value</para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_d0509e45-f804-454f-b1fe-c2f8cbc4e529">AIM Element or Attribute</para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_794b0741-9a66-4442-83c4-e7820a6390a0">AIM Data Type</para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_65681b31-d83d-4d15-9632-f8f027766595">AIM Multiplicity</para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_7ea924dd-19b8-47d7-820e-3f433e67b2be">Comment</para>
                  </th>
                </tr>
              </thead>
              <tbody>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_7c4a3be9-08d6-48d5-9851-b8641fb880b4">Manufacturer</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_b37c7a1d-112f-4be5-90ba-eee31f007e55">LO</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_29afb152-01c1-4e28-9a91-65cb2e8a6d30">1</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_5e80eec4-1d5f-495e-b9e3-5fa9ca6dabb5">2</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_f1110c7c-338f-49df-b4fe-76d392928aed"/>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_e9afca03-b3b0-48e2-ba68-fc51e87537c1">Image​Annotation​​Collection/​equipment/​manufacturer/​@value</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_4e4e09a8-908f-4350-b9c4-dac522637344">ST</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_015a6f28-159f-48f7-a169-9212ff41c2b9">0..1</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_e3366965-31d8-4900-a673-8d16dd7f4519"/>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_390143ad-bfb7-4bc3-a4d6-6b867e242c86">ManufacturerModelName</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_7cfcdd46-9b2e-4687-bb7a-0fe7086ebbd9">LO</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_bacd7d1d-dd36-40d2-b2ec-60e756b11d9f">1</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_47e7c7ea-782b-40a8-877f-aff7227fe13c">3</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_091cb281-3921-4e3e-b39c-75448d6a29be"/>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_1e16aaf7-e8cc-4365-9ee3-a33605493a73">Image​Annotation​​Collection/​equipment/​manufacturerModelName/​@value</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_bb152305-6ada-4276-a0b0-abc070cd4d55">ST</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_cc9711e5-456a-451b-8d86-5ac1644311ac">0..1</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_556bcb5a-5c00-4503-b1c9-8f2b345f0875"/>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_3400688f-4cdc-41b1-af1d-da3e7382e11e">SoftwareVersions</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_9a02c3c2-1d06-415d-af5d-bf137df21d49">LO</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_a863c5d3-bfc8-4bd7-a96e-83514a89f965">1-n</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_aea316f1-94be-4ec5-a821-bbfc7a9ded6d">3</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_9a3981cd-e19d-414a-85b3-299ffec77b76"/>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_4d18b910-ab7e-4a5c-8dec-49c7b34e27ba">Image​Annotation​​Collection/​equipment/​softwareVersion/​@value</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_b54370af-b05b-42a9-bd1b-b973be9090c6">ST</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_d71f42b5-dbdf-46ed-8369-4f6dcbc7373b">0..1</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_aa0728a8-696d-493f-808b-16a88876b950"/>
                  </td>
                </tr>
              </tbody>
            </table>
          </section>
          <section label="A.6.1.1.9" status="5" xml:id="sect_A.6.1.1.9">
            <title>Mapping of DICOM SR Document General Module</title>
            <table frame="box" label="A.6.1.1.9-1" rules="all" xml:id="table_A.6.1.1.9-1">
              <caption>Mapping of DICOM SR Document General Module</caption>
              <thead>
                <tr valign="top">
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_e1a5f6c8-3c87-4467-8d4a-06c14524c6f8">DICOM Attribute</para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_0a64ec92-9883-4b55-a963-8add5a87e9e7">DICOM VR</para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_8a4fe74d-4870-4133-bcd1-e526db89ef33">DICOM VM</para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_0b0b769d-f8c4-439b-8562-dc3da77cb2ff">DICOM Usage Type</para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_390e2816-5ad2-4011-bf69-8a78b134aa4b">Generated Value</para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_1bb01481-e88a-4fc8-a3a4-22cec222f8de">AIM Element or Attribute</para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_e2f917e4-da01-42cd-a162-10447c5a7502">Data Type</para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_859abab5-12c5-4ff5-a242-6427edcc5f0d">Multiplicity</para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_5db7f42e-71d7-4484-87cd-fe48032300c1">Comment</para>
                  </th>
                </tr>
              </thead>
              <tbody>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_292faf4f-47fb-4a92-a3c9-eb4db7fb928a">InstanceNumber</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_cc14fc90-fd8c-490d-9657-df89642427ff">IS</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_7b4d45b3-fd83-4c97-8f6a-b6b0a903040a">1</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_315dd634-bf82-44b6-bfcf-54871d2cf637">1</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_46ccc1a8-643d-4e96-b7de-b33642f21b18">New</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_b46542cc-4cf4-46a8-b798-68fa73190203"/>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_f4a1fc8f-3959-4b11-a7c3-4bdedd0da7b2"/>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_7071a70a-7954-4a36-a789-9a62c5ddb21f"/>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_d293d5bc-f5c8-44ff-8b52-8e63be6bdc1e"/>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_3af54432-db2a-4856-b5c8-48744f6df750">CompletionFlag</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_386355fd-4221-4ed1-b913-3b655ac3f0d1">CS</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_bad5b957-b025-4971-bc2c-0a664bffd1aa">1</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_0f359f64-18dd-424c-b8bb-197e3b753371">1</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_693dc614-f1ce-4165-8fca-69ae68ea8868">"COMPLETE"</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_ccd0d7ab-3db3-4781-b9db-3ebfed953db3"/>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_e417a373-a4e9-4884-9ba6-5c1a1105c29c"/>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_3ba731e3-64d2-41ee-9886-667c776dbc4f"/>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_e89da920-d724-4399-ba5e-02f5a7626d68"/>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_4123a11a-8aa1-403e-972a-6cd67f285bf2">VerificationFlag</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_5f2c090d-efb1-486d-b385-1c4d5eab6d41">CS</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_fd96aa9f-1649-45f8-b126-685576641301">1</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_ac6562cd-34f1-4d0c-9dfc-b356520a3ade">1</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_db2a5f41-90ed-43de-b94e-950064047ed9">"UNVERIFIED"</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_e8bda4a7-8bbb-4f58-876b-6e2527ecbeb2"/>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_3bace083-0f55-4646-a7e1-c4bc8043b4ab"/>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_54ee0a14-bd47-4336-98b9-aa2e34db42dc"/>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_8455a6f9-7d8e-4df9-aa83-cefd7a34af93">Sending a value of "VERIFIED" would trigger the need to send the Verifying Observer Sequence, which is not described in this mapping.</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_68a4811d-ddb2-4606-b7b7-33d9f4c393de">ContentDate</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_cae6e42d-9e9b-484e-bbfd-1e5cc81b891a">DA</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_3b8f1a6b-ccbe-4300-9c3a-76c2cf679a46">1</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_5c789528-7a4c-454b-bd90-c61804731920">1</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_cba4b9e5-98bd-43b7-a990-41c73a7d9d88"/>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_0edeae55-637c-4193-80e8-2c389fa35564">Image​Annotation​​Collection/​dateTime/​@value</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_64dbd195-4d91-476e-b186-ac00b455cb2a">TS</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_782e5114-ab0f-42d3-b265-1e1628f1b716">1..1</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_deb966b3-1046-47b3-abf6-d29fffa762d5"/>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_fd35a35d-454b-4593-9e32-1a6e9ca6b80e">ContentTime</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_8b250cb4-a514-46ce-9f72-0627f861bf9c">TM</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_9bce0483-7548-4e0b-898e-e114d5125a1c">1</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_e9d2af58-45de-4bab-b080-241151dd0847">1</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_2a044c66-730c-4928-891b-e49a4dd9baf7"/>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_734d213a-ebcb-497a-b1ec-bdb13bbd0de0">Image​Annotation​​Collection/​dateTime/​@value</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_0c4d872f-9010-403c-a865-af40714ccd1d">TS</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_9cf78668-53f9-4167-8fbe-7db2979e34ef">1..1</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_02ca2fce-9a00-4123-a43d-65499c756b75"/>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_cdc77c74-b07c-420c-8819-55f88b4bc392">Author​Observer​Sequence</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_f91ced57-138e-4358-890b-c87c47860b2f">SQ</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_ab64d045-0c67-4337-9c34-03865fd42d87">1-n</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_ddb05d2d-c50b-484d-b2a6-fd9848a1e992">3</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_c456d324-73e6-4218-8cf3-b13390f46aae"/>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_214b7183-c649-42ce-89e5-45e50b6ed3aa"/>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_82a5d447-e30f-4682-9ffc-ff2dea489a9c"/>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_e2e6066c-72b0-4b0e-9886-6daacfcf4c8b"/>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_138d7524-34de-4860-9584-16d28fd06f78">Don't send sequence at all if observer is a device rather than a human since the device is described in the General Equipment Module.</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_618c9139-b283-488b-b215-9c4b8bd6c72e">Author​Observer​Sequence &gt; Observer​Type</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_ca370eca-0077-4ff9-b09b-8f5485dec8a2">CS</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_1f7e9623-3e91-4266-8938-24f1c35e43f5">1</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_f559795c-11a5-4bfa-b021-4d472652a184">1</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_bd9efcbe-422a-47a6-8eb1-8ec3b9369b51">PSN</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_1676bf38-36ba-4ad9-93c9-06f77bd40168"/>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_5dfed721-c49c-400c-a8a1-584bcd97e33f"/>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_0222db0c-5757-4483-a1a6-3a3bd405e95f"/>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_435ca69f-d264-4cda-804e-1c29e69ce4b0">DICOM allows PSN or DEV.</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_ac10ed4e-386c-4b0f-ae2c-8d9b1f1251f8">Author​Observer​Sequence &gt; Person​Name</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_f7408f5d-2593-4b8a-a49c-4e297050b9fb">PN</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_c6e9a07c-6b25-4cbc-a0da-fb82738bba36">1</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_c47c94df-0380-4dd1-9fbf-3dc486c23c38">1C</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_d4e3d6fa-e641-4d88-a003-f9c64dbc0cd0"/>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_e1e676fb-224c-499d-a1ce-040dd72d8613">Image​Annotation​​Collection/​user/​​name/​@value</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_d157fb0f-900c-43d4-9d52-221f7e4cfc4f">ST</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_aeb5a1a1-25bb-46d8-ba46-d176121b37eb">1..1</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_26b73c61-af64-41be-9820-4028b5e91cc6">If PSN.</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_6015c471-3cc6-48fe-b0d6-803abb478d19">Author​Observer​Sequence &gt; Person​Identification​Code​Sequence</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_c67bdccf-5170-4975-8864-afc4633f9aa5">SQ</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_c910ef5c-9750-4e04-84d5-8fe0c5485df8">1</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_f4d1ebb3-2045-4961-b62a-9f748d58b464">2C</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_33149043-b45a-4156-ab51-00fdbd19704a">Empty</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_a498c609-b248-4ce8-b08d-501d99776177"/>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_9b5bb4fa-7f7b-4a67-8435-c09c42efdbf7"/>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_99aa8e83-c6ae-4657-b4d1-c557cf23b1d9"/>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_ce795b37-57e8-41fb-99fa-969dff9d4d3a">If PSN.</para>
                    <para xml:id="para_295e1db2-c086-4286-9e49-f078e9feb3d7">Not in AIM.</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_89b597c2-60f0-48ee-adbd-d02b5b510196">Author​Observer​Sequence &gt; Institution​Name</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_98c98043-5852-414d-9068-84a4557d8a26">SQ</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_1412f641-7df6-49b5-b4ee-7257594c9a96">1</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_d5e32f02-108e-4ec2-a5ea-cf3197b8f8ef">2</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_5dd3db77-a7b3-4959-9a01-835d3f09a3a9">Empty</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_2cb23c3f-0db0-47f4-b6e5-c5676ac8be1d"/>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_3cacc34d-7c32-47c2-bed9-d201e5a17930"/>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_b0f76a80-56a8-4b25-babd-f9337397aa5a"/>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_1326bf43-ac8d-4145-91fb-5a02d85defe0">Not in AIM.</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_89158e25-fe07-4e82-bf4b-153498ed5188">Author​Observer​Sequence &gt; Institution​Code​Sequence</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_d3197c08-1083-4d5f-b5b8-ae68ebf08917">SQ</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_a4c4dd47-a86b-45af-8a1c-6f6cab522e3b">1</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_cfd4a651-289e-43d8-823f-52b450842db6">2</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_35672355-ffed-408e-85a8-baba6381cedd">Empty</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_debe1ef2-4fa1-4e1b-8953-61adf4a8d66e"/>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_2bfdfccf-3897-485e-b7d2-c6c9309515dd"/>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_02ced999-8e21-4a0a-aff0-a1f988e81b28"/>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_dd682388-606e-43fe-a289-ecc99d7b8e13">Not in AIM.</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_0a33adc9-4596-43d9-964a-86e2913c95e0">Identical​Documents​​Sequence</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_bbe92366-8217-41ad-9861-0b7433f53956">SQ</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_12d82c9f-fc54-4e14-87d8-aa149d3abb79">1</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_592e2d03-bb81-4b1a-89bd-2d856f174ee2">1C</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_008f79bc-9af2-4f28-a02f-b77756c0527b">UIDs of other identical converted instances.</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_1c2d73fb-928d-4b00-80fe-9b70663f6ca1"/>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_c0b68101-d991-40a4-a400-d6cce8561e56"/>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_833a50cb-7ed3-42c3-a93e-e5bf946e4605"/>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_30d29c03-a4ac-4c34-9c41-c3018e2d0a35">Required if this document is stored with different SOP Instance UIDs in one or more other Studies</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_35078849-68af-4950-9d7c-2df1f2bb73ed">Current​Requested​Procedure​Evidence​Sequence</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_e529a497-acdd-42f3-8ac5-a5f498e8cdea">SQ</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_208eb6dc-9960-4e2f-9ac2-bc07d57a7ecf">1</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_0136430c-0c27-4c25-a45d-aa44f160dca1">1</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_bcaa137f-e53a-4fb6-a93a-0cbb9916c99e"/>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_08362c79-ee34-4501-b38a-0c1527b31eef"/>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_e81c8ac7-ac5f-4e1d-bd4c-d073faaae19c"/>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_e9985b93-3833-47c5-a734-45c50613e10d"/>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_f7b09a90-f5ca-457d-83e8-89e080842c94"/>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_8ef26da6-634e-49e0-8b0e-faaa5627ba56">Current​Requested​Procedure​Evidence​Sequence&gt;​StudyInstanceUID</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_7985df9f-32e2-49c4-92c5-66c91ce3efbc">UI</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_ca4cd9de-26a1-4d17-a29e-094ea77bebde">1</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_52c97b10-4597-4978-95ce-63c8eb5865bf">1</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_f8d572b6-682e-4133-9312-3c9fa7981700"/>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_adaeaa01-9903-4a26-8af9-c781871d4cd6">Image​Annotation​​Collection/​image​Annotations/​image​Reference​Entity​​Collection/​ImageReferenceEntity/​imageStudy/​instanceUid/​@root</para>
                      <para xml:id="para_d86bc5a0-b7f3-452c-9af3-899d0383b538">Image​Annotation​​Collection/​image​Annotations/​Image​Annotation/​​segmentation​Entity​Collection/​SegmentationEntity/​studyInstanceUid/​@root</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_4824b215-9e55-4ec1-9370-c7e58429a10d">II, II</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_11e0ba39-3396-4e34-bd5e-14556cb417ef">1..1, 0..1</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_81c8b90b-e751-4a1e-83fd-5f98db97f2a3"/>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_fbd75b73-72bc-4c36-a148-9817db807e03">Current​Requested​Procedure​Evidence​Sequence&gt;​ReferencedSeriesSequence</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_accce09d-37c0-4bf0-8995-7fb4d5c2dabc">SQ</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_4fbbddb5-88ad-4635-a3ff-11f1ab9469f0">1</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_58f8bd56-21e7-4ab9-9ec5-5f923d5c8da5">1</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_44f70c84-3249-49d0-a534-766f8559f542"/>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_59045b91-c2b7-48b6-84ab-f5ecb9fa8385"/>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_4a0f8f35-ed13-457a-864d-d5fb7a234df1"/>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_16b56ac3-4feb-4f9e-a988-9249b74826b7"/>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_c4a85d5a-c289-44f5-958a-e388315e25f6"/>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_28b7b7c3-d9d5-4492-936e-40f6accab5c5">Current​Requested​Procedure​Evidence​Sequence&gt;​ReferencedSeriesSequence&gt;​SeriesInstanceUID</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_579544b0-73ab-4751-9593-6ca19d5a02a5">UI</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_a677675b-9084-420a-9041-4ecee67b7968">1</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_dcc5a340-2b5b-4f7b-b419-261744d57a28">1</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_d0cd35bf-fd6f-49e5-a0f8-15deb397ec82"/>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_cb8488af-60f4-4bce-84f1-fe0e93d2b828">Image​Annotation​​Collection/​image​Annotations/​image​Reference​Entity​​Collection/​ImageReferenceEntity/​imageStudy/​imageSeries/​instanceUid/​@root</para>
                      <para xml:id="para_f2c748c8-27e6-4b5f-a27f-3b130ab5dbee">Image​Annotation​​Collection/​image​Annotations/​Image​Annotation/​​segmentation​Entity​Collection/​SegmentationEntity/​seriesInstanceUid/​@root</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_495ad730-a9f6-4082-9e5e-5a8a5a232413">II, II</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_b74d25ff-3373-4da3-92aa-f3812e6e04ca">1..1, 0..1</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_f9a258da-359c-4230-9a43-b06744d79b89"/>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_e07d4b58-62ce-4afb-a9de-421b953de66f">Current​Requested​Procedure​Evidence​Sequence&gt;​ReferencedSeriesSequence&gt;​ReferencedSOPSequence</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_3266e19e-8f47-4763-8b98-21b17bc37181">SQ</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_cec22da8-4956-45e2-bd3f-7a7952c98a00">1</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_6e5bab24-47d6-4c44-a44f-ed7be5cb488c">1</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_9ab2639f-99f2-4704-8187-78b105be2074"/>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_35ee5a01-1205-408c-9b02-f90278d130b7"/>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_dcadedd5-a7a7-41f4-9b9a-04be638fa443"/>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_19f0bfe5-c37a-42fb-9378-dcfb7689f780"/>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_e83f6a2f-6b44-4387-a34c-06f5ef198605"/>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_ceb8d0be-1156-417b-baa1-0e95996efff3">Current​Requested​Procedure​Evidence​Sequence&gt;​ReferencedSeriesSequence&gt;​ReferencedSOPSequence&gt;​ReferencedSOPClassUID</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_36bb910b-3b86-4b9d-a092-7b9047a4a5d8">UI</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_b603abf9-0c48-486f-ad56-48f883fac312">1</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_2c91c68a-ca86-4244-b9eb-892dd7092a30">1</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_3905bd9f-1ab1-4987-8988-4c7b1772f56b"/>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_710d43a1-96e3-403f-afb9-23845275e721">Image​Annotation​​Collection/​image​Annotations/​image​Reference​Entity​​Collection/​ImageReferenceEntity/​imageStudy/​imageSeries/​image​Collection/​Image/​sopClassUid/​@root</para>
                      <para xml:id="para_92ad1d2e-2395-46c7-9a11-1a9bf1ada0c9">Image​Annotation​​Collection/​image​Annotations/​Image​Annotation/​​segmentation​Entity​Collection/​SegmentationEntity/​sopClassUid/​@root</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_cfdcbc64-fb14-4699-8ec0-1af84eb06809">II, II</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_b83a3d4c-8c5f-431c-bec1-83facc6da531">1..1, 0..1</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_63c6a45f-7b06-4202-a170-563869cd70ae">If the studyInstanceUid or seriesInstanceUid of a SegmentationEntity are absent, this reference cannot be included for that instance.</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_b138911f-26b4-4dfc-8a55-e590c0a06ede">Current​Requested​Procedure​Evidence​Sequence&gt;​ReferencedSeriesSequence&gt;​ReferencedSOPSequence&gt;​ReferencedSOPInstanceUID</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_29ff95f3-6b07-4359-8405-b7f00fc6b55b">UI</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_abcf74e5-f8e3-4bde-85d3-18fe6f4bf384">1</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_75559f86-1a0a-4371-898b-084804415387">1</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_ab3259fb-5bcc-47ca-938d-82000657b6cf"/>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_52b4900d-1ec3-4f00-8096-8914329a755f">Image​Annotation​​Collection/​image​Annotations/​image​Reference​Entity​​Collection/​ImageReferenceEntity/​imageStudy/​imageSeries/​image​Collection/​Image/​sopInstanceUid/​@root</para>
                      <para xml:id="para_b365a108-562d-4148-a8ab-4f68c1479c01">Image​Annotation​​Collection/​image​Annotations/​Image​Annotation/​​segmentation​Entity​Collection/​SegmentationEntity/​sopInstanceUid/​@root</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_9cbe3197-829d-4430-b30c-6c61f8c9b4bd">II, II</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_b0da5b3e-eae0-41b7-87dd-73abfb6e46d3">1..1, 0..1</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_d9d8aaa9-08bf-45f1-852b-f1389fe4135c">If the studyInstanceUid or seriesInstanceUid of a SegmentationEntity are absent, this reference cannot be included for that instance.</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_a03bd971-89eb-4166-a52f-cb269ea138be">Performed​Procedure​Code​Sequence</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_aafcbe55-7d75-4590-a5b5-7a3c0c4cd667">SQ</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_e4067a51-7c43-469f-9af3-ea0e92db0abb">1</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_aefd8831-009c-465d-810e-a828bfbb3d80">2</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_17df769d-35d1-4a8a-b8af-1a3b70f5df5d">Empty</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_dd62376f-42a6-4371-b144-1d09c72c655b"/>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_792a55e0-376c-4eee-a433-da6fa2b8f916"/>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_2cdab1af-103a-42e2-a050-81ca2833c9e7"/>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_0382e5e2-462f-4037-bdcc-49971285bf2a">Not in AIM.</para>
                  </td>
                </tr>
              </tbody>
            </table>
          </section>
          <section label="A.6.1.1.10" status="5" xml:id="sect_A.6.1.1.10">
            <title>Mapping of DICOM SR Document Content Module</title>
            <para xml:id="para_f61a7562-35e3-4abd-ad38-6ed04a6361f2">All the Attributes in the SR Document Content Module are transformed by processing the DICOM SR Content Tree, and accordingly are not described in the same tabular manner as the "header" Attributes, since the mapping depends on the DICOM SR template structure. The Attributes common to each Content Item of the Content Tree are:</para>
            <itemizedlist>
              <listitem>
                <para xml:id="para_c24c0c3f-a412-4150-9b9c-63235ae2d999">ValueType</para>
              </listitem>
              <listitem>
                <para xml:id="para_84e2040c-cebc-4b0f-a146-9202a08fda83">ConceptNameCodeSequence</para>
              </listitem>
              <listitem>
                <para xml:id="para_cfe7846c-efff-4ec8-984c-a9480fcf7793">ObservationUID</para>
              </listitem>
              <listitem>
                <para xml:id="para_eeac07f9-9f76-48b0-8cae-d710e49f569b">ContentSequence</para>
              </listitem>
            </itemizedlist>
            <para xml:id="para_126250ee-d7d4-435a-ba8e-8b7caad9a691">Each child Content Item with a "by-value" relationship with its parent also contains:</para>
            <itemizedlist>
              <listitem>
                <para xml:id="para_1724111e-5fc1-449f-af6e-580e303ff3cc">RelationshipType</para>
              </listitem>
            </itemizedlist>
            <para xml:id="para_7b41e53e-ce49-458c-90f6-762d951f2020">The additional required Attributes in each Content Item depend on the ValueType:</para>
            <itemizedlist>
              <listitem>
                  <para xml:id="para_6fdf624f-9742-49eb-916d-8f8d118ba58c">TEXT - Text​Value</para>
              </listitem>
              <listitem>
                  <para xml:id="para_115c716e-42bf-4bfb-a95a-721d4dd10a81">DATETIME - Date​Time</para>
              </listitem>
              <listitem>
                <para xml:id="para_1075a781-8edf-4f74-b6f3-5013127a5b0b">DATE - Date</para>
              </listitem>
              <listitem>
                <para xml:id="para_abc7f433-c3fe-4424-8de7-cf1c817f7a4f">TIME - Time</para>
              </listitem>
              <listitem>
                  <para xml:id="para_8490272f-a22c-44e2-afd5-864a58239b56">PNAME - Person​Name</para>
              </listitem>
              <listitem>
                <para xml:id="para_43406bcc-1de1-4113-b2ec-55f26d333993">UIDREF - UID</para>
              </listitem>
              <listitem>
                  <para xml:id="para_42c4efc3-4b19-4738-a9f1-5e199879b0cc">NUM - Measured​Value​Sequence, Measured​Value​Sequence&gt;Numeric​Value, Measured​Value​Sequence&gt;Measurement​Units​Code​Sequence</para>
              </listitem>
              <listitem>
                <para xml:id="para_5f2e20d3-d0f3-40b7-8f24-1ee5bbefcb82">CODE - ConceptCodeSequence</para>
              </listitem>
              <listitem>
                  <para xml:id="para_0018e0ce-f97c-4a54-a495-dfe60a2957e2">COMPOSITE - Referenced​SOP​Sequence, Referenced​SOP​Sequence&gt;Referenced​SOP​Class​UID, Referenced​SOP​Sequence&gt;Referenced​SOP​Instance​UID</para>
              </listitem>
              <listitem>
                  <para xml:id="para_6ea4adf3-a6db-495e-b274-48b93dcf91e2">IMAGE - Referenced​SOP​Sequence, Referenced​SOP​Sequence&gt;Referenced​SOP​Class​UID, Referenced​SOP​Sequence&gt;Referenced​SO​PInstance​UID, Referenced​SOP​Sequence&gt;Referenced​Frame​Number, Referenced​SOP​Sequence&gt;Referenced​Segment​Number</para>
              </listitem>
              <listitem>
                  <para xml:id="para_3e5f89cd-bac6-4748-a10a-d26932eeee8e">SCOORD - Graphic​Data, Graphic​Type</para>
              </listitem>
              <listitem>
                  <para xml:id="para_506a05c2-f0a0-4a59-b538-3cee1729dd8d">SCOORD3D - Referenced​Frame​Of​Reference​UID, Graphic​Data, Graphic​Type</para>
              </listitem>
              <listitem>
                  <para xml:id="para_3733e495-f682-40db-84c9-1cd9f90a2562">CONTAINER - Continuity​Of​Content, Content​Template​Sequence, Content​Template​Sequence&gt;Mapping​Resource, Content​Template​Sequence&gt;Template​Identifier</para>
              </listitem>
            </itemizedlist>
              <para xml:id="para_428647da-fd25-4a55-8c3e-9c53c4cb3c9d">Observation​UID is required for the following Content Items in order to propagate the aim:unique​Identifer information:</para>
              <itemizedlist>
                  <listitem>
                      <para xml:id="para_4469a725-969c-44bb-833d-4585393d0f02">IMAGE - for (121191, DCM, "Referenced Segment") Content Item corresponding to aim:Segmentation​Entity</para>
                  </listitem>
                  <listitem>
                      <para xml:id="para_28d24cb3-3ea1-48c3-98f0-1003589e8149">CONTAINER - for (126200, DCM, "Image Library Group") Content Item corresponding to aim:Image​Reference​Entity</para>
                  </listitem>
                  <listitem>
                      <para xml:id="para_c85da26c-91cd-4b3b-9f2c-e9a7f080c678">CONTAINER - for (125007, DCM, "Measurement Group") Content Item corresponding to aim:Image​Annotation</para>
                  </listitem>
                  <listitem>
                      <para xml:id="para_b9f2ea8f-957f-467a-9df6-0a4fbf83a3c4">NUM - for any Content Item corresponding to aim:Calculation​Entity</para>
                  </listitem>
                  <listitem>
                      <para xml:id="para_7304ca18-e97b-464f-a8d2-0e0ae7352500">SCOORD - for any Content Item corresponding to aim:MarkupEntity</para>
                  </listitem>
              </itemizedlist>
              <para xml:id="para_658c21d6-df99-4dd8-8f3c-a39037ee116d">Observation​DateTime is required for the following Content Items in order to propagate the aim:dateTime information:</para>
              <itemizedlist>
                  <listitem>
                      <para xml:id="para_6f39f523-ec97-4357-9f56-156580a0736e">CONTAINER - for (125007, DCM, "Measurement Group") Content Item corresponding to aim:Image​Annotation</para>
                  </listitem>
              </itemizedlist>
          </section>
          <section label="A.6.1.1.11" status="5" xml:id="sect_A.6.1.1.11">
            <title>Mapping of DICOM SOP Common Module</title>
            <table frame="box" label="A.6.1.1.11-1" rules="all" xml:id="table_A.6.1.11-1">
              <caption>Mapping of DICOM SOP Common Module</caption>
              <thead>
                <tr valign="top">
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_ff9b2ca2-6cbc-4310-b1c4-148c5c50c51e">DICOM Attribute</para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_7ec5c798-ba2e-42af-a9fb-ff7bee0b9818">DICOM VR</para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_61ce7ee9-bbcd-40e4-b56a-1df9a3c15c1e">DICOM VM</para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_350efd92-0ae9-43ee-8f9f-6f7fdb61b4e5">DICOM Usage Type</para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_576a9e06-ec24-494b-aefe-dfffc026b2cd">Generated Value</para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_1d1db16f-5db8-4ba4-82b1-4b6ef51fdd56">AIM Element or Attribute</para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_66271f38-9b18-448e-8748-a8da0210f633">AIM Data Type</para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_29761da5-9def-467c-9d01-e088dcd66a7f">AIM Multiplicity</para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_97952708-1fa8-4871-9461-399fa915c2be">Comment</para>
                  </th>
                </tr>
              </thead>
              <tbody>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_116c0a15-d57c-4958-85b0-280647de6107">SOPClassUID</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_680debe4-df1f-422d-b001-5361cc99aafb">UI</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_87167da3-fca1-4b06-a33d-5204a43725a4">1</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_f0b6497b-5852-4fc7-8169-a6d07e76bbdf">1</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_94787066-7c86-4f45-ba9b-7ee6be2c059b">"1.2.840.10008.​5.​1.​4.​1.​1.​88.​22"</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_1793ba33-fd37-40b1-a7a9-a56d4542d327"/>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_a9fb6e13-fe96-4713-9eef-a92619bd3876"/>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_497e40aa-0b39-459e-80e6-429b9f41d3f1"/>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_611c3209-ba5a-4e09-9218-bce3dbfad6c9">The fixed value is the SOP Class UID of the Enhanced SR Storage SOP Class, which is sufficient unless SCOORD3D are used, in which case the Comprehensive 3D SR Storage SOP Class is required, which has a UID of "1.2.840.10008.​5.​1.​4.​1.​1.​88.​34".</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_cc117342-db7a-484e-b1b7-15cb3a755823">SOPInstanceUID</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_cde7acad-d413-4da0-b1ec-465e224b3455">UI</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_25b5dd09-7bf0-4ddf-9c04-78679cce0f7c">1</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_cbd8d113-20e3-440a-8911-70d32cf6d4ad">1</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_c0dffc5e-749d-40f5-9d32-f26ebb553c4c">Generated if more than one converted instance in separate studies.</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_6a3581dc-b18e-45f7-aa6e-8d366dbf016b">Image​Annotation​​Collection&gt;uniqueIdentifier/​@root</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_335b887a-1af7-4d41-a25b-a95c9d6f5d76">II</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_eb56ce5a-c303-45f8-ba98-1399eb88ccd0">1..1</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_805f8876-d00b-4c3e-bca7-8df6aded35eb"/>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_0269e8c8-4689-49a6-90d0-e7a16a94b45b">SpecificCharacterSet</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_1de49c1c-ddc4-49c7-8f9c-8fbbfbcd4db3">CS</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_e4cfe8a0-367f-40bd-b351-8bcfaa414d06">1</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_22968b49-294b-4525-a5bc-306f84c3c6df">1</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_9d56edeb-bc3d-4d64-ad04-4c59b41e910a">"ISO_IR 192"</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_8705d876-6efa-4ed3-92c7-664d3dac0b4e"/>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_c8f70b4f-87c3-48eb-b45e-10fc3f5e3be7"/>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_7bfa24f8-9685-496f-b323-40f21580a564"/>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_025396d7-c177-46a0-a209-8254bb4dfaef">The fixed generated value corresponds to the normal UTF-8 specified in the XMLDecl of the XML prolog <xref linkend="biblio_XML"/> in the AIM instance. Other values may be used if they are sufficient to describe all the encoded characters in the transformed instance. E.g., it may be omitted entirely if all the characters are US-ASCII.</para>
                  </td>
                </tr>
              </tbody>
            </table>
          </section>
          
          
          
          
          
          
          
          
          
          
          
          
          
          
          
          
          
          
          
          
          
          
          
          
          
          
          
          
          
          
          
          
          
          
          
          
          
          
          
          
          
          
          
          
          
          
          
          
          
          
          
          
          
          
          
          
          
          
          
          
        </section>

        <section label="A.6.1.2" status="4" xml:id="sect_A.6.1.2">
          <title>Content Tree</title>
          <section label="TID 1500" status="5" xml:id="sect_Mapping_TID_1500">
            <title>Mapping of Measurement Report</title>
            <para xml:id="para_fc34e5cc-1f6b-4543-85ab-fdd9294d1fb1">This section describes the mapping of <olink targetdoc="PS3.16" targetptr="sect_TID_1500" xrefstyle="select: docname labelnumber quotedtitle"/>.</para>
            
            <table frame="box" label="TID 1500" rules="all" xml:id="table_Mapping_TID_1500">
            <caption>Mapping of Measurement Report</caption>
               <thead>
                  <tr valign="top">
                     <th align="center" colspan="1" rowspan="1">
                       <para xml:id="para_043b0cea-15c2-4584-b520-1b66d62a142d">DICOM SR Path</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_4e216d58-166e-462c-a45d-01d51b2dfac1">DICOM VT</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_e4a66c96-e498-44b0-9549-7242ff178546">DICOM VM</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_d7c9cdf9-75e5-4b44-99ab-19f68e8ac390">DICOM Usage Type</para>
                     </th>
                    <th align="center" colspan="1" rowspan="1">
                      <para xml:id="para_0221b393-976a-4056-86e7-868aa5297975">Generated Value</para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                      <para xml:id="para_86741d66-c0d2-4b02-81a5-ad54f3361210">AIM Element or Attribute</para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                      <para xml:id="para_61dc7897-4548-4c42-9a61-52557b204d0c">AIM Data Type</para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                      <para xml:id="para_360e3c0e-bdf5-4bd5-8902-631eaec70380">AIM Multiplicity</para>
                    </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_dc79d176-fc3f-4ceb-8668-a35b2da69a83">Comment</para>
                     </th>
                  </tr>
               </thead>
               <tbody>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_768d2567-3037-4692-92d4-2dfa3130e99f">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink>
</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_ab2127a7-7dbd-4d94-ab73-5412ae1a9077">CONTAINER</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_78e7780a-b26c-4868-8c1c-0f39ea0a2250">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_644c9238-a551-489b-a6ee-7cbf84762012">M</para>
                     </td>
                  <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_a991fcd9-7826-4a24-a9f3-ac9d205b4280"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_6c6fe36a-d1f9-4afb-9ba7-f6db075c42dc"/>
                     </td>
                  <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_5eb6e185-ccbe-4f2f-8f86-64d44580bd04"/>
                     </td>
                  <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_339bc02e-f17a-46fa-b880-4fd94bb42666"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                       <para xml:id="para_b11185b8-9b6b-4ba4-b4d4-8bd2bd9d8753">The fixed Concept Name code is an appropriate choice selected from <olink targetdoc="PS3.16" targetptr="sect_CID_7021" xrefstyle="select: labelnumber quotedtitle"/> in the absence of a "document title" concept in AIM.</para>
                    <para xml:id="para_13e5ce97-4501-44c8-ac13-f6ebbac979e1">If out of band information is available, a more appropriate code can be used.</para>
                     </td>
                  </tr>
                  <tr valign="top">
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_4a5e4966-5e71-48a1-a6ff-d3f9e3f384b3">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt;</para>
                    </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_741a742c-fdac-4871-8a14-f05a6b7d8ca5">INCLUDE</para>
                     </td>
                    <td align="center" colspan="1" rowspan="1">
<para xml:id="para_7449ccdf-ccd8-4f5e-9902-31df3e541633">1</para>
</td>
                    <td align="center" colspan="1" rowspan="1">
<para xml:id="para_2f4ae6db-af1c-40e0-96ab-a63f3f64b785">M</para>
</td>
                    <td align="left" colspan="5" rowspan="1">
                        <para xml:id="para_b4b12429-0b76-44c3-a2d5-38c3b002dace">
<xref linkend="sect_Mapping_TID_1204" xrefstyle="select: labelnumber quotedtitle nopage"/>
                    </para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_c68fcd0d-078f-4340-9b2c-77192aa19aca">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt;</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_2dd5655f-edda-4867-9095-15c108763d4e">INCLUDE</para>
                     </td>
                    <td align="center" colspan="1" rowspan="1">
<para xml:id="para_0b8b2dfd-fe77-41b2-a45a-987b811a128a">1</para>
</td>
                    <td align="center" colspan="1" rowspan="1">
<para xml:id="para_1e09a0b6-f954-443b-87ce-7128e77c9bb0">M</para>
</td>
                    <td align="left" colspan="5" rowspan="1">
                        <para xml:id="para_26b2aadc-0a46-4f56-9d58-f7084d3f88dd">
<xref linkend="sect_Mapping_TID_1001" xrefstyle="select: labelnumber quotedtitle nopage"/>
                    </para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_ab0a6fa1-3a24-4dc2-86e4-1ec43fb8b7ec">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_121058">(121058, DCM, "Procedure reported")</olink>
</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_c4ba0598-5afa-48a5-8b8b-e4e05ba9ed16">CODE</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_e8cd25b1-180d-4fa1-bd9f-c2d189f17734">1-n</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_ce295d50-6a3d-4923-a810-0ac8a4163ba5">M</para>
                     </td>
                  <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_6be68148-b281-44ae-b366-24b7a8173700">
<link xl:href="http://snomed.info/id/363679005" xl:show="new">(363679005, SCT, "Imaging procedure")</link>
</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_7d202807-5c6a-4dff-9b4f-bc0a01a0073a"/>
                     </td>
                  <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_89891acf-edf7-4e2c-9ec2-a7ff81a0404c"/>
                     </td>
                  <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_9db6f891-984f-4d80-80ba-c3a9a373820c"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                       <para xml:id="para_7228461f-091b-4656-b4d5-28bb2e2d4513">The fixed generic code value is suggested in lieu of AIM containing any information about the imaging procedure, so more specific procedure codes such as those from <olink targetdoc="PS3.16" targetptr="sect_CID_100" xrefstyle="select: labelnumber quotedtitle"/> cannot be used.</para>
                    <para xml:id="para_df4430d7-a4b3-4d8f-8a79-6b31fd4d7511">If out of band information is available, a more appropriate code can be used.</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_c4392c98-6b39-4852-a77c-72a6da33e07e">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; </para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_061079fd-9fe4-4175-82ea-389888fe4eec">INCLUDE</para>
                     </td>
                    <td align="center" colspan="1" rowspan="1">
<para xml:id="para_ea875fc9-b017-4da4-955d-5bd99dfb679a">1</para>
</td>
                    <td align="center" colspan="1" rowspan="1">
<para xml:id="para_2a892365-dcde-46ae-864c-8e027f9aed8b">M</para>
</td>
                    <td align="left" colspan="5" rowspan="1">
                        <para xml:id="para_805f6b7a-234f-42b3-b16a-9f1f7fea4892">
<xref linkend="sect_Mapping_TID_1600" xrefstyle="select: labelnumber quotedtitle nopage"/>
                    </para>
                     </td>
                 </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_f87c3624-f355-44fa-9c20-7013d7f5c1c2">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126010">(126010, DCM, "Imaging Measurements")</olink>
</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_562da109-62fd-4265-9c34-cf9aca43d2c8">CONTAINER</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_de0dcfb6-61bd-4808-86a8-505dbfc758fd">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_cbc16254-ed3d-45f2-b533-af26ae0f9899">C</para>
                     </td>
                  <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_ab618c3b-489e-47f8-98f5-5bb1b02b0158"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_f3e40f2d-be0f-448b-b3b8-7e603f59a678"/>
                     </td>
                  <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_7443d5bc-831a-4c7a-94a4-aa200b177ac5"/>
                     </td>
                  <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_d43f1585-5078-44a6-b604-e0e32bd173a3"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_def76c39-5095-42a6-87ca-a89bdc925f35">IFF measurements are present in the source AIM object</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_c8ff3323-2de0-4cca-b946-e82b7b71cfa8">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126010">(126010, DCM, "Imaging Measurements")</olink> &gt;</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_f5e0187b-e48d-4118-9bfb-b4132df8ed13">INCLUDE</para>
                     </td>
                    <td align="center" colspan="1" rowspan="1">
<para xml:id="para_e187ae90-f9ad-47ec-90aa-c3eb188e202a">1-n</para>
</td>
                    <td align="center" colspan="1" rowspan="1">
<para xml:id="para_90df9a0f-67e8-47b7-b136-41f8583c2e33">U</para>
</td>
                    <td align="left" colspan="5" rowspan="1">
                        <para xml:id="para_3905cc3c-4613-4f8b-9194-b6adf33e77ab">
<xref linkend="sect_Mapping_TID_1410" xrefstyle="select: labelnumber quotedtitle nopage"/>
                    </para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_d3c5e53d-8dd7-4e1e-a772-850f44a1c83c">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126010">(126010, DCM, "Imaging Measurements")</olink> &gt;</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_5342d6d1-c3f3-4eb3-95d2-4be9ccbede50">INCLUDE</para>
                     </td>
                    <td align="center" colspan="1" rowspan="1">
<para xml:id="para_8b422123-2bf9-4e81-b0d5-d35d96521f1b">1-n</para>
</td>
                    <td align="center" colspan="1" rowspan="1">
<para xml:id="para_dad5376a-2099-4e85-bf87-88d3d1bce0bb">U</para>
</td>
                    <td align="left" colspan="5" rowspan="1">
                        <para xml:id="para_ec8098ef-3114-4df2-b8b5-a7384e0fa9b8">
<xref linkend="sect_Mapping_TID_1411" xrefstyle="select: labelnumber quotedtitle nopage"/>
                    </para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_3410fb19-6119-43a0-a373-4794e832da9c">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126010">(126010, DCM, "Imaging Measurements")</olink> &gt;</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_1f93d17b-897a-434e-9ee6-af7bce27b561">INCLUDE</para>
                     </td>
                    <td align="center" colspan="1" rowspan="1">
<para xml:id="para_e610515b-624e-429a-a3d3-2512c884bc44">1-n</para>
</td>
                    <td align="center" colspan="1" rowspan="1">
<para xml:id="para_5e4c5b41-4b43-4324-a20f-7aa04f56fbca">U</para>
</td>
                    <td align="left" colspan="5" rowspan="1">
                        <para xml:id="para_81807a5d-f29c-4060-a6ce-c4ff3bf7fc8d">
<xref linkend="sect_Mapping_TID_1501" xrefstyle="select: labelnumber quotedtitle nopage"/>
                    </para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_8ecf8b48-b5d9-413a-a027-76d0d53ce002">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink>  &gt; <olink targetdoc="PS3.16" targetptr="DCM_126011">(126011, DCM, "Derived Imaging Measurements")</olink>
</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_3fc73237-6a34-448a-970d-ffc2e8264519">CONTAINER</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_0cad8e93-f1d6-4f5c-93cd-036e940b1795">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_5be15017-ae25-40f2-b756-8c39d6617c8c">C</para>
                     </td>
                  <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_b3f21b4e-a856-41db-996a-c5795cb1e85a"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_9770a9d7-9493-4420-8a5e-73e65e780f3f"/>
                     </td>
                  <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_fbe2c3b8-fba7-4b9c-8fae-5937c8aaaced"/>
                     </td>
                  <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_f6a7d515-b675-4e52-97c5-84244b1e555d"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_f92ab8f2-d020-4837-9d8e-125b2de40ccb">Not applicable to current use cases</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td>
                    <para xml:id="para_085cd99c-2bef-49ab-afee-f33b1b62322c">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126011">(126011, DCM, "Derived Imaging Measurements")</olink> &gt;</para>
                  </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_81384a38-4220-4b9e-a3b8-a753831ea9f3">INCLUDE</para>
                     </td>
                    <td align="center" colspan="1" rowspan="1">
<para xml:id="para_60f0e8a0-d99f-4518-ae75-246d34b6122f">1-n</para>
</td>
                    <td align="center" colspan="1" rowspan="1">
<para xml:id="para_3d08e3f5-eff8-4db4-ae66-679fb70653bf">U</para>
</td>
                    <td align="left" colspan="5" rowspan="1">
                        <para xml:id="para_0b5e3ae2-f8c5-4877-8f72-3cca0490f4ca">
<xref linkend="sect_Mapping_TID_1420" xrefstyle="select: labelnumber quotedtitle nopage"/>
                    </para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_d9b9f4f4-ee5b-464c-8475-503ba2895953">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <link xl:href="https://uts.nlm.nih.gov/metathesaurus.html?cui=C0034375" xl:show="new">(C0034375, UMLS, "Qualitative Evaluations")</link>
</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_ab48dfe8-144f-49e9-b42b-f351bf923e1e">CONTAINER</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_7ef8e785-55be-4718-bcec-91537a3a5e43">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_4dd2140c-9612-49a5-b5d6-c0fbaedaacf7">C</para>
                     </td>
                  <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_6f02df20-b688-497c-91c4-fab73ce9d09c"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_1f8859b2-811f-488a-8169-da612b320f3a"/>
                     </td>
                  <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_1f8b0a25-a2cf-4bc1-adc8-27f0b014222c"/>
                     </td>
                  <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_9a1a1ec1-0f67-4684-b2d3-5c8f2bb74013"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_26377bcd-a17d-4750-940a-4c2b441db72e">IFF Imaging​Observation​Entity elements are present in the source AIM object</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_6b84d4db-efdd-490c-b9e5-840ef745bdc4">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <link xl:href="https://uts.nlm.nih.gov/metathesaurus.html?cui=C0034375" xl:show="new">(C0034375, UMLS, "Qualitative Evaluations")</link> &gt; CODE</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_2048148c-a30d-427f-a4d2-57b4ab1dad08">CODE</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_4be4f838-0161-4296-be42-a082bf020589">1-n</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_51549c83-5f92-4044-a154-5e11a5e5804c">U</para>
                     </td>
                  <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_f4ae4e88-7566-49ae-b3eb-9d97b3394e5c"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                         <para xml:id="para_9df75c68-9ff6-4288-84b4-cd57fd0b897e">NAME = Image​Annotation​​Collection/​image​Annotations/​Image​Annotation/​imaging​Observation​Entity​Collection/​Imaging​Observation​Entity​/​imaging​Observation​Characteristic​Collection​/​Imaging​Observation​Characteristic​​/​questionType​Code</para>
                         <para xml:id="para_bb29bc9b-9e9b-4456-aa85-d4724348dc7d">or NAME = Image​Annotation​​Collection/​image​Annotations/​Image​Annotation/​imaging​Observation​Entity​Collection/​Imaging​Observation​Entity​/type​Code</para>
                         <para xml:id="para_398c51d2-5d57-4def-9045-9a8b3fa54319">VALUE = Image​Annotation​​Collection/​image​Annotations/​Image​Annotation/​imaging​Observation​Entity​Collection/Imaging​Observation​Entity/imaging​Observation​Characteristic​Collection​/​Imaging​Observation​Characteristic​/type​Code</para>
                     </td>
                  <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_ee631245-37c1-4132-917f-b480ec736089">CD, CD</para>
                     </td>
                  <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_1cd88a1b-e2ef-4e58-a6a0-875bf4c894ca">1, 0..1</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                         <para xml:id="para_29752944-a84a-478f-b7a8-ac5d0b256f1f">The concept name may be encoded as a specific question​Type​Code for the Imaging​Observation​Characteristic​ or inherited from the type​Code of the parent Imaging​Observation​Entity.</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_6bb5ab99-c630-4c85-961d-bbcc5eca7fdb">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <link xl:href="https://uts.nlm.nih.gov/metathesaurus.html?cui=C0034375" xl:show="new">(C0034375, UMLS, "Qualitative Evaluations")</link> &gt; TEXT</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_093e5b62-411c-4728-b7fa-ac99ebffec95">TEXT</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_c651019d-f774-4e9c-b426-6432a676ba9c">1-n</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_60230835-43af-4004-b332-8fcfd4b9b686">U</para>
                     </td>
                  <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_926d0167-1d3a-4dc1-9b9a-ba4b44a6dc78"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_c021858a-0d02-48fb-80ac-4b4b2ee8744b"/>
                     </td>
                  <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_809f888c-de0c-488f-ac83-53922b3b9167"/>
                     </td>
                  <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_d0fcaefc-7405-4963-8cea-8f0a1e2fd35e"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_1a56c436-acbf-4e32-b486-2d9247df86cb">Not used in AIM</para>
                     </td>
                  </tr>
               </tbody>
            </table>
         </section>
          
                   <section label="TID 1501" status="3" xml:id="sect_Mapping_TID_1501">
            <title>Mapping of Measurement Group</title>
            <para xml:id="para_6640d68a-105d-4c93-a679-a7c8b4953c94">This section describes the mapping of <olink targetdoc="PS3.16" targetptr="sect_TID_1501" xrefstyle="select: docname labelnumber quotedtitle"/>.</para>
            <table frame="box" label="TID 1501" rules="all" xml:id="table_Mapping_TID_1501">
            <caption>Mapping of Measurement Group</caption>
              <thead>
                <tr valign="top">
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_d49007cd-9414-4005-845d-512ee3afd6df">DICOM SR Path</para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_187d0e68-5592-4050-8b72-3efe408e7df3">DICOM VT</para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_56a7020f-1782-4f8b-acca-b93c87c7fd16">DICOM VM</para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_aa2d2b8f-5ade-4dee-8156-590d484d00bb">DICOM Usage Type</para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_62ff628f-99ef-4c62-a7b7-e427ca5e4195">Generated Value</para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_b5810386-dd35-4f0b-a421-9c1554c14c8d">AIM Element or Attribute</para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_ce678e7e-c048-4ab9-850d-8eb657c779b0">AIM Data Type</para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_251b6c65-8c2f-4f64-b0c5-d7d26c1494f6">AIM Multiplicity</para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_1b5c58f6-e9ff-4619-b7c7-1f7efcc86a2b">Comment</para>
                  </th>
                </tr>
              </thead>
              <tbody>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_d79d3a4e-3120-4e5d-b35c-a519f0ffe038">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126010">(126010, DCM, "Imaging Measurements")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_125007">(125007, DCM, "Measurement Group")</olink>
</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_74f0eb23-1f33-4176-af64-701d09da492a">CONTAINER</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_d7f92f3c-a426-48a1-b5d9-ea06a3133a7b">1</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_9c620ded-6a36-4807-a369-b06458b3e7b8">M</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_08470585-9008-402e-a2f9-18c5df158856"/>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_3b6fc81e-dbea-4fd4-8920-4954689ca3d1">Image​Annotation​​Collection/​image​Annotations/​Image​Annotation</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_c72b086a-9c9c-4c02-9f74-a0ee67d001f0"/>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_99011044-1358-4672-a1dc-e3a770ad7f11"/>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_9d4630a1-e81e-4180-ad09-8f204870986c">The value of aim:unique​Identifier/@root is mapped to the Observation​UID Attribute of the CONTAINER.</para>
                      <para xml:id="para_43184a3c-709b-4539-a601-598f5736d9e4">The value of aim:date​Time/@root is mapped to the Observation​Date​Time​Attribute of the CONTAINER.</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_c77e8c01-7792-4f6c-9f23-7db2c605522f">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126010">(126010, DCM, "Imaging Measurements")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_125007">(125007, DCM, "Measurement Group")</olink> &gt; (C67447, NCIt, "Activity Session")</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_7fec71d3-5ad4-47ac-b01d-1b79c36012b9">TEXT</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_0482d3f7-fa41-4a51-bf04-26335db47988">1</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_19c8dbcd-4108-421a-b6eb-3af6fab6d07b">U</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_5b69c5a1-9610-46d8-9a41-e6a93b8615ff"/>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_bb1adbc9-fbce-4cf4-92bc-ed20056ea13b"/>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_10761d18-f3f9-4e12-9e23-1cd12e0ab35d"/>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_82098f55-3b3d-432d-ad30-f505221ab256"/>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_8c03188f-ae36-4533-8bb1-1cb5b34aeb63">Not present in AIM.</para>
                    <para xml:id="para_4e29c7c6-c33f-4e68-8b17-ff2a521cad8f">Identifies the session during which the measurements were made. The NCI Thesaurus definition is "time, period, or term devoted to some activity".</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_a79990e3-7785-4866-a2be-4721de47a84f">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126010">(126010, DCM, "Imaging Measurements")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_125007">(125007, DCM, "Measurement Group")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_112039">(112039, DCM, "Tracking Identifier")</olink>
</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_e8256222-5fe9-4b2a-94da-e09a251d3a5f">TEXT</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_aeb9aecb-d927-48dd-afb4-a5faf88f5f2a">1</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_0fedf2ba-4b9d-4991-9880-a8914396ef11">M</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_88f808dd-79f1-43a7-b3cd-a3f968df96be"/>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_a18d6106-3643-4f00-a0ea-ebcb8ced8b15">Image​Annotation​​Collection/​image​Annotations/​Image​Annotation/​name/​@value</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_2ccf65df-4f77-4374-81e3-7c664cbb283e">ST</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_1c30e9a5-3acf-44ca-9436-0a198598a332">1</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_0eea0561-60bb-4947-86b5-8a360f38ac96"/>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_cb699ebd-6ba9-4bf0-8a63-9a96a8903131">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126010">(126010, DCM, "Imaging Measurements")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_125007">(125007, DCM, "Measurement Group")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_112040">(112040, DCM, "Tracking Unique Identifier")</olink>
</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_b7a1fdfb-989e-4492-9d5d-c4fece99bd46">UIDREF</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_a817773d-c600-498d-93ff-03b183245b36">1</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_b10ee8b3-b4fa-4527-8061-e8666ef9ea5d">M</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_f8f1e300-dba0-4f5b-9359-9158a67d77d4"/>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_489b5c85-4195-4053-80b4-f2cb7d2d4ce9">Image​Annotation​​Collection/​image​Annotations/​Image​Annotation/​tracking​UniqueIdentifier/​@root</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_b8d292a7-ad91-4f0c-b733-5a2567db0cbf">II</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_22eda7d1-1e9b-42d1-ad61-ad1ada461d4d">1</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_9bf004e6-978f-47a8-9469-7b6e2c3061c4">If tracking​Unique​Identifier is absent,
                          then Image​Annotation/​uniqueIdentifier may be used as a proxy for Tracking Unique Identifier,
                          but this does not allow longitudinal identification of the same lesion because Image​Annotation/​uniqueIdentifier must be unique to each AIM annotation file,
                          in which case only Image​Annotation/​name can be used to recognize commonality.</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_53e4f014-edd5-4f73-8dce-cfd2acfa805d">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126010">(126010, DCM, "Imaging Measurements")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_125007">(125007, DCM, "Measurement Group")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_121071">(121071, DCM, "Finding")</olink>
</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_ccf3d80b-a07d-4576-a951-f76fb3855516">CODE</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_cb19dd37-45aa-4f6e-a2e3-b7657e9681d4">1</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_1c20d6dc-a5ae-4afd-8357-d4cef15b5190">U</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_7c5a53b6-13cc-48e1-85c1-72d7e8cfa350"/>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_92b84b67-3a32-4816-a789-a1e946f072a3">Image​Annotation​​Collection/​image​Annotations/​Image​Annotation/​typeCode</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_3fe0cf48-bdca-47ef-b14b-63c56ee60932">CD</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_966d7570-6aea-41f0-9c2a-aeac700d8f06">1..n</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_0205313d-689f-4538-addb-c23e63b40533">Only a single aim:typeCode value can be mapped.</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_7a5e3eea-726a-42bc-9cbe-44afe946319c">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126010">(126010, DCM, "Imaging Measurements")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_125007">(125007, DCM, "Measurement Group")</olink> &gt; </para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_74bacb20-9b6d-4f85-8bb8-55f8ec9a6da0">INCLUDE</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_9c0659bd-033f-4b13-8c8a-61b140d92ddf">1</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_2dd5682a-14fe-4e9d-b5a2-0c0bb385fc1f">U</para>
                  </td>
                  <td align="left" colspan="4" rowspan="1">
                    <para xml:id="para_2588552d-a895-4bd4-9db8-85e20704e701">
<xref linkend="sect_Mapping_TID_1502" xrefstyle="select: labelnumber quotedtitle nopage"/>
                    </para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_696f86d3-e634-4d70-ac06-4b2eff37c57c">Not used in AIM.</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_a2ab127e-f788-4f3f-91c3-455c6c58512a">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126010">(126010, DCM, "Imaging Measurements")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_125007">(125007, DCM, "Measurement Group")</olink> &gt; <link xl:href="http://snomed.info/id/370129005" xl:show="new">(370129005, SCT, "Measurement Method")</link>
</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_6aec8392-caba-48e8-b44b-cbda49c84446">CODE</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_22c89503-dff7-44df-9757-6ab783734c94">1</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_64a87c75-2484-4c17-8413-0317502db8dc">U</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_4ea2a2e4-5b28-4703-83e3-1e85aeb26e76"/>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_2f7343bd-b6e8-45ff-9776-38844562bd39"/>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_3e4ed8fc-9e2c-4342-830f-621f397c77ef"/>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_bfe06fd1-ef79-49fd-81bb-cb21d7d1f9dc"/>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_e3045ce9-4b2e-44dc-a830-5ba875ca95df">Not used in AIM at this level.</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_8c1dd64e-92ff-47bb-89d9-55c33dcca93f">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126010">(126010, DCM, "Imaging Measurements")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_125007">(125007, DCM, "Measurement Group")</olink> &gt; <link xl:href="http://snomed.info/id/363698007" xl:show="new">(363698007, SCT, "Finding Site")</link>
</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_5d80c40d-6cea-44c0-b803-ead0721157d1">CODE</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_b45f5ce0-f934-4cfc-8dec-444416680953">1-n</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_ee8b9ab3-0f03-4f1d-8083-429c489a0b1c">U</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_dd74dd9b-78b6-43f1-b2be-81d83c45c8ea"/>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_71b842f5-7a3c-4619-bb0c-50945ca8e7a1">Image​Annotation​​Collection/​image​Annotations/​Image​Annotation/​imagingPhysical​Entity​Collection/​ImagingPhysicalEntity[label/​@value='Location' or label/​@value='Lobar Location' or label/​@value='Segmental Location' or label/​@value='Organ Type']/typeCode</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_9461e65b-9b52-4ded-9269-58ce31d47b7a">CD</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_2162122b-6a33-4333-b785-561d303fdc05">1..n</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_0443f875-5f0f-49e9-b730-38cfa5935b6c">If a paired structure in AIM, this entry will pre-coordinate the laterality with the site.</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_73792436-5d61-4db9-ba26-15ce0aa845ae">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126010">(126010, DCM, "Imaging Measurements")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_125007">(125007, DCM, "Measurement Group")</olink> &gt; <link xl:href="http://snomed.info/id/363698007" xl:show="new">(363698007, SCT, "Finding Site")</link> &gt; <link xl:href="http://snomed.info/id/272741003" xl:show="new">(272741003, SCT, "Laterality")</link>
</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_bd4f79aa-2f06-4321-9104-282dbbd3e86f">CODE</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_3a8bb7b9-2a53-4246-860c-b46384e91588">1</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_252b6626-9556-40b9-aabc-d7ce4fb96e11">U</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_f768dd2c-db74-4a3d-b2f1-7866a25905b4"/>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_52ecf76e-26bf-416d-97fc-b9c6d523e59c"/>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_0ca07deb-41ad-448e-8b10-81ba298e5e9a"/>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_e78f2a31-5c1c-4800-95a7-8189c789a1d8"/>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_eadfce8d-03a2-4683-ba45-3a9fc7e07d94">
<olink targetdoc="PS3.16" targetptr="sect_CID_244" xrefstyle="select: labelnumber quotedtitle"/>
</para>
                    <para xml:id="para_0d41395b-4235-4022-a681-70e18136f713">Not used in AIM. </para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_a8e02b16-89d1-4574-87e2-5dc9f212faa9">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126010">(126010, DCM, "Imaging Measurements")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_125007">(125007, DCM, "Measurement Group")</olink> &gt; <link xl:href="http://snomed.info/id/363698007" xl:show="new">(363698007, SCT, "Finding Site")</link> &gt; <link xl:href="http://snomed.info/id/106233006" xl:show="new">(106233006, SCT, "Topographical modifier")</link>
</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_3dab7832-6034-4ac5-bd28-8fce38aeeb8e">CODE</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_266a29b9-56be-4a45-99da-9815bd4a4834">1</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_89cfd8b3-02c1-4631-9033-e423c6edd76b">U</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_8eb05f12-ff83-4a35-96f1-2a1220281d0e"/>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_cc846f75-8405-4059-8846-b109671e935d"/>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_e35c41d8-e979-481b-b6ea-0b65ab7c948f"/>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_184a4427-f5a5-4271-a2b8-a5c05b574813"/>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_59426e00-93e7-412b-8873-cb9917e6312a">Not used since AIM does not have a mechanism for post-coordinating the location.</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_e0f4b04f-811d-431b-8539-866bfcd95002">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126010">(126010, DCM, "Imaging Measurements")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_125007">(125007, DCM, "Measurement Group")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126100">(126100, DCM, "Real World Value Map used for measurement")</olink>
</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_ac578953-652b-4973-8731-40690e653bde">COMPOSITE</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_0378db2b-a82f-4f6a-8ab6-ff4833661f71">1</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_b3aad688-bb27-48b9-aaf4-1b06d82ad035">U</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_9dcfe2f1-ce82-48f6-9ea0-f8dda19f5a72"/>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_338d119b-cfa2-4b37-bfe4-42579b0359bd"/>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_3a2717cd-018b-4e1b-bf75-d0931d8566f8"/>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_354b5583-9099-4ea5-a768-22a8a6f330a9"/>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_62959436-149c-499c-815b-d5247cebfbe8">Not used in AIM.</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_618cfc7e-96ba-4512-a50c-1b28cf7cfe31">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126010">(126010, DCM, "Imaging Measurements")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_125007">(125007, DCM, "Measurement Group")</olink> &gt; </para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_1a53f0d7-a525-4bbe-925f-471f4f2b7c2a">INCLUDE</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_c669ad5b-737d-4450-8d42-2d7955cf8b53">1-n</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_0e9e5988-3b53-4aa3-b412-e0141c9e6c33">M</para>
                  </td>
                  <td align="left" colspan="5" rowspan="1">
                    <para xml:id="para_b2bbd6ed-0c5e-4b25-9f52-9932985d2c4e">
<xref linkend="sect_Mapping_TID_300" xrefstyle="select: labelnumber quotedtitle nopage"/>
</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_a25e5b3e-24ee-4876-ad32-a6bc6125af2b">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126010">(126010, DCM, "Imaging Measurements")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_125007">(125007, DCM, "Measurement Group")</olink> &gt; $QualitativeEvaluations</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_4c41379a-bfe3-4ecb-9634-621fa5c5e9d3">CODE</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_b2f29034-9183-46d8-ba1f-0b9f9d0c86e2">1-n</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_a80ce333-ef7f-4bc1-b681-879344e29422">U</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_000cc901-4c7f-41f5-bdd3-1efee32bfa80"/>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_f4de8602-6e33-45f5-8d15-2f9c7d337f08"/>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_c7877275-1c0a-4119-a125-8a3b5a5dd1e1"/>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_fba3b61c-94d6-450a-a41d-9d6013f4a461"/>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_526b54cf-4593-4d1a-a6ee-70106358af21">Not used in AIM.</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_78aabc2a-e32c-4b96-8a4a-2c68a95991f3">
                        <olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126010">(126010, DCM, "Imaging Measurements")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_125007">(125007, DCM, "Measurement Group")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_121106">(121106, DCM, "Comment")</olink>
</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_36d3744d-abc0-4fe6-892a-905a93e732d1">TEXT</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_7bc8377b-7e48-42e8-aa66-73f2a7894acb">1-n</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_5fe895f5-ac8f-48e9-89eb-0a6782be8219">U</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_1d92b4da-0ab3-489b-95cb-dabe7ce7e50d"/>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_f07e4030-2863-4c63-9b8c-5d7cd16e5bab">Image​Annotation​​Collection/​image​Annotations/​Image​Annotation/​comment/​@value</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_c663344d-a6c6-4ad7-8cd6-200cdd77b112"/>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_1383e447-663a-472e-a7d1-25779966585d"/>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_af3be6e7-3b01-4dcf-9ab0-b7fe4e456eee">The parameter $Qualitative​Evaluations is not used in AIM, but this TEXT content item is used to map the AIM comment as if it were a Qualitative Evaluation.</para>
                  </td>
                </tr>
              </tbody>
            </table>
         </section>
         <section label="TID 1502" status="3" xml:id="sect_Mapping_TID_1502">
            <title>Mapping of Time Point Context</title>
            <para xml:id="para_c6422d82-0d41-413a-a678-1484a8601a64">This section describes the mapping of <olink targetdoc="PS3.16" targetptr="sect_TID_1502" xrefstyle="select: docname labelnumber quotedtitle"/>.</para>
            <table frame="box" label="TID 1502" rules="all" xml:id="table_Mapping_TID_1502">
            <caption>Mapping of Time Point Context</caption>
               <thead>
                  <tr valign="top">
                     <th align="center" colspan="1" rowspan="1">
                       <para xml:id="para_85113944-2d5b-4e47-babf-003e0a8f9212">DICOM SR Path</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_87b94184-5a79-4bd1-ba42-4b6782440e61">DICOM VT</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_ba1f53b3-475c-4437-a5f4-13df41a632f4">DICOM VM</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_128612de-fa4c-4c28-8f9e-56560e64dcd7">DICOM Usage Type</para>
                     </th>
                    <th align="center" colspan="1" rowspan="1">
                      <para xml:id="para_2f0911a4-68e0-46d0-8d89-d56a0338e7aa">Generated Value</para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                      <para xml:id="para_6c4641fe-a7cd-4260-a7e8-ab000e4ae2a9">AIM Element or Attribute</para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                      <para xml:id="para_2154281c-d31a-4e18-a84b-1e94775ec3ed">AIM Data Type</para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                      <para xml:id="para_3f8dfc01-c3c4-45cd-a8b0-163cafb1b78d">AIM Multiplicity</para>
                    </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_b81ca0a4-5e6c-43f1-8dad-71c5835a6b7d">Comment</para>
                     </th>
                  </tr>
               </thead>
               <tbody>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_67b2f166-8caa-4903-8e4d-c45872edf763">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126010">(126010, DCM, "Imaging Measurements")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_125007">(125007, DCM, "Measurement Group")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126070">(126070, DCM, "Subject Time Point Identifier")</olink>
</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_a607a897-ce4f-4279-8125-8e267e4ee2f2">TEXT</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_787d6e81-8036-47f8-a298-aadb468ef704">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_33d1f8de-3283-47a2-8995-965e717b7919">U</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_f5bf6c8f-6ada-4746-8162-236816e232b4"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_fe175b1d-2239-4a5a-8cfc-ba4285ba3176"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_1cc74d3c-3041-47c6-b54e-b8a5e131f16f"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_7ae69808-1622-4a3e-8751-384280213155"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_016a8836-cb56-4faa-967e-c73128877e67">Not used in AIM.</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_0138b6e6-da2e-4228-9111-6af571cce4b9">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126010">(126010, DCM, "Imaging Measurements")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_125007">(125007, DCM, "Measurement Group")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126071">(126071, DCM, "Protocol Time Point Identifier")</olink>
</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_1043bd26-40e8-4a4f-a561-d6034677a4cb">TEXT</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_c6a80f45-9713-4017-b27c-9cac4a4a2ffa">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_77c291c4-5e9a-486d-909a-c617154fd7d6">U</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_a48013cc-d4f1-42c6-9e83-ac136a9d3c06"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_d6b15d58-9009-4c2e-99be-51064c052663"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_ebdf1535-fd81-4d7c-a2d7-de1eca82a225"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_cdc21550-b901-46c3-ab07-0ad40b1fe6a7"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_28098f01-eae0-45ec-8993-799d127d5264">Not used in AIM.</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_c4543190-7254-4d7b-ba74-b3d2dbd86cea">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126010">(126010, DCM, "Imaging Measurements")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_125007">(125007, DCM, "Measurement Group")</olink> &gt; <link xl:href="https://uts.nlm.nih.gov/metathesaurus.html?cui=C2348792" xl:show="new">(C2348792, UMLS, "Time Point")</link>
</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_a05464ce-5c5e-4499-bcc6-f82ecfcaed49">TEXT</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_18d9ca2a-ed75-4770-ac23-77e285fbcf03">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_5a02ed96-6f10-451a-b99d-ec8d4064cc1b">M</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_4a6ecf13-a04b-4c83-a08b-3902c155f6be"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_2376175f-b8e3-46df-aa82-102fc4a269e1"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_4af56775-bc66-42f8-bbcc-7c5b339ece18"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_35206843-1204-4d9f-aa8b-85c992b906b7"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_1fe43eb5-841e-4c3a-b183-2dcb4ab30cdd">Not used in AIM.</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_bdf460f4-ab6b-4d51-b11b-49517df70331">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126010">(126010, DCM, "Imaging Measurements")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_125007">(125007, DCM, "Measurement Group")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126072">(126072, DCM, "Time Point Type")</olink> </para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_6de28612-345a-40d4-8f7d-1643a72e623c">CODE</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_97df12ba-7fff-409d-b249-394aaf7213d9">1-n</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_6d14cb0d-c790-4ae3-bfe8-86e06eedc96c">U</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_dc5e53c4-fed4-4bb7-a5f9-23885573a000"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                       <para xml:id="para_b738df90-b329-47fc-96f2-7905f33522bf"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_ffe56243-b5eb-4d3d-9f6c-3da50ecfa095"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_f126c8b1-0ad8-4783-9524-b720b31a5d79"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_cb18258e-31f1-4a3e-8ebc-535b61ec3ac8">B<olink targetdoc="PS3.16" targetptr="sect_CID_6146" xrefstyle="select: labelnumber quotedtitle"/>
                    </para>
                        <para xml:id="para_d920dcae-d110-4a52-bc9d-f917cb507535">Not used in AIM.</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_e838c76f-4d11-46ef-906c-df564477884c">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126010">(126010, DCM, "Imaging Measurements")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_125007">(125007, DCM, "Measurement Group")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126073">(126073, DCM, "Time Point Order")</olink>
</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_a9a0736f-0ec0-49b2-9940-49dbdae8687c">NUM</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_505e14b6-cf2b-42fc-a118-44551ff4abef">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_45b85718-9716-4572-b6a5-98ba8844880e">U</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_672c2e1c-143d-4125-b15a-ed738515e2a2"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_76aebe97-bde2-4433-b309-9505d5b672d9"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_8b69652a-fc0d-4ef0-b869-6a8f4729251b"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_9703a0e3-1a6e-421a-967a-7053e6e1a99d"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_7443fedd-aa60-4215-a778-b1a0fece7082">Not used in AIM.</para>
                     </td>
                  </tr>
                 <tr valign="top">
                   <td align="left" colspan="1" rowspan="1">
                     <para xml:id="para_9ba865a6-9cb0-408c-8821-0717c6f4e7d6">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126010">(126010, DCM, "Imaging Measurements")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_125007">(125007, DCM, "Measurement Group")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_128740">(128740, DCM, "Longitudinal Temporal Offset from Event")</olink>
</para>
                   </td>
                   <td align="left" colspan="1" rowspan="1">
                     <para xml:id="para_2d3fc46c-113b-4782-8633-a460f0ccc52b">NUM</para>
                   </td>
                   <td align="center" colspan="1" rowspan="1">
                     <para xml:id="para_a4f06d19-dea1-4d43-bb28-e70af84baa71">1</para>
                   </td>
                   <td align="center" colspan="1" rowspan="1">
                     <para xml:id="para_fe6424f4-c866-43a0-b801-57fdc5e8893a">U</para>
                   </td>
                   <td align="left" colspan="1" rowspan="1">
                     <para xml:id="para_5b0cc5fd-63bb-49bf-bed9-7405eccc18f2"/>
                   </td>
                   <td align="left" colspan="1" rowspan="1">
                     <para xml:id="para_5b53942c-fe8d-4331-ba69-7bfce262daf5"/>
                   </td>
                   <td align="left" colspan="1" rowspan="1">
                     <para xml:id="para_65078650-d0a6-4bd8-b5ba-1fa28e49d2fa"/>
                   </td>
                   <td align="left" colspan="1" rowspan="1">
                     <para xml:id="para_a9b8adf2-e62d-45d5-b464-57b2fb3cdedf"/>
                   </td>
                   <td align="left" colspan="1" rowspan="1">
                     <para xml:id="para_04cf772f-a3ac-47b1-ae2a-8772deec1884">Not used in AIM.</para>
                   </td>
                 </tr>
                 <tr valign="top">
                   <td align="left" colspan="1" rowspan="1">
                     <para xml:id="para_a0f066e8-0466-4474-bb28-b88ac2ca91ae">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126010">(126010, DCM, "Imaging Measurements")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_125007">(125007, DCM, "Measurement Group")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_128740">(128740, DCM, "Longitudinal Temporal Offset from Event")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_128741">(128741, DCM, "Longitudinal Temporal Event Type")</olink>
</para>
                   </td>
                   <td align="left" colspan="1" rowspan="1">
                     <para xml:id="para_853bfc5d-d031-4ab3-864d-ab91bb3491b7">CODE</para>
                   </td>
                   <td align="center" colspan="1" rowspan="1">
                     <para xml:id="para_7f44f13c-d242-4cbf-a9e9-f3f900df07d2">1</para>
                   </td>
                   <td align="center" colspan="1" rowspan="1">
                     <para xml:id="para_b58e65bc-0a86-417e-83ad-1218c653d4dc">U</para>
                   </td>
                   <td align="left" colspan="1" rowspan="1">
                     <para xml:id="para_dac3aaab-2d8e-4e5d-a5c9-a7b9482b6d07"/>
                   </td>
                   <td align="left" colspan="1" rowspan="1">
                     <para xml:id="para_d708d100-b24c-435a-8258-e3da983ea62f"/>
                   </td>
                   <td align="left" colspan="1" rowspan="1">
                     <para xml:id="para_46758425-cf5e-4b34-850b-511e553779a1"/>
                   </td>
                   <td align="left" colspan="1" rowspan="1">
                     <para xml:id="para_16d85ba4-a16d-4c88-9a94-defa3072444a"/>
                   </td>
                   <td align="left" colspan="1" rowspan="1">
                     <para xml:id="para_c99f64fd-b6a3-4ccd-979c-0f92af3c9582">Not used in AIM.</para>
                   </td>
                 </tr>
               </tbody>
            </table>
         </section>

          
                   <section label="TID 1410" status="3" xml:id="sect_Mapping_TID_1410">
            <title>Mapping of Planar ROI Measurements</title>
            <para xml:id="para_159eeb60-cbc3-41fd-b7b8-8a8e9667e2e1">This section describes the mapping of <olink targetdoc="PS3.16" targetptr="sect_TID_1410" xrefstyle="select: docname labelnumber quotedtitle"/>.</para>
            <table frame="box" label="TID 1410" rules="all" xml:id="table_Mapping_TID_1410">
            <caption>Mapping of Planar ROI Measurements</caption>
               <thead>
                  <tr valign="top">
                     <th align="center" colspan="1" rowspan="1">
                       <para xml:id="para_21927832-a0ff-4b84-8b63-ef8aa33dae0b">DICOM SR Path</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_6f6f355e-9363-4b6f-b4cb-e6a01843e274">DICOM VT</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_39c8976a-3ee8-4418-8854-1430bc198b5f">DICOM VM</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_5b4f0455-416b-44ce-97c0-548c5c2290bb">DICOM Usage Type</para>
                     </th>
                    <th align="center" colspan="1" rowspan="1">
                      <para xml:id="para_d309071f-8a06-4366-9e30-378079e79310">Generated Value</para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                      <para xml:id="para_96bbf08b-6220-4f26-8573-06f9047f5b6d">AIM Element or Attribute</para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                      <para xml:id="para_acd7799d-9a16-4f58-903f-017c6f9824f5">AIM Data Type</para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                      <para xml:id="para_5f946d18-1932-4a28-acd8-a072283a2a95">AIM Multiplicity</para>
                    </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_e6ab6bd6-aaac-4cfe-b945-14c9e81968c8">Comment</para>
                     </th>
                  </tr>
               </thead>
               <tbody>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_c987c086-f795-422c-8196-a1c0a642db5c">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126010">(126010, DCM, "Imaging Measurements")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_125007">(125007, DCM, "Measurement Group")</olink>
</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_e1ac0f8d-655b-4d4e-b8a3-be9c9ea6b405">CONTAINER</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_b1507fc0-c110-4e02-8344-170da7087890">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_ef1efb5a-843a-4695-aa3f-582894a6c9ce">M</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_6ff2503a-6dc2-4e18-93d5-442fdfc96c8d"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                         <para xml:id="para_643974c2-fe59-48e4-a5fe-bc0a67954671">Image​Annotation​​Collection/​image​Annotations/​Image​Annotation</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_e096ef54-ef6a-4e5d-9b4c-7106e30fb5d2"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_927f010b-91d3-456c-ae0d-e44939ba9b47"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                         <para xml:id="para_7ce69c15-286f-4bad-a42f-599f04c43968">The value of aim:unique​Identifier/@root is mapped to the Observation​UID Attribute of the CONTAINER.</para>
                         <para xml:id="para_887081f9-fbcc-4dac-bfdd-da9145a4fa02">The value of aim:date​Time/@root is mapped to the Observation​Date​Time​Attribute of the CONTAINER.</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_d6910845-ec0d-4921-8b68-4d83d2d01883">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126010">(126010, DCM, "Imaging Measurements")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_125007">(125007, DCM, "Measurement Group")</olink> &gt; (C67447, NCIt, "Activity Session")</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_4bdd6323-6d11-49f3-a011-405af705480b">TEXT</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_836fd5c5-eb79-4bb5-bf34-6f03061e691f">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_d279fc4f-8b83-457b-8a9e-d80ac2cff826">U</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_8472e81d-b0bd-4176-9d9d-3151db4c231e"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_2259e2bd-9ce9-473d-95f9-5ae2be636b48"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_c7890ccf-d988-4a5b-8ff9-709ccd9e298b"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_b5687265-5857-4e8e-9413-0615a1b91ff5"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                       <para xml:id="para_56b411b0-88da-4679-ba1a-696d056699a8">Not present in AIM.</para>
                       <para xml:id="para_361db87e-9118-4b1a-a339-a2ea38fa3782">Identifies the session during which the measurements were made. The NCI Thesaurus definition is "time, period, or term devoted to some activity".</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_2305126b-bd39-43e1-ab62-29cce8d5127d">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126010">(126010, DCM, "Imaging Measurements")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_125007">(125007, DCM, "Measurement Group")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_112039">(112039, DCM, "Tracking Identifier")</olink>
</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_0d31e476-607b-4b8f-a6bd-f87090f16b61">TEXT</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_81203299-364d-46ae-8d87-98e804afdd06">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_7567ad86-88ca-4479-8bf6-035412a74bd0">M</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_d0953888-d655-44b0-8ebc-a553ccf86250"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_0c633a1d-c949-41e7-96f5-0e66534c97f4">Image​Annotation​​Collection/​image​Annotations/​Image​Annotation/​name/​@value</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_168c49dc-3c3c-416a-b9f4-428c27a5c236">ST</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_4664a709-7121-4ac5-9880-8a09e45db0ae">1</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_6cbb6588-ac55-42ec-99e5-d4fe0eb2b1ba"/>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_4363b569-f374-4492-8611-4f0d125f843b">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126010">(126010, DCM, "Imaging Measurements")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_125007">(125007, DCM, "Measurement Group")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_112040">(112040, DCM, "Tracking Unique Identifier")</olink>
</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_84d21c0a-ffa1-4545-abcd-b91ac38e17fc">UIDREF</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_fea0b394-93c7-492e-92b2-1f998cd74dd9">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_8490f25f-7767-4f60-bb70-36450362388d">M</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_9781d039-d16f-4827-931b-88c828b4fd1f"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                         <para xml:id="para_18a021c3-8fab-4f64-a0c0-6512163cf0b1">Image​Annotation​​Collection/​image​Annotations/​Image​Annotation/​tracking​UniqueIdentifier/​@root</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_518f1466-a4bb-4b0e-a640-c1e500a3f46e">II</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_99c2ca49-db6e-4a4d-9b29-053af8c8faec">1</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                         <para xml:id="para_db966c10-8a4e-4da3-9c12-d66a8bb2a896">If tracking​Unique​Identifier is absent,
                             then Image​Annotation/​uniqueIdentifier may be used as a proxy for Tracking Unique Identifier,
                             but this does not allow longitudinal identification of the same lesion because Image​Annotation/​uniqueIdentifier must be unique to each AIM annotation file,
                             in which case only Image​Annotation/​name can be used to recognize commonality.</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_35bdd2a5-c236-4fb3-993d-2aad3ac89039">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126010">(126010, DCM, "Imaging Measurements")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_125007">(125007, DCM, "Measurement Group")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_121071">(121071, DCM, "Finding")</olink>
</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_f671866f-422c-4b39-82d5-b21931077013">CODE</para>
                  </td>
                    <td align="center" colspan="1" rowspan="1">
                      <para xml:id="para_319125e7-e19b-422b-8b72-720dd7d8d972">1</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_3b5a2025-292a-4ca1-9bf6-67a2361eb7cb">U</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_9e1ec8a2-236c-45d3-bdf2-7c5c851b7668"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_bd1a8682-333b-44fd-9364-906013a17872">Image​Annotation​​Collection/​image​Annotations/​Image​Annotation/​typeCode</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_de89a54d-f2bf-41df-9552-20b10dd19db0">CD</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_0ed0ba93-3087-45fd-b3cc-5aad54348030">1..n</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_974d63f5-02fa-4a59-88ab-8c196a655711">Only a single aim:typeCode value can be mapped.</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_841e6155-654d-4a73-97f1-19a79b6045b8">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126010">(126010, DCM, "Imaging Measurements")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_125007">(125007, DCM, "Measurement Group")</olink> &gt;</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_78132495-fb50-40cc-8468-814aa42ab0b4">INCLUDE</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_ac61aa73-f2d7-4b3b-b6a2-279d1bb53b1c">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_3cccdb6e-8259-42b0-854e-00befe0f9716">U</para>
                     </td>
                     <td align="left" colspan="4" rowspan="1">
                        <para xml:id="para_5cb5390b-ecb2-4e5a-97ed-57b457769e91">
<xref linkend="sect_Mapping_TID_1502" xrefstyle="select: labelnumber quotedtitle nopage"/>
</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                       <para xml:id="para_359948f7-cab2-4137-99eb-d5ed0c42b6e6">Not used in AIM.</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_e7be05c9-6c47-4d03-a6c2-c9348fa5db07">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126010">(126010, DCM, "Imaging Measurements")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_125007">(125007, DCM, "Measurement Group")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_111030">(111030, DCM, "Image Region")</olink>
</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_156039e3-f9ed-4f8d-a00f-4184899d323b">SCOORD</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_f778774d-3444-4708-83a7-1e621463f577">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_cfc45b06-60a4-494a-9836-05731010c7eb">MC</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_1cdd10d8-4a81-4954-b604-05fefa96d719"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                       <para xml:id="para_88156561-30e4-4c1e-858f-ea19e84c9e13">Image​Annotation​​Collection/​image​Annotations/​Image​Annotation/​​markup​Entity​Collection/​MarkupEntity/​twoDimension​SpatialCoordinate​​Collection/​TwoDimension​SpatialCoordinate</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_1426fb6d-5a62-42f2-9aff-9243da6802a5"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_6a8a75c6-5489-4921-be3f-70255bf98668"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_c7dbf47c-60aa-4ad5-900e-66806186e53d">A Graphic Type of MULTIPOINT is not permitted in the DICOM template.</para>
                         <para xml:id="para_148b4942-0363-4a51-8b78-ea92c2941bd1">One or more MarkupEntity instances within an Image​Annotation instance may be constrained to be associated with particular NUM measurement(s) in the included TID 1419 by a Calculation​Entity​References​Markup​Entity​Statement.</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_20506aad-cc2d-430a-94d6-ce378b54203a">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126010">(126010, DCM, "Imaging Measurements")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_125007">(125007, DCM, "Measurement Group")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_111030">(111030, DCM, "Image Region")</olink> &gt; IMAGE</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_0aaa799e-8f8d-46c7-ab7d-501fe8498901">IMAGE</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_a6bf6cd1-6d64-4e1e-9a65-25e15a77ed9a">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_855b07cf-e4dc-4b02-9ab1-c19e12299233">M</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_680d06ec-59b7-4895-a535-45b5e42ec679"/>
                     </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_b9dfa3d6-fc3b-44ce-a0df-85cc08d2f715">Image​Annotation​​Collection/​image​Annotations/​Image​Annotation/​​markup​Entity​Collection/​MarkupEntity/​imageReferenceUid/​@root, referencedFrameNumber/​@value</para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_284e4e07-6a96-498a-950c-b03b172ccfb5">II, INT</para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_9e2e9ec1-ba4b-4dad-ac4b-3f7dd7b7bc7c">1, 0..1</para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_1d5a809d-cd66-4d52-9a94-3b0855b6198c">The Referenced SOP Class UID is obtained from imageReference​Entity​Collection; see <xref linkend="table_A.8-5" xrefstyle="select: label"/>
</para>
                    </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_c261769f-99bb-4b39-960c-65b290bbae71">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126010">(126010, DCM, "Imaging Measurements")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_125007">(125007, DCM, "Measurement Group")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_121214">(121214, DCM, "Referenced Segmentation Frame")</olink>
</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_6330bc68-c81d-4557-9c89-2836639f1903">IMAGE</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_6026572a-2df4-428c-a8a5-e1c3291f802e">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_c45b729d-32c4-406a-9bcd-c37afd0b8fff">MC</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_eea0ee14-a572-47fe-927f-c4e34d6f5b49"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                       <para xml:id="para_e5ba4322-ec0c-4d50-9974-e828084a9013">Image​Annotation​​Collection/​image​Annotations/​Image​Annotation/​​segmentation​Entity​Collection/​SegmentationEntity</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_f4ae363b-cccf-4868-820a-689ae3e06962"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_63943c70-0a40-4118-a333-38ee2acb0a63"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_e183d2d9-d28c-4f41-8651-407601f73510">Reference shall be to a Segmentation Image, with a single value specified in Referenced Frame Number, and with a single value specified in Referenced Segment Number.</para>
                         <para xml:id="para_d987ff82-5dac-4e61-8767-ae2979fecea9">The value of aim:unique​Identifier/@root is mapped to the Observation​UID Attribute of the IMAGE.</para>
                         <para xml:id="para_e4a96743-8e06-47cc-97cd-b14bf3ef60b9">One or more Segmentation references within an Image​Annotation instance may be constrained to be associated with particular NUM measurement(s) in the included TID 1419 by a Calculation​Entity​References​Segmentation​Entity​Statement.</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_f2a0a34f-0e46-4d42-a348-27778f5fb396">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126010">(126010, DCM, "Imaging Measurements")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_125007">(125007, DCM, "Measurement Group")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_121233">(121233, DCM, "Source image for segmentation")</olink>
</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_abdd40d3-a260-4109-badc-67988ac7c3ab">IMAGE</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_82b1c353-8748-4c29-8b4f-0e8e78712915">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_38a98f6b-1df2-4167-aa3d-e6654f47a523">MC</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_4c6aec67-5b16-4266-90da-fe6756af7740"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_cb972b6d-9a23-4fee-89b8-09a98d4a3b43">Image​Annotation​​Collection/​image​Annotations/​Image​Annotation/​​segmentation​Entity​Collection/​SegmentationEntity/​referencedSopInstanceUid/​@root</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_3ce6f609-fa1b-49a1-81e9-f569db58de40"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_0feab670-d133-4909-bc25-670e07f5eaa3"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                       <para xml:id="para_e1fb4ba7-9416-4237-b3c6-aa82a75de1e5"/>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_6fc29904-e74d-49ec-bb2a-c93334f7e405">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126010">(126010, DCM, "Imaging Measurements")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_125007">(125007, DCM, "Measurement Group")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_121200">(121200, DCM, "Illustration of ROI")</olink>
</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_9de605d1-5cd8-4841-b884-91800016b252">IMAGE</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_77196d73-d22b-4ed8-ba4e-4a4ac97961cb">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_71544d8f-c3e1-4dc8-bb31-162f11bc7f3e">U</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_b7751aff-5157-4c5b-a74b-212f8513104f"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_32fc4558-16f2-4193-a9b2-dcf4b7cec2b1"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_164d93d0-9e67-4ea1-82b9-eebd66438189"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_5a263514-3eef-44c3-9a76-3b38f2defdf3"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                       <para xml:id="para_d5c959a9-bfe5-45d6-8cf5-5d46d416a1a9">Not used in AIM.</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_983cfa06-50d9-47c2-9ae7-2e3b85e6eaa2">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126010">(126010, DCM, "Imaging Measurements")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_125007">(125007, DCM, "Measurement Group")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126100">(126100, DCM, "Real World Value Map used for measurement")</olink>
</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_830a823d-6fe2-4d57-96c2-fa2cd9d030ec">COMPOSITE</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_693da8f4-7ab0-4883-b3ff-44185b23fb8c">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_6dbf231b-68dd-4f34-9dc0-d319292e6d35">U</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_8ea0c02e-bd74-404e-821c-345253aa380d"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_1c87da56-1b1b-4b4b-ac82-e3a8ae53c4cd"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_a261a280-2210-4c6f-85c1-06720e123a9a"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_87e128ec-8117-4d62-9dba-b6feeb0d1cc4"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_58899384-bc7e-44a0-97d5-6b4bf1e77f13">Not used in AIM.</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_698bc2bc-27fe-4c23-ba89-1d799b6f1865">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126010">(126010, DCM, "Imaging Measurements")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_125007">(125007, DCM, "Measurement Group")</olink> &gt; </para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_8012a639-c38a-4366-9115-d98eda141bc4">INCLUDE</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_b25a6bf7-1055-42cb-930e-6ad22a137daf">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_d17a68cd-da2a-445d-a291-96e0a44175f8">M</para>
                     </td>
                     <td align="left" colspan="4" rowspan="1">
                       <para xml:id="para_939c229f-5fe9-44d2-9615-9ffc0e6b3f9e">
<xref linkend="sect_Mapping_TID_1419" xrefstyle="select: labelnumber quotedtitle nopage"/>
</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_593ec1c9-144d-4958-9ab6-fe9ccc0ebafe"/>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_7918e78f-966d-418e-920b-8d925abc04b7">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126010">(126010, DCM, "Imaging Measurements")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_125007">(125007, DCM, "Measurement Group")</olink> &gt; $QualitativeEvaluations</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_d30992fd-5f9a-4bd3-9b42-2756886537c3">CODE</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_731ae518-d1f8-476d-8b97-3f522a178b2e">1-n</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_f0ddba63-ba7b-4d0f-a151-d4e3c94a0021">U</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_1cba5ae0-d6c0-4999-bddd-561cbcad16b5"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_62e2b07c-1b78-4186-b36a-10344beea651"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_7703c2c8-bd25-44af-8a81-054de7117012"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_c8a671fd-45cf-491c-b41b-3d2ad92140c2"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_732599a9-0f64-43d8-84fe-7ff378d236f9">Not used in AIM.</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_b093b700-1618-4eb4-97f1-100088f19abe">
                            <olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126010">(126010, DCM, "Imaging Measurements")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_125007">(125007, DCM, "Measurement Group")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_121106">(121106, DCM, "Comment")</olink>
</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_6107311d-e4d6-46c8-9f6d-b725df7013f2">TEXT</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_2d0a4c8a-846f-4d35-b23d-e7e340b309fb">1-n</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_738c5a9d-42c8-487b-8925-75999542cd2c">U</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_7e461f5d-9672-488a-a8c8-097e767badb5"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                         <para xml:id="para_121a8b21-d96a-4d0f-8c98-ccc8406a77f2">Image​Annotation​​Collection/​image​Annotations/​Image​Annotation/​comment/​@value</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_97b64285-004a-4a80-9387-de88189eefaf"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_0e543612-af31-47ca-8b53-8e134d574b04"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                         <para xml:id="para_bc490048-f65b-41d6-98f1-ffd3fc056f99">The parameter $Qualitative​Evaluations is not used in AIM,
                             but this TEXT content item is used to map the AIM comment as if it were a Qualitative Evaluation</para>
                     </td>
                  </tr>
               </tbody>
            </table>
         </section>
         <section label="TID 1411" status="3" xml:id="sect_Mapping_TID_1411">
            <title>Mapping of Volumetric ROI Measurements</title>
            <para xml:id="para_10a22ca8-48f4-4c06-a268-ffc1a967c175">This section describes the mapping of <olink targetdoc="PS3.16" targetptr="sect_TID_1411" xrefstyle="select: docname labelnumber quotedtitle"/>.</para>
            <table frame="box" label="TID 1411" rules="all" xml:id="table_Mapping_TID_1411">
            <caption>Mapping of Volumetric ROI Measurements</caption>
               <thead>
                  <tr valign="top">
                     <th align="center" colspan="1" rowspan="1">
                       <para xml:id="para_db7d7a78-b031-4209-84da-e95af2c34e0d">DICOM SR Path</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_35b448c9-3d2c-402e-889e-753c3641bfc6">DICOM VT</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_a75122c5-cc44-4b53-a5b3-858987254e20">DICOM VM</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_6c9fbc98-2a37-4bb4-80ee-ef475d592a1c">DICOM Usage Type</para>
                     </th>
                    <th align="center" colspan="1" rowspan="1">
                      <para xml:id="para_d4132ef5-9c6c-418c-ad74-ab82ead8ae06">Generated Value</para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                      <para xml:id="para_cab11705-7620-4f66-9669-f29715842810">AIM Element or Attribute</para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                      <para xml:id="para_d6969bac-e67e-4c92-8860-9301a4c17d8a">AIM Data Type</para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                      <para xml:id="para_a552f7b5-d028-4c8c-a31b-063352d4d156">AIM Multiplicity</para>
                    </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_2d2a259e-657a-4d53-b871-057194dc6e05">Comment</para>
                     </th>
                  </tr>
               </thead>
               <tbody>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_e094d380-67f8-46bc-bc2a-f4296e9d7fb6">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126010">(126010, DCM, "Imaging Measurements")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_125007">(125007, DCM, "Measurement Group")</olink>
</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_85d92b96-1bc9-4d00-a8d1-bf9e1a37e52d">CONTAINER</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_d3cf1a6e-1e67-450f-8a09-edc6cb662295">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_9d34caf7-3719-4395-b397-f416f10d3ca1">M</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_951d3924-7d69-422a-8d5a-166b2c8af781"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                         <para xml:id="para_4b25d15f-a68b-4a66-997c-301a6be18a93">Image​Annotation​​Collection/​image​Annotations/​Image​Annotation</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_9739d74a-5e71-4d62-aa3a-1109c3641dc4"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_2e78fa7c-b710-4243-895a-4c6289605b3f"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                         <para xml:id="para_184bb699-31b7-4bf5-a47f-010e0236cc48">The value of aim:unique​Identifier/@root is mapped to the Observation​UID Attribute of the CONTAINER.</para>
                         <para xml:id="para_8560b27a-cd48-4ace-9820-16afcbd9944f">The value of aim:date​Time/@root is mapped to the Observation​Date​Time​Attribute of the CONTAINER.</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_5270a9ba-52f5-4ca6-bba8-3f14c7bbd9c3">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126010">(126010, DCM, "Imaging Measurements")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_125007">(125007, DCM, "Measurement Group")</olink> &gt; (C67447, NCIt, "Activity Session")</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_9cb09402-8061-4b6c-a0d8-c9131c5f6194">TEXT</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_998bd93c-4665-43dc-bd66-e1299c21b3c2">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_8e39b8ef-1467-4ca6-8218-973e86cde260">U</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_0910409b-50fd-40c9-ab90-fcb03de30211"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_c812f642-7bad-4874-be4e-3a7d40da8dee"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_99c90b4c-b8a4-45ba-a25f-a11fc1f9bdc1"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_dd6e2d62-6ac4-4e17-b8d9-d7c2ee66ce44"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                       <para xml:id="para_e3cf4cb3-2d05-49c7-b61f-ac16f2654c8c">Not present in AIM.</para>
                       <para xml:id="para_ae82de45-8a13-4c32-87be-9ea72e826b80">Identifies the session during which the measurements were made. The NCI Thesaurus definition is "time, period, or term devoted to some activity".</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_ef240c96-ff30-403f-ad3a-bf1d52a23082">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126010">(126010, DCM, "Imaging Measurements")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_125007">(125007, DCM, "Measurement Group")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_112039">(112039, DCM, "Tracking Identifier")</olink>
</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_05a71610-c371-4b73-9947-0feb9fee5d4f">TEXT</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_dd4bb872-8358-4fc0-942c-526756ad2421">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_f792d623-2e70-4559-a090-565a998d4c0f">M</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_abe37899-e46f-48d5-bc20-2b8fc79ac415"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_08af278c-4ade-4b24-b573-d1be0ec45336">Image​Annotation​​Collection/​image​Annotations/​Image​Annotation/​name/​@value</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_5b370822-99d9-426d-8a9a-1ccbcade0542">ST</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_2f268b4e-1e72-49aa-a4e8-ae54079d9fbf">1</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_5596ec1e-42b9-440f-8b29-43321485d6b0"/>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_a57d8566-0a67-4962-a11c-d4a191dcf06d">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126010">(126010, DCM, "Imaging Measurements")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_125007">(125007, DCM, "Measurement Group")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_112040">(112040, DCM, "Tracking Unique Identifier")</olink>
</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_006ed1cf-95ba-4a81-9233-c0ec4c1671c2">UIDREF</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_749a7996-6a75-4bee-a50c-6d49a68fd24e">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_1e96a817-2aa7-44a7-9727-734b66b47e73">M</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_a7ebf523-fdd5-47fd-8e71-774485d63edf"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                         <para xml:id="para_3e7905cb-51de-4399-ac76-9ac990dcd7ed">Image​Annotation​​Collection/​image​Annotations/​Image​Annotation/​tracking​UniqueIdentifier/​@root</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_8b03176c-690a-4c65-b9e8-62751d53f61e">II</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_9c1fff4b-ea69-409d-bdba-593298214d7c">1</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                         <para xml:id="para_611174bd-0a5d-4c2d-8339-a63373bd6f4e">If tracking​Unique​Identifier is absent,
                             then Image​Annotation/​uniqueIdentifier may be used as a proxy for Tracking Unique Identifier,
                             but this does not allow longitudinal identification of the same lesion because Image​Annotation/​uniqueIdentifier must be unique to each AIM annotation file,
                             in which case only Image​Annotation/​name can be used to recognize commonality.</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_78e1dc11-170e-4c57-8faf-6c4f3731ea11">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126010">(126010, DCM, "Imaging Measurements")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_125007">(125007, DCM, "Measurement Group")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_121071">(121071, DCM, "Finding")</olink>
</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_2d685f01-79d4-47db-a487-bc2b9117bb6e">CODE</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_291c27ad-cc9f-489b-8661-baf0fa4ddd27">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_6f63c41c-5f1a-4155-99f3-5fd4f41af927">U</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_3845967d-9eb5-4803-9885-85e5593b0e4a"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_dfada371-ac05-4c1e-a6ea-5c9f348c0675">Image​Annotation​​Collection/​image​Annotations/​Image​Annotation/​typeCode</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_5916d7ba-5c24-4bc6-b1a6-f07dee4c2635">CD</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_b767b365-5fe2-4f60-a2bb-ce4229d96173">1..n</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_d1931ecf-1caa-45a4-9e53-82d510157982">Only a single aim:typeCode value can be mapped.</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_ceb74c1c-4cd2-4838-8680-ff9bc4acb75b">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126010">(126010, DCM, "Imaging Measurements")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_125007">(125007, DCM, "Measurement Group")</olink> &gt; </para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_c1e2320c-15a3-401a-b6ce-3ff031387586">INCLUDE</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_86f945c2-bb22-4c5b-a263-090b7599927d">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_20f3f9e9-c924-4ca8-b860-de466d55f926">U</para>
                     </td>
                     <td align="left" colspan="4" rowspan="1">
                        <para xml:id="para_bfe2e857-e72d-4743-a77d-101a8d438014">
<xref linkend="sect_Mapping_TID_1502" xrefstyle="select: labelnumber quotedtitle nopage"/>
</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                       <para xml:id="para_c5cdce14-c9e8-4350-b785-a5db1cf003e7">Not used in AIM.</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_1eb5b7c7-448d-477d-a5e6-76a6fe4f96b9">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126010">(126010, DCM, "Imaging Measurements")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_125007">(125007, DCM, "Measurement Group")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_111030">(111030, DCM, "Image Region")</olink>
</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_df75b533-ba06-4232-b8f1-d8638b4d6623">SCOORD</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_58cc967c-5008-4663-a659-27aa6d55bf7b">1-n</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_c7b72938-84a5-4201-be83-8216cfe73633">MC</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_d36da637-d1f7-4463-bc63-cc76a974a782"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                       <para xml:id="para_3cf736c2-51e6-423b-abea-883fd593c9fd">Image​Annotation​​Collection/​image​Annotations/​Image​Annotation/​​markup​Entity​Collection/​MarkupEntity/​twoDimension​SpatialCoordinate​​Collection/​TwoDimension​SpatialCoordinate</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_8c0736b5-687d-4b8d-bce5-b3bf682e1fb5">REAL</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_95a09b00-b8a2-4fd2-8b7b-7dc7be33b247">1..n</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_1519b6a2-5f97-41c6-ac43-a18dbe32695a">A Graphic Type of MULTIPOINT is not permitted in the DICOM template.</para>
                         <para xml:id="para_1124a7b0-f3c5-48d0-b977-710ff0697f70">The value of aim:​Markup​Entity/​aim:​unique​Identifier/​@root is mapped to the Observation​UID Attribute of the SCOORD.</para>
                         <para xml:id="para_35cb3bd7-6636-4fa6-90e8-80f9363095c4">One or more MarkupEntity instances within an Image​Annotation instance may be constrained to be associated with particular NUM measurement(s) in the included TID 1419 by a Calculation​Entity​References​Markup​Entity​Statement.</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_a13360f0-f562-4875-a9e8-603c5d277586">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126010">(126010, DCM, "Imaging Measurements")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_125007">(125007, DCM, "Measurement Group")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_111030">(111030, DCM, "Image Region")</olink> &gt; IMAGE</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_6f7e11c6-e0e9-4d3b-9927-66ccd54ddf52">IMAGE</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_1180b58e-8a59-4e3a-809a-acbf431df9bb">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_004df329-43ed-4e4a-b654-a05b6c508963">M</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_05e0a642-4cad-4f73-948f-e07ccb0aacec"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                       <para xml:id="para_cb8379a9-6b9c-4e3c-85d7-99e5bc574fe7">Image​Annotation​​Collection/​image​Annotations/​Image​Annotation/​​markup​Entity​Collection/​MarkupEntity/​imageReferenceUid/​@root, referencedFrameNumber/​@value</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_c45cc2f7-393e-461c-831b-71c1a5a9af0b">II, INT</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_0e076300-22c6-4e2d-ba59-b964786e3ed9">1, 0..1</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                       <para xml:id="para_595626ec-3a9b-4b0d-8463-0d9c9cebd5b6">The Referenced SOP Class UID is obtained from imageReference​Entity​Collection; see <xref linkend="table_A.8-5" xrefstyle="select: label"/>
</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_d3c1045f-69b7-4951-a6e0-db205d2ece92">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126010">(126010, DCM, "Imaging Measurements")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_125007">(125007, DCM, "Measurement Group")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_121191">(121191, DCM, "Referenced Segment")</olink>
</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_66e8cae8-379e-41c1-a37a-98e03316e5f1">IMAGE</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_428e53cd-5c9b-4c94-8f31-b99ba612eff4">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_ac6f30a6-00c0-4f42-b563-eb7907f4118b">MC</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_67872521-4f20-4438-b2d8-fb7d034f621f"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_080cc582-7f4e-4fc7-a6dd-e0abbe580010">Image​Annotation​​Collection/​image​Annotations/​Image​Annotation/​​segmentation​Entity​Collection/​SegmentationEntity</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_f4ab3819-8093-45d2-9de9-f34fe5cdba83">INT</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_6caba6f1-3936-4827-86b4-8a7bac32a0b0">0..1</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_dc3b935b-c51a-47b0-a2ff-061622ca684c">Reference shall be to a Segmentation Image or Surface Segmentation object, with a single value specified in Referenced Segment Number.</para>
                         <para xml:id="para_03fe23f8-3172-4587-8231-cc45acb0aa4a">One or more Segmentation references within an Image​Annotation instance may be constrained to be associated with particular NUM measurement(s) in the included TID 1419 by a Calculation​Entity​References​Segmentation​Entity​Statement</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_f68d6c8e-ae3f-46af-8289-8e3ec34f3bf2">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126010">(126010, DCM, "Imaging Measurements")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_125007">(125007, DCM, "Measurement Group")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_121231">(121231, DCM, "Volume Surface")</olink>
</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_aa656a67-7386-47f9-be90-7e6735e51be5">SCOORD3D</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_e7613514-238e-4268-908f-c538f4e4fe71">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_b710c00b-b06b-4784-83a6-329eeea28a26">MC</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_bececb39-3ac1-4459-94f4-ad5465b08d23"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_1edb8186-b562-43c4-96ad-c3fed772e0b9"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_779a04e8-0521-4795-a6e5-7dc7a3f32bad"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_df380b9d-0d0f-426f-9d36-cd15800ccd22"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_ebfbc1f9-fadf-46d0-a491-2407ccab90e2">Not used in AIM.</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_31a12578-f60f-41c7-a4d5-0acd514ab809">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126010">(126010, DCM, "Imaging Measurements")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_125007">(125007, DCM, "Measurement Group")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_121233">(121233, DCM, "Source image for segmentation")</olink>
</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_e671fc9a-2eaf-48d5-a9f3-51afdf0365c7">IMAGE</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_0113e8ad-4fdb-4673-98ca-99ab09120af9">1-n</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_307a199a-3385-42d9-afad-8d5bb10af5d9">MC</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_9317fb0e-c0d7-4f75-9578-aa94804cdf8b"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_48e3147e-035a-4f06-8862-fab52595f3a9">Image​Annotation​​Collection/​image​Annotations/​Image​Annotation/​​segmentation​Entity​Collection/​SegmentationEntity/​referencedSopInstanceUid/​@root</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_ba09febb-08c1-42b3-bfce-20f02daaba2e"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_70671862-8e9e-43b2-a3e0-9c362cb85fa9"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_293e1645-7ac0-4a85-aced-bea93b700882"/>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_334cbf51-a6b5-4f79-bd4b-bb1ae0a65fb1">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126010">(126010, DCM, "Imaging Measurements")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_125007">(125007, DCM, "Measurement Group")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_121232">(121232, DCM, "Source series for segmentation")</olink>
</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_8bacc5e4-ddb8-4d07-9c0f-857ea7ab65fa">UIDREF</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_562f6f9b-8fcf-497e-9847-c99cf57bef45">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_fd805d2b-2b56-4db4-bf8e-d4240ab3ecad">MC</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_fb5b6559-1ce9-470d-8917-c42f577dafd9"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_61061b62-7d30-4f2b-8880-1e9500b1cf88"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_547f8c81-8ae5-4f00-92ae-c474ccf70d4a"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_38e9d0ad-cfb1-47e2-b9c0-82bedc70f6cf"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_f212d77e-542e-4e23-8529-8247a8132cef">Not used in AIM</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_47d92ec9-1122-4836-9be3-7be2c0fc195b">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126010">(126010, DCM, "Imaging Measurements")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_125007">(125007, DCM, "Measurement Group")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_121200">(121200, DCM, "Illustration of ROI")</olink>
</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_8668cccf-ac90-4c7f-a05b-14b8f6827f60">IMAGE</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_b8b6852d-809a-44ac-a66e-bec61437b1b3">1-n</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_79ee6616-9f38-45e1-b8df-223e6f89f36d">U</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_66d81de2-4ec4-402c-912a-8bebd8737f3e"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_2d167cbb-8da6-429b-8590-4c3b5df7aed8"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_27af4826-9efe-4d2c-8151-d1e5c4bbfc94"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_ba1e1a96-91eb-4bf6-99cd-2264749ba9fb"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_f771b1d4-9ad0-47a8-b202-56205ad2122b">Not used in AIM.</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_31ed261a-41fa-4687-b2f8-f93bf84a3a7a">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126010">(126010, DCM, "Imaging Measurements")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_125007">(125007, DCM, "Measurement Group")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126100">(126100, DCM, "Real World Value Map used for measurement")</olink>
</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_b192d89c-32ed-4cdc-9b9e-ef0b446907c7">COMPOSITE</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_3788cdbb-9895-4c41-bfc4-d02986874139">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_f5ad85b4-e59b-4e1e-a4f9-8da2eef17ae0">U</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_c80ce927-ece2-4a24-8f1a-a2331451534e"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_0c151326-d7d8-430c-a56a-50ab7a810f22"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_29ee425c-1510-43b2-972b-e49192856f1c"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_f5539a00-d5e9-4552-992f-95d0ac8057e2"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_ac4e9f9a-5b59-44d6-8dc9-a98de2299281">Not used in AIM.</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_db6a491e-74a0-491d-b2fc-5e63050c5ec7">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126010">(126010, DCM, "Imaging Measurements")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_125007">(125007, DCM, "Measurement Group")</olink> &gt; </para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_47d8e379-96c6-4926-9d36-a689f1af7a2a">INCLUDE</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_a46d9333-7136-4d71-863b-55d0d3952108">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_01a68bb8-a34e-49de-af43-526bf6d7f8d8">M</para>
                     </td>
                     <td align="left" colspan="4" rowspan="1">
                        <para xml:id="para_2bcd245b-fccd-4fa2-935d-9a072cc77e08">
<xref linkend="sect_Mapping_TID_1419" xrefstyle="select: labelnumber quotedtitle nopage"/>
                    </para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_10c52253-f4f4-4500-8a3c-96924e78179e"/>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_d669b63b-1341-4b3a-a296-84de777495d7">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126010">(126010, DCM, "Imaging Measurements")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_125007">(125007, DCM, "Measurement Group")</olink> &gt; $QualitativeEvaluations</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_f246bd46-1b8c-4d90-9059-6a289caac441">CODE</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_357c563f-f824-4b53-9507-5b39fd8b414e">1-n</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_432362a8-dea5-4fce-8e89-b8d46061c57e">U</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_8ab5ce04-ade2-4afd-be61-5dd8f8736299"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_70dda449-fce5-4faf-a75f-1342ac1070be">Image​Annotation​​Collection/​image​Annotations/​Image​Annotation/​imaging​Observation​Entity​Collection/​Imaging​Observation​Entity/​imaging​Observation​Characteristic​Collection/​Imaging​Observation​Characteristic​/​questionType​Code,type​Code</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_12b38dc1-cf02-43d7-b369-83eeb06f35af">CD, CD</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_e815cabf-4102-460a-b2ed-fd0d3c86a2f3">1, 0..1</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_44ee3d10-771a-4168-868a-94bca2ea2cea"/>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_461d4a03-525f-4f53-b4d2-cd215e0cf716">
                            <olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126010">(126010, DCM, "Imaging Measurements")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_125007">(125007, DCM, "Measurement Group")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_121106">(121106, DCM, "Comment")</olink>
</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_9b2557b2-6ff1-4f83-b62e-6b2b26fe9118">TEXT</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_bcf4b48a-f6a5-4df3-b0cc-c5b31814c0c3">1-n</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_f3d16e3a-68f7-4622-b208-902ea6fc3894">U</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_4a00fc8e-1043-4c37-8b3f-6558ac4ced71"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                         <para xml:id="para_6cd443b8-6e5f-49de-b8b3-c222f214848b">Image​Annotation​​Collection/​image​Annotations/​Image​Annotation/​comment/​@value</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_833f6403-525c-449e-9306-c1f1b06c8338"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_979841f8-b276-4275-a258-5a025b2f7209"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                         <para xml:id="para_afc72fe1-11d4-4903-98a3-3a1ab8a3cdd2">The parameter $Qualitative​Evaluations is not used in AIM,
                             but this TEXT content item is used to map the AIM comment as if it were a Qualitative Evaluation.</para>
                     </td>
                  </tr>
               </tbody>
            </table>
         </section>
         <section label="TID 1419" status="3" xml:id="sect_Mapping_TID_1419">
            <title>Mapping of ROI Measurements</title>
            <para xml:id="para_f8d69ac5-cca8-4e1e-99d6-9ba5353c57c2">This section describes the mapping of <olink targetdoc="PS3.16" targetptr="sect_TID_1419" xrefstyle="select: docname labelnumber quotedtitle"/>.</para>
            <table frame="box" label="TID 1419" rules="all" xml:id="table_Mapping_TID_1419">
            <caption>Mapping of ROI Measurements</caption>
               <thead>
                  <tr valign="top">
                     <th align="center" colspan="1" rowspan="1">
                       <para xml:id="para_7fb3708a-0871-432d-ad92-8a37365a75e3">DICOM SR Path</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_98309402-6b33-4a03-a74d-c368073a78ca">DICOM VT</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_18a49caa-cdf7-4671-8734-426f14ab12c3">DICOM VM</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_fffe89bb-9db2-450a-9d47-e4e4a3909114">DICOM Usage Type</para>
                     </th>
                    <th align="center" colspan="1" rowspan="1">
                      <para xml:id="para_ff41a835-3f43-47d6-9bc5-f0cd967aeec9">Generated Value</para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                      <para xml:id="para_0d26a397-11bd-4c37-bf18-5dc19f28da45">AIM Element or Attribute</para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                      <para xml:id="para_880a3b67-9448-40a6-8135-27e824a83bde">AIM Data Type</para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                      <para xml:id="para_9c0b13d9-d94e-414a-88e7-1b07efc69b1f">AIM Multiplicity</para>
                    </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_490e60b2-b450-45ac-b0ab-c091bddcc23b">Comment</para>
                     </th>
                  </tr>
               </thead>
               <tbody>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_e5f2772b-d4b0-4176-8009-429881e04281">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126010">(126010, DCM, "Imaging Measurements")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_125007">(125007, DCM, "Measurement Group")</olink> &gt; <link xl:href="http://snomed.info/id/370129005" xl:show="new">(370129005, SCT, "Measurement Method")</link>
</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_d9a6c762-6fda-4e27-a312-3892f9d1d44c">CODE</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_4194239c-a7da-45a3-8c15-de5f13d763be">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_dd223dd3-2e0c-486c-a7bc-a365f38a95c6">U</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_28be1fa4-3126-43d0-8821-3feeb90e772b"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_8e50fcef-86ac-43c5-9f1e-3fdb78cf2c08"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_a5d06a48-ab24-4ef1-8198-552eb3e6f7b4"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_fc515059-14b7-46d2-8e51-6387f20b6f3d"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_3ab1d8bd-ef51-4ef3-94b7-77a41b7e78f1">Not encoded in AIM at this level.</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_4aaaff04-d258-4a66-aaee-1fd67cf5a97a">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126010">(126010, DCM, "Imaging Measurements")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_125007">(125007, DCM, "Measurement Group")</olink> &gt; <link xl:href="http://snomed.info/id/363698007" xl:show="new">(363698007, SCT, "Finding Site")</link>
</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_c961abb0-8c34-462a-80d2-c12391579f7b">CODE</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_a70d4206-482c-48c3-8a8c-d3a695a16c14">1-n</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_c43a832c-ba10-44fd-851e-7f5ef719fbf2">U</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_2f7ada30-eda1-4c0a-80eb-7027187bf441"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_53ea55b7-55f0-49c9-90f7-3ee24d5fa13f">Image​Annotation​​Collection/​image​Annotations/​Image​Annotation/​imagingPhysical​Entity​Collection/​ImagingPhysicalEntity​[label/​@value='Location' or label/​@value='Lobar Location' or label/​@value='Segmental Location' or label/​@value='Organ Type']/typeCode</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_89db9a88-a5c4-43f0-a0a5-013f47c1a3af">CD</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_2213501b-3c77-4977-8496-6b57fec5ecd7">1..n</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_faab830d-e75d-4045-b4dc-5dbb01809a2f">The finding site is factored out since it is common to all measurements in this group.</para>
                    <para xml:id="para_0071a89d-3939-45a8-bda4-1dbda6fffcc6">If a paired structure in AIM, this entry will pre-coordinate the laterality with the site.</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_600926cf-b262-42eb-a73f-c911db75e02d">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126010">(126010, DCM, "Imaging Measurements")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_125007">(125007, DCM, "Measurement Group")</olink> &gt; <link xl:href="http://snomed.info/id/363698007" xl:show="new">(363698007, SCT, "Finding Site")</link> &gt; <link xl:href="http://snomed.info/id/272741003" xl:show="new">(272741003, SCT, "Laterality")</link>
</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_0b4903e4-a668-4c23-8d2c-c5e246f634ec">CODE</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_01bbcb58-3c5f-43a7-8aeb-db60b4aad915">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_9847a5e0-156a-4a56-bad2-31c79053ea2b">U</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_cb222c00-a201-4d28-b305-4f47255fb50c"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_a6148855-998f-4e49-b3ae-70623b941ce1"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_7e8314c5-55f5-432f-8a17-e65fe6e48e6f"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_6d441d41-0c65-4c26-8db9-36bb65a1dbee"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_bdfed09e-8fde-4f43-a15d-2c20612f0c3e">Not used since AIM does not have a mechanism for post-coordinating the location.</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_500139ae-ad36-489f-b70e-538b85313ff0">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126010">(126010, DCM, "Imaging Measurements")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_125007">(125007, DCM, "Measurement Group")</olink> &gt; <link xl:href="http://snomed.info/id/363698007" xl:show="new">(363698007, SCT, "Finding Site")</link> &gt; <link xl:href="http://snomed.info/id/106233006" xl:show="new">(106233006, SCT, "Topographical modifier")</link>
</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_523b3ebf-db7c-41dc-890e-7e3988897e60">CODE</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_b6475b0c-7650-4584-af17-d4f831c0624d">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_358323f1-71e3-428f-9695-ca5b96f5c90d">U</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_de738f1b-0d3c-49c8-8673-2394698f5164"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_b133a106-5552-4553-b9c0-e40bf0ac6098"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_e35ac2b4-9eec-44fb-869a-c025277c1f6b"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                       <para xml:id="para_fde1be8a-8340-485e-bb8f-d50542e25ec4"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_66a3253d-daa2-463d-abce-bd777561d8fe">Not used since AIM does not have a mechanism for post-coordinating the location.</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_4ec43574-db5e-42fa-8693-0b112778a479">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126010">(126010, DCM, "Imaging Measurements")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_125007">(125007, DCM, "Measurement Group")</olink> &gt; $Measurement</para>
                     </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_47c4e516-60df-4362-99ac-099676437471">NUM</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_5e2b1f65-da01-4dc4-b130-31d5c9a2fb66">1-n</para>
                  </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_58012abb-31e1-488f-8cee-d2f4c23997f6">M</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_ccd5ea59-84d0-4f6d-a2d9-5ce38f563c75"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_17c2e059-71bb-4895-b007-f5b162cb333e">NAME = Image​Annotation​​Collection/​image​Annotations/​Image​Annotation/​calculation​Entity​Collection/​CalculationEntity/​typeCode[1]</para>
                    <para xml:id="para_a33c25c7-5240-4396-b635-257c20127ccf">VALUE = Image​Annotation​​Collection/​image​Annotations/​Image​Annotation/​calculation​Entity​Collection/​CalculationEntity/​calculationResult​Collection/​CalculationResult/​​@value, calculationData​Collection/​CalculationData/​@value</para>
                        <para xml:id="para_57a4ba00-998b-4bda-be73-0e4fcd4ab972">UNITS = Image​Annotation​​Collection/​image​Annotations/​Image​Annotation/​calculation​Entity​Collection/​CalculationEntity/​calculationResult​Collection/​CalculationResult/​unitOfMeasure</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_b55af109-cdbf-4513-89b3-9ea49513cec2">CD, ST, CD</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_39de319d-5729-46ff-a69c-ff1c10b9fa98">1..n, 0..n, 1</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_5c8dab78-0df8-4a9d-9779-399ac71c0496">The first typeCode entry is assumed to be the primary concept. Other typeCode entries may be considered as modifiers.</para>
                       <para xml:id="para_8c35a732-3f30-411e-88aa-a7742cfa53fc">Value may be found in either Compact​Calculation​Result (i.e., value child of CalculationResult) or first value of Extended​Calculation​Result (i.e., nested within calculation​Result​Collection).</para>
                       <para xml:id="para_87e523de-b6b1-48c0-a922-ad0ef4c6ca6e">Only mapping of a single value from Extended​Calculation​Result is supported.</para>
                    <para xml:id="para_29c6bf31-0761-41ef-8428-37c4ef3f6fdf">If no measurement is present in AIM (0..n) then do not include the template in the first place.</para>
                         <para xml:id="para_95f816dd-5237-413d-9058-e6eaeb84549c">The value of aim:unique​Identifier/@root is mapped to the Observation​UID Attribute of the NUM Content Item.</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_b2ea64c7-6887-4c0e-b2f9-31a2e607aeec">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126010">(126010, DCM, "Imaging Measurements")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_125007">(125007, DCM, "Measurement Group")</olink> &gt; $Measurement &gt; $ModType</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_61f009cf-1bce-4b6e-bddc-3b7822932f7f">CODE</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_6268b647-af7e-43a5-9a02-65cc5ed156c7">1-n</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_142cfa61-e807-4800-abdd-5726b1d0141b">U</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_24822154-ef07-43cb-a496-d1a7748d98cf"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_bf836de6-dc8c-4f83-9ebd-d7f706f527bb">Image​Annotation​​Collection/​image​Annotations/​Image​Annotation/​calculation​Entity​Collection/​CalculationEntity/​typeCode​[position() != 1]</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_2aff4918-7e25-4c0f-badd-b294560c9f5a">CD</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_8625db48-5b08-4e86-950d-e7881cc56536">1..n</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_f539b0b8-aa17-4c34-b36d-4cee60841644">The first typeCode entry is assumed to be the primary concept. This row can be used if successive typeCode entries are recognized, and not a method or derivation, and a name-value pair can be constructed from the typeCode entry (which is a single code).</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_3acf0714-aff5-4541-aea0-d2aff908380d">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126010">(126010, DCM, "Imaging Measurements")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_125007">(125007, DCM, "Measurement Group")</olink> &gt; $Measurement &gt; <link xl:href="http://snomed.info/id/370129005" xl:show="new">(370129005, SCT, "Measurement Method")</link>
</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_ee499e79-d6cf-4329-b108-6def764eb628">CODE</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_1055c78d-efc9-431e-8478-3058a0574cf3">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_636f32b6-f04b-44ea-9242-19dbd64257ae">U</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_17fe1ce1-59ca-4eec-abbb-43ffb7fe21c8"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_37c6335a-2d8b-4020-8857-7de9fafdf008">Image​Annotation​​Collection/​image​Annotations/​Image​Annotation/​calculation​Entity​Collection/​CalculationEntity/​typeCode​[position() != 1]</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_f55c18e6-8d09-4050-a8c7-90a1417b48a2">CD</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_7789b06b-9d74-487d-ad14-8bcf663ff010">1..n</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                       <para xml:id="para_9e73fd92-3932-4295-9e8e-dffbc39cf383">The first typeCode entry is assumed to be the primary concept. Other typeCode entries may be considered as modifiers, but there is no standard order, so recognition as a "method" depends on recognition of specific codes.</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_88a395b9-c2c3-49da-95dd-4bdf3bd94b13">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126010">(126010, DCM, "Imaging Measurements")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_125007">(125007, DCM, "Measurement Group")</olink> &gt; $Measurement &gt; <olink targetdoc="PS3.16" targetptr="DCM_121401">(121401, DCM, "Derivation")</olink>
</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_31745f74-1c3a-4bc7-aa38-f5243fb7d76b">CODE</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_11ef53d3-4a31-4b9a-97ec-6bdcbbe51922">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_aad96f78-c9cc-41e4-a161-334f98d8ebe1">U</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_d5998ab2-67a1-4427-90ca-4aebb37ec4c5"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_1af7b057-e4f9-4f5f-a852-542617380dd6">Image​Annotation​​Collection/​image​Annotations/​Image​Annotation/​calculation​Entity​Collection/​CalculationEntity/​typeCode​[position() != 1]</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_bd1abe3c-ca76-43d0-96bc-30a351f1734e">CD</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_7a4a5153-fdaa-4499-a12a-88f8f88ba5c9">1..n</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                       <para xml:id="para_9eab8407-1d0e-456f-8203-e7c5710b5cc2">The first typeCode entry is assumed to be the primary concept. Other typeCode entries may be considered as modifiers, but there is no standard order, so recognition as a "derivation" depends on recognition of specific codes.</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_12eb8b6b-30de-488a-9e5e-d2470dfbceed">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126010">(126010, DCM, "Imaging Measurements")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_125007">(125007, DCM, "Measurement Group")</olink> &gt; $Measurement &gt; <link xl:href="http://snomed.info/id/363698007" xl:show="new">(363698007, SCT, "Finding Site")</link>
</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_e5ebdd46-00f3-4e58-8dde-074731cea545">CODE</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_dc9dddcf-f458-4ed1-b00a-48064711dac0">1-n</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_ff8c5f35-f2a2-48d1-aed0-e2800cb454fe">U</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_eea85213-2f90-4f5f-b8c0-94cce4200d98"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_1a1df84f-3f23-490a-85c6-b7658e29ff11"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_39da9c90-6f20-45cb-9ea0-d050a32546dc"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_06a706a7-8dc7-4e0b-aa0a-8586dc3149ee"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_01ee20ca-59ee-480c-a950-b5a37d23dff7">Not used, since is sent at measurement group level since common to all measurements in annotation.</para>
                    <para xml:id="para_0df95aa4-0837-4cf3-8f6c-78fbb4bd1223">If a paired structure in AIM, this entry will pre-coordinate the laterality with the site.</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_b28ca836-8f7d-4700-aef9-109b0f2b71cf">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126010">(126010, DCM, "Imaging Measurements")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_125007">(125007, DCM, "Measurement Group")</olink> &gt; $Measurement &gt; <link xl:href="http://snomed.info/id/363698007" xl:show="new">(363698007, SCT, "Finding Site")</link> &gt; <link xl:href="http://snomed.info/id/272741003" xl:show="new">(272741003, SCT, "Laterality")</link>
</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_f42d9d25-fecb-4db2-8447-ec059f7d2f53">CODE</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_63e43127-b79a-4c47-b86f-9a20b53abc68">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_fa0affec-21f6-46ca-88f8-a7a6f82b6f69">U</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_1242551f-2f7f-4786-8a6d-780a60420945"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                       <para xml:id="para_ebe2367c-39b2-4136-b80c-ac8ff04824d3"> </para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_21625b05-a1a9-49c1-916b-f9634ec65b3e"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_6c2bc7c9-146f-4f66-93e9-c5b47cf4e5c8"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_e9f17317-29ad-408e-93dc-95dfb9aa8f5a">
<olink targetdoc="PS3.16" targetptr="sect_CID_244" xrefstyle="select: labelnumber quotedtitle"/>
</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_916f6ce3-bf17-44e5-8c9e-459ba31663f2">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126010">(126010, DCM, "Imaging Measurements")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_125007">(125007, DCM, "Measurement Group")</olink> &gt; $Measurement &gt; <link xl:href="http://snomed.info/id/363698007" xl:show="new">(363698007, SCT, "Finding Site")</link> &gt; <link xl:href="http://snomed.info/id/106233006" xl:show="new">(106233006, SCT, "Topographical modifier")</link>
</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_d011bbd2-4fe1-4b7d-bdd8-9c232053d71d">CODE</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_651fdc9c-936d-47f4-ba11-60d866cf3c11">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_cc527ba3-9010-411c-a3b0-969581c57cfc">U</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_a57b3b25-dc55-4e4f-a48b-0d138824f283"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_17c36059-da4f-4eb5-aa9f-af13ba299c2f"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_7eb34840-a57a-4444-8af5-a5e8f0e1d694"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_16261bbf-a079-4664-88e0-055e631a3d45"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                       <para xml:id="para_10c39410-0a09-4951-bed7-333113d6a683">Not used since AIM does not have a mechanism for post-coordinating the location.</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_39233b51-f2d6-4fc7-9ff9-61b33abd0694">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126010">(126010, DCM, "Imaging Measurements")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_125007">(125007, DCM, "Measurement Group")</olink> &gt; $Measurement &gt; </para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_2c4ad608-cfba-4626-bb26-12628b90bf02">INCLUDE</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_57641630-e372-4330-a260-5e69cc0dc313">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_80f48292-b342-40ee-9397-52b0ce5da828">U</para>
                     </td>
                     <td align="left" colspan="4" rowspan="1">
                       <para xml:id="para_0fab0712-2c9a-4874-b396-8eff3ea11722">
<olink targetdoc="PS3.16" targetptr="sect_TID_310" xrefstyle="select: docname labelnumber quotedtitle"/>
</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_f47c476a-a70f-407f-93d0-319990f1aa4b">Statistical and normal range properties are not used in AIM use cases for this mapping.</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_afc483f4-b9df-4496-972c-1e7ef4de0bf4">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126010">(126010, DCM, "Imaging Measurements")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_125007">(125007, DCM, "Measurement Group")</olink> &gt; $Measurement &gt; $DerivationParameter</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_484c65fb-79ef-4ba6-a991-b9dcd441470c">NUM</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_0e07eb8f-1900-440a-a3b4-acaae5cf4d75">1-n</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_e95b6367-0cf1-4b2a-bee8-a9dd39f66d5a">UC</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_88a0e5a7-1ede-488f-9ef9-7d701311620a"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_1b267bb6-040a-4e5e-93c9-c42ef33a388e"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_ac4f9440-eef9-4d0e-bcdf-35ea90300e42"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_a89f7d9c-6232-4516-9d30-c27edfa060ee"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_829e94a5-014b-40ad-bae2-f4f869d53cbc">Not used in AIM</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_a5c2a1a0-30a3-4840-9cb8-8dac3126e974">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126010">(126010, DCM, "Imaging Measurements")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_125007">(125007, DCM, "Measurement Group")</olink> &gt; $Measurement &gt; $DerivationParameter</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_013bdfa6-ca4d-40b8-8abd-1a15a2fd6f6a">NUM</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_213d2b61-70a4-4e32-a722-aa6b7902f602">1-n</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_a0514a73-11e6-4d47-91ea-e733c5430fc8">UC</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_5e04329c-9f7c-448b-bd3d-dabedc8d6ed5"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_20116b19-661d-48a1-98bb-a76825922a8c"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_531db059-91b4-4d3f-b991-31302e67b4e9"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_48965751-8a02-4321-adb8-784cbe72b1b4"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_d8e4eeaf-fa81-4c7e-b4ac-2d2f579d3a68">Not used in AIM</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_bde1e23a-48bd-409c-bae0-04ad8c3644e9">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126010">(126010, DCM, "Imaging Measurements")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_125007">(125007, DCM, "Measurement Group")</olink> &gt; $Measurement &gt; </para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_8483ab5c-a9de-4c6b-8766-aeb7772e9324">INCLUDE</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_ce827363-0228-44b1-a0d9-6e86f5cb7a9a">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_31e60763-d590-424c-9377-08349d6c4915">UC</para>
                     </td>
                     <td align="left" colspan="4" rowspan="1">
                        <para xml:id="para_3a6d1620-6f8a-4f8c-bacb-5011211d0080">
<olink targetdoc="PS3.16" targetptr="sect_TID_315" xrefstyle="select: docname labelnumber quotedtitle"/>
</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                       <para xml:id="para_b006ef28-753c-48b0-a686-e73ae964f4cc">Not used in AIM</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_87a4fac2-b704-4ae6-a301-e28fe4d56d60">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126010">(126010, DCM, "Imaging Measurements")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_125007">(125007, DCM, "Measurement Group")</olink> &gt; $Measurement &gt; TEXT</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_a97541a5-26e4-4f44-bd94-25e739f3059c">TEXT</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_229215ec-d96a-458a-a540-dae576700e11">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_058f41bb-9a4c-4e20-b568-4dc50c2720f8">UC</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_185d1848-67f0-427f-99d2-efee9faf66e2"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_cc67b50c-0179-4267-aa9e-26c278a7ef94"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_65a23054-7d75-4429-a84b-849857ee2762"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_8a8d35b8-07ec-472c-a8bc-4066118f7d54"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                       <para xml:id="para_074cc028-6537-48f2-a61f-9241153847d4">
<olink targetdoc="PS3.16" targetptr="sect_CID_228" xrefstyle="select: labelnumber quotedtitle"/>
</para>
                    <para xml:id="para_0359dcb9-71bf-4790-91f4-273fb93530cf">Not used in AIM</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_94d59b8b-a1c1-4062-b9bb-3196e31733e1">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126010">(126010, DCM, "Imaging Measurements")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_125007">(125007, DCM, "Measurement Group")</olink> &gt; $Measurement &gt; </para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_4ce7d6f3-8adf-4594-af20-a6974dc8e2c3">INCLUDE</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_4a97fcd6-6ab9-43c7-9475-6e864a16dbbb">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_5b7160ad-fd5c-44ac-84b5-4f49d9b698c8">U</para>
                     </td>
                     <td align="left" colspan="4" rowspan="1">
                        <para xml:id="para_60eaa484-2eda-47fe-816b-33f810501749">
<olink targetdoc="PS3.16" targetptr="sect_TID_1000" xrefstyle="select: docname labelnumber quotedtitle"/>
</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_b752621d-4648-4f2c-b098-16e1a697dfa1">Quotations are not used in AIM.</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_6936635a-9a1a-477a-a08b-4b8151931e9e">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126010">(126010, DCM, "Imaging Measurements")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_125007">(125007, DCM, "Measurement Group")</olink> &gt; $Measurement &gt; <olink targetdoc="PS3.16" targetptr="DCM_121050">(121050, DCM, "Equivalent Meaning of Concept Name")</olink>
</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_27be5412-0ff9-4164-80ed-e9f09f6c42a0">TEXT</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_ddd3e983-c126-4f06-a9f4-d587208670e9">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_89ef59dc-5573-42fd-ab23-c5ff6f660565">U</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_629f78f2-56ac-441e-8c17-ad09531e7080"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_d7cbd026-2dc4-4821-b1ca-525ee5851dcd"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_20593d21-4038-461a-9cd4-00979b44261e"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_b3a2ae09-72aa-457b-82a1-3907de81adae"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_c687eaae-7dcc-4148-a1b4-170f95c18f30">Not used in AIM.</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_4236688e-3ed9-478a-b990-2c32c6f51510">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126010">(126010, DCM, "Imaging Measurements")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_125007">(125007, DCM, "Measurement Group")</olink> &gt; $Measurement &gt; <olink targetdoc="PS3.16" targetptr="DCM_126100">(126100, DCM, "Real World Value Map used for measurement")</olink>
</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_a2b46ffc-4020-4849-b160-19631179e1c4">COMPOSITE</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_7c5c7b0c-4f04-4e92-b661-751afa952c13">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_d16289c7-060c-4cb6-b04e-49f6a9ea704e">U</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_42a88da4-b93e-48a5-a70d-5cda4414446a"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_6ea6a523-9bf9-4784-9e73-a110f033d2f5"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_ccec620e-9f9e-4b31-a66d-d56483763757"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_232d3695-dd28-43b7-86b1-168ee1c9c5a4"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_93e4026a-e745-47f8-8c76-298b5344e2cc">Not used in AIM.</para>
                     </td>
                  </tr>
                 <tr valign="top">
                   <td align="left" colspan="1" rowspan="1">
                     <para xml:id="para_8b8c68b3-6df1-49ee-b1f4-825853c9f0c7">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126010">(126010, DCM, "Imaging Measurements")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_125007">(125007, DCM, "Measurement Group")</olink> &gt; $Measurement &gt; </para>
                   </td>
                   <td align="left" colspan="1" rowspan="1">
                     <para xml:id="para_e11b17ea-f4c7-4fb5-b072-1d5e86cfb8d3">INCLUDE</para>
                   </td>
                   <td align="center" colspan="1" rowspan="1">
                     <para xml:id="para_9afdbec6-eb3e-45a1-8373-c1ed9dd4a95e">1</para>
                   </td>
                   <td align="center" colspan="1" rowspan="1">
                     <para xml:id="para_6f8eac7b-e931-4763-aead-34e6c5fdb4d0">U</para>
                   </td>
                   <td align="left" colspan="4" rowspan="1">
                     <para xml:id="para_9b3e3da5-43b3-4001-b23d-6ece8203397b">
<xref linkend="sect_Mapping_TID_4019" xrefstyle="select: labelnumber quotedtitle nopage"/>
                     </para>
                   </td>
                   <td align="left" colspan="1" rowspan="1">
                     <para xml:id="para_28770775-95dd-43dc-b321-7571a4a87451"/>
                   </td>
                 </tr>
               </tbody>
            </table>
         </section>
         <section label="TID 1420" status="3" xml:id="sect_Mapping_TID_1420">
            <title>Mapping of Measurements Derived From Multiple ROI Measurements</title>
            <para xml:id="para_a643ffde-3f76-4065-bd5b-1f684474af8c">This section describes the mapping of <olink targetdoc="PS3.16" targetptr="sect_TID_1420" xrefstyle="select: docname labelnumber quotedtitle"/>.</para>
            <table frame="box" label="TID 1420" rules="all" xml:id="table_Mapping_TID_1420">
            <caption>Mapping of Measurements Derived From Multiple ROI Measurements</caption>
               <thead>
                  <tr valign="top">
                     <th align="center" colspan="1" rowspan="1">
                       <para xml:id="para_43c648d0-6925-4166-bf64-474136aa8c64">DICOM SR Path</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_4cbd4d20-08b6-4b74-842b-eb8f65a293fb">DICOM VT</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_8972eef2-124a-4ce8-be08-78a90a60562e">DICOM VM</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_c9db8ef5-55d3-49de-a704-5364d1fa2afe">DICOM Usage Type</para>
                     </th>
                    <th align="center" colspan="1" rowspan="1">
                      <para xml:id="para_0cf56a06-aa84-4dee-8f10-ebb04f2851e6">Generated Value</para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                      <para xml:id="para_b69b654b-1d70-4e4a-9a4d-448bf799682a">AIM Element or Attribute</para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                      <para xml:id="para_fcfe4f69-f9a4-4e51-bf6f-35d47d7574b5">AIM Data Type</para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                      <para xml:id="para_37437996-bdd2-4ac9-b4e3-5b20b794c06e">AIM Multiplicity</para>
                    </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_6c0055d9-3352-4d3d-94cd-566ba2ac1f73">Comment</para>
                     </th>
                  </tr>
               </thead>
               <tbody>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_4b8f33fb-186d-4691-bd0f-4182c449bf39">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126011">(126011, DCM, "Derived Imaging Measurements")</olink> &gt; NUM</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_edadb859-86c3-4eb3-9f32-f0b39d70b549">NUM</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_106bafb4-b8eb-43dc-889e-ac951e42022f">1-n</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_f7b11904-85a8-44d8-b86b-ffdf201024ff">M</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_9f324b98-a92f-4962-b872-6d614756b7e4"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_9cf88586-4d66-4466-903a-47561dd0c478"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_f628c7b5-168c-4755-91a8-b3d7b9a7ef3c"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_f8709ceb-1b9b-4b69-83b3-37d0d6e76f9d"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                       <para xml:id="para_2279e49f-bfc5-4a24-8738-8029366c179d">
<olink targetdoc="PS3.16" targetptr="sect_CID_7465" xrefstyle="select: labelnumber quotedtitle"/>
</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_c72d7909-832b-4330-be95-1e057bbef7ae">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126011">(126011, DCM, "Derived Imaging Measurements")</olink> &gt; NUM &gt; </para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_fc359f3a-225a-4148-bed4-0bd995d42d9a">INCLUDE</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_6b473fca-d377-40f0-88c9-ca9e88ee3911">1-n</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_c6f18629-a511-4829-9347-5defe28897e1">MC</para>
                     </td>
                     <td align="left" colspan="4" rowspan="1">
                        <para xml:id="para_bdba5deb-1cca-404e-8162-65c0bbcf3a7d">
<xref linkend="sect_Mapping_TID_1410" xrefstyle="select: labelnumber quotedtitle nopage"/>
                    </para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_bd91111e-6db5-4700-a119-e3441bf7d5ab"/>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_54558c9d-7c17-4224-8bea-4bd9655f72d3">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126011">(126011, DCM, "Derived Imaging Measurements")</olink> &gt; NUM &gt; </para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_4360e358-9c77-484e-97f3-cbb3a31e22be">INCLUDE</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_3908d9ec-8634-4792-8de0-4671a21cc263">1-n</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_d3e4adb1-69a0-482e-b962-3b6b35d053e4">MC</para>
                     </td>
                     <td align="left" colspan="4" rowspan="1">
                        <para xml:id="para_eb8a67c7-2d7e-4b9e-a28e-888f1c04215b">
<xref linkend="sect_Mapping_TID_1411" xrefstyle="select: labelnumber quotedtitle nopage"/>
                    </para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_0a20712a-de94-411e-8786-3f1b4a8752a6"/>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_c6f21dca-3d31-43ae-8fbe-c1311e738371">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126011">(126011, DCM, "Derived Imaging Measurements")</olink> &gt; NUM &gt; </para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_6b1d0845-f00f-40db-a54b-6cdb8fefdf4b">INCLUDE</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_d0d781ef-3638-40c7-870b-a99060bd80a6">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_4e23818b-65f9-4afe-9e5c-756f2e28541b">U</para>
                     </td>
                     <td align="left" colspan="4" rowspan="1">
                       <para xml:id="para_af6b7058-7199-475e-88e5-1c67d9eccfce">
<olink targetdoc="PS3.16" targetptr="sect_TID_310" xrefstyle="select: docname labelnumber quotedtitle"/>
</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                       <para xml:id="para_993d75f8-240b-4eeb-9458-41e442cddcbe">Statistical and normal range properties are not used in AIM use cases for this mapping.</para>
                     </td>
                  </tr>
               </tbody>
            </table>
         </section>
          
                   <section label="TID 300" status="3" xml:id="sect_Mapping_TID_300">
            <title>Mapping of Measurement</title>
            <para xml:id="para_70b6ca7c-916b-4e84-9ffd-5ee66fa96567">This section describes the mapping of <olink targetdoc="PS3.16" targetptr="sect_TID_300" xrefstyle="select: docname labelnumber quotedtitle"/>.</para>
            <table frame="box" label="TID 300" rules="all" xml:id="table_Mapping_TID_300">
            <caption>Mapping of Measurement</caption>
              <thead>
                <tr valign="top">
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_f9e21000-de84-4f17-9b40-34aa589e5ede">DICOM SR Path</para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_087334aa-7d47-4f47-be1c-6123fc4d2a24">DICOM VT</para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_c5d5feb4-6dc7-4e56-a013-727a7a3b7c36">DICOM VM</para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_3f365770-3afb-491e-a491-4576ac222826">DICOM Usage Type</para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_61bd4dc3-8b20-4322-9545-c5605d18c4c8">Generated Value</para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_3d412feb-d3f4-4be2-9f97-9d0d95850fab">AIM Element or Attribute</para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_4423606a-3055-4aa2-93fb-570e55df913e">AIM Data Type</para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_e3502b04-012b-4741-977d-65417630ea6a">AIM Multiplicity</para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_459a02d8-2eee-4cb6-8f91-32bfef42337d">Comment</para>
                  </th>
                </tr>
              </thead>
               <tbody>
                 <tr valign="top">
                   <td align="left" colspan="1" rowspan="1">
                     <para xml:id="para_09bba8fc-4914-4a59-a7ab-4b4ee0e768f5">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126010">(126010, DCM, "Imaging Measurements")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_125007">(125007, DCM, "Measurement Group")</olink> &gt; $Measurement</para>
                   </td>
                   <td align="left" colspan="1" rowspan="1">
                     <para xml:id="para_a03455ea-de08-4a0a-b91c-52d24a5cef05">NUM</para>
                   </td>
                   <td align="center" colspan="1" rowspan="1">
                     <para xml:id="para_d732d01a-752f-4f54-b876-4af726025039">1</para>
                   </td>
                   <td align="center" colspan="1" rowspan="1">
                     <para xml:id="para_640260a4-347d-42a0-b112-6335ec76434b">M</para>
                   </td>
                   <td align="left" colspan="1" rowspan="1">
                     <para xml:id="para_674b5089-a13a-48bd-9613-f15b0641e7c9"/>
                   </td>
                   <td align="left" colspan="1" rowspan="1">
                     <para xml:id="para_98c1b3b8-f036-4eac-9702-b3498cb6283c">NAME = Image​Annotation​​Collection/​image​Annotations/​Image​Annotation/​calculation​Entity​Collection/​CalculationEntity/​typeCode[1]</para>
                     <para xml:id="para_d828f626-a30c-44fe-813a-a3f91b871e1d">VALUE = Image​Annotation​​Collection/​image​Annotations/​Image​Annotation/​calculation​Entity​Collection/​CalculationEntity/​calculationResult​Collection/​CalculationResult/​​@value, calculationData​Collection/​CalculationData/​@value</para>
                     <para xml:id="para_432a3a30-2d99-4054-aafb-8006b502f2bd">UNITS = Image​Annotation​​Collection/​image​Annotations/​Image​Annotation/​calculation​Entity​Collection/​CalculationEntity/​calculationResult​Collection/​CalculationResult/​unitOfMeasure</para>
                   </td>
                   <td align="left" colspan="1" rowspan="1">
                     <para xml:id="para_3a7ac115-4e09-4178-88a9-f4f71e3b43d3">CD, ST, CD</para>
                   </td>
                   <td align="left" colspan="1" rowspan="1">
                     <para xml:id="para_8d5164d9-50b3-41de-ab22-a66e0a9016a8">1..n, 0..n, 1</para>
                   </td>
                   <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_544dac74-9581-4a61-8b76-ea06324d27c5">The first typeCode entry is assumed to be the primary concept. Other typeCode entries may be considered as modifiers.</para>
                     <para xml:id="para_2f2d47e0-0a7b-455d-a67a-7ddf77fda348">Value may be found in either Compact​Calculation​Result (i.e., value child of CalculationResult) or first value of Extended​Calculation​Result (i.e., nested within calculation​Result​Collection).</para>
                     <para xml:id="para_5086358b-834f-4017-9000-a66489784719">Only mapping of a single value from Extended​Calculation​Result is supported.</para>
                       <para xml:id="para_6c7b5af0-40d2-416b-9cf4-57d02f3c3360">The value of aim:unique​Identifier/@root is mapped to the Observation​UID Attribute of the NUM Content Item.</para>
                   </td>
                 </tr>
                 <tr valign="top">
                   <td align="left" colspan="1" rowspan="1">
                     <para xml:id="para_347dade7-d205-4da7-8634-24e7acdd72df">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126010">(126010, DCM, "Imaging Measurements")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_125007">(125007, DCM, "Measurement Group")</olink> &gt; $Measurement &gt; $ModType</para>
                   </td>
                   <td align="left" colspan="1" rowspan="1">
                     <para xml:id="para_c690f85a-6b6f-4df0-b252-4042939f3e89">CODE</para>
                   </td>
                   <td align="center" colspan="1" rowspan="1">
                     <para xml:id="para_b7eadee4-3ab1-40f2-acf2-7a1756920586">1-n</para>
                   </td>
                   <td align="center" colspan="1" rowspan="1">
                     <para xml:id="para_2dd24aa8-ee68-49be-9162-a299f7270a40">U</para>
                   </td>
                   <td align="left" colspan="1" rowspan="1">
                     <para xml:id="para_a890e975-fedd-48ae-a17f-80fff3b72f67"/>
                   </td>
                   <td align="left" colspan="1" rowspan="1">
                     <para xml:id="para_ba642c1d-f787-44f7-801a-ad0a2ba2cc19">Image​Annotation​​Collection/​image​Annotations/​Image​Annotation/​calculation​Entity​Collection/​CalculationEntity/​typeCode​[position() != 1]</para>
                   </td>
                   <td align="left" colspan="1" rowspan="1">
                     <para xml:id="para_5f7e7769-0887-4550-915b-2f9cc782963d">CD</para>
                   </td>
                   <td align="left" colspan="1" rowspan="1">
                     <para xml:id="para_25dff7e4-898f-4235-b1a8-216635e848a1">1..n</para>
                   </td>
                   <td align="left" colspan="1" rowspan="1">
                     <para xml:id="para_38e64c85-424f-4bbc-99ce-1d99fd278d0e">The first typeCode entry is assumed to be the primary concept. This row can be used if successive typeCode entries are recognized, and not a method or derivation, and a name-value pair can be constructed from the typeCode entry (which is a single code).</para>
                   </td>
                 </tr>
                 <tr valign="top">
                   <td align="left" colspan="1" rowspan="1">
                     <para xml:id="para_b1277580-60a0-49fb-a9fb-a56c970caea2">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126010">(126010, DCM, "Imaging Measurements")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_125007">(125007, DCM, "Measurement Group")</olink> &gt; $Measurement &gt; <link xl:href="http://snomed.info/id/370129005" xl:show="new">(370129005, SCT, "Measurement Method")</link>
</para>
                   </td>
                   <td align="left" colspan="1" rowspan="1">
                     <para xml:id="para_ab667b00-aa2b-475f-818b-5c8d6bbebd70">CODE</para>
                   </td>
                   <td align="center" colspan="1" rowspan="1">
                     <para xml:id="para_1767d96c-cd8d-4616-adc2-27fdd1e39e69">1</para>
                   </td>
                   <td align="center" colspan="1" rowspan="1">
                     <para xml:id="para_8bc88f90-c297-46d7-aa39-de0c05321053">U</para>
                   </td>
                   <td align="left" colspan="1" rowspan="1">
                     <para xml:id="para_8afc2627-7956-4cee-981b-a3641cbe084b"/>
                   </td>
                   <td align="left" colspan="1" rowspan="1">
                     <para xml:id="para_007ed9cb-747a-4c3e-a637-27468bf40ed4">Image​Annotation​​Collection/​image​Annotations/​Image​Annotation/​calculation​Entity​Collection/​CalculationEntity/​typeCode​[position() != 1]</para>
                   </td>
                   <td align="left" colspan="1" rowspan="1">
                     <para xml:id="para_3da0e8a3-c5ef-4128-a094-a52f055a69b2">CD</para>
                   </td>
                   <td align="left" colspan="1" rowspan="1">
                     <para xml:id="para_3388373a-cd31-4c04-9db9-a52db1e77419">1..n</para>
                   </td>
                   <td align="left" colspan="1" rowspan="1">
                     <para xml:id="para_c10230d3-7f7e-4ab5-a7c6-b0310a9203de">The first typeCode entry is assumed to be the primary concept. Other typeCode entries may be considered as modifiers, but there is no standard order, so recognition as a "method" depends on recognition of specific codes.</para>
                   </td>
                 </tr>
                 <tr valign="top">
                   <td align="left" colspan="1" rowspan="1">
                     <para xml:id="para_e8fee1fa-f466-4f84-b5b1-0b3630e04312">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126010">(126010, DCM, "Imaging Measurements")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_125007">(125007, DCM, "Measurement Group")</olink> &gt; $Measurement &gt; <olink targetdoc="PS3.16" targetptr="DCM_121401">(121401, DCM, "Derivation")</olink>
</para>
                   </td>
                   <td align="left" colspan="1" rowspan="1">
                     <para xml:id="para_eb8da72f-de40-4ff8-bb46-f63bd9ed8bdf">CODE</para>
                   </td>
                   <td align="center" colspan="1" rowspan="1">
                     <para xml:id="para_e6af1871-5e45-4268-a6b5-13c0834aaed9">1</para>
                   </td>
                   <td align="center" colspan="1" rowspan="1">
                     <para xml:id="para_c3322668-ee55-425c-aea6-d551ac191a67">U</para>
                   </td>
                   <td align="left" colspan="1" rowspan="1">
                     <para xml:id="para_eb3a815b-d9f0-4873-b295-561d996fb136"/>
                   </td>
                   <td align="left" colspan="1" rowspan="1">
                     <para xml:id="para_e8158a74-c270-444b-90b9-7e47ef641be0">Image​Annotation​​Collection/​image​Annotations/​Image​Annotation/​calculation​Entity​Collection/​CalculationEntity/​typeCode​[position() != 1]</para>
                   </td>
                   <td align="left" colspan="1" rowspan="1">
                     <para xml:id="para_eea20018-ea2f-4ee6-9670-ae53ed51eff5">CD</para>
                   </td>
                   <td align="left" colspan="1" rowspan="1">
                     <para xml:id="para_3d2ce027-87f6-47d0-9e53-b8a7bafe29e4">1..n</para>
                   </td>
                   <td align="left" colspan="1" rowspan="1">
                     <para xml:id="para_321d4684-f7a3-42f9-8571-036e39f5bfc9">The first typeCode entry is assumed to be the primary concept. Other typeCode entries may be considered as modifiers, but there is no standard order, so recognition as a "derivation" depends on recognition of specific codes.</para>
                   </td>
                 </tr>
                 <tr valign="top">
                   <td align="left" colspan="1" rowspan="1">
                     <para xml:id="para_14180469-233b-4a53-898f-d44954f7d2cf">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126010">(126010, DCM, "Imaging Measurements")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_125007">(125007, DCM, "Measurement Group")</olink> &gt; $Measurement &gt; <link xl:href="http://snomed.info/id/363698007" xl:show="new">(363698007, SCT, "Finding Site")</link>
</para>
                   </td>
                   <td align="left" colspan="1" rowspan="1">
                     <para xml:id="para_4650e58d-27b7-49fe-803b-822eb72428d2">CODE</para>
                   </td>
                   <td align="center" colspan="1" rowspan="1">
                     <para xml:id="para_f0111033-2e3d-4670-aa1d-eff7e4a828f2">1-n</para>
                   </td>
                   <td align="center" colspan="1" rowspan="1">
                     <para xml:id="para_22da2cdb-6b8c-448e-964c-3c9c0eb8c4e4">U</para>
                   </td>
                   <td align="left" colspan="1" rowspan="1">
                     <para xml:id="para_b2a55fc4-66b4-4193-883f-3a23d9419073"/>
                   </td>
                   <td align="left" colspan="1" rowspan="1">
                     <para xml:id="para_a1670c7d-d0b9-4e7c-8744-cca79e5096f0"/>
                   </td>
                   <td align="left" colspan="1" rowspan="1">
                     <para xml:id="para_7f73eb95-ba8d-4015-9d32-1c1a4a1770f6"/>
                   </td>
                   <td align="left" colspan="1" rowspan="1">
                     <para xml:id="para_67f9b3ae-bec8-4202-a2c7-d3871e4dcd47"/>
                   </td>
                   <td align="left" colspan="1" rowspan="1">
                     <para xml:id="para_6b9f624a-b526-43de-b37e-b706fa149722">Not used in AIM, since it is sent at measurement group level because it is common to all measurements in annotation.</para>
                   </td>
                 </tr>
                 <tr valign="top">
                   <td align="left" colspan="1" rowspan="1">
                     <para xml:id="para_0396f6da-f59b-4aed-9e95-3fdd455b3058">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126010">(126010, DCM, "Imaging Measurements")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_125007">(125007, DCM, "Measurement Group")</olink> &gt; $Measurement &gt; <link xl:href="http://snomed.info/id/363698007" xl:show="new">(363698007, SCT, "Finding Site")</link> &gt; <link xl:href="http://snomed.info/id/272741003" xl:show="new">(272741003, SCT, "Laterality")</link>
</para>
                   </td>
                   <td align="left" colspan="1" rowspan="1">
                     <para xml:id="para_c92a1d11-c128-45f1-962c-1c0b8f7fce63">CODE</para>
                   </td>
                   <td align="center" colspan="1" rowspan="1">
                     <para xml:id="para_2e9d7b63-2853-415a-90dd-b2711c74be88">1</para>
                   </td>
                   <td align="center" colspan="1" rowspan="1">
                     <para xml:id="para_1f98656b-a25c-412b-8e94-c6cfe8815dfe">U</para>
                   </td>
                   <td align="left" colspan="1" rowspan="1">
                     <para xml:id="para_d16cb8e7-b37d-402d-b000-27f8e1f1f76f"/>
                   </td>
                   <td align="left" colspan="1" rowspan="1">
                     <para xml:id="para_0af96515-aad4-4753-b214-29454ea4c135">
<olink targetdoc="PS3.16" targetptr="sect_CID_244" xrefstyle="select: labelnumber quotedtitle"/>
                     </para>
                   </td>
                   <td align="left" colspan="1" rowspan="1">
                     <para xml:id="para_aa96800c-d048-4d89-8f7b-ed1206efde23"/>
                   </td>
                   <td align="left" colspan="1" rowspan="1">
                     <para xml:id="para_9d40a024-f7d8-4be8-9e49-a2f9d90a8602"/>
                   </td>
                   <td align="left" colspan="1" rowspan="1">
                     <para xml:id="para_5adc99b0-7e13-4d52-9adc-63618d6c1d12">Not used in AIM.</para>
                   </td>
                 </tr>
                 <tr valign="top">
                   <td align="left" colspan="1" rowspan="1">
                     <para xml:id="para_870bc79d-d712-46a4-af2e-288f9be6d15c">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126010">(126010, DCM, "Imaging Measurements")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_125007">(125007, DCM, "Measurement Group")</olink> &gt; $Measurement &gt; <link xl:href="http://snomed.info/id/363698007" xl:show="new">(363698007, SCT, "Finding Site")</link> &gt; <link xl:href="http://snomed.info/id/106233006" xl:show="new">(106233006, SCT, "Topographical modifier")</link>
</para>
                   </td>
                   <td align="left" colspan="1" rowspan="1">
                     <para xml:id="para_3d785f25-1b7a-4b6c-9ec4-7ffad7854103">CODE</para>
                   </td>
                   <td align="center" colspan="1" rowspan="1">
                     <para xml:id="para_31f637a9-1c9d-4a05-9eb5-4f22f0fb73e6">1</para>
                   </td>
                   <td align="center" colspan="1" rowspan="1">
                     <para xml:id="para_d7cdcd82-f036-4ac7-90a3-fea9b641b326">U</para>
                   </td>
                   <td align="left" colspan="1" rowspan="1">
                     <para xml:id="para_dde543d4-dc86-4fd5-a67b-30a38b010a51"/>
                   </td>
                   <td align="left" colspan="1" rowspan="1">
                     <para xml:id="para_fdb83fa4-4442-431d-a97d-0814d13aeb7e">$TargetSiteMod</para>
                   </td>
                   <td align="left" colspan="1" rowspan="1">
                     <para xml:id="para_53f0b4f2-2ee8-45f1-89f5-4351bb8e6e9b"/>
                   </td>
                   <td align="left" colspan="1" rowspan="1">
                     <para xml:id="para_69492702-accb-4d15-8f71-093151a29a62"/>
                   </td>
                   <td align="left" colspan="1" rowspan="1">
                     <para xml:id="para_ea12b8a0-c835-47fa-9230-468b190c982c">Not used since AIM does not have a mechanism for post-coordinating the location.</para>
                   </td>
                 </tr>
                 <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_2fc0abe8-0d9d-43a1-971e-9a96d797c282">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126010">(126010, DCM, "Imaging Measurements")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_125007">(125007, DCM, "Measurement Group")</olink> &gt; $Measurement &gt;</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_c61d2bd5-5f7c-44df-a73e-933535c4c4b3">INCLUDE</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_b09a17cb-13cc-49fb-bb3c-3c0df3423f23">1</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_bfcb3cda-38bc-48f4-a454-90f9db411610">U</para>
                     </td>
                     <td align="left" colspan="4" rowspan="1">
                       <para xml:id="para_44861ee8-2862-4ca9-90f2-d1df6b70e889">
<olink targetdoc="PS3.16" targetptr="sect_TID_310" xrefstyle="select: docname labelnumber quotedtitle"/>
                        </para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                       <para xml:id="para_2a24331e-ed0f-4960-9e67-bbdda588a567">Statistical and normal range properties are not used in AIM use cases for this mapping.</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_1e9acc82-214d-43dd-bf24-008d2e68e2de">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126010">(126010, DCM, "Imaging Measurements")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_125007">(125007, DCM, "Measurement Group")</olink> &gt; $Measurement &gt; $DerivationParameter</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_8bd3761a-10af-4d50-abcf-c6568c85806f">NUM</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_ae84ac05-d79c-4f98-b3e3-19cf98ecb3ba">1-n</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_f59daf62-a974-45f1-a296-f8cb8a70b2fa">UC</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_acc64694-c500-434a-8903-28032d851149"/>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_f3b5a159-974d-47be-90da-87c8b43524f9"/>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_f0fadb6e-08ce-4459-9700-4fadc3709b69"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_ddd1f8a8-328f-4b6e-a7c0-71382f0f9013"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                       <para xml:id="para_65966215-ca5f-4ae4-814e-1476730678bb">Not used in AIM for our use cases.</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_bc80afbc-0e9a-43f8-bfbc-114233997038">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126010">(126010, DCM, "Imaging Measurements")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_125007">(125007, DCM, "Measurement Group")</olink> &gt; $Measurement &gt; $DerivationParameter</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_be9c37c6-e565-4c4e-9530-7fb98e4bf40c">NUM</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_d039f854-bedc-4302-ac8e-1dcaaf26fcaa">1-n</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_d93af869-ecb7-409c-80da-4addd19e1c52">UC</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_9f655590-3ba1-4354-807e-4bca780bfe92"/>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_3e7ddf28-4873-4833-9381-1f9dd92a8cb6"/>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_3321460d-5b53-4b69-b578-4e7572582783"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_f534fa4e-8464-43a3-ae76-d697d0558ad6"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                       <para xml:id="para_b5b74187-efed-4ee5-bf41-ce064497d52c">Not used in AIM for our use cases.</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_46a75716-2e93-45d8-8844-a5c92d298352">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126010">(126010, DCM, "Imaging Measurements")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_125007">(125007, DCM, "Measurement Group")</olink> &gt; $Measurement &gt;</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_e54d3cc3-5916-42b3-8113-a0e8b10d6cf7">INCLUDE</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_744368ce-ecfa-4e44-950f-20409129dc11">1</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_9cb5d4f7-a917-44bf-90d0-a334669da6b7">UC</para>
                     </td>
                     <td align="left" colspan="4" rowspan="1">
                       <para xml:id="para_12a39dd1-522f-4113-bb00-7aab59df8d72">
<olink targetdoc="PS3.16" targetptr="sect_TID_315" xrefstyle="select: docname labelnumber quotedtitle"/>
                        </para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                       <para xml:id="para_11b254e6-e81d-4e78-b0e5-42f9802dbdc4">Beyond the scope of our use cases to map.</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_0662f8d8-496f-425e-bfe3-221cf1916edf">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126010">(126010, DCM, "Imaging Measurements")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_125007">(125007, DCM, "Measurement Group")</olink> &gt; $Measurement &gt; TEXT</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_02757111-ef51-4693-ade8-2d87d3e9b0dd">TEXT</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_fa6f778a-fc48-4484-a86d-e125d40d6c33">1</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_7cfa14c5-ef6d-427e-b0aa-bb66e600beb7">UC</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                       <para xml:id="para_6180c859-7631-4a10-9490-705d585d93ad"> </para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_d458cb3d-a110-4a6c-8d89-5067534005a0"/>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_2baa7042-b01f-4533-b42c-d0a7d3aa37b1"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_648b8889-fb27-44ca-ad4c-9f6a1133e4c2"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_5bacb0a7-cea1-43ec-949e-35d607665ed2">
<olink targetdoc="PS3.16" targetptr="sect_CID_228" xrefstyle="select: docname labelnumber quotedtitle"/>
</para>
                    <para xml:id="para_14a03123-e1cf-42bd-96a9-acc562be1a3b">Beyond the scope of our use cases to map.</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_e17b4189-d80c-4463-bea5-0f2673ca3b2c">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126010">(126010, DCM, "Imaging Measurements")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_125007">(125007, DCM, "Measurement Group")</olink> &gt; $Measurement &gt; $ImagePurpose</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_bf8149c2-67d2-4d7a-9164-77dedd20ce57">INCLUDE</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_1ac6c17c-7b5b-4ba5-9543-95bfa6b7ecb8">1-n</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_a68baba5-b8cf-419f-b754-b6f2411d4396">U</para>
                     </td>
                     <td align="left" colspan="4" rowspan="1">
                        <para xml:id="para_120cc28c-3c85-4952-98d9-9f653a92ff96">
<xref linkend="sect_Mapping_TID_320" xrefstyle="select: labelnumber quotedtitle nopage"/>
                        </para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_2ffef9c6-6134-4e8c-a925-6b2f39c126a2"/>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_61727218-698e-458a-9d30-bf8e0213cb52">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126010">(126010, DCM, "Imaging Measurements")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_125007">(125007, DCM, "Measurement Group")</olink> &gt; $Measurement &gt; $WavePurpose</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_b6321a5c-9156-4b7d-b034-e0db188812da">INCLUDE</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_c4a1b4ce-2d7f-48a4-87a2-87182458d2da">1-n</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_84c28190-3ef4-41d6-93d4-f05f88218f52">U</para>
                     </td>
                     <td align="left" colspan="4" rowspan="1">
                       <para xml:id="para_58deaa89-d1f5-4e52-b810-5bdc9200f4c6">
<olink targetdoc="PS3.16" targetptr="sect_TID_321" xrefstyle="select: docname labelnumber quotedtitle"/>
                    </para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_a2c27d04-eb6c-4193-87ac-2f0c9a0c8201">Not used in our AIM use cases.</para>
                  </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_4884ee49-72d5-4212-906f-758611cf3511">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126010">(126010, DCM, "Imaging Measurements")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_125007">(125007, DCM, "Measurement Group")</olink> &gt; $Measurement &gt;</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_6bbed568-cde9-4bfa-b267-00066f75a65f">INCLUDE</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_b252d931-0d0c-4225-9d51-276e2d18c9ea">1</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_b6a1299a-c224-40c4-9172-b21c070668c6">U</para>
                     </td>
                     <td align="left" colspan="4" rowspan="1">
                       <para xml:id="para_397d0127-7423-48c7-88e8-ab7217ff9d6b">
<olink targetdoc="PS3.16" targetptr="sect_TID_1000" xrefstyle="select: docname labelnumber quotedtitle"/>
                        </para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_deb3ac97-4c3c-45f3-b350-2efd222cf32d">Not used in our AIM use cases.</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_08bad3f2-234f-4145-8a86-bb1715e1d60d">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126010">(126010, DCM, "Imaging Measurements")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_125007">(125007, DCM, "Measurement Group")</olink> &gt; $Measurement &gt; <olink targetdoc="PS3.16" targetptr="DCM_121050">(121050, DCM, "Equivalent Meaning of Concept Name")</olink>
</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_f2239d5d-651b-4748-9ec8-2a3235aa8092">TEXT</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_be03b3c3-00ae-40d5-b5b4-ba031dc4a26b">1</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_8d17a17e-ac8b-4819-8139-4369381bfacb">U</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_cd2aed7b-ec22-47e3-a009-dafa2ed74a53"/>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_6f5eef72-5d38-4322-8822-2a7161d1fdac"/>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_d67c8a3a-0405-4df6-9fdc-b4f0ad9531fc"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_b3c15d82-e19c-4f57-b404-223f9b9af36d"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_3c7b2791-3787-4941-802d-a34c2927af6a">Not used in AIM</para>
                     </td>
                  </tr>
                 <tr valign="top">
                   <td align="left" colspan="1" rowspan="1">
                     <para xml:id="para_eaf3196a-550b-47fc-81e7-fd90fc767420">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126010">(126010, DCM, "Imaging Measurements")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_125007">(125007, DCM, "Measurement Group")</olink> &gt; $Measurement &gt;</para>
                   </td>
                   <td align="left" colspan="1" rowspan="1">
                     <para xml:id="para_b43d4e08-a504-4b83-b818-5425049348d2">INCLUDE</para>
                   </td>
                   <td align="center" colspan="1" rowspan="1">
                     <para xml:id="para_b275fdba-87d6-4905-96a2-b3e75781d8a1">1</para>
                   </td>
                   <td align="center" colspan="1" rowspan="1">
                     <para xml:id="para_b5f30a94-a8b2-41f9-965b-aaf6ea2e5163">U</para>
                   </td>
                   <td align="left" colspan="4" rowspan="1">
                     <para xml:id="para_8ab6072c-47ef-4090-94b2-ffb20cd3426f">
<xref linkend="sect_Mapping_TID_4108" xrefstyle="select: labelnumber quotedtitle nopage"/>
                    </para>
                   </td>
                   <td align="left" colspan="1" rowspan="1">
                     <para xml:id="para_84db875c-2243-42ce-8cdd-416c61128b0a">Not mapped at this level for TID 1500, but rather at the Measurement Group level in TID 1501.</para>
                   </td>
                 </tr>
                 <tr valign="top">
                   <td align="left" colspan="1" rowspan="1">
                     <para xml:id="para_75d7b267-14e1-4f72-9903-98e8a8d1ba99">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126010">(126010, DCM, "Imaging Measurements")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_125007">(125007, DCM, "Measurement Group")</olink> &gt; $Measurement &gt; <olink targetdoc="PS3.16" targetptr="DCM_126100">(126100, DCM, "Real World Value Map used for measurement")</olink>
</para>
                   </td>
                   <td align="left" colspan="1" rowspan="1">
                     <para xml:id="para_967001d1-eb45-4dee-afc3-ecfa1e7fed43">COMPOSITE</para>
                   </td>
                   <td align="center" colspan="1" rowspan="1">
                     <para xml:id="para_248718fa-b62a-4c55-a463-14ad0f49db69">1</para>
                   </td>
                   <td align="center" colspan="1" rowspan="1">
                     <para xml:id="para_66ecf93a-ff01-4838-becd-7d6c0212c8ec">U</para>
                   </td>
                   <td align="left" colspan="1" rowspan="1">
                     <para xml:id="para_b2d5bbc4-a8cd-4b55-8aab-b40265a769aa"/>
                   </td>
                   <td align="left" colspan="1" rowspan="1">
                     <para xml:id="para_7cc916b7-a41d-488f-8556-41b4f25b7e27"/>
                   </td>
                   <td align="left" colspan="1" rowspan="1">
                     <para xml:id="para_b6891652-fe20-4285-95e0-a6b02598f888"/>
                   </td>
                   <td align="left" colspan="1" rowspan="1">
                     <para xml:id="para_1035a06a-9af5-4fae-9571-ed82e45d9909"/>
                   </td>
                   <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_b8c38e2f-ac58-4c76-9f48-d9e7409579fd">Not used in AIM.</para>
                   </td>
                 </tr>
                 <tr valign="top">
                   <td align="left" colspan="1" rowspan="1">
                     <para xml:id="para_63ec45a7-5bb6-4a5c-b3b8-f47b82b60ecc">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126010">(126010, DCM, "Imaging Measurements")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_125007">(125007, DCM, "Measurement Group")</olink> &gt; $Measurement &gt; </para>
                   </td>
                   <td align="left" colspan="1" rowspan="1">
                     <para xml:id="para_363a85b8-8419-4414-8bf6-a7c22314b037">INCLUDE</para>
                   </td>
                   <td align="center" colspan="1" rowspan="1">
                     <para xml:id="para_7c509859-9527-434d-b2eb-f06991ac26ff">1</para>
                   </td>
                   <td align="center" colspan="1" rowspan="1">
                     <para xml:id="para_91d43460-fa33-4ea3-91b2-b90926b89a7f">U</para>
                   </td>
                   <td align="left" colspan="4" rowspan="1">
                     <para xml:id="para_5f1bee04-6d53-4f98-8a89-6c953e010620">
<xref linkend="sect_Mapping_TID_4019" xrefstyle="select: labelnumber quotedtitle nopage"/>
                     </para>
                   </td>
                   <td align="left" colspan="1" rowspan="1">
                     <para xml:id="para_6fd0041f-73fa-4cb1-92b7-a8ebae8f01ba"/>
                   </td>
                 </tr>
               </tbody>
            </table>
         </section>

          <section label="TID 320" status="3" xml:id="sect_Mapping_TID_320">
            <title>Mapping of Image or Spatial Coordinates</title>
            <para xml:id="para_0f508e07-b687-44c5-a8a5-9e106affa1f9">This section describes the mapping of <olink targetdoc="PS3.16" targetptr="sect_TID_320" xrefstyle="select: docname labelnumber quotedtitle"/>.</para>
            
            
            
            <table frame="box" label="TID 320" rules="all" xml:id="table_Mapping_TID_320">
            <caption>Mapping of Image or Spatial Coordinates</caption>
              <thead>
                <tr valign="top">
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_e9851800-8cdb-4cb3-a0fe-423d9f80bf32">DICOM SR Path</para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_d17c2b6d-8611-468f-b13b-a0b41daece17">DICOM VT</para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_f4fe63ed-b60e-4ad7-adaf-266f3dcc5fc6">DICOM VM</para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_80f641fe-bd25-455b-9e65-c770702452fd">DICOM Usage Type</para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_1f59ffc0-8bd3-4d41-9790-9f6ce0ef4926">Generated Value</para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_ea071136-d1bf-4d42-ab1e-59ca5fff0c30">AIM Element or Attribute</para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_26ec7a54-d2e0-47a5-a25d-adab121d52b1">AIM Data Type</para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_3549a2b0-e99b-4af5-ad9f-d157bf473c0d">AIM Multiplicity</para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_f7652208-f37f-4abf-8de7-8f42327a50e6">Comment</para>
                  </th>
                </tr>
              </thead>
              <tbody>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_eff456ea-aabb-491d-ab50-660ee7b09d57">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126010">(126010, DCM, "Imaging Measurements")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_125007">(125007, DCM, "Measurement Group")</olink> &gt; $Purpose</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_724be64a-374f-4286-a18c-cf89312aafd5">IMAGE</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_b0958d5d-ffe5-4dd9-abf4-99710c50eebc">1</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_39d340cd-6dee-4e43-88d8-37b81761c400">MC</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_84cb9150-c5b4-4248-944f-25622ef0f567"/>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_90af4b99-6674-4f93-87fa-b1d04b5f7c01">Image​Annotation​​Collection/​image​Annotations/​Image​Annotation/​​markup​Entity​Collection/​MarkupEntity/​imageReferenceUid/​@root, referencedFrameNumber/​@value</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_8e0d43ae-dc1e-4947-a883-27c275dd23b3">II, INT</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_9d862a40-535d-44f6-9b4d-868c4b8baab6">1, 0..1</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_76a73dee-a766-4495-89a4-469661ac29f3">An entire image reference without spatial coordinates</para>
                    <para xml:id="para_ed0b05f0-af95-4018-92b6-569bfbdf92fc">The Referenced SOP Class UID is obtained from imageReference​Entity​Collection; see <xref linkend="table_A.8-5" xrefstyle="select: label"/>
</para>
                    <para xml:id="para_b3c73016-341c-4d35-8219-55fe356e0e55">Only the by-value (SELECTED FROM) relationship is used, not the by-reference (R-SELECTED FROM) relationship.</para>
                  </td>
                </tr>                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_6e17b0a1-4ded-4a1f-a016-177e396cd94e">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126010">(126010, DCM, "Imaging Measurements")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_125007">(125007, DCM, "Measurement Group")</olink> &gt; $Purpose</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_1a747957-5ebb-45df-b0ef-78fe8a9a1882">SCOORD</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_9c0e7ba8-4f8a-4c4d-aff6-1c708dfa420d">1</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_80020ddc-d4b9-43ab-8571-ca5094035797">MC</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_6d3e83ef-3acc-496f-83a1-8c72435fb42c"/>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_014542d0-a40a-4b97-90d5-47cd83b2d29c">Image​Annotation​​Collection/​image​Annotations/​Image​Annotation/​​markup​Entity​Collection/​MarkupEntity/​twoDimension​SpatialCoordinate​​Collection/​TwoDimension​SpatialCoordinate</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_65ef9091-f0e9-482f-b3db-21ec98ec0674">REAL</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_69401148-18d6-470a-8a29-7f735494f669">1..n</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_23a2ff24-560b-43f2-85e0-23c01db7c30e">A reference to coordinates on an image.</para>
                      <para xml:id="para_8f46a58b-b174-43d6-add1-b4570a3f52a7">The value of aim:​Markup​Entity/​aim:​unique​Identifier/​@root is mapped to the Observation​UID Attribute of the SCOORD.</para>
                      <para xml:id="para_b815d145-e523-4f12-aeb4-f8b245404e1b">One or more MarkupEntity instances within an Image​Annotation instance may be constrained to be associated with particular NUM measurement(s) in the included TID 1419 by a Calculation​Entity​References​Markup​Entity​Statement.</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_26ee1546-fd06-4d8a-a36f-e1756efd8d94">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126010">(126010, DCM, "Imaging Measurements")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_125007">(125007, DCM, "Measurement Group")</olink> &gt; $Purpose &gt; IMAGE</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_d61cf4a1-5b91-4248-a893-d353bba4e693">IMAGE</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_09e2d050-851b-48de-b9ea-1c763b8fdabb">1</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_c5fecc94-8347-4efd-bdc4-5953e701257c">M</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_8777b4bc-9768-4ef4-a9d0-d980c1043ef2"/>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_e10a9a43-2a8b-452a-a65a-34ba47cb8112">Image​Annotation​​Collection/​image​Annotations/​Image​Annotation/​​markup​Entity​Collection/​MarkupEntity/​imageReferenceUid/​@root, referencedFrameNumber/​@value</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_2c644c2e-5b9a-42b7-95f8-c87ce3fc0703">II, INT</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_612cb80d-3d27-4087-b65b-1adddae3bf26">1, 0..1</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_47dd6570-bc6d-4a71-a346-a80e01fa4779">The Referenced SOP Class UID is obtained from imageReference​Entity​Collection; see <xref linkend="table_A.8-5" xrefstyle="select: label"/>
</para>
                    <para xml:id="para_751b5b73-5058-4d29-862d-7958cac5570f">Only the by-value (SELECTED FROM) relationship is used, not the by-reference (R-SELECTED FROM) relationship.</para>
                  </td>
                </tr>
              </tbody>
            </table>
          </section>
          
          <section label="TID 4019" status="3" xml:id="sect_Mapping_TID_4019">
            <title>Mapping of Algorithm Identification</title>
            <para xml:id="para_b00e27d1-f0b3-4fd3-9d8a-72f1feaceebd">This section describes the mapping of PS3.16 <olink targetdoc="PS3.16" targetptr="sect_TID_4019" xrefstyle="select: docname labelnumber quotedtitle"/>.</para>
            <table frame="box" label="TID 4019" rules="all" xml:id="table_Mapping_TID_4019">
              <caption>Mapping of Algorithm Identification</caption>
              <thead>
              <tr valign="top">
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_089f2847-d6ab-436d-9c9f-b3a75a6f3d9d">DICOM SR Path</para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_c4f1c4db-15bc-4375-a677-ce87c10bc98d">DICOM VT</para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_dee1783f-7f3f-4a3f-b9f6-32ff67a400ea">DICOM VM</para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_8230863a-a73c-4c76-ae28-cf19c06941d3">DICOM Usage Type</para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_eef6b5fd-4f25-4a7a-9a36-7208ec1ed030">Generated Value</para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_78197b17-bfea-40a1-ada6-ff8ce8a128a7">AIM Element or Attribute</para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_d46deaab-9924-47e7-8cd2-07cd636060e8">AIM Data Type</para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_6db44c3a-c96c-411c-9add-8496b89cd604">AIM Multiplicity</para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_689d418a-6b25-4d99-b510-09af0dad3baf">Comment</para>
                </th>
              </tr>
              </thead>
              <tbody>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_9f91dcd3-f50e-40d8-801a-9358b1d9fc95">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126010">(126010, DCM, "Imaging Measurements")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_125007">(125007, DCM, "Measurement Group")</olink> &gt; $Measurement &gt; <olink targetdoc="PS3.16" targetptr="DCM_111001">(111001, DCM, "Algorithm Name")</olink>
</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_a0be002a-b839-4d61-9940-ca85861db4c0">TEXT</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_49a7d5a1-d50c-48d0-9068-9f56e7e5f207">1</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_dd5e2ff0-de46-4a16-b607-2fa7201186d4">U</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_1c1dc651-a9ca-4cdf-bec6-a6ee972c5b75"/>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_9f299f66-50cc-4f91-a7b7-f539803229d2">Image​Annotation​​Collection/​image​Annotations/​Image​Annotation/​calculation​Entity​Collection/​CalculationEntity/​algorithm/​name/​@value</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_2ff3c5dd-0a1e-44bf-b09b-6cd390a19a02">ST</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_96557701-e480-40b6-a5cf-19cdb0e04b18">1</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">The  type attribute (CD 1..n) is not supported by CID 4019</td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_49ec4804-18e7-4647-8b44-ff74dd4e2de9">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126010">(126010, DCM, "Imaging Measurements")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_125007">(125007, DCM, "Measurement Group")</olink> &gt; $Measurement &gt; <olink targetdoc="PS3.16" targetptr="DCM_111003">(111003, DCM, "Algorithm Version")</olink>
</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_6485a135-79f5-481c-9686-cba8a8b42eeb">TEXT</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_efe9466a-3436-40ed-8934-8f7f8e31b720">1</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_68ec0f3d-fd58-4c82-839f-902f1ed80e68">U</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_ff880dde-7043-4b6c-beac-fc82b3f27227"/>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_86d07c95-f96e-498e-9f38-e7b2b8d8abc0">Image​Annotation​​Collection/​image​Annotations/​Image​Annotation/​calculation​Entity​Collection/​CalculationEntity/​algorithm/​version/​@value</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_69a90dfc-1d4f-4725-bd26-85b22e6708bb">ST</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_0d27de56-653a-4715-ba8a-a51287f22e1a">0..1</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_e3c70a05-b209-4769-8a8f-4134100d2308"/>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_44bfa161-99bc-4620-af16-77a5ffaef824">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126010">(126010, DCM, "Imaging Measurements")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_125007">(125007, DCM, "Measurement Group")</olink> &gt; $Measurement &gt; <olink targetdoc="PS3.16" targetptr="DCM_111002">(111002, DCM, "Algorithm Parameters")</olink>
</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_fd6cae6c-cf93-4701-a96c-5bc142b9a6b4">TEXT</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_1098e676-b761-449f-9f89-f2aa31ced932">1-n</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_2a90f4e0-7c69-48e4-a2ef-8b41a187dbd3">U</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_17e73441-5c68-44bb-ab4d-e1367147534f"/>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_330047b1-27ca-4b43-a0b1-a6af8d8ba696">Image​Annotation​​Collection/​image​Annotations/​Image​Annotation/​calculation​Entity​Collection/​CalculationEntity/​algorithm/​parameter/​@dataType,@name,@value</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_e31754f1-f1f4-48a3-9364-7dda64d2fb5f">CD,ST,ST</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_ff1971c0-3414-478d-96a2-b267a2a23f25">0..n</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_58c09bcd-041d-4c54-9802-ab9095c0e52c">Separate mapping of dataType, name and value attributes is not supported by CID 4019</para>
                  </td>
                </tr>
                <tr>
                  <td align="left" colspan="1"/>
                  <td align="left" colspan="1"/>
                  <td align="center" colspan="1"/>
                  <td align="center" colspan="1"/>
                  <td align="left" colspan="1"/>
                  <td align="left" colspan="1">
                    <para xml:id="para_6904d4e5-9e71-4099-bc3d-4baca543ed48">Image​Annotation​​Collection/​image​Annotations/​Image​Annotation/​calculation​Entity​Collection/​CalculationEntity/​algorithm/​type</para>
                  </td>
                  <td align="left" colspan="1">CD</td>
                  <td align="left" colspan="1">1..n</td>
                  <td align="left" colspan="1">No corresponding mapping in DICOM SR template at this time.</td>
                </tr>
              </tbody>
            </table>
          </section>
          
          <section label="TID 4108" status="3" xml:id="sect_Mapping_TID_4108">
            <title>Mapping of Tracking Identifier</title>
            <para xml:id="para_359c7e16-7fe9-4bf5-b6b2-cf0f34e5fd74">This section describes the mapping of <olink targetdoc="PS3.16" targetptr="sect_TID_4108" xrefstyle="select: docname labelnumber quotedtitle"/>.</para>
            <para xml:id="para_a2739af6-ac92-4d21-91f9-dbfa8f483875">For the purpose of this mapping, this template is not used to track individual measurements; rather, the corresponding content items defined in <xref linkend="sect_Mapping_TID_1501" xrefstyle="select: labelnum quotedtitle"/> are mapped at the Measurement Group level instead. </para>
            <table frame="box" label="TID 4108" rules="all" xml:id="table_Mapping_TID_4108">
            <caption>Mapping of Tracking Identifier</caption>
              <thead>
                <tr valign="top">
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_79c9a959-48a8-4f2c-8b67-d890da2e2c94">DICOM SR Path</para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_0db7ac2f-14e4-4885-8a2f-f7b1bf499117">DICOM VT</para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_e11c1b3d-580c-49b3-be44-7bc6dd0f9cb8">DICOM VM</para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_25d40012-486c-4e05-bc77-e602b037e4e0">DICOM Usage Type</para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_940459ff-89e7-4fbc-bf52-437f0a6f72da">Generated Value</para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_df7e0955-e469-4f63-8c54-ad397d977c19">AIM Element or Attribute</para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_ba9e1605-d79e-48df-88ff-29e0f046d6dd">AIM Data Type</para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_c227ca0f-8763-4af6-9e7c-075049a54a47">AIM Multiplicity</para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_a4338ddc-8a6c-4df4-8ffc-6c0b1ad49173">Comment</para>
                  </th>
                </tr>
              </thead>
              <tbody>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_32f0dbb3-e530-42db-a64f-677dcd78ba35">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126010">(126010, DCM, "Imaging Measurements")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_125007">(125007, DCM, "Measurement Group")</olink> &gt; $Measurement &gt; <olink targetdoc="PS3.16" targetptr="DCM_112039">(112039, DCM, "Tracking Identifier")</olink>
</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_bc3f117e-2035-4e65-86cf-d4a9efeeb9b8">TEXT</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_f966099f-6c29-41ca-85a6-cb916e5bd829">1</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_5b45d14a-ae85-4330-a3cd-c232c0a88bd6">U</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_a3f03d80-8e1a-45b5-b934-09b8137553b2"/>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_c0933f75-9430-4d71-9c17-f8575caefea8"/>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_2bc207c9-0ece-49c7-b222-007384898533"/>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_1b0516dc-c647-43b0-9072-5672caac272f"/>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_08db1916-3901-4bd2-8ea7-06fdf9db2da5">MC but U in parent TID 300.</para>
                    <para xml:id="para_7fddd65d-0b26-4510-8a98-85cff044a45c">Not mapped at this level for TID 1500, but rather at the Measurement Group level in TID 1501.</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_315a9c30-ff80-42c4-8b6a-934265befcae">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126010">(126010, DCM, "Imaging Measurements")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_125007">(125007, DCM, "Measurement Group")</olink> &gt; $Measurement &gt; <olink targetdoc="PS3.16" targetptr="DCM_112040">(112040, DCM, "Tracking Unique Identifier")</olink>
</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_377adc03-21c7-4956-9986-702016e0169e">UIDREF</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_f82f1fc9-8068-4fb8-b1dd-7841be9feba6">1</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_6067d86a-af07-4f83-8e1e-055fb63416e8">U</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_4e2bfc02-3777-4951-bea1-206ac55c3cf7"/>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_99990495-1151-471c-bcf8-5492ac90fdf8"/>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_6cc77081-5782-4896-af03-d957fa72fed3"/>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_94569094-7112-4acb-978a-cdd4d230bf83"/>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_1ab1efda-7ab4-41cb-b406-2f75d0643ce5">MC but U in parent TID 300.</para>
                    <para xml:id="para_ffe5dfab-f00e-460a-8621-b0918001e114">Not mapped at this level for TID 1500, but rather at the Measurement Group level in TID 1501.</para>
                  </td>
                </tr>
              </tbody>
            </table>
          </section>
          
          <section label="TID 1204" status="3" xml:id="sect_Mapping_TID_1204">
            <title>Mapping of Language of Content Item and Descendants</title>
            <para xml:id="para_eb6db83e-e70d-4f63-83ee-d5b99ef4c9c2">This section describes the mapping of <olink targetdoc="PS3.16" targetptr="sect_TID_1204" xrefstyle="select: docname labelnumber quotedtitle"/>.</para>
            <table frame="box" label="TID 1204" rules="all" xml:id="table_Mapping_TID_1204">
            <caption>Mapping of Language of Content Item and Descendants</caption>
              <thead>
                <tr valign="top">
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_a2875a9c-5e94-4119-b100-11fbf19cc62f">DICOM SR Path</para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_10163747-b29f-4a70-be04-67aa0f6edc2b">DICOM VT</para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_9eaab593-52be-4271-b976-a28ca36ab065">DICOM VM</para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_7d3f12be-6785-4da9-91e4-e38d6c49b476">DICOM Usage Type</para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_46738eb1-e917-4560-9822-077c085f1aba">Generated Value</para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_9c524082-aaa6-47a2-b76e-cf4c265d01fa">AIM Element or Attribute</para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_1a2968b2-da9e-4830-8aa8-cb3b559ef39c">AIM Data Type</para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_47edb6a8-b451-4ac9-b864-0ebb9f468069">AIM Multiplicity</para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_17ba7ea0-b6a2-49f3-9e1e-81c4e5e6bffa">Comment</para>
                  </th>
                </tr>
              </thead>
              <tbody>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_41583901-34d2-45ab-8c40-f247fb945469">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_121049">(121049, DCM, "Language of Content Item and Descendants")</olink>
</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_03fe73a1-2976-4f5f-ba50-c93e37a5f91c">CODE</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_bfa69005-4fce-43ba-bc89-16a7d97ef339">1</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_6a3f4ceb-7380-480c-b66f-3b23eb4135bb">M</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_a334f27a-3922-412c-8b60-f5cad3867046">(eng,RFC5646,"English")</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_ec624794-32e0-48cf-9242-2d0dc2b5168c"> </para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_b8187f1b-8dbc-4c69-8277-05efbfea0bbb"/>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_3d0d8cd6-d793-4529-922d-5a3d3d4085bc"/>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_7df04e4b-e115-48a3-874a-f6d60791b365">Not used in AIM; discarded if present in DICOM SR. <olink targetdoc="PS3.16" targetptr="sect_CID_5000" xrefstyle="select: labelnumber quotedtitle"/>
</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_163fc809-a624-46c9-8b35-e938fe2659f9">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_121049">(121049, DCM, "Language of Content Item and Descendants")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_121046">(121046, DCM, "Country of Language")</olink>
</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_70eb360c-b9ce-47ff-87d8-cea63199dfe3">CODE</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_f7858942-6643-4126-a170-1409f30741bf">1</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_555f9524-c650-4bec-bd8c-f98d30bbdd68">U</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_97fe988c-16bc-4513-b18c-62935c1406ea">(US,ISO3166_1,"United States")</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_77d28695-c1c9-4e51-bcb7-591fa30422fd"> </para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_033d6437-d289-4509-8282-1d1f3d80d571"/>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_b11efa0f-51f9-4517-913f-1ab5d8e7b3c2"/>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_e4bec887-52fd-49d3-811c-c68c5ef90046">Not used in AIM; discarded if present in DICOM SR. <olink targetdoc="PS3.16" targetptr="sect_CID_5001" xrefstyle="select: labelnumber quotedtitle"/>
</para>
                  </td>
                </tr>
              </tbody>
            </table>
          </section>
          
          <section label="TID 1001" status="3" xml:id="sect_Mapping_TID_1001">
            <title>Mapping of Observation Context</title>
            <para xml:id="para_6fd4e6b6-bf85-4f7a-aab1-0df945206ca5">This section describes the mapping of <olink targetdoc="PS3.16" targetptr="sect_TID_1001" xrefstyle="select: docname labelnumber quotedtitle"/>.</para>
            <table frame="box" label="TID 1001" rules="all" xml:id="table_Mapping_TID_1001">
            <caption>Mapping of Observation Context</caption>
              <thead>
                <tr valign="top">
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_a4cb05f6-5fe1-46eb-b774-761cb361e6da">DICOM SR Path</para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_de20ad09-fdbf-428b-98c3-20927ac65a32">DICOM VT</para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_d8e4a98e-5608-4444-9821-aa6c7c0b1357">DICOM VM</para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_f5ee5168-3923-4052-abf4-bd088bd18032">DICOM Usage Type</para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_063b9cfc-b98c-402b-a358-d347397dd520">Generated Value</para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_44ad8a24-98bb-466b-b48f-a8209b6c84ac">AIM Element or Attribute</para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_00d11958-cc1c-43ce-9621-3f9f523a6d33">AIM Data Type</para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_b2205c8c-daa0-4350-98ec-6ce4a18adb21">AIM Multiplicity</para>
                  </th>
                  <th align="center" colspan="1" rowspan="1">
                    <para xml:id="para_9c375800-f378-4727-b396-85d23a1d824a">Comment</para>
                  </th>
                </tr>
              </thead>
              <tbody>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_2510293d-8617-4c3b-9bd0-a12ab03bed3f">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt;</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_87976a08-4656-4063-9cda-39424cab7702">INCLUDE</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_fa1146b5-65a3-4f4a-a148-a919d6207880">1-n</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_43cfc7cf-666a-47b7-a8df-eabd4041ea4e">MC</para>
                  </td>
                  <td align="left" colspan="4" rowspan="1">
                    <para xml:id="para_180064f9-e3fb-446e-8448-2011342218f0">
<xref linkend="sect_Mapping_TID_1002" xrefstyle="select: labelnumber quotedtitle nopage"/>
</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_f568d261-2d88-49cd-bd99-8797b892de24">Only required for AIM if Attributes of the Author Observer Sequence (0040,A078) are insufficient to describe the person observer; not used for a device observer.</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_71de902d-b2dd-40c7-831c-a3b747e5d9bb">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt;</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_e9c99437-e6f7-470f-ba10-2d7be016da35">INCLUDE</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_f62c2e5d-c3ff-474b-9534-cfa9630de775">1</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_888860e9-0854-4926-a783-97551675bea2">MC</para>
                  </td>
                  <td align="left" colspan="4" rowspan="1">
                    <para xml:id="para_0e17b8de-ae85-4b28-be3b-1ee04039020f">TID 1005 "Procedure Context"</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_84f26cea-3faf-4d38-9b9a-ac4e61efa478">Not used in AIM since redundant with header information; discarded if present in DICOM SR.</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_63494bf8-41e4-4cfa-bb37-0c97750dbedf">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt;</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_798111e8-8ed5-4395-acb7-b3bd981333c3">INCLUDE</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_0041c177-4b9d-4074-969f-697333be193e">1</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_0f2e088e-e1a8-4743-aa73-f78759fbeeaf">MC</para>
                  </td>
                  <td align="left" colspan="4" rowspan="1">
                    <para xml:id="para_aa0eac98-04fb-4549-bed9-cbfc1898189d">TID 1006 "Subject Context"</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_556ae35a-9aa9-4bf5-a9f4-5547071a15a7">Not used in AIM since redundant with header information; discarded if present in DICOM SR.</para>
                  </td>
                </tr>
              </tbody>
            </table>
          </section>
          
          <section label="TID 1002" status="3" xml:id="sect_Mapping_TID_1002">
            <title>Mapping of Observer Context</title>
            <para xml:id="para_e223e264-12f5-4e96-bfee-9034e0c3dad6">This section describes the mapping of <olink targetdoc="PS3.16" targetptr="sect_TID_1002" xrefstyle="select: docname labelnumber quotedtitle"/>.</para>
            <table frame="box" label="TID 1002" rules="all" xml:id="table_Mapping_TID_1002">
            <caption>Mapping of Observer Context</caption>
              <thead>
                  <tr valign="top">
                     <th align="center" colspan="1" rowspan="1">
                       <para xml:id="para_7a8f5774-ed50-4504-8d0c-437531052439">DICOM SR Path</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_be2ff171-ec49-445d-8b06-6534f1e88fb7">DICOM VT</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_868875cd-260a-41d6-bd34-4356a01c01a6">DICOM VM</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_ae5373e2-c468-41c4-b6eb-fe7b0ae5b0b6">DICOM Usage Type</para>
                     </th>
                    <th align="center" colspan="1" rowspan="1">
                      <para xml:id="para_b85ec6dd-60ff-4057-b0f8-cba231255af9">Generated Value</para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                      <para xml:id="para_0e6b2dc6-890f-4141-8fcc-145a93316a4b">AIM Element or Attribute</para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                      <para xml:id="para_bf5c55ed-4d82-4a15-a5c5-0bbaf0036bbe">AIM Data Type</para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                      <para xml:id="para_803251d1-3548-47c6-8ea8-dd73863f5ba0">AIM Multiplicity</para>
                    </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_f3ea3121-2637-4e9b-8224-f25c07086bda">Comment</para>
                     </th>
                  </tr>
               </thead>
              <tbody>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_3bba8a55-3854-4e49-89fb-fef4622f0876">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_121005">(121005, DCM, "Observer Type")</olink>
</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_b006ab62-3ace-4c5a-b1de-43bd3185d479">CODE</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_69f24cce-ea7a-4c0b-886d-b427df73e578">1</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_2a527a39-be1d-4540-bf87-8b9da9fabbe1">MC</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_a659b83b-95f5-449c-8650-3173f7110099"/>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_3caf0410-fa49-4700-bbd9-39afa0a0e168"/>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_681ca281-ee48-4f05-a673-b22120a5e970"/>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_cf9a9a27-239d-441d-b639-c9c18da02936"/>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_6abe37db-126f-4d36-af9f-1efb0837f8d4">Since this template is only used for AIM for person observers, which is the default, it may be omitted or explicitly sent as <olink targetdoc="PS3.16" targetptr="DCM_121006">(121006, DCM, "Person")</olink>; see also <olink targetdoc="PS3.16" targetptr="sect_CID_270" xrefstyle="select: labelnumber quotedtitle"/>
                    </para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_0e30d9f6-65dc-41c0-912b-6d5d2d3933e4">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt;</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_3df5f61e-6a9a-4aff-b6ef-bdfa6b3b2834">INCLUDE</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_7577fcdd-1a7d-46d7-9a7a-14a1f2cf15c8">1</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_b7a26b57-1d63-4da0-8c1b-d24cd01ae8d7">MC</para>
                  </td>
                  <td align="left" colspan="4" rowspan="1">
                    <para xml:id="para_c1ba22d1-7348-496a-ba11-df2a894638a9">
<xref linkend="sect_Mapping_TID_1003" xrefstyle="select: labelnumber quotedtitle nopage"/>
</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_8dde2d10-99a4-4124-b35e-8689d8d3f887">IFF Row 1 value = <olink targetdoc="PS3.16" targetptr="DCM_121006">(121006, DCM, "Person")</olink> or Row 1 is absent</para>
                  </td>
                </tr>
                <tr valign="top">
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_8ad9055a-1fbd-4a6e-b842-ecff7918b85d">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt;</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_ed624f42-019e-4500-a367-5b37560cd92b">INCLUDE</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_89935b0a-4f88-4216-9ecc-61c1fea3369f">1</para>
                  </td>
                  <td align="center" colspan="1" rowspan="1">
                    <para xml:id="para_87c3685d-f41a-4bcd-9f0c-ba4f35c9806b">MC</para>
                  </td>
                  <td align="left" colspan="4" rowspan="1">
                    <para xml:id="para_b2dfa196-508b-4b72-b958-0e72242ae11a">TID 1004 "Device Observer Identifying Attributes"</para>
                  </td>
                  <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_1738e0df-b343-43dc-b01a-938205935681">Not used in AIM; IFF Row 1 value = <olink targetdoc="PS3.16" targetptr="DCM_121007">(121007, DCM, "Device")</olink>
</para>
                  </td>
                </tr>
              </tbody>
            </table>
          </section>
          
                   <section label="TID 1003" status="3" xml:id="sect_Mapping_TID_1003">
            <title>Mapping of Person Observer Identifying Attributes</title>
            <para xml:id="para_1de30482-63d2-4cde-8e62-771d39f5e4cf">This section describes the mapping of <olink targetdoc="PS3.16" targetptr="sect_TID_1003" xrefstyle="select: docname labelnumber quotedtitle"/>.</para>
            <table frame="box" label="TID 1003" rules="all" xml:id="table_Mapping_TID_1003">
            <caption>Mapping of Person Observer Identifying Attributes</caption>
               <thead>
                  <tr valign="top">
                     <th align="center" colspan="1" rowspan="1">
                       <para xml:id="para_44a783b0-c834-4663-8a7b-d1dd5000d912">DICOM SR Path</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_c254fd1c-036d-42f9-bb11-ab2d7856d725">DICOM VT</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_a92a1f73-eae8-4ab1-8d68-a66259ced520">DICOM VM</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_c75f93f3-ba46-49da-b6f9-1faad29a4f0e">DICOM Usage Type</para>
                     </th>
                    <th align="center" colspan="1" rowspan="1">
                      <para xml:id="para_965944e5-6a78-4950-b050-15536a8eb62f">Generated Value</para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                      <para xml:id="para_21a79f0c-ae9d-404d-a90a-6e33055ce3eb">AIM Element or Attribute</para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                      <para xml:id="para_e1a67298-39f7-46b7-b517-609e6d98d956">AIM Data Type</para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                      <para xml:id="para_03be0abc-022d-4e9a-8620-4d6c73752b11">AIM Multiplicity</para>
                    </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_92397fae-17be-405d-8f97-a07d5bf1760a">Comment</para>
                     </th>
                  </tr>
               </thead>
               <tbody>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_0941316d-1c41-46f0-9983-e577b13c9a7d">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_121008">(121008, DCM, "Person Observer Name")</olink>
</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_3acccb8d-597b-4aed-ad43-5ffb3ecac651">PNAME</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_7b23afe7-1ce5-4f52-b9df-2fee17612301">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_f48b4fe3-7529-4681-9d77-8578a72f82a7">M</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_fa438234-9445-44e3-903d-838074d51ae6"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                       <para xml:id="para_adc1a8b2-1b53-4c31-a95b-f1a41f502e55">Image​Annotation​​Collection/​user/​name/​@value</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_9a238d2d-aa03-4889-baf0-4cb877b4822a">ST</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_98c852f5-065e-4bcb-9d24-b513d53ab98a">1..1</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_cc7cfe6b-308b-476c-ad6b-939597417f4e"/>
                     </td>
                  </tr>
                 <tr valign="top">
                   <td align="left" colspan="1" rowspan="1">
                     <para xml:id="para_e92a634e-2711-4290-847e-3b7d5721c7fe">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_128774">(128774, DCM, "Person Observer's Login Name")</olink>
</para>
                   </td>
                   <td align="left" colspan="1" rowspan="1">
                     <para xml:id="para_419b9a80-ee3f-428e-8536-ad047d7585ef">PNAME</para>
                   </td>
                   <td align="center" colspan="1" rowspan="1">
                     <para xml:id="para_f2495bbd-c559-4ceb-8268-e5daa75c7304">1</para>
                   </td>
                   <td align="center" colspan="1" rowspan="1">
                     <para xml:id="para_063e1238-987a-464b-a0ab-7d7ef7dc3a87">M</para>
                   </td>
                   <td align="left" colspan="1" rowspan="1">
                     <para xml:id="para_c202d66c-8fab-4a37-a51a-22779434b3c1"/>
                   </td>
                   <td align="left" colspan="1" rowspan="1">
                     <para xml:id="para_13219e0b-60d4-4f40-b570-f6531eece2d4">Image​Annotation​​Collection/​user/​loginName/​@value</para>
                   </td>
                   <td align="left" colspan="1" rowspan="1">
                     <para xml:id="para_6f3889ba-474b-4738-9e03-12cfdb19e7d3">ST</para>
                   </td>
                   <td align="left" colspan="1" rowspan="1">
                     <para xml:id="para_79ee3c31-484f-4f16-a764-c7d55fa3373f">1..1</para>
                   </td>
                   <td align="left" colspan="1" rowspan="1">
                     <para xml:id="para_fcd7deea-e718-41f5-bc7b-d570f6923903"/>
                   </td>
                 </tr>
                 <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_5ab785a0-143a-4ef1-9748-449b9749a559">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_121009">(121009, DCM, "Person Observer's Organization Name")</olink>
</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_9d74559c-b896-4f83-a6c3-923e419b13d9">TEXT</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_244b4d1a-f19c-4236-9754-c789c103db11">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_99bf65cf-9c02-418d-bc91-f13bb921bb79">U</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_b2905d6d-189e-4daf-b340-f301d0138cac"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_278688b7-dbd0-421b-ae5e-cf366232becf"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_b52afe38-615f-4c14-ba7b-715f345ba6fb"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_30dc3664-e59c-4ac0-950b-10f55749463e"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_240e2cd2-6ad6-4f8c-ab96-6f80f9428c0c">Not used in AIM.</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_1323b967-5adf-457a-9097-b299b4907885">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_121010">(121010, DCM, "Person Observer's Role in the Organization")</olink>
</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_c20ee8bf-47f4-43db-8d18-bc2b836cb974">CODE</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_498e16c2-06dd-4261-8b55-ebd771cb587c">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_0ca61610-c847-4179-9de6-4e9b8eb54ca2">U</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_fd5c9086-6310-4289-8fd1-fb32acebf9bc"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                       <para xml:id="para_dff25304-1883-412e-900a-3948417cd51b"> </para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_2fd0e432-1df5-4052-ba45-c70c76361109"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_cafede24-a2e2-4c56-bf4b-b370f5ab21e0"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_16bd5b6c-d8e3-42c2-9587-c986e5fe7745">Not used in AIM.</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_01a75035-2a53-46d1-a82c-cea86380ac04">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_121011">(121011, DCM, "Person Observer's Role in this Procedure")</olink>
</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_12089639-2f75-4a98-bbd4-958152a936b2">CODE</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_b3a8f26b-595f-408a-9569-560733368d04">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_e525eec0-8373-4d06-9379-659cd630d8b5">U</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_8bc015b9-774e-4209-84f9-5d51d4edf263"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                       <para xml:id="para_ea0ee5db-cc99-4dc6-bd7e-9c826f568d91">Image​Annotation​​Collection/​user/​roleInTrial/​@value</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_fc4ecc99-d106-42f8-8acc-b5d7c2a4c625">ST</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_8291ee5e-155e-4270-a1db-836f301681f1">0..1</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_9bc650ba-d7c7-4b5b-94ea-be876b9db08d">B<olink targetdoc="PS3.16" targetptr="sect_CID_7453" xrefstyle="select: labelnumber quotedtitle"/>
</para>
                       <para xml:id="para_81841d4a-a1cd-4411-9257-d13aab1950be">Values mapped are amended in CP 1734.</para>
                    <para xml:id="para_e2ec0814-eca1-47aa-a5ca-da7776686235">AIM does not define a value set for the roles, so no standard mapping to the DICOM codes is defined.</para>
                     </td>
                  </tr>
                 <tr valign="top">
                   <td align="left" colspan="1" rowspan="1">
                     <para xml:id="para_50b6706d-9d3c-4c1d-b7b4-50f98bf75b30">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_121011">(121011, DCM, "Person Observer's Role in this Procedure")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_128775">(128775, DCM, "Identifier within Person Observer's Role")</olink>
</para>
                   </td>
                   <td align="left" colspan="1" rowspan="1">
                     <para xml:id="para_09699e54-854e-4c49-b02f-9c41bfb13ef0">TEXT</para>
                   </td>
                   <td align="center" colspan="1" rowspan="1">
                     <para xml:id="para_a33c8fbd-241f-428c-a1da-286d802271e7">1</para>
                   </td>
                   <td align="center" colspan="1" rowspan="1">
                     <para xml:id="para_54564482-4cdc-4900-9c4d-4a8d62d5a6af">U</para>
                   </td>
                   <td align="left" colspan="1" rowspan="1">
                     <para xml:id="para_6cf26c05-b282-4620-bbc9-81136bf7c118"/>
                   </td>
                   <td align="left" colspan="1" rowspan="1">
                     <para xml:id="para_f0b705d5-34ab-4ec6-8960-d0bcde1cc8cb">Image​Annotation​​Collection/​user/​numberWithinRoleOfClinicalTrial/​@value</para>
                   </td>
                   <td align="left" colspan="1" rowspan="1">
                     <para xml:id="para_4cf40dd5-0b30-41f2-b2e5-18dfddc5770a">INT</para>
                   </td>
                   <td align="left" colspan="1" rowspan="1">
                     <para xml:id="para_a76b9956-2afa-4be7-9a7e-c569d349ba50">0..1</para>
                   </td>
                   <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_7c40c7cf-6235-4e85-b264-c45fb95db332">DICOM allows for alphanumeric, whereas AIM is INT only.</para>
                   </td>
                 </tr>
               </tbody>
            </table>
         </section>
          
         <section label="TID 1600" status="3" xml:id="sect_Mapping_TID_1600">
            <title>Mapping of Image Library</title>
            <para xml:id="para_61ae2047-1ae3-4c45-a738-39ced94dd5ce">This section describes the mapping of <olink targetdoc="PS3.16" targetptr="sect_TID_1600" xrefstyle="select: docname labelnumber quotedtitle"/>.</para>
            <table frame="box" label="TID 1600" rules="all" xml:id="table_Mapping_TID_1600">
            <caption>Mapping of Image Library</caption>
               <thead>
                  <tr valign="top">
                     <th align="center" colspan="1" rowspan="1">
                       <para xml:id="para_0b438193-283e-4e9b-abb8-95fcc6c8afaf">DICOM SR Path</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_654bbd57-c5d6-4492-8693-c131f449c8f5">DICOM VT</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_159c42ab-2740-4381-b096-8597b05a95d1">DICOM VM</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_94f4c451-b0d8-47b2-84fc-cfb4a04b2ba1">DICOM Usage Type</para>
                     </th>
                    <th align="center" colspan="1" rowspan="1">
                      <para xml:id="para_6eb67db3-5a9c-4db8-b00f-5e2faf2160a9">Generated Value</para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                      <para xml:id="para_82e1be6f-bb86-435e-a629-81ae2ca07627">AIM Element or Attribute</para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                      <para xml:id="para_f6ab1655-68bc-49f4-978e-1eb29222ba9a">AIM Data Type</para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                      <para xml:id="para_8ce8015a-4233-4762-ae8d-fbb12a00cc77">AIM Multiplicity</para>
                    </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_b058b532-0941-4c59-b02d-487c399f0794">Comment</para>
                     </th>
                  </tr>
               </thead>
               <tbody>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_4a9a940d-d492-40a6-bb05-17021e7c5d42">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_111028">(111028, DCM, "Image Library")</olink>
</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_814639a8-e2aa-48cf-b667-c042c72d3daf">CONTAINER</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_9b6632db-2636-4978-8740-f0e488020778">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_fb1b48d1-7d10-489d-ad6c-970d320a37b3">M</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_5a85408b-13a3-4d9c-8114-c2c78c5296f8"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_23ee84d8-5283-49c3-bbd4-45e6918d3481"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_2a3228a0-b613-4ef7-ad70-2456233cdbdf"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_7a9b6d19-da38-46f0-8601-8e66c13c29bb"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_426596a5-87bf-42af-bdf7-95b1d89a642c"/>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_0a12cc68-e5ac-48f9-9f90-4a7fcb6086f6">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_111028">(111028, DCM, "Image Library")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126200">(126200, DCM, "Image Library Group")</olink>
</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_5bd3331b-edd7-4b4a-9691-4de6be0d8b32">CONTAINER</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_fb76a76f-1f72-4175-8e1c-959e97693acd">1-n</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_b42e52f2-43c8-431f-b833-ecdb451074db">U</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_20f5a748-2e0f-412d-82c7-5a3ed3406704"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                         <para xml:id="para_1103889f-9940-4d41-ac0a-9052374fbbe2">/Image​Annotation​Collection/​image​Annotations/​Image​Annotation/​imageReferenceEntityCollection/​ImageReferenceEntity</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_cd290c13-adfe-4297-a44d-c48137875685"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_f7748499-4e90-488b-95f0-ff6105b232c2"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                         <para xml:id="para_f08d99ae-e62a-4d8e-9ec8-6e8aef27a2c9">The value of aim:unique​Identifier/@root is mapped to the Observation​UID Attribute of the CONTAINER.</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_adefd0c2-043a-4640-bf7c-00527b9def5d">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_111028">(111028, DCM, "Image Library")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126200">(126200, DCM, "Image Library Group")</olink> &gt; </para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_12df68c4-b260-4ea5-b0a4-92e0c74d3fd7">INCLUDE</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_9eb34836-7620-479c-88f1-3d0ea78fc187">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_f3766a90-81f8-4772-8259-d944eda4aede">U</para>
                     </td>
                     <td align="left" colspan="4" rowspan="1">
                        <para xml:id="para_37344090-02d3-4d56-b1bd-c26a9212c04a">
<xref linkend="sect_Mapping_TID_1602" xrefstyle="select: labelnumber quotedtitle nopage"/>
                    </para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_c867ade4-fff9-410a-bd51-2d1a0738a60b"/>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_767cce57-0354-4364-985c-0cb7ef434ea6">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_111028">(111028, DCM, "Image Library")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126200">(126200, DCM, "Image Library Group")</olink> &gt; </para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_42b431e3-5164-4159-a4ee-880ce298af28">INCLUDE</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_34766fcb-6772-40c0-8285-1236f79b8b49">1-n</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_b7ec7a1f-460b-482e-aed9-3254d0383c5b">U</para>
                     </td>
                     <td align="left" colspan="4" rowspan="1">
                        <para xml:id="para_a4b1cf49-9671-41bf-92d2-c464938088b4">
<xref linkend="sect_Mapping_TID_1601" xrefstyle="select: labelnumber quotedtitle nopage"/>
                    </para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_00cb7a4d-5f06-4460-9d0a-fbce4c67d0bf"/>
                     </td>
                  </tr>
               </tbody>
            </table>
         </section>
         <section label="TID 1601" status="3" xml:id="sect_Mapping_TID_1601">
            <title>Mapping of Image Library Entry</title>
            <para xml:id="para_49c3cf1c-04d8-4858-aaaf-bd1f9be8f073">This section describes the mapping of <olink targetdoc="PS3.16" targetptr="sect_TID_1601" xrefstyle="select: docname labelnumber quotedtitle"/>.</para>
            <table frame="box" label="TID 1601" rules="all" xml:id="table_Mapping_TID_1601">
            <caption>Mapping of Image Library Entry</caption>
               <thead>
                  <tr valign="top">
                     <th align="center" colspan="1" rowspan="1">
                       <para xml:id="para_b125a405-0e33-4929-8655-3018ae3401be">DICOM SR Path</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_095553ea-2a2b-4e9b-836f-da06a062221f">DICOM VT</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_c731719d-723e-4d1f-8d58-0942e3a891d6">DICOM VM</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_dfb280d7-ac50-43b2-88c6-4fd595a73896">DICOM Usage Type</para>
                     </th>
                    <th align="center" colspan="1" rowspan="1">
                      <para xml:id="para_24f5117c-8885-4c24-8cf0-bf9c6d1cae52">Generated Value</para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                      <para xml:id="para_e063c052-9033-4e8b-9f9d-45d23915a1a4">AIM Element or Attribute</para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                      <para xml:id="para_7209a5a2-6b64-4fb4-9d7e-c5e5123e3a53">AIM Data Type</para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                      <para xml:id="para_3a7bf180-ec42-4fb7-b275-059a090e90ab">AIM Multiplicity</para>
                    </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_09904d88-8071-4163-b825-1fe73e7bed98">Comment</para>
                     </th>
                  </tr>
               </thead>
               <tbody>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_0100e482-42a1-43eb-b9ca-72fd476e1228">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_111028">(111028, DCM, "Image Library")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126200">(126200, DCM, "Image Library Group")</olink> &gt; IMAGE</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_15cad349-18da-423a-996c-2ff2000f8024">IMAGE</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_f5331669-9f3a-478b-9450-75b82fe63554">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_074523bf-eb90-4906-8a7a-946a94d8a1db">M</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_9650efad-3eaf-4dd4-b44e-779ebb2dec01"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_7f7219b4-aab4-4941-b022-005d50150c0f">/Image​Annotation​Collection/​image​Annotations/​Image​Annotation/​imageReferenceEntityCollection/​ImageReferenceEntity/​imageStudy/​imageSeries/​image​Collection/​Image/​sopInstanceUid/​@root​, sopClassUid/​@root​</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_6676b333-9e50-48b4-9058-a9152317cf5b">II, II</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_3a470511-448c-4f4c-bdef-467f7cf60621">1, 1</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_14507e1e-ea7e-4d39-a1f9-842d12d460b3"/>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_46988a38-d1f3-428b-9b5f-a5752bc5e16e">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_111028">(111028, DCM, "Image Library")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126200">(126200, DCM, "Image Library Group")</olink> &gt; IMAGE &gt;</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_37e84f95-2e25-4585-a1f8-6fb8dd324f1e">INCLUDE</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_a106564c-af3d-4adc-bfd4-e1af29f0d2c9">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_7ca671f8-336c-4666-a92d-2c138e8df311">U</para>
                     </td>
                     <td align="left" colspan="4" rowspan="1">
                        <para xml:id="para_dd31769d-2c36-4d74-adac-7a8df467fea7">
<xref linkend="sect_Mapping_TID_1602" xrefstyle="select: labelnumber quotedtitle nopage"/>
                    </para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_278d77a6-efdf-4262-836b-33502638679b"/>
                     </td>
                  </tr>
               </tbody>
            </table>
         </section>
         <section label="TID 1602" status="3" xml:id="sect_Mapping_TID_1602">
            <title>Mapping of Image Library Entry Descriptors</title>
            <para xml:id="para_de6c2cfe-f110-4a8f-884b-f8de2a37a16c">This section describes the mapping of <olink targetdoc="PS3.16" targetptr="sect_TID_1602" xrefstyle="select: docname labelnumber quotedtitle"/>.</para>
            <table frame="box" label="TID 1602" rules="all" xml:id="table_Mapping_TID_1602">
            <caption>Mapping of Image Library Entry Descriptors</caption>
               <thead>
                  <tr valign="top">
                     <th align="center" colspan="1" rowspan="1">
                       <para xml:id="para_928a047d-6450-4f94-8201-e485112c102b">DICOM SR Path</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_b2d427ba-0d79-4bf5-bb22-ec4b96f2bb4f">DICOM VT</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_500b7a2e-9133-4bd6-9117-e51f6976ac86">DICOM VM</para>
                     </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_8fe311c2-39ee-439a-82db-8a83cae3b571">DICOM Usage Type</para>
                     </th>
                    <th align="center" colspan="1" rowspan="1">
                      <para xml:id="para_0aefcabf-75b8-4eec-ae7e-105591033e70">Generated Value</para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                      <para xml:id="para_ca75011a-4d08-45ec-ab60-be19a2e6c2f7">AIM Element or Attribute</para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                      <para xml:id="para_8c2e4575-06e8-4cb0-a1f2-3e2305473a32">AIM Data Type</para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                      <para xml:id="para_407e7480-d8c0-47f1-b307-ad4e20d5c0a2">AIM Multiplicity</para>
                    </th>
                     <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_5b7b9560-40fd-4446-a32c-9c7be631b53d">Comment</para>
                     </th>
                  </tr>
               </thead>
               <tbody>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_186d6a9a-60fa-4d05-b06d-0c8d8bca30a0">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_111028">(111028, DCM, "Image Library")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126200">(126200, DCM, "Image Library Group")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_121139">(121139, DCM, "Modality")</olink>
</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_9550d9d2-f179-4516-a2e8-7085e57667ee">CODE</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_5beb076c-b8ff-40f4-91f1-137c010492be">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_73ca6cad-224a-4f26-82d9-923a79ddd416">U</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_70b173be-1170-4dab-a18f-2e20f74df1a8"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                       <para xml:id="para_8eded41b-0a1e-4585-9a6f-e4823d3812cb">/Image​Annotation​Collection/​image​Annotations/​Image​Annotation/​imageReferenceEntityCollection/​ImageReferenceEntity/​imageStudy/​imageSeries/​modality/​@code</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_1994d791-aceb-4baa-adf8-aedd0ec74167">CD</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_84f578ef-c1df-4599-a166-d8b8dfcc4b7a">1..1</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                    <para xml:id="para_e348485b-1229-47d2-a3b4-ccce03f7ae49">AIM does not formally define a value set but common usage is the set of code strings defined for the image Modality Attribute, and these have a 1:1 correspondence with the code values of <olink targetdoc="PS3.16" targetptr="sect_CID_29" xrefstyle="select: labelnumber quotedtitle"/>.</para>
                  </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_9469640a-00c8-4b24-9386-abe0ba221bea">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_111028">(111028, DCM, "Image Library")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126200">(126200, DCM, "Image Library Group")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_123014">(123014, DCM, "Target Region")</olink>
</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_7dac670f-9804-4c78-ac24-2e9eb88b2922">CODE</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_ad7090c4-7b2a-46ad-8eb2-bb0a1ebf7108">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_c884a5ea-85cf-48b6-b2a0-5d55923cc9cd">U</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_73740e09-312b-4a2b-962b-e3e4acf5b216"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                       <para xml:id="para_0cb80542-e835-4d35-8af2-5c4186ccef13"> </para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_16e85589-6b8b-4210-89a2-c41c10f0158a"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_994be73d-532c-4239-b8e3-860eb5e9de1e"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_51205b5f-462e-4f91-8971-ba511743646a">Not used in AIM; discarded if present in DICOM SR.</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_ddfc90a7-6854-4f2c-bd88-3d1d14ac4f04">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_111028">(111028, DCM, "Image Library")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126200">(126200, DCM, "Image Library Group")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_111027">(111027, DCM, "Image Laterality")</olink>
</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_b87582f0-0499-495a-addb-c6f6160af2c2">CODE</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_69f5a2c7-5628-4c5f-b171-230433162e89">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_623a9770-a1b7-4e12-a629-6a65afb3f888">U</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_c8f7b11b-5d4d-4d5e-a280-c7e723a38a77"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                       <para xml:id="para_a265dae4-0e78-41d4-91c3-f1f26d34b305"> </para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_6c90afc4-3690-4720-9a1f-8cb6977cf3c8"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_eb96ae08-8a7d-4b1e-ab0c-707bb905bc65"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_3cece162-acfe-48aa-a6ba-d5423530d8d4">Not used in AIM; discarded if present in DICOM SR.</para>
                     </td>
                  </tr>
                   <tr valign="top">
                       <td align="left" colspan="1" rowspan="1">
                           <para xml:id="para_025205a8-bd89-4725-adc1-91536119daf4">126000, DCM, "Imaging Measurement Report") &gt; (111028, DCM, "Image Library") &gt; (126200, DCM, "Image Library Group") &gt;(121022, DCM, "Accession Number")</para>
                       </td>
                       <td align="left" colspan="1" rowspan="1">
                           <para xml:id="para_833408ce-8ec3-435f-915c-4d24d9cb2466">TEXT</para>
                       </td>
                       <td align="center" colspan="1" rowspan="1">
                           <para xml:id="para_c161e2d8-f617-41cf-88c5-dc1dad060b4e">1</para>
                       </td>
                       <td align="center" colspan="1" rowspan="1">
                           <para xml:id="para_4785ffa7-3392-4a93-82ca-0014b794b2df">U</para>
                       </td>
                       <td align="left" colspan="1" rowspan="1">
                           <para xml:id="para_ac762881-cdce-48ef-a5b8-ecb9c2bcc5ca"/>
                       </td>
                       <td align="left" colspan="1" rowspan="1">
                           <para xml:id="para_f49dffc6-aaf6-46a6-a3af-dc9c5eb7a757">/Image​Annotation​Collection/​image​Annotations/​Image​Annotation/​image​Reference​Entity​Collection/​Image​Reference​Entity/​image​Study/​accession​Number/​@value</para>
                       </td>
                       <td align="left" colspan="1" rowspan="1">
                           <para xml:id="para_40942fc8-6d16-4a45-ae56-3e25649bc20c">ST</para>
                       </td>
                       <td align="left" colspan="1" rowspan="1">
                           <para xml:id="para_9dc1b9f9-70a9-4bd8-9d61-00bbcb9655ce">0..1</para>
                       </td>
                       <td align="left" colspan="1" rowspan="1">
                           <para xml:id="para_61cd6bd6-bb09-4694-a05b-05e356211163"/>
                       </td>
                   </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_7806d53e-3b3c-41e4-9998-cea021b14ae5">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_111028">(111028, DCM, "Image Library")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126200">(126200, DCM, "Image Library Group")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_111060">(111060, DCM, "Study Date")</olink>
</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_bdf85b99-56ef-44ab-8b45-1325715db4d3">DATE</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_173cbbb2-a9d9-459d-ba75-4a5a40af4c37">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_747ef794-f3e6-4693-8194-3f60d62245c3">U</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_aa53d6fa-cb88-4138-b436-6a16e86ccc2d"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_a0cdace6-5a43-4791-a4e7-4cd70accd34d">/Image​Annotation​Collection/​image​Annotations/​Image​Annotation/​imageReferenceEntityCollection/​ImageReferenceEntity/​imageStudy/​startDate/​@value</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_0a21660d-8f79-4b82-92ac-1032ebd3dfe8">TS</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_c2c709bf-8820-4d2b-82a0-b68ed11dc270">1..1</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_83283d82-ca46-4b96-82f6-0d3ea43011de"/>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_d1dc9d00-d4e9-46b2-965e-e277127b6188">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_111028">(111028, DCM, "Image Library")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126200">(126200, DCM, "Image Library Group")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_111061">(111061, DCM, "Study Time")</olink>
</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_9905044c-7fe5-46ee-b126-be36708e689a">TIME</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_78673d2c-82ca-4957-b763-d61e3c537cb5">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_896e7cc0-c19f-4ac1-ad7c-7662ade5359d">U</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_b70e73cb-3a26-4986-9f85-6bca4468300f"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_1319866b-842b-412a-a288-70f2c5a69633">/Image​Annotation​Collection/​image​Annotations/​Image​Annotation/​imageReferenceEntityCollection/​ImageReferenceEntity/​imageStudy/​startTime/​@value</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_8ad879fe-37ce-433a-bbaa-82f460d0323c">TS</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_d50af4bf-a6c3-43c8-bb30-d4786ff3dabd">1..1</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_f0bdb38c-d015-43d2-8759-ae68648dfc49"/>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_09b8bf40-e3b0-415a-a9d2-6f4573220aa7">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_111028">(111028, DCM, "Image Library")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126200">(126200, DCM, "Image Library Group")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_111018">(111018, DCM, "Content Date")</olink>
</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_0f1c6690-8392-42f3-bfff-1c74bf49c4ae">DATE</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_96920b10-f4c4-438d-a548-708b349cda33">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_57db5934-7c16-4059-9fb5-af7eaaaf6387">U</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_b4aa9963-30f8-444d-b907-cb2785525698"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_cb76e648-f978-4d6b-be97-582f2aa7814b"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_d0afcbde-f455-44b9-8461-0d52a261bb74"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_e9752218-61b9-495b-9f7c-e51856204cd7"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_acc0745b-cf29-4c49-af2c-aa07b0c3c774">Not used in AIM; discarded if present in DICOM SR. </para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_00f8b972-a307-4951-830f-786451d0dd7c">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_111028">(111028, DCM, "Image Library")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126200">(126200, DCM, "Image Library Group")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_111019">(111019, DCM, "Content Time")</olink>
</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_951cb7e8-93ce-403d-99ca-87388f02994b">TIME</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_b88da640-1f7e-4806-b1c8-601cacc1a3d7">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_a7c2c48f-b203-4a67-8132-ba4b190a3ebf">U</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_e9fe4c6d-5c1f-4be7-9ff8-395b3fa02663"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_a44edf1a-60cb-4fbf-89d9-cb134d34209e"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_b0f8cb0c-0a4b-4161-a78f-a83a013fb878"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_203adddb-e06e-41d3-9ed4-cc14737b755c"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_7f63e7a5-4652-4521-8210-a9845385fbc4">Not used in AIM; discarded if present in DICOM SR. </para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_ad6e3a86-7b14-4adf-ae49-e6e3cb3d9cbe">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_111028">(111028, DCM, "Image Library")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126200">(126200, DCM, "Image Library Group")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126201">(126201, DCM, "Acquisition Date")</olink>
</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_8de4dc90-9d03-4c96-828d-2144bf567229">DATE</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_080852b9-10f3-4dcb-9e0c-8ee922a026b5">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_476039b9-21a9-4e36-ac3c-8a7d758f8764">U</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_b1cf441c-4aa7-4a51-a2ac-6be7328d8a90"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_6127f6ea-575f-4704-b48f-078b3e94bfdc"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_b73690ca-d577-44a9-873e-0d73b6ca17f7"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_cbfb0467-474d-49e6-8f08-782449b3c971"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_1ca6bc40-1eba-4e8f-b838-cc4cce57b74c">Not used in AIM; discarded if present in DICOM SR. </para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_1a85f687-dc5f-4a2d-84e5-d8cc1ffbe2b8">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_111028">(111028, DCM, "Image Library")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126200">(126200, DCM, "Image Library Group")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126202">(126202, DCM, "Acquisition Time")</olink>
</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_27b173ee-574f-444b-b6ff-49f984eb3760">TIME</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_3fc82ebd-df2a-44e5-bfdc-ca697ceb1c26">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_0e38d966-54be-4375-9b37-183e833b678a">U</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_d407dcd5-4379-43fd-b703-86a7366ea826"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_61f03fce-d637-4203-921f-ababdb69393d"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_e4c2fdf2-3317-4303-af09-a8b4ec832c1b"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_58faefea-e959-4235-a541-bbb4dbbfa6d3"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_b9577991-09e3-40a7-b923-a402e64cb3b8">Not used in AIM; discarded if present in DICOM SR. </para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_a295c3e8-d257-4a42-a1d3-0c45b06464e7">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_111028">(111028, DCM, "Image Library")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126200">(126200, DCM, "Image Library Group")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_112227">(112227, DCM, "Frame of Reference UID")</olink>
</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_0f250087-37b5-4d10-a217-c82d5d7c6765">UIDREF</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_2bf3a2c1-1d91-4f46-8fd1-80bce50a749f">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_ed08813f-1628-4ebc-a82d-f00969d38be8">U</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_a5528d9d-0b83-4969-8031-df4429f9040b"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_42e61e66-7f23-46bb-b962-e23888b615d0"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_6edc8c88-d0da-4bcb-94c8-16984d0eec78"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_58b80ded-3e6c-4f4a-a7eb-e5f2dfa93883"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_7c02a4e2-7e45-4434-8e99-68bf1d858cee">Not used in AIM; discarded if present in DICOM SR. </para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_1a8d9ec1-d4e0-4e1e-bb2a-d31971e740e4">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_111028">(111028, DCM, "Image Library")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126200">(126200, DCM, "Image Library Group")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_110910">(110910, DCM, "Pixel Data Rows")</olink>
</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_ea272ab0-d799-477c-a2e5-4bf8262d463e">NUM</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_0c9b49dc-bff6-4474-99bd-fec238671016">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_a2eac657-77e2-4ccd-8c76-129e2f5a7e9f">U</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_24b0b18e-2d7c-4b84-a963-3cfbcce44235"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_e757f46a-791d-4eb6-9baf-041d04c2236f"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_4744980f-8a6c-4fca-8c26-a0db162e781c"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_24208aed-0bd4-4cec-acca-8c765815abd2"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_97c3158f-380c-4f66-b12c-ea2e9cce961d">Not used in AIM; discarded if present in DICOM SR.</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_33818608-f772-42d9-8345-74d10b11afc4">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_111028">(111028, DCM, "Image Library")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126200">(126200, DCM, "Image Library Group")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_110911">(110911, DCM, "Pixel Data Columns")</olink>
</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_95550d0e-9baa-4b45-b710-444d85b38f82">NUM</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_a44bb575-a26d-4541-8b79-0a11ecef0551">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_78a66041-5eb3-475d-9fb4-81316ad32ef0">U</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_7fcf4c58-a1d0-4c13-bf8a-b726673da3ac"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_cb6fd122-ea36-44ef-ae94-954ac2cb448b"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_022d8e16-218a-42b7-9d3b-d5a004333089"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_e4310b03-7700-4328-b548-7f213783e82f"/>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_70b0b8c0-9e9d-4da2-ad3c-1b36bc77f43a">Not used in AIM; discarded if present in DICOM SR.</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_5caf3472-214a-4076-aa8a-531c3fc44736">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_111028">(111028, DCM, "Image Library")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126200">(126200, DCM, "Image Library Group")</olink> &gt;</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_94f59676-a88b-4640-8e41-7fbf7d2b5e11">INCLUDE</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_33af1578-4249-4a97-8360-92c93aa034d4">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_6cae3a05-fadf-4942-9710-23ba5f9a192a">U</para>
                     </td>
                     <td align="left" colspan="4" rowspan="1">
                        <para xml:id="para_71cb3fac-15ad-485d-b42f-1695fe2a048a">TID 1603 "Image Library Entry Descriptors for Projection Radiography"</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_97123a3e-7817-44e4-93de-3ac17b1f358d">Not used in AIM; discarded if present in DICOM SR. </para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_9efbea1b-9f5c-49c0-b3a6-65f7cf66cee2">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_111028">(111028, DCM, "Image Library")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126200">(126200, DCM, "Image Library Group")</olink> &gt;</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_61e174b0-0ab2-48d1-b472-1fe8f70e2e74">INCLUDE</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_8c7ef19e-d6fe-42be-b8d0-cedf79a33ff3">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_e1c89a8b-50d4-4b80-ab97-f05c3040dbbb">U</para>
                     </td>
                     <td align="left" colspan="4" rowspan="1">
                        <para xml:id="para_4218026f-cdf2-4f0b-95c7-d275477e2042">TID 1604 "Image Library Entry Descriptors for Cross-Sectional Modalities"</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_66f7a3f3-14fe-4526-aa74-df7739782fc1">Not used in AIM; discarded if present in DICOM SR.</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_98ad9212-8bc9-4bf1-ac0c-9687f5d6c43b">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_111028">(111028, DCM, "Image Library")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126200">(126200, DCM, "Image Library Group")</olink> &gt;</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_1b223702-8567-4fcd-812d-9654e6280b84">INCLUDE</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_43f4f27d-6731-4ec5-9866-0b73efa27825">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_f50bc954-f4da-4b11-ad90-17d5acdbd48b">U</para>
                     </td>
                     <td align="left" colspan="4" rowspan="1">
                        <para xml:id="para_566e7d3b-81c5-41a1-83da-fece25c74736">TID 1605 "Image Library Entry Descriptors for CT"</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_b10c8eed-8a41-4e07-9402-a639ebc40f85">Not used in AIM; discarded if present in DICOM SR.</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_f2a60a4a-2247-4763-996c-b0b9caea7d13">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_111028">(111028, DCM, "Image Library")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126200">(126200, DCM, "Image Library Group")</olink> &gt;</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_399b4968-467c-49b7-b0c1-eeab09f26db9">INCLUDE</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_4f41b03d-ad4a-47b9-a0ab-3148f61b1b81">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_88da2557-c7f7-403f-9fd5-6beb98bd287c">U</para>
                     </td>
                     <td align="left" colspan="4" rowspan="1">
                        <para xml:id="para_d7ef3941-ad4d-4c03-9d84-17b95ba8d3d3">TID 1606 "Image Library Entry Descriptors for MR"</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_47007d6e-acb7-47e3-bfc8-1ad1799361f3">Not used in AIM; discarded if present in DICOM SR.</para>
                     </td>
                  </tr>
                  <tr valign="top">
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_6f15e93f-a77c-46a2-a671-3557f8a7c02b">
<olink targetdoc="PS3.16" targetptr="DCM_126000">(126000, DCM, "Imaging Measurement Report")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_111028">(111028, DCM, "Image Library")</olink> &gt; <olink targetdoc="PS3.16" targetptr="DCM_126200">(126200, DCM, "Image Library Group")</olink> &gt;</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_eac4e227-d347-48d1-94c5-d5715553e932">INCLUDE</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_3ed582f6-19a6-4b02-931c-bb0314dc687e">1</para>
                     </td>
                     <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_9bb7dba1-e27b-4fb6-94b0-7a6588f33630">U</para>
                     </td>
                     <td align="left" colspan="4" rowspan="1">
                        <para xml:id="para_4eed6532-a04f-4de0-88f6-b2818435261b">TID 1607 "Image Library Entry Descriptors for PET"</para>
                     </td>
                     <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_a5a2827a-a612-4d13-ab4f-58ec44a7426c">Not used in AIM; discarded if present in DICOM SR.</para>
                     </td>
                  </tr>
               </tbody>
            </table>
         </section>


          
        </section>
      </section>
    </section>
      
      <section label="A.7" status="2" xml:id="sect_A.7">
      <title>Sample Documents</title>
      <para xml:id="para_10475b60-1070-477e-97f5-6683db3e19c1">This section describes a sample AIM v4.2 instance and the same content transformed into a DICOM SR TID 1500 instance.</para>
      <section label="A.7.1" status="3" xml:id="sect_A.7.1">
        <title>Source AIM v4 Instance</title>
        <programlisting><![CDATA[<?xml version="1.0" encoding="UTF-8"?>
<ImageAnnotationCollection xmlns="gme://caCORE.caCORE/4.4/edu.northwestern.radiology.AIM" xmlns:rdf="http://www.w3.org/1999/02/22-rdf-syntax-ns#" xmlns:xsi="http://www.w3.org/2001/XMLSchema-instance" aimVersion="AIMv4_2" xsi:schemaLocation="gme://caCORE.caCORE/4.4/edu.northwestern.radiology.AIM AIM_v4.2_rv2_XML.xsd">
    <uniqueIdentifier root="2.25.224793923339609181243139195858254344686"/>
    <studyInstanceUid root="2.25.80159168229010751652502576830057032194"/>
    <seriesInstanceUid root="2.25.323817225444021135415209334192751441320"/>
    <accessionNumber value="AN5678AIM"/>
    <dateTime value="20170201180043"/>
    <user>
        <name value="Doe^Jane"/>
        <loginName value="jdoe"/>
        <roleInTrial/>
    </user>
    <equipment>
        <manufacturerName value="Acme Medical Systems"/>
        <manufacturerModelName value=""/>
        <softwareVersion value="36.00"/>
    </equipment>
    <person>
        <name value="CM-1-111-000000"/>
        <id value="293761767066931586407385203810190772174"/>
        <birthDate value="19600101000000"/>
        <sex value="M"/>
        <ethnicGroup/>
    </person>
    <imageAnnotations>
        <ImageAnnotation>
            <uniqueIdentifier root="2.25.56002466128627498886935079903172938041"/>
            <typeCode code="52988006" codeSystemName="SCT">
                <iso:displayName xmlns:iso="uri:iso.org:21090" value="Lesion"/>
            </typeCode>
            <dateTime value="20170201180043"/>
            <name value="Lesion1"/>
            <comment value="PT / WB NAC P600 / 0"/>
            <trackingUniqueIdentifier root="2.25.165294254063588909770717555738008800301"/>
            <calculationEntityCollection>
                <CalculationEntity>
                    <uniqueIdentifier root="2.25.51420968257530981243824658943871973198"/>
                    <typeCode code="126401" codeSystemName="DCM">
                        <iso:displayName xmlns:iso="uri:iso.org:21090" value="SUVbw"/>
                    </typeCode>
                    <typeCode code="255605001" codeSystemName="SCT">
                        <iso:displayName xmlns:iso="uri:iso.org:21090" value="Minimum"/>
                    </typeCode>
                    <description value="SUVbw Minimum"/>
                    <mathML/>
                    <calculationResultCollection>
                        <CalculationResult type="Scalar" xsi:type="CompactCalculationResult">
                            <unitOfMeasure value="g/ml{SUVbw}"/>
                            <dataType code="C48870" codeSystemName="NCI">
                                <iso:displayName xmlns:iso="uri:iso.org:21090" value="Double"/>
                            </dataType>
                            <dimensionCollection>
                                <Dimension>
                                    <index value="0"/>
                                    <size value="1"/>
                                    <label value="Minimum"/>
                                </Dimension>
                            </dimensionCollection>
                            <value value="1.98024"/>
                        </CalculationResult>
                    </calculationResultCollection>
                    
                </CalculationEntity>
                <CalculationEntity>
                    <uniqueIdentifier root="2.25.205292243885258032428819330909580896146"/>
                    <typeCode code="126401" codeSystemName="DCM">
                        <iso:displayName xmlns:iso="uri:iso.org:21090" value="SUVbw"/>
                    </typeCode>
                    <typeCode code="56851009" codeSystemName="SCT">
                        <iso:displayName xmlns:iso="uri:iso.org:21090" value="Maximum"/>
                    </typeCode>
                    <description value="SUVbw Maximum"/>
                    <mathML/>
                    <calculationResultCollection>
                        <CalculationResult type="Scalar" xsi:type="CompactCalculationResult">
                            <unitOfMeasure value="g/ml{SUVbw}"/>
                            <dataType code="C48870" codeSystemName="NCI">
                                <iso:displayName xmlns:iso="uri:iso.org:21090" value="Double"/>
                            </dataType>
                            <dimensionCollection>
                                <Dimension>
                                    <index value="0"/>
                                    <size value="1"/>
                                    <label value="Maximum"/>
                                </Dimension>
                            </dimensionCollection>
                            <value value="5.68816"/>
                        </CalculationResult>
                    </calculationResultCollection>
                    
                </CalculationEntity>
                <CalculationEntity>
                    <uniqueIdentifier root="2.25.70160252080234577167847509948368893276"/>
                    <typeCode code="126401" codeSystemName="DCM">
                        <iso:displayName xmlns:iso="uri:iso.org:21090" value="SUVbw"/>
                    </typeCode>
                    <typeCode code="373098007" codeSystemName="SCT">
                        <iso:displayName xmlns:iso="uri:iso.org:21090" value="Mean"/>
                    </typeCode>
                    <description value="SUVbw Mean"/>
                    <mathML/>
                    <calculationResultCollection>
                        <CalculationResult type="Scalar" xsi:type="CompactCalculationResult">
                            <unitOfMeasure value="g/ml{SUVbw}"/>
                            <dataType code="C48870" codeSystemName="NCI">
                                <iso:displayName xmlns:iso="uri:iso.org:21090" value="Double"/>
                            </dataType>
                            <dimensionCollection>
                                <Dimension>
                                    <index value="0"/>
                                    <size value="1"/>
                                    <label value="Mean"/>
                                </Dimension>
                            </dimensionCollection>
                            <value value="2.329186593407"/>
                        </CalculationResult>
                    </calculationResultCollection>
                    
                </CalculationEntity>
                <CalculationEntity>
                    <uniqueIdentifier root="2.25.140657026119469861895824082767088344984"/>
                    <typeCode code="126401" codeSystemName="DCM">
                        <iso:displayName xmlns:iso="uri:iso.org:21090" value="SUVbw"/>
                    </typeCode>
                    <typeCode code="386136009" codeSystemName="SCT">
                        <iso:displayName xmlns:iso="uri:iso.org:21090" value="Standard Deviation"/>
                    </typeCode>
                    <description value="SUVbw Standard Deviation"/>
                    <mathML/>
                    <calculationResultCollection>
                        <CalculationResult type="Scalar" xsi:type="CompactCalculationResult">
                            <unitOfMeasure value="g/ml{SUVbw}"/>
                            <dataType code="C48870" codeSystemName="NCI">
                                <iso:displayName xmlns:iso="uri:iso.org:21090" value="Double"/>
                            </dataType>
                            <dimensionCollection>
                                <Dimension>
                                    <index value="0"/>
                                    <size value="1"/>
                                    <label value="Standard Deviation"/>
                                </Dimension>
                            </dimensionCollection>
                            <value value="1.8828952323684"/>
                        </CalculationResult>
                    </calculationResultCollection>
                    
                </CalculationEntity>
            </calculationEntityCollection>
            <segmentationEntityCollection>
                <SegmentationEntity xsi:type="DicomSegmentationEntity">
                    <uniqueIdentifier root="2.25.318310842062810077214341266367812728264"/>
                    <sopInstanceUid root="2.25.134884066033959077306435705240550195701"/>
                    <studyInstanceUid root="2.25.19202292006231006756726546749423641172"/>
                    <seriesInstanceUid root="2.25.225493840038502954753967211679094249480"/>
                    <sopClassUid root="1.2.840.10008.5.1.4.1.1.66.4"/>
                    <referencedSopInstanceUid root="2.25.319214308104243787945491694789635628411"/>
                    <segmentNumber value="1"/>
                </SegmentationEntity>
            </segmentationEntityCollection>
            <imageReferenceEntityCollection>
                <ImageReferenceEntity xsi:type="DicomImageReferenceEntity">
                    <uniqueIdentifier root="2.25.239108061065263370785162033783811931375"/>
                    <imageStudy>
                        <instanceUid root="2.25.52186905385055707830834793159643714079"/>
                        <startDate value="20170113"/>
                        <startTime value="070844"/>
                        <accessionNumber value="AN1234IMG"/>
                        <imageSeries>
                            <instanceUid root="2.25.263500776851326986665835510707132143772"/>
                            <modality code="PT" codeSystemName="DCM">
                                <iso:displayName xmlns:iso="uri:iso.org:21090" value="Positron emission tomography"/>
                            </modality>
                            <imageCollection>
                                <Image>
                                    <sopClassUid root="1.2.840.10008.5.1.4.1.1.128"/>
                                    <sopInstanceUid root="2.25.319214308104243787945491694789635628411"/>
                                </Image>
                            </imageCollection>
                        </imageSeries>
                    </imageStudy>
                </ImageReferenceEntity>
            </imageReferenceEntityCollection>
        </ImageAnnotation>
    </imageAnnotations>
</ImageAnnotationCollection>
]]></programlisting>
      </section>
      <section label="A.7.2" status="3" xml:id="sect_A.7.2">
        <title>Target DICOM SR "Measurement Report" (TID 1500)</title>
         <para xml:id="para_570fdbd3-3bea-4c47-abde-f36986175626">A compact representation of the semantic content of the transformed DICOM SR tree is shown here:</para>
        <programlisting><![CDATA[1: : CONTAINER: (126000,DCM,"Imaging Measurement Report")  [SEPARATE] (DCMR,1500)
	>1.1: HAS CONCEPT MOD: CODE: (121049,DCM,"Language of Content Item and Descendants")  = (eng,RFC5646,"English")
		>>1.1.1: HAS CONCEPT MOD: CODE: (121046,DCM,"Country of Language")  = (US,ISO3166_1,"United States")
	>1.2: HAS OBS CONTEXT: PNAME: (121008,DCM,"Person Observer Name")  = "Doe^Jane"
	>1.3: HAS OBS CONTEXT: TEXT: (128774,DCM,"Person Observer's Login Name")  = "jdoe"
	>1.4: HAS CONCEPT MOD: CODE: (121058,DCM,"Procedure reported")  = (44136-0,LN,"PET unspecified body region")
	>1.5: CONTAINS: CONTAINER: (111028,DCM,"Image Library")  [SEPARATE]
		>>1.5.1: CONTAINS: CONTAINER: (126200,DCM,"Image Library Group")  [SEPARATE] (,2.25.239108061065263370785162033783811931375)
			>>>1.5.1.1: CONTAINS: IMAGE:  = (1.2.840.10008.5.1.4.1.1.128,2.25.319214308104243787945491694789635628411)
				>>>>1.5.1.1.1: HAS ACQ CONTEXT: CODE: (121139,DCM,"Modality")  = (PT,DCM,"Positron emission tomography")
				>>>>1.5.1.1.2: HAS ACQ CONTEXT: TEXT: (121022,DCM,"Accession Number")  = "AN1234IMG"
				>>>>1.5.1.1.3: HAS ACQ CONTEXT: DATE: (111060,DCM,"Study Date")  = "20170113"
				>>>>1.5.1.1.4: HAS ACQ CONTEXT: TIME: (111061,DCM,"Study Time")  = "070844"
	>1.6: CONTAINS: CONTAINER: (126010,DCM,"Imaging Measurements")  [SEPARATE]
		>>1.6.1: CONTAINS: CONTAINER: (125007,DCM,"Measurement Group")  [SEPARATE] (20170201180043,2.25.56002466128627498886935079903172938041)
			>>>1.6.1.1: HAS OBS CONTEXT: TEXT: (112039,DCM,"Tracking Identifier")  = "Lesion1"
			>>>1.6.1.2: HAS OBS CONTEXT: UIDREF: (112040,DCM,"Tracking Unique Identifier")  = "2.25.165294254063588909770717555738008800301"
			>>>1.6.1.3: CONTAINS: CODE: (121071,DCM,"Finding")  = (52988006,SCT,"Lesion")
			>>>1.6.1.4: CONTAINS: IMAGE: (121191,DCM,"Referenced Segment")  = (1.2.840.10008.5.1.4.1.1.66.4,2.25.134884066033959077306435705240550195701) [Segment 1] (,2.25.318310842062810077214341266367812728264)
			>>>1.6.1.5: CONTAINS: IMAGE: (121233,DCM,"Source image for segmentation")  = (1.2.840.10008.5.1.4.1.1.128,2.25.319214308104243787945491694789635628411)
			>>>1.6.1.6: CONTAINS: NUM: (126401,DCM,"SUVbw")  = 1.98024 (g/ml{SUVbw},UCUM,"g/ml{SUVbw}") (,2.25.51420968257530981243824658943871973198)
				>>>>1.6.1.6.1: HAS CONCEPT MOD: CODE: (121401,DCM,"Derivation")  = (255605001,SCT,"Minimum")
			>>>1.6.1.7: CONTAINS: NUM: (126401,DCM,"SUVbw")  = 5.68816 (g/ml{SUVbw},UCUM,"g/ml{SUVbw}") (,2.25.205292243885258032428819330909580896146)
				>>>>1.6.1.7.1: HAS CONCEPT MOD: CODE: (121401,DCM,"Derivation")  = (56851009,SCT,"Maximum")
			>>>1.6.1.8: CONTAINS: NUM: (126401,DCM,"SUVbw")  = 2.329186593407 (g/ml{SUVbw},UCUM,"g/ml{SUVbw}") (,2.25.70160252080234577167847509948368893276)
				>>>>1.6.1.8.1: HAS CONCEPT MOD: CODE: (121401,DCM,"Derivation")  = (373098007,SCT,"Mean")
			>>>1.6.1.9: CONTAINS: NUM: (126401,DCM,"SUVbw")  = 1.8828952323684 (g/ml{SUVbw},UCUM,"g/ml{SUVbw}") (,2.25.140657026119469861895824082767088344984)
				>>>>1.6.1.9.1: HAS CONCEPT MOD: CODE: (121401,DCM,"Derivation")  = (386136009,SCT,"Standard Deviation")
			>>>1.6.1.10: CONTAINS: TEXT: (121106,DCM,"Comment")  = "PT / WB NAC P600 / 0"
]]></programlisting>
        <para xml:id="para_6ebc416f-eb33-4822-9013-4cc0a39551a1">The AIM sample transformed into SR illustrated at the Attribute encoding level shown in <xref linkend="table_A.7.2-1" xrefstyle="select: label"/> includes information on the SR document body tree depth (column 1: SR Tree Depth), nesting level for nested artifacts such as sequences and sequence items (column 2: Nesting), DICOM Attribute names (column 3: Attribute), DICOM tag (column 4: Tag), the DICOM Attribute value representation (Column 5: VR as specified in <olink targetdoc="PS3.5" targetptr="PS3.5" xrefstyle="select: labelnumber"/>), the hexadecimal value of value length (column 6: VL (hex)) and the sample document Attribute values (column 7: Value).</para>
            <table frame="box" label="A.7.2-1" rules="all" xml:id="table_A.7.2-1">
      <caption>Transformed SR document encoding at the Attribute
      level</caption>
      <thead>
        <tr valign="top">
          <th align="center" colspan="1" rowspan="1">
            <para xml:id="para_e97043be-2dc9-482d-a9e0-66c7530b948f">SR Tree Depth</para>
          </th>
          <th align="center" colspan="1" rowspan="1">
            <para xml:id="para_eb30b405-2dfb-4c3c-bffc-64a8957466a0">Nesting</para>
          </th>
          <th align="center" colspan="1" rowspan="1">
            <para xml:id="para_99757ee3-28c5-45a5-bbc2-dfc6a3029b3c">Attribute</para>
          </th>
          <th align="center" colspan="1" rowspan="1">
            <para xml:id="para_8264ff77-9c70-42f2-a343-6aa43aec8f15">Tag</para>
          </th>
          <th align="center" colspan="1" rowspan="1">
            <para xml:id="para_d9dd3b06-eb18-4585-8b59-fc0651f985d6">VR</para>
          </th>
          <th align="center" colspan="1" rowspan="1">
            <para xml:id="para_d3223435-4629-45fd-b8c4-64ed8430de04">VL (hex)</para>
          </th>
          <th align="center" colspan="1" rowspan="1">
            <para xml:id="para_e1b9e687-771c-4e5a-8a25-4042ce2253bd">Value</para>
          </th>
        </tr>
      </thead>
      <tbody>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_31f86638-fceb-45c3-ba92-88d92a169dcd"/>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_9bfb425d-f7b0-4254-8d2c-1472e29944d5"/>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_188b4471-c8b0-4f51-afe2-866e72bd6fbd">File Meta Information Group Length</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_388e2a25-8b60-4e2a-9404-ceae13deb00e">(0002,0000)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_cb23b722-5ab1-4fd7-92e9-dd6ac29adb19">UL</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_b0884542-59b9-4a97-9131-8689f04657cc">0004</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_bb16fb19-a32d-41cd-868d-c5a5619a7131">0x000000ba</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_78a69ce0-8bbc-4385-8ef8-eda1cacf3a11"/>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_8c2b0370-f91a-4a0d-bbdc-60c6419c7f32"/>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_55c8cdfb-035e-421d-abb2-959a755faae6">File Meta Information Version</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_bbc95679-88a2-40ed-a884-0b4fb98efc4d">(0002,0001)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_e5fbc6cd-9c57-4615-9f91-395b183f8503">OB</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_f62bf280-03da-4f33-91a2-e25ac3945a98">0002</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_77be7f42-dcc8-404a-b4f6-b487be4309a8">0x00,0x01</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_6e28fec8-8cb8-44bb-abf7-2660fae4cb87"/>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_6bb1e38e-c1db-4f9f-ab54-e8a99cb99da9"/>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_8c5dfc0a-1cfa-4613-b4cb-8bdcf98cc6e5">Media Storage SOP Class UID</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_85e7f5cc-7263-4d74-ae63-35843664e15a">(0002,0002)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_1de167f3-5992-4c1f-8fc4-adbff0829dcc">UI</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_6ea1995b-831e-4c6e-b49b-a55985ec710c">001e</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_74e06ffb-075e-46cf-86c7-34fd057efd11">1.2.840.10008.5.1.4.1.1.88.22</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_dc035cc3-46fe-4c4f-9bd8-30caa2c45cf6"/>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_8b489e14-ffa0-4548-8f9a-cc21204e8f33"/>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_bf89b2c3-e696-4102-b587-321bf9897dbe">Media Storage SOP Instance UID</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_40acaee4-5961-48cf-a90f-eeb3e0529ad0">(0002,0003)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_8063d572-1b5e-407b-9a32-6bcef935f749">UI</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_8fd46934-628a-4ff1-a6fb-9ac7223e221f">002c</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_9967a411-ff62-497c-acab-f1a94cf9eeac">
            2.25.224793923339609181243139195858254344686</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_6aec40ad-e6f6-4f90-b2e4-38f0e46770d1"/>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_8a273e01-0b4a-46e6-bc13-98c59a039a68"/>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_d30d96ed-9b53-404e-a437-f3aa2e1adfaa">Transfer Syntax UID</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_57ceeccf-7558-4773-b937-7f6795f24b89">(0002,0010)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_0b518804-637b-4f04-aba5-a8ef9b4fba28">UI</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_c0586add-3c5d-471f-9b4d-87a1695663c2">0014</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_0ed6a2e8-4b0c-45f9-847b-c5d9eaf65547">1.2.840.10008.1.2.1</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_3af3e044-22f1-4ed8-bd45-3d9e5af5614f"/>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_b3587c8b-6dfd-41af-96d3-42fca6d968f6"/>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_96d5f5b6-4c1f-449f-8e45-6f185ae63c99">Implementation Class UID</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_4de82f2e-6eee-4bfc-8b9e-d370a1e07ad5">(0002,0012)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_2866f88a-7a77-439d-982b-7e8bd0459bd2">UI</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_de2eed3c-5593-4885-a4a2-4de1a62cfe45">0016</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_8b11c541-c197-49ff-a362-12ce9bfddbaf">1.3.6.1.4.1.5962.99.2</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_dee919ba-2802-4b5f-aef4-03fd09368b01"/>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_5ef32048-0cf5-4444-87e9-01af27b13961"/>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_a09c17b6-d02b-4d5f-b194-2b1d772e286f">Implementation Version Name</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_bfbdeb7e-b51f-49eb-b545-a9f2cce37212">(0002,0013)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_45248f08-3155-4867-9216-809af3f316a7">SH</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_bd3a7862-02d0-48e9-a43d-a89e23c9e0f9">0010</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_f80446c9-2e78-436f-bf41-def61d14a3cc">PIXELMEDJAVA001</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_2bf30fcb-c259-4d4f-b69f-63bac66c689f"/>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_9e3952da-1008-4b52-adc6-f56d79a960e5"/>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_36c9f5da-1743-48c2-938b-1685a4ced59e">SOP Class UID</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_51e70746-2c9e-461b-afda-ee341831d125">(0008,0016)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_5205209e-1c58-4b34-94d9-3fa8ba12e4c7">UI</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_0e31836b-8242-4683-a840-db7f5032c9dc">001e</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_5be42864-2dcc-4097-9640-b278ae63243f">1.2.840.10008.5.1.4.1.1.88.22</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_69de8974-d0e3-457d-b67c-7963ffc7eb97"/>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_388d75f0-d420-469a-97c5-f3554dccd227"/>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_351824e2-438a-415b-a69c-65eceb72fa2f">SOP Instance UID</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_075016ad-4651-4862-9e86-020302846db8">(0008,0018)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_bb384601-a194-402d-8253-3e1c96bbac88">UI</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_04ba1a5d-fd67-4edf-9e0f-b5749a051f76">002c</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_0302bf50-c647-4ef1-bc95-8caf7ef79cc4">
            2.25.224793923339609181243139195858254344686</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_ef272deb-7bf7-4f23-b907-1a18f07c6df8"/>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_50dfc5f2-9b4e-46d9-9da4-a748073527a8"/>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_f8c3f9b3-a6e3-4fdd-a530-e8699ca05935">Study Date</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_04d52bf9-10c6-49de-af06-f5e5ea3f5ceb">(0008,0020)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_47839e54-a21f-4e37-ad3c-470bd4ca403b">DA</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_a2b639a3-d860-45ba-874f-f7f7dbf9279f">0000</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_9e0ee0cf-42de-4618-b458-a96b33d31b54"/>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_db9187a7-496b-4383-9699-1b2bd7eeb331"/>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_1616bb81-0867-45a0-a13e-0e59f6afacb1"/>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_eee034a4-0a54-4f75-bd43-84e0b7efcc09">Content Date</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_2553999c-02fc-4549-b102-feacf1ed2e70">(0008,0023)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_3b766c35-0837-4ce2-aeec-dfd2e2e662f2">DA</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_a068cc50-1750-416d-bdcf-b63ada3ca734">0008</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_d6d74b02-09f6-4887-bcc4-d712d1f4a84a">20170201</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_62ff92ad-83f5-4f01-9133-65093837aaf6"/>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_9b20bac0-17c9-4dae-a2b8-00105a149138"/>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_d51040e8-11a5-4ad5-8352-e51f32c73531">Study Time</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_c1209928-58fa-4806-bff1-177c5c1e32bf">(0008,0030)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_ae37bf80-23f5-49bf-b22b-7dfe451f9253">TM</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_70602bca-968b-4007-8b1a-832287a4052b">0000</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_c4cacde8-87f0-4790-b832-77a5618f19de"/>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_846886ec-2206-4323-9b78-dabd92f6b1f2"/>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_6ec3331d-00a1-47ed-80e6-19125332f4e0"/>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_2dee9753-6999-41ec-b688-8a5d8646cae6">Content Time</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_66c83816-e827-4478-9e44-ef962212d8e6">(0008,0033)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_bd240295-b02d-4be0-956e-6179507a9105">TM</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_84724b4b-4449-48e7-ac21-744c739e8285">0006</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_169ae2d0-47b7-4b0a-9ccf-ccd32c44dda0">180043</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_332c2a81-fd65-4aad-b65f-10666a5f6a0c"/>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_2381eca6-5ce7-4a68-b97a-1bc7a61da2cd"/>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_f16cb4f3-1265-4c48-91fe-d958dff35a04">Accession Number</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_27ed0520-40b1-4d6f-acf4-5fc72a51b038">(0008,0050)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_0d2ee0e6-f91e-449d-896d-9cbd0f3d4a9d">SH</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_70f8b5ef-66ff-4eb9-9d48-9ecaaf2fb972">000a</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_b0385632-ba20-41c1-ba3c-e7534dcd1f72">AN5678AIM</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_6c8bfaec-0b5e-4fe8-a3f7-420c6dc0dcb8"/>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_c29c5e95-df60-4ec7-bd3e-9f2bb76cfc66"/>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_1df74c77-86e5-47ad-b8b8-42e1d745b16b">Modality</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_3b5b6917-4873-4aff-9ee8-b677ac42ab75">(0008,0060)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_fe21260d-e6a9-4c18-8094-93b8f4efc323">CS</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_0981ca85-1483-44eb-8e59-8eda6270dd9c">0002</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_b82f4d3f-2bcd-439d-a80c-2dc678aea260">SR</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_83f4e04a-3d58-4810-87fd-2f760874dc34"/>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_1ebbb318-2bd9-4907-9519-bfed6b16e7ff"/>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_5b6d2406-7c57-4aba-b68f-624bacce573a">Manufacturer</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_abd5a12e-a6a5-4c15-a682-5104e5ac053b">(0008,0070)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_ca21f92c-2ee7-4654-b401-cff9d88e054d">LO</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_8470ac62-14b3-4393-8a4c-7184bfc44c3a">0014</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_d032c27b-a637-4e51-87be-184010646147">Acme Medical Systems</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_47bbe176-e145-4220-8f12-ef385e7f345d"/>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_0fbf2857-f0bd-4539-9dbf-7dedd424cb5f"/>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_b8a6c366-50a7-4f9b-b4bf-b937921e44e5">Referring Physician's Name</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_81d5b7f5-e7b1-4885-b738-c316ef19884e">(0008,0090)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_4d790595-1da3-481f-8d81-1918d3aac2f8">PN</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_f25418f6-38e2-4424-bb3b-f67f91fbd5fc">0000</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_1b56f7a4-e7ae-4435-bc58-4023490ae885"/>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_78f24519-5de9-4dcc-af9e-7f6d7e90675a"/>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_afedcc32-c523-41e9-ac0e-9343dae246fd"/>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_b059be46-4551-4d6a-b431-403c70050bbb">Manufacturer's Model Name</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_e20ffa2b-f3ed-411c-bf00-f7d0b6740492">(0008,1090)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_647792ab-e7fd-4266-93bd-dc3dde9560ca">LO</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_3f3ffb7f-ada7-45b9-9f93-5c8edcf68ca2">0000</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_da040867-aa49-4e82-ac7f-46b0d014664a"/>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_32ca5640-a74e-41f6-893d-a068a8a7ff96"/>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_7aaeb96d-3ebe-4aec-a270-5a0fa1369b28"/>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_5bdba557-4da3-4e77-85cf-e8c7f4a094bd">Referenced Performed Procedure Step
            Sequence</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_3db3a97b-5435-4cf4-9173-8a7241b0fb8c">(0008,1111)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_5c8cf916-2336-4cec-a5fb-33f8a94c8cfa">SQ</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_18e93cd3-5ce6-43c0-805a-e235714b850d">ffffffff</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_bf88f222-6089-4bea-bda1-11e8163b92b6"/>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_5458402c-41a2-4dd8-9670-b494c90e9a35">%endseq</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_3669f627-3992-4167-8a80-fdf8df2d99d9"/>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_92236359-f352-4380-b906-435bc4c838c2"/>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_3f512f15-5d9b-4a4d-af2f-8f0335ba5091">Patient's Name</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_778a4d43-0252-4e25-9985-6c265b0a9a1e">(0010,0010)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_cc576caa-b4c0-44fe-9cfa-76657b1400a2">PN</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_16122468-4fc8-4b2f-9e83-3f176bfcb530">0010</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_582be214-7202-44e8-b9cb-273224bfea47">CM-1-111-000000</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_98d27924-11af-4b74-8ab0-4d1aa60582b6"/>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_17f5e430-6542-4d9f-8974-6ac0603077e3"/>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_c7380bfa-fe62-4fad-b61e-16d1cc2b709d">Patient ID</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_535e3823-1d95-4e6b-94d3-917ee99c7de9">(0010,0020)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_ce7bcae2-41bf-4f68-a5dd-2febe516faf6">LO</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_d42e9671-77a6-4458-83e7-3b1ac60ff02b">0028</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_af132dbf-781b-4d9a-9bcc-c1f82a46e558">293761767066931586407385203810190772174</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_206c471b-615d-40fa-bd8a-75b803d4caab"/>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_451f0f0b-da4c-41d9-b618-9dbbff43a35c"/>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_d85048e2-2a45-4a92-812e-0995fe41bdd5">Patient's Birth Date</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_6c4fd608-0717-4714-a18d-28fe7948798c">(0010,0030)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_ff360ff4-4916-460b-a71a-c1d4b7c19466">DA</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_262779cd-1b8f-4c22-a964-71966db5e4c3">0008</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_01d852d3-fd7c-436f-b584-2e409efa8693">19601000</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_71820179-3e37-4438-92ad-6e23072033fe"/>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_e8611cd0-7083-4e11-8132-c6ce41713d62"/>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_8d1c3cb5-441c-460c-92fd-7dbc15c3160d">Patient's Sex</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_fd6d807b-eae1-455b-b564-2ca34c8be80d">(0010,0040)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_7f357131-f8e7-4226-8ba5-3064b27e9059">CS</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_2be2e7ae-5b70-4090-8026-8d89ac148b01">0002</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_9720aa50-6b99-457f-bcde-805c482d3418">M</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_c95d8cf0-cbb8-44d0-8114-95663c9f8c37"/>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_4c447d96-15a3-45f2-bd54-259e463f5091"/>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_d9d33301-0a4b-47fa-870a-7423de3e84d4">Ethnic Group</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_2b8fbd98-f58c-42e6-b723-a42843fdadf2">(0010,2160)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_fb77fd40-c6d5-4b7f-bb1c-0e9d203d4349">SH</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_04304493-fb42-473a-8358-74cb57f7e291">0000</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_c0588e9a-02ec-4c82-84b3-81678c0276ba"/>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_9c067209-a3fe-4f49-834d-c51a947fe0de"/>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_dc782d08-a4c2-4409-9656-715a7ad91887"/>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_52694cef-0c8d-47e3-a8ff-8b5411668c65">Software Versions</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_faa429dc-97f9-456d-a53a-9f2cfeec0daf">(0018,1020)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_36d477ff-7aa4-43a9-8a7e-c92d934b2461">LO</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_1da14dfa-7784-4499-9169-c424443da3c3">0006</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_9ab6078e-2ab9-4d53-bb0f-63ca54177dfd">36.00</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_bbce4db6-1b97-4db7-b17c-9af84c5045e1"/>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_b247abad-df4c-4109-9813-a832247674bd"/>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_de44eb6e-d1ca-463c-89eb-bd286644d915">Study Instance UID</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_eaaeeafd-0aee-4732-830f-8fa2e01853ee">(0020,000d)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_34b3c348-9170-4be3-8756-1bc6ec821599">UI</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_f2ec6052-5a7e-4e41-8ba2-ec17120b32ab">002c</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_9bf54760-77b3-4299-8949-fd8f8bc05ba4">
            2.25.80159168229010751652502576830057032194</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_1877fc1b-c1e8-4871-8735-ba060ba296ac"/>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_ed066eea-f0d8-487b-b118-b67aace3511f"/>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_634261e9-1f83-459c-b2f2-96b69d5810c1">Series Instance UID</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_46a0e9ec-4a9e-4a75-a54e-595f74f9929c">(0020,000e)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_70339735-495f-4a41-9062-c300a569383c">UI</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_625d2be4-3648-445e-9bc2-af8230fd4d71">002c</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_2f9990b9-429f-4f0c-9d7d-7aca14accfb3">
            2.25.323817225444021135415209334192751441320</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_22e99ec1-0eaa-434f-b54e-04c068b774b0"/>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_ea853bf7-cb39-4bcc-87e0-bc621528fd71"/>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_8ab809e8-8e6c-470e-9bf9-da3e7ef0d647">Study ID</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_e68c3593-4895-447f-8b9c-4ac9d8784823">(0020,0010)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_cfc5d55a-3998-4207-a1af-3185cf6a5fa1">SH</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_a1bc798d-3f06-4cb2-8ae4-2665e5b3da5e">0000</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_2cf11aeb-9f82-4e10-b3e5-5d722c6b71ce"/>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_1969797e-592f-4766-b71a-8034099d3f40"/>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_bcf53b1b-184a-4984-a565-f5341499cb11"/>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_6a3d3c88-1f99-46d2-a580-a985b45403bf">Series Number</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_ed642bf0-902e-400c-8b2b-c323439ab054">(0020,0011)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_b002a08f-d8ee-4578-99d9-2e2cf7048393">IS</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_90d7a6f7-1dab-413e-888c-db1e3f8f37d5">0004</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_51067773-e1a7-4765-b500-1cbd33cddcf0">7291</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_33bf8680-8014-492b-8a5a-9f0b206847fc"/>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_e56534fd-7542-4783-b5f1-99857c69e19d"/>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_941b80b3-f2d6-492c-83d7-5cbfdf3b6695">Instance Number</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_a6567903-8a64-4e5d-a234-01002f74b0e5">(0020,0013)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_1fe4544c-2059-45d3-ba35-2142a792b8d9">IS</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_bb7455ec-d537-4299-b0dd-be99df378907">0002</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_7f751da1-8a80-473d-990d-2a1964084564">1</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_c1d05729-9ed3-48fc-b312-4f4874ab09b7">1</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_e9d9bcb7-f02c-4fe6-ba75-bde83a7cb957"/>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_8860cd93-9802-4c59-af8a-01e141b78d90">Value Type</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_8094e7cb-4745-4922-9000-ab28581175a7">(0040,a040)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_85259ff3-4366-4066-bbab-e8af61057787">CS</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_26d43193-1372-447b-a7b5-8739540cf83f">000a</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_45ef3ec2-bd59-47a2-a9ff-66aea4f6850b">CONTAINER</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_f3599626-1844-4be0-bd2f-ec767a718abd">1</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_07fc8489-2a20-4a6e-843e-f74369663eb1"/>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_679ae7d7-9bd8-4904-99cb-16f303ea0cc9">Concept Name Code Sequence</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_982fa59e-6d0a-4df1-9aa1-4c9ba90d376f">(0040,a043)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_218342b9-a019-48e4-afce-bca5d5bb42a2">SQ</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_dcc668f0-d53f-4f29-9e4e-3f131276c2ce">ffffffff</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_053a777b-45ea-4d4a-960d-b58c3e6c6e99"/>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_0dfe8058-cd0d-4c68-8328-ad3d840b6327">%item</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_20c24ab2-129d-40e5-9df0-df364de0b7e3"/>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_788d096c-6661-4364-b9d0-fbef3eb14590">&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_0ea4a0d0-bde9-4892-9f80-04280c403c7f">Code Value</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_7eb35839-5924-47d7-8993-a25ee1a9c9ba">(0008,0100)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_31600ecf-d18d-4263-a60b-9a5ec78a3a98">SH</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_08570bb1-5790-49f8-a186-7dff332acb1a">0006</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_17224660-100c-4cfc-8ff1-a8d2ba297606">126000</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_51a126c1-f4b7-46a1-bc8c-69fd8b9746ff"/>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_4ecdd417-bafb-4140-ad5c-5b7f0ee97f7b">&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_09e8e7bc-a284-4a79-bd33-970590d9c7d2">Coding Scheme Designator</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_33a7936b-0eba-4a23-94dd-e53966d8eacb">(0008,0102)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_f0365939-7985-4e26-9dc7-0a032329abe8">SH</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_348abfa1-3ebc-41ee-bd80-20948bc112c0">0004</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_7bdfb915-53b2-4746-adc8-c73abd606184">DCM</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_245fd46f-9c4e-4356-8239-7f9f16a5ec8d"/>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_7fea1922-86f1-47b5-99d1-4c305c1fe646">&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_c8a6dfbb-86f2-44d1-8864-3136c0bd9ad6">Code Meaning</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_ae6c727f-f21f-4cd8-9b3b-2c5b6446acee">(0008,0104)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_d45f3d48-5577-4029-a43e-a96e43c89222">LO</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_94e3df7c-ee93-43c4-bd76-6c564804da4a">001a</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_42384a71-1916-4606-8fad-2c5f8ddc8d9d">Imaging Measurement Report</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_780db99b-9dbc-460c-9efc-e2dbf4269cad">%enditem</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_7d6ab410-6ab9-43bd-afb4-4c4ad2ec519d">%endseq</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_13bedbe3-b9d3-4e5f-9d63-6b22e9cff4a5">1</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_b7717de6-9b40-4f70-b7e4-effb6812967a"/>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_1c8e277d-b481-4121-92eb-f377e94a1ccd">Continuity Of Content</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_0c21ae3b-10bb-49b1-a07d-65b16243ccd0">(0040,a050)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_fad6228b-36a8-4407-ac74-4b4773246595">CS</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_5b647bca-1ebe-43f2-8911-86cb0e538198">0008</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_36b10aa4-aa47-4811-ae2d-edeca60f3e51">SEPARATE</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_eb47f8e1-8ea3-4a37-a682-ca027a7a21d4"/>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_69206a71-d1d6-4598-81ac-7ff30efc0996"/>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_002dacd5-da1e-438e-b6b6-60e2d25c41fd">Author Observer Sequence</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_7c25be34-488d-4829-a59c-684871e57779">(0040,a078)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_0dc9c6e5-b036-41de-885d-255cc2f3ccc2">SQ</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_7ca2b87e-ebb9-4ce8-9001-ae4851b6d2b1">ffffffff</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_67a7d7d0-a1e6-4136-bd94-f5f5b37eca20"/>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_25293246-fe16-46d0-ab2a-a032b0c33ffd">%item</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_c8da7ecf-a8f3-4a73-8f8f-e475849b8c6d"/>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_eb1d45f0-21cd-4379-a3c2-c478d7e221a7">&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_9997decf-4418-4a27-92c3-5b295a45f74a">Institution Name</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_53462312-705b-434b-b01e-0e4d07672572">(0008,0080)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_0f4281e0-09d9-427f-82c3-2ef3f7a90bf1">LO</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_01a3900f-0e0a-4d8e-b1dd-672135ef9fc8">0000</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_509c6c8c-490f-42da-99b4-68e00e85d936"/>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_2ef85901-e0bd-4b80-afbe-5a688fb5d11f"/>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_0d557c3b-95b3-4ac3-9334-a1dc0c8467a2">&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_20006e97-4bdc-4f15-b8cb-924e2c794fc1">Institution Code Sequence</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_48ff43ab-0127-47a3-956d-c0f45459e135">(0008,0082)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_e82e71cd-b36d-46d0-ac7c-35218843a4d3">SQ</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_ec1a3216-f1d1-4af0-8eb3-72036ba799a8">ffffffff</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_782443fe-83d1-487c-9e68-9bbf6f82f659"/>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_906e9118-b115-44f0-a98c-e602eab8cfb1">%endseq</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_4a59a260-c3c7-4240-af33-b4a4e8a17f59"/>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_31314936-3dbd-4136-9d93-37a2d5a57eb6">&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_158be43a-24d1-49d3-9f14-5d3d29227deb">Person Identification Code Sequence</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_b4235df9-8a2d-476d-abbd-177601b4918e">(0040,1101)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_1d6530b5-6415-4db3-8bc9-94028ede31dc">SQ</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_1136c4d4-658c-4290-af6c-b32c5fca9958">ffffffff</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_ba3965d5-fedc-4394-818b-df04d1d56541"/>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_55cb1769-e315-4a0a-aec9-1adbe559acbb">%endseq</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_376c73b0-c72d-4993-9d8e-9eca5716342f"/>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_abb7f2e1-4d9a-49d9-8846-d43f05502c16">&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_3c216c6c-bcda-45bb-8764-7eeee2251aac">Observer Type</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_92d0b3b9-ccce-4e14-83b2-370a10c1c2a9">(0040,a084)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_286dbfe8-8205-4992-8efb-14d45eaea5b0">CS</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_a5a0dc3b-3595-4c49-bd33-8391c0080964">0004</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_3310a3fa-15ee-49e4-af10-cdbb7797335a">PSN</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_1e0f0160-560d-468d-ac9d-288a9b25f409"/>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_6a514378-614a-40ac-b218-ca515c2a124b">&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_7b15c7d2-5122-47e7-b4aa-5e2feaaf12bc">Person Name</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_a5f40309-d90c-4bea-af02-9f5305f6adf0">(0040,a123)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_58483af8-6ff6-431d-9e92-f5f381ac9bba">PN</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_dfecf10a-7463-4656-b43a-6ebe718eb84c">0008</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_5cce044d-033b-4db8-9e6d-e86340118ef8">Doe^Jane</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_21d9eec3-581c-49ad-a9b7-9f05ff9b2a47">%enditem</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_ebc993b9-f5a6-4878-a770-d0fc8ba932c3">%endseq</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_6faaae36-cb3f-4899-b2b5-8633d4798704"/>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_76580425-8c99-4945-af0f-78cf02c04a8d"/>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_3cb7a207-e3db-4e5d-b4fd-4da859ce40a6">Performed Procedure Code Sequence</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_4c14dea4-b87d-46e1-a381-dde6242cbcef">(0040,a372)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_d16cc906-7fc2-4e96-9c41-2bd91e7b0f41">SQ</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_dff6d774-198f-42ef-8f26-ed1893286109">ffffffff</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_d44f3490-918d-45a5-aad8-8cbe39c3bee4"/>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_91e26fb7-4e43-4e23-b66b-ac651744dc06">%endseq</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_a4527065-aacc-434d-9c5a-a16c8eceb06e"/>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_cab034ec-5c6f-423d-a3de-30474e2a86d4"/>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_0dec459c-0648-4d64-93d8-ca98b955bc6e">Current Requested Procedure Evidence
            Sequence</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_4cf1e523-e365-4dda-8123-0be3412d787d">(0040,a375)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_5464eef4-b978-4e0e-bc4a-0cce26246a6e">SQ</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_c4b92b69-8bdf-4ff2-b8d5-63e2da5833ce">ffffffff</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_b879e612-74df-4f4c-b1b7-99b27c62e353"/>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_249f7e71-5423-4a09-bec9-28d61b3a7b1c">%item</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_78f9c753-6ccb-471f-b2db-866ab6dae00c"/>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_c1d71c1d-e1cd-4af2-9832-b002ba81e170">&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_bd71d5d7-d463-425f-ac0e-60d6d42df979">Referenced Series Sequence</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_3a6b494a-3630-42c8-bc58-fcac5e5eb9f0">(0008,1115)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_1890a48c-3dd9-46ef-983b-3459caa5e384">SQ</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_641f2e73-d634-4d1a-aa25-0bfb63fdbf50">ffffffff</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_a70f8769-105d-4c24-9aa1-fe5e7601198d"/>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_4c33306f-7a87-4dd5-9478-a30008c2956d">%item</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_30c5d5fb-29ac-4c5e-b084-f426f4758d71"/>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_4c4ac241-a658-4d25-9a0e-b1536baaeac4">&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_47f0a0f5-eff1-459c-a868-3957fa833d94">Referenced SOP Sequence</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_48119efd-eb74-40bb-b843-0fdf2ecd2b31">(0008,1199)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_803ea9a7-b4e5-4b85-ae42-b6cac5b21b6c">SQ</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_bbecd563-45eb-4331-80e1-35f103c707cb">ffffffff</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_bbcc37a2-2111-4d12-bd90-7d39b704f4b4"/>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_39d9185b-fd23-44ce-9ea9-5079bf430769">%item</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_fffee30c-da78-4c72-8bb4-0395982c8e08"/>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_7f921cd8-b7a1-4809-9798-0866b1d6de19">&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_4a8fb0b4-7888-4050-a0fd-e702e5655322">Referenced SOP Class UID</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_46e53e98-4e84-449c-af45-ac842dd1ae54">(0008,1150)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_a393bf6f-9cdc-40eb-91b7-4b4e575b397c">UI</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_4bdd798e-27b4-40d3-9426-10192a15d84c">001c</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_8014b997-00b3-4fdd-9c5e-f352373c00b3">1.2.840.10008.5.1.4.1.1.128</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_923a28e4-5fba-4b8b-80bc-8ba9815cfa1f"/>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_3747ba05-2b05-4974-8866-9ada21dbd33f">&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_1bb356c3-acc7-4cfd-bce8-fbd26d7729c7">Referenced SOP Instance UID</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_31e052eb-ecc3-4dda-a0bf-407a6ed1a610">(0008,1155)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_d2c97771-fc8d-4118-b40d-4556bceaa6c0">UI</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_adeab3bf-b7a1-4ab0-b633-41c05af6eaa5">002c</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_dc67e070-aa79-4747-a44d-d3e8e1b8f832">
            2.25.319214308104243787945491694789635628411</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_e753dccf-c007-486f-ab6a-bac1b19110e7">%enditem</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_7e35e014-6012-48d3-8435-1c1a476b8e67">%endseq</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_368ec7ba-91f0-459f-9daa-11c862448563"/>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_061f8dbc-d48c-4ba2-80d4-f5193f7b0f28">&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_66b215bb-6f72-4398-ad3e-28d19ec44f5a">Series Instance UID</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_81dd4dd8-534d-46de-a8f9-224516318628">(0020,000e)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_9499a17a-1dc9-4cdf-bdad-76f7cddb6436">UI</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_332d3bf7-d445-4c33-8132-a0969ce4c3bf">002c</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_c48a77d5-8c6c-48ca-8cf5-9bbc6b19836d">
            2.25.263500776851326986665835510707132143772</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_1d5ede93-60e1-4fc2-b8f7-0ee407c4cdb0">%enditem</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_16df8084-f0a5-47c0-98e4-144b89b07941">%endseq</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_443c8c23-3ab3-42de-baf5-216a4a1e1ca6"/>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_015e4b10-861b-4e97-9c8b-1911629a22c9">&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_208f303f-6aa3-4e31-95c5-174e2492ffed">Study Instance UID</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_598a9db6-7578-43ad-8609-c7e1fd93eedc">(0020,000d)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_9cb6aab0-3fc7-4920-aa6d-d57882b98b01">UI</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_3e533026-7fbe-42c3-85bc-e18570f8f776">002c</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_3234ef32-9c01-4e2c-9671-8f93053c086c">
            2.25.52186905385055707830834793159643714079</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_c787fd5b-6e22-40dc-a8f5-f0d009ce7ccf">%enditem</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_533f7d5d-4a46-429d-b15a-e002e5c663b4">%item</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_885a9c63-c4a0-4017-b4ec-9ae586a880d1"/>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_e08269e9-5c15-4f11-a5ed-8ecf761d9135">&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_14ad56f2-f474-45bd-bd4d-28ce1b39182e">Referenced Series Sequence</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_818903eb-8875-42c4-84b8-95e2146915db">(0008,1115)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_3c697922-1f8c-45c6-956a-434a45784056">SQ</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_f39971e1-e49d-4ba5-8234-3c8b740e9a95">ffffffff</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_be252cc1-5768-4f86-b788-1a0265acc21e"/>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_186157fb-4d38-46f0-9081-1fe6b4294991">%item</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_8cf3e115-ba96-4403-bda8-4fc94fbf24b5"/>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_79041488-e028-4a11-8323-be25a293d46f">&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_c547b5d9-bebb-47ee-b74b-ecf5f9e035ac">Referenced SOP Sequence</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_369f44b8-8326-456f-8fa2-1711a8216ae3">(0008,1199)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_26093d0f-04b1-4011-90a9-e34ce5640fa4">SQ</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_542ff2aa-1dec-4090-ac1a-c417dd2a2392">ffffffff</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_875336af-210d-4e15-97bf-3e969879b3f4"/>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_c5c845fa-5f9a-4218-a317-995cd6cbd52a">%item</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_443adab5-639c-4199-99cb-13c9653181bf"/>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_f6c0fe58-a1e0-4bd2-b1a4-4340823e52dc">&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_6df8cef0-5c34-498d-95ad-84fe79054733">Referenced SOP Class UID</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_e47e3978-072c-42e3-82b1-3c6bd38338ea">(0008,1150)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_68c6b5a0-adcc-4337-b62a-1a0bb75ee73f">UI</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_2411ae57-765a-4a22-a04c-e9ee39538d64">001c</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_f57574b4-2f5d-48a1-afe8-252647d0c595">1.2.840.10008.5.1.4.1.1.66.4</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_1f719731-af5c-4bf0-b810-58f4b31e25bc"/>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_0caf18f9-23d6-4fbf-ac39-a955dba84029">&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_79c495c2-ac09-4981-a3bd-270a46e60e40">Referenced SOP Instance UID</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_a71f33e3-7836-4769-9aaf-6c8aaf4dd2a1">(0008,1155)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_d86a6419-be92-4826-8a9a-0d017f179723">UI</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_8606270d-8e72-4519-b70d-4e685a258381">002c</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_a296a1fb-966a-4335-8038-9c666f62a487">
            2.25.134884066033959077306435705240550195701</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_487e4af6-5004-4751-bd6c-d8d8d1ec2194">%enditem</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_262a4edd-cf18-4f51-b328-eae1f5110e38">%endseq</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_0d3f1f70-da32-458c-bd71-2ed69d0ea1da"/>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_0d1c82e0-d335-4041-9caf-bc50daf2a512">&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_ab2790bd-46b5-4073-a9e2-4b0fd79ec906">Series Instance UID</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_3ba5e985-ae4d-4fdc-8db1-18926890d5bc">(0020,000e)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_f0d7be15-3b40-4457-b5e3-b08d6dd64794">UI</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_90164bdb-9911-4025-acfd-53a5124ea572">002c</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_30b8fd60-8543-4d08-ab22-9f03fe45d001">
            2.25.225493840038502954753967211679094249480</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_8e0c0802-bcb3-45df-93a6-eac618397be0">%enditem</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_fb0077e0-2a32-4a38-bfc0-306018b3563e">%endseq</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_bc653e5a-de2d-4d48-b72b-f133100213b3"/>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_0e614353-5fb6-4a33-b309-319e2d849a2d">&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_13fba727-b4a8-4486-afc8-970ae55721d3">Study Instance UID</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_4ee0549f-e544-44fa-9474-87018c47ed4f">(0020,000d)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_9e03c85e-c7fe-4885-a66e-365416bf4050">UI</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_9ca2a458-31c4-42cd-942a-d76a3cc21944">002c</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_e5fd72d5-4662-413b-9f6d-c920e8cc7ee0">
            2.25.19202292006231006756726546749423641172</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_071cea85-1fe6-457f-b3c7-d66117055cd0">%enditem</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_8b247b75-bca7-4064-a6f7-4dbe3508b4c7">%endseq</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_4c0c33ce-ef7f-4494-858e-50a1c82bcdda"/>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_72b65291-320b-4233-8aed-a9989146d1f7"/>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_da59684c-12ac-4582-ba60-e8f93b66b130">Completion Flag</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_53e70cbf-9e38-44bc-abfc-7d64fb79a45a">(0040,a491)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_c68edbc4-cdff-4b72-bc8d-d1b70bd0fd78">CS</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_9ee94323-5956-47f1-8de3-70e248da22a2">0008</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_21f80729-88f3-4ca6-acc1-d97d71a14c54">COMPLETE</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_b082c87c-458d-498f-9f58-71befeda1541"/>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_95e5e48f-079c-4902-981b-c93b69b4a46c"/>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_d3ac9578-1965-4341-af12-cabac7eeddb4">Verification Flag</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_5fa2197a-3580-4704-90f3-be24d939a63e">(0040,a493)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_a86ea542-70d3-4916-9305-431d69a8ffa0">CS</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_aba6de8d-1c4d-42e0-a1f5-5d7bcbb0947f">000a</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_e535a96f-2fe3-4a0f-8db1-0cf7aa6c1bf5">UNVERIFIED</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_778b2339-e2e0-4d26-8afd-aaca24e223d8"/>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_a982e809-e555-4492-8b2a-7c1f522e1d36"/>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_cb1fa7fe-d0f5-480b-9030-cb65a521014e">Content Template Sequence</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_0558fad3-a978-47ac-a310-0ec5a19098dd">(0040,a504)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_a7520f5f-55b1-476d-a4bb-7aa5af12ad9e">SQ</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_4e006242-30ad-414b-99a1-2b641b5ac00b">ffffffff</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_7c2be9ff-9c01-4dd1-b516-bc104222b95a"/>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_e3050c68-6409-455f-b144-0587f20647be">%item</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_e34dcb3e-d771-41ed-9567-3261da82a591"/>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_030952c0-fe1c-4784-afe8-6a7cfc23dafa">&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_2dc0598b-0d22-4645-a380-52e128ef43f5">Mapping Resource</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_63fdd957-7fb7-40dc-9244-e24563299f7a">(0008,0105)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_c15c3fce-16c8-460a-b0df-72dab0a67862">CS</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_5e3110a1-4bc0-40c4-bd7c-366b888459a1">0004</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_4a267d1f-1885-4aae-ac76-6810e0510561">DCMR</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_468ff9e0-396c-452b-b4b8-8ec6fcb2db15"/>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_ce819edf-1f8d-40fe-9d87-99bca80f67d8">&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_385efdec-e931-4579-a1a7-c6eefdd152f3">Template Identifier</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_454ed35b-7de9-40d5-81a0-cac0644a560e">(0040,db00)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_b83f2ac0-0276-4a61-bcf6-b6eb004e9f53">CS</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_5ed37d38-28d4-44e8-ba53-3f4239f010d9">0004</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_3aaede9e-4071-4d7b-9868-72c53eca015f">1500</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_3365e8dd-7e74-4435-9ace-4c365f79d8ff">%enditem</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_6298fb8b-c281-40f1-ac1d-b0f81780b3c6">%endseq</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_62ac18db-37d9-4f18-8b12-28c5b79a03d4">1</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_1d06402e-a0da-4209-9064-2381f49dc77c"/>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_cd794687-3900-4529-b7b6-50c8b328bddc">Content Sequence</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_87fdbbba-cc61-4d70-b3ff-004e960d215d">(0040,a730)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_a9162d7c-4019-4962-9628-90c9568240dd">SQ</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_9eadd590-a625-4d0d-9448-6041b4cafb9e">ffffffff</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_5cf6b423-dd3e-4f36-b581-1beb3b4027fb"/>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_96a15433-fbd7-403a-9449-a3dde4791162">%item</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_d78f2139-8d30-4910-82b1-cedb0f640308">1.1</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_837bdb2a-19a4-46a9-86b0-253bd9326640">&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_50c14460-6c9e-4f87-a74e-3d00253cf45d">Relationship Type</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_0fdce3cb-bb7b-4e2e-8b48-4241510bc208">(0040,a010)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_6aa1baf0-ebbf-43c3-be32-c934303ca294">CS</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_cdcc9acc-e498-413d-8f13-6595d4262662">0010</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_7be37bc7-b900-472d-8e8e-c2631febca79">HAS CONCEPT MOD</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_26708060-e527-4fe6-a484-5ec46a7fc030">1.1</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_2150d8b9-cd43-4fc7-9140-ad7dc8c40990">&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_3a2e45be-9ad8-4488-9293-bc5a6169d37b">Value Type</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_85770959-4562-4f12-b45d-6ab6e7fccad7">(0040,a040)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_bf1a50a3-f9fd-41ee-b1e7-15ee14d35e9a">CS</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_03aaf8d6-2fbb-4bfb-aa19-66087e5b2af5">0004</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_bf97aad5-1c23-4f79-aa16-1cee398cc179">CODE</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_a40e982d-acbf-4997-bf6c-09d9f6d0a7d1">1.1</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_f1a5b50a-7407-42b2-90a5-bf56e96ae7ea">&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_508cff5d-d8d1-4af9-a174-ef1cecd817af">Concept Name Code Sequence</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_fb34c21f-289b-4c1f-94c9-ba2d8c9381c7">(0040,a043)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_8fcba6e1-bbcc-4810-bd16-21ff26173516">SQ</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_5a536562-0184-406c-9cd5-104f6874beb5">ffffffff</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_b486d757-0b93-4f8a-8040-4bcd96972785"/>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_3b192166-3a0a-429e-b296-a661adad4a1f">%item</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_992c4829-fc60-4f3c-9859-79d8e3e16e64">1.1</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_5e6c7ae5-95af-4447-8a49-92d8cfcc7215">&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_4b7426ba-91b8-4521-8dba-156a6c2f885c">Code Value</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_bfda6992-2f7c-4ee2-b3e3-8d69c098dc22">(0008,0100)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_d3fa3920-1be5-4d99-bda8-897c4aa5b9fa">SH</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_910c7a94-15dd-4fee-8d57-d597f749f0b0">0006</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_a28a8221-f284-4599-b8dc-8cefa8becf92">121049</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_bf3736a1-ea82-4e8a-84a7-c5d67879cb63">1.1</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_d2092959-fc56-421d-8c56-d11ed8fdf175">&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_5a648a7a-8b09-4488-90c7-08429e0afa7e">Coding Scheme Designator</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_ecdfe73b-269e-4906-9272-cd1a42535b61">(0008,0102)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_1f0bcbf3-9fd9-4292-b337-aef740a212a2">SH</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_2e9c4d47-3842-4d43-bf9f-bfd448f27cb7">0004</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_cc21a373-70fb-49d7-8955-eaa48e3c5352">DCM</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_a64b629d-a80b-4753-be45-97af366b2a63">1.1</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_6f340de5-c866-4497-aa09-826ff45cbacd">&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_4a1e52de-1aa9-4bad-b3c7-754e4a6a05d5">Code Meaning</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_dcd65903-5489-4957-90de-e939790a72ec">(0008,0104)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_eec2f0b9-770b-4f11-9924-b75c8d896102">LO</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_7c5560dd-7a60-4f6d-a947-e917da712f02">0028</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_f9d53f06-5088-4b40-a9a0-9097e03bc228">Language of Content Item and Descendants</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_bea606a0-29be-432e-9227-878bf9b4ee9d">%enditem</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_2f1ca886-c062-4e13-b428-d124253ffa8a">%endseq</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_3e16738e-4930-42fa-b42a-2808ebc514ff">1.1</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_9507e529-2eab-441a-a86f-21b7b782d5fa">&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_8539b1e3-3f36-43d9-9fd3-601e3c8a2484">Concept Code Sequence</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_60aec71c-a944-4142-9c9d-26ba67b3ca5d">(0040,a168)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_efb55b7c-ec44-45f4-96b1-45e0c9579d8a">SQ</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_ebccc964-7667-433e-bcdb-7e953bacbdb2">ffffffff</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_290cb004-5710-46f9-ab75-bd2ac99134c9"/>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_b008a906-3ac1-4045-b529-5af16081b3c7">%item</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_4c41e04b-d2a7-42ca-8105-81388bdecb28">1.1</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_e0e0f0f8-97da-42d6-9c90-072b9811b511">&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_6ee5c977-3fe9-4283-9627-e48e5365dae8">Code Value</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_9849be65-36be-4607-a0c1-33161a06f8a4">(0008,0100)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_3656e86b-e250-480c-8d3f-1fe804cb0c4e">SH</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_a9961d97-687a-40be-bf96-2eb52125e6e7">0004</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_5126df2e-fbfc-40b3-8e66-768c5b4e8e3b">eng</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_2f6e8d9b-28c1-41bb-a24f-ed4e047144c6">1.1</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_34cde413-26a7-4943-aa89-fa9c0e7c1327">&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_e5952a29-9189-47d1-97a5-b622f10f7881">Coding Scheme Designator</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_0c09d54d-dde0-450f-bf0f-fc33400d3588">(0008,0102)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_2578d8f3-2153-4b76-b4ea-cf611100a7bc">SH</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_0f2dff07-98fd-4436-9f46-d4137a062682">0008</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_46bacebf-633e-4ad1-a969-0519aa4fd636">RFC5646</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_fbcd8c68-3788-4422-a4e0-5ccca0215000">1.1</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_a78d5677-0974-4139-9061-43332a463dd5">&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_99aac93f-d556-4e32-85a2-d11ece1da02a">Code Meaning</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_3e0a8810-4776-4d22-95c4-de4f6e296343">(0008,0104)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_a243d745-8cd9-4ab7-905d-11d0acd0a41c">LO</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_fd934bc2-c376-4f52-b52c-e7057bc52647">0008</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_0b8b1510-1b14-45dc-a09c-d493de9874f5">English</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_50e846e5-f841-4f96-9720-3c9228916854">%enditem</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_378bd172-4f34-482c-a843-5ed468efd19b">%endseq</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_aa61325c-e1a4-4131-b62f-915250ec17f0">1.1</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_8b8d9e45-6681-437f-af10-cf85555accd5">&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_59e43d52-c987-4388-83e4-1622b3a57b4e">Content Sequence</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_47fbbe18-c84d-4563-a2af-b5f3fb6d1f72">(0040,a730)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_6df41bd3-1a51-4dbb-b43d-c424778969a2">SQ</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_dd5a3a89-f75a-4e56-aac0-658e59e434b9">ffffffff</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_585732e2-673d-499b-adf4-4a48ce634695"/>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_9b4039ee-b570-42b9-9383-be6d900ea075">%item</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_a8390bb3-8774-4a98-94ea-14da31705e10">1.1.1</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_22170786-e548-4943-867d-4791cb146dee">&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_c6649fc2-31d5-4099-9d67-11425774fdcc">Relationship Type</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_b26ec39b-28c2-4355-a5e2-a9cdfe396263">(0040,a010)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_48674c76-77eb-41ab-96da-24310d20b7fa">CS</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_1881a5de-ced2-4c5e-9f35-a0345cdd2d96">0010</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_e066fac8-3af6-42e7-9aec-73be6ce8e1f1">HAS CONCEPT MOD</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_fd318f21-96ab-449f-b25c-50ebb23ad33e">1.1.1</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_05e14134-d8a4-49e8-985b-c5d1ab5358ec">&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_044311c8-d3e8-4c78-9f3d-d7f043a5cd6f">Value Type</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_c6e61ae7-3f13-4d51-b3d4-fe75db1cd47c">(0040,a040)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_a102e2fe-587a-4f3b-a376-4d347eaa7f86">CS</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_fb621efa-158d-4380-acbe-d298e1f72d61">0004</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_c5afa7e2-92fb-4e02-acc2-74ff36f11025">CODE</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_fc797c2d-650c-4fad-8559-89591bd62a32">1.1.1</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_333172d8-4c0f-463c-971a-1ad203aae5a1">&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_92d3b33c-296c-483c-af68-9f89d7a0dc9a">Concept Name Code Sequence</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_b7cd83ed-41f4-4e1f-b33d-27e80fb68766">(0040,a043)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_64052395-8f27-4443-b75a-14bc77cd687b">SQ</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_25d131fe-93f9-4efc-a4b5-a398179f363e">ffffffff</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_a6049b65-d6d9-4254-9086-5b6a7b56bf07"/>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_f65619a3-5615-435c-a047-84a125fb38ac">%item</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_e4170fb1-a612-49a8-bc33-43a6b528ec03">1.1.1</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_82ac258f-1927-4432-8b06-52ea149a422e">&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_bbc3b275-4c72-441f-bcb2-4cb643ffb44e">Code Value</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_0bb81bb4-389d-47ce-a600-6bfc0d1c4ace">(0008,0100)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_2410531f-7d9b-4a92-bb04-59f8aa9910aa">SH</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_43724afe-d5e1-436a-9c28-2ffbf8f5b728">0006</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_446bb3ba-b112-4137-a6c1-84fb6c02bd35">121046</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_d82ab12d-083a-43a1-bb05-b6f62c7cacf8">1.1.1</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_550bc42b-6425-45d8-a92d-45ed2aba153e">&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_6d313ff6-8924-461f-84c7-1a031a8b2852">Coding Scheme Designator</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_91636536-3966-4a27-971f-beacd4af1374">(0008,0102)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_9233098b-0c3f-4533-b9c6-4eef1a4634c3">SH</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_b9439091-aca1-4312-9b86-5dba81d4bf05">0004</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_4ce47f7f-3104-488b-9c53-ab611ff51fb1">DCM</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_880f6ee9-eb20-4059-ac16-4cf11e48702f">1.1.1</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_797d04ee-542d-4820-a10e-3c79a9d9a4e1">&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_90a3fb71-57ea-4d52-b636-a35a4c6363fd">Code Meaning</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_dd4734aa-2ae1-4b9d-b744-b6cb803c1c76">(0008,0104)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_4db034b5-f354-4faf-8f10-783b07a46282">LO</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_9cbc623b-31ee-4295-9600-714d4d54f6fc">0014</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_609332a3-6383-4072-b839-f9d1818b08f0">Country of Language</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_8732c093-ef25-4287-a314-9d97685f957c">%enditem</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_0662eed1-6893-4d31-a9c5-858d30e810f2">%endseq</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_d18e073e-7fb2-4039-ab9a-6b67f8dd1d0f">1.1.1</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_55807aff-9c16-4e65-862d-ab82c7cce70f">&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_113cde4f-2413-411c-b84c-068e5389f063">Concept Code Sequence</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_f9f877d5-6165-49b7-abbb-9cde189f20ad">(0040,a168)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_9b086ece-0e4e-40b1-ada7-23343fdd71d0">SQ</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_2fd6bdd6-6582-4986-9f4f-8e680657bf1a">ffffffff</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_0353c6f7-8226-4ba2-afb9-95945f515f70"/>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_15d9917b-0ac8-423f-8c68-97a2c99ddc93">%item</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_47393bd4-c5e8-49f8-9930-35a8bad8e317">1.1.1</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_6abc059a-5a12-44e2-b545-b4b2f68338e7">&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_31dc532b-d9b6-4459-b152-2b8221d49855">Code Value</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_e0f11852-2d63-400a-8184-b2d1ae589913">(0008,0100)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_efbe58a0-ead9-40d1-9d91-9f228c102680">SH</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_c2f780b3-1182-4964-9a15-b89a8decf33e">0002</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_7f5e8fde-3101-40d5-bc54-9e83108854ed">US</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_7aff930e-1b15-4fd7-b8f3-4d9caaad28b5">1.1.1</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_a79ba4a3-6e90-416b-aa0c-9fe36147e8ab">&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_8bf93c21-f9f3-42b7-969f-d46aa7db1147">Coding Scheme Designator</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_18318566-97fc-46f1-ac32-09b88b02ce70">(0008,0102)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_9299ecc3-7d29-4cb3-bc6b-2fd8c8cbcc2a">SH</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_edaded58-b343-492c-861a-addbf6cb3207">000a</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_46062676-180a-462d-89d3-c6227d742a79">ISO3166_1</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_cb5e3090-c68b-41a8-b1e6-0d9efc1bfd0f">1.1.1</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_3dd2936b-a2c2-46f6-93f8-5b33155f5b20">&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_b254718e-de73-44cf-b9bf-f05b1e5d29e8">Code Meaning</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_2e113dfd-e832-4b06-9865-75202d416b5f">(0008,0104)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_ea48f33a-4165-4166-84d9-0be2dcf517a1">LO</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_eff7e0d9-4a90-4fba-a08a-543455271725">000e</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_1340e312-43f1-446b-8d0f-ec0f63051639">United States</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_a48e5175-fcf1-4ccd-871b-8628b0ad12ae">%enditem</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_8d3f0345-1ee2-4728-a19b-e0c8f5232a50">%endseq</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_7b4e7ab8-10da-4697-ad6a-69527d99d2e4">%enditem</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_363d7f0a-c249-4afe-9117-f32bc7bffd82">%endseq</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_2a16a9e9-2316-4a6a-87e7-74ae72526d2a">%enditem</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_b6bb9cfe-16a5-40eb-b309-3c38fed445c9">%item</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_91192c6a-7a83-4a78-b175-f985b608f681">1.2</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_8807fb3d-83d7-4834-8b81-57a1a7361f73">&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_f5fa2f62-8343-40d7-a38b-a344277cc3ad">Relationship Type</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_75ad4dfb-5351-4852-82a3-e24002a8ccee">(0040,a010)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_c4566b68-ce47-4551-8280-0160e38259da">CS</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_6b410e09-478a-40ae-8b60-87f032833c24">0010</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_8406e282-cf05-4e4c-b03e-2ef04c6a5df0">HAS OBS CONTEXT</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_9812cf1a-6232-49c0-879e-339ecd4bfb44">1.2</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_3277c1ce-e70b-4e9b-964a-de217478c335">&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_4da0427d-2c15-4aad-8f9c-32dbf29b3bab">Value Type</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_cd95eeb3-f640-4bdb-a47c-a034c18d4b97">(0040,a040)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_a953f141-66d1-41ae-82c1-2e67bdf18458">CS</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_c54e7f1f-7db0-4770-8ad4-a386c57eba7e">0006</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_d5d71d46-69af-44bf-ba6a-dc7ab57c6d35">PNAME</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_9d2675fc-1943-4fb4-ba9a-87a1b4c349b4">1.2</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_829e1ddd-f7e7-4d89-8bac-a026fa0e243e">&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_26d19ce1-59dd-4edf-8074-4754686e347a">Concept Name Code Sequence</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_968b5a84-e93a-4f8c-b797-6c142deb6f19">(0040,a043)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_4070c116-98a2-4a25-90f2-2a8cfbbed43b">SQ</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_c2d31455-fabe-4610-be27-fdd029df6efb">ffffffff</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_8776d8e9-e524-46eb-8356-cd3c568bd24e"/>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_c52db4d3-f584-497e-b363-335a50213b55">%item</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_3e1831f3-1496-4f8e-bc52-a05f1d301eca">1.2</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_af42bafd-ddd4-4c24-8f3c-030e8fdb9964">&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_9e847dd9-b793-441b-8eb2-e7ad6a735f47">Code Value</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_a30baa77-fedc-4403-bc64-d938a310c418">(0008,0100)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_562e3f3e-c96c-4665-899f-72158b266e7a">SH</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_e7b2aa81-bfd9-4851-b2f0-975860380eba">0006</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_57ef8cd3-8d66-4f24-b3f3-45ad3bdf9d44">121008</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_aae09ad2-f270-42ff-8539-c7ea0140295d">1.2</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_320712f2-6305-43c8-b129-31af554b1a38">&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_e2d71600-1a5a-4ce6-8d3b-fe32a66fff91">Coding Scheme Designator</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_75568a2f-1c77-4354-9bda-26023e53c95e">(0008,0102)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_534a9f8e-3bd3-4484-829a-fcc7ca45f8ec">SH</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_4e379024-11a1-4815-ab94-d0f9f6478921">0004</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_21212207-69c0-4d01-91b4-2fc1b860baf9">DCM</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_b7f273c9-c4ab-44c1-8df8-ecf517a715e7">1.2</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_bc64fcf4-2724-42f4-a1c4-6a88de25089c">&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_077c4bfa-3f5f-48f6-95d9-b1710bc34000">Code Meaning</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_bdfa9a74-068c-417a-b849-2716b1b0a84c">(0008,0104)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_5ce2677b-fec9-4cf9-872b-8b2e3a3057e5">LO</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_0badb2eb-bd3f-4b28-a0d5-374ebf0ea464">0014</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_2238727b-a676-4b9b-a9b6-17c0bbcafc61">Person Observer Name</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_875c572f-305d-4452-8517-f2a03063e489">%enditem</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_5ce18f99-5092-40c1-8d83-d33cbab55a3a">%endseq</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_2366f4dc-7c20-4c66-89ec-67c89e0623c6">1.2</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_ec626d03-4197-4d30-a4de-70b2ac4826de">&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_d00a7c05-8140-409b-9040-09f26d095230">Person Name</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_10b7df89-b49d-42b6-807d-2592be34a893">(0040,a123)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_525f627c-7ec9-4c5d-814f-f759b23691bb">PN</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_64aea2e0-1370-401d-99bf-710d02d8f252">0008</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_e0d66daa-eb93-40d3-a71e-30d8677f4d9b">Doe^Jane</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_5c522b48-0ab3-4550-8e0e-95611c742e7e">%enditem</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_ada18daf-eda2-4b07-8921-359c0c9b72ff">%item</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_abd71551-a7bd-4e3f-b9cf-25ce342abb0a">1.3</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_acf533e1-197c-409a-8c55-f0819cbcf785">&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_f15a8407-8e52-4af9-acaa-688bea666c15">Relationship Type</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_08dbc038-ddeb-4e59-99f0-38a3d8ee7ed1">(0040,a010)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_dbab778f-ae1f-42c7-b980-5da61e40ce85">CS</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_9d61261b-5b46-45ed-abd2-15457f7ff18c">0010</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_f1c35951-cf77-4cf5-ac50-6de28be3ca2d">HAS OBS CONTEXT</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_3b988e39-0772-4bd6-aacc-7bad823d3d14">1.3</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_2e86769d-a951-4fa1-8154-2e20fc137fdd">&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_2317a304-715c-4254-90fc-1ecc46db52df">Value Type</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_1267b071-e095-43aa-863c-7b18fe5261cc">(0040,a040)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_2ac45aad-50b5-4793-b368-58453833e3bd">CS</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_c170695e-c72a-48df-b773-10c3f0c7c327">0004</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_6927d379-1e0d-430a-b6c8-d8ec547ff2ca">TEXT</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_7f1cf168-526f-4d79-80cc-9c59c12581fe">1.3</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_52a69de1-0198-4cc8-81d1-5a3add2ebf9b">&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_43aa80e1-8471-4c4d-b2bb-1847563febdb">Concept Name Code Sequence</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_f4c4f554-dd85-497f-9f54-ee7349c21421">(0040,a043)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_3e29ba44-1add-45d4-b9af-9daaf0811043">SQ</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_d5117ee0-ab77-4f33-8398-a9e8ae6d4813">ffffffff</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_d126b1d8-1c43-4146-96c0-bc9bde2daa09"/>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_5de1e3eb-59c5-4ed1-976a-b7832dd228d1">%item</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_01ce3227-6ebc-446a-b089-c247dc258f79">1.3</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_876f0fa1-f228-40b1-a28a-7b0d15384171">&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_5b535bad-1619-4263-94b3-7db335ef6902">Code Value</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_844785bf-6745-4df8-bf25-3fc439c4f630">(0008,0100)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_0134b5fa-314b-4b09-b1c2-3abcacc05451">SH</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_578a6bf2-88b0-45fe-a0b6-30833c73c113">0006</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_105a91c8-d986-45de-86ca-5e11c5f53772">128774</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_c8fd44ed-14c2-42f4-9bf4-5aa2a17ad0da">1.3</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_876ef6af-1831-4dee-97c3-2f6e00ecec9e">&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_451d6653-d587-4d3e-9946-0b8df2eaefef">Coding Scheme Designator</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_71c2cf08-a0ce-43cf-a931-0a5d5661f9b2">(0008,0102)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_4c16847f-7abc-4dbc-8b61-733db6ad3732">SH</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_31646070-fe4b-4515-bf8e-5d766b4a8585">0004</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_05839f3c-ef86-4a86-bfde-4f54564d8368">DCM</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_0ff72541-84fc-4f15-81ed-97a2acb1c8a0">1.3</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_d526777c-6673-4935-8975-839576c8a72f">&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_815e11ec-fed1-4492-9407-32f90add54e3">Code Meaning</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_79e2ae81-0bb4-4685-8bf7-3ec072d34df7">(0008,0104)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_b07f9a03-1322-4370-8b82-564b8547219d">LO</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_1f92b58d-2811-4f8b-9c87-28b51b01d9c6">001c</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_600240f3-33c7-44d6-b14a-fa03b8009406">Person Observer's Login Name</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_a9a52b94-d45d-45d9-8945-d42128fd8f68">%enditem</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_04633fa3-ccdd-4c9b-ac9c-a1cd40c44282">%endseq</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_13941dbd-f748-4037-8f3c-24f87dbf886d">1.3</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_f92d1965-0b7a-4f6e-846a-0b6ba54d6b08">&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_ab7f3ad9-b4e8-4695-886f-84c5f58662ab">Text Value</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_0c7a9bf6-e872-4dc9-958c-02d720b10b63">(0040,a160)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_5655d26c-b40e-4b3d-a1bd-31ded758cbda">UT</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_c2a0cd52-b9ff-406f-8ebc-4303442850d7">0004</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_2138d6c7-9731-4b41-8c24-c04d92e10c16">jdoe</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_eb81fec2-a5df-4c97-9f9a-8b9fd55809b2">%enditem</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_46ce8d41-2053-40fe-9c78-addc635e7e5e">%item</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_53be5083-c6ae-47cd-8895-7c577efdb5fc">1.4</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_b2d08ac1-685f-42f1-91be-5782469073c4">&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_32e32df9-d7f1-469f-812b-ce06f9685ede">Relationship Type</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_9b9658b1-2ed0-4b74-b3a6-0020a2f6192b">(0040,a010)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_e0e3d835-d5cc-47a5-ad86-5ba8687989ea">CS</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_78c175b7-180d-49c9-b096-e92df893d6db">0010</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_30649799-28e0-4431-9f80-be11b486c151">HAS CONCEPT MOD</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_6fc9e4b3-6186-4cc5-8761-63f5bc5db9f0">1.4</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_54d195cc-daeb-4bae-9d07-a45c9635029d">&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_470f6bc5-99e7-4567-ae61-a991317ccde4">Value Type</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_07896c6d-9f9e-42bf-80ff-3313aa94752d">(0040,a040)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_8c827fb9-5e24-443a-9bab-8a286c767da9">CS</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_ba1917cb-0cef-4203-a631-01cabf6b1a8d">0004</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_680b295a-760b-429a-81e4-06065a1df176">CODE</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_60742cae-ef7e-4cac-9615-97166ab0b259">1.4</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_737db9ba-6d39-47c4-8e33-aaa1606e71f5">&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_71867d0f-114b-402a-9555-9d7ce286483c">Concept Name Code Sequence</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_904bef6b-69bd-47e8-abb4-85c9f952728d">(0040,a043)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_baae085d-eb99-44bd-93b1-6848ded87318">SQ</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_0a33e855-64b1-46c8-ac91-003cfca1fb5d">ffffffff</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_e0e3cb2e-4a03-4c9b-9926-3871d9da2817"/>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_38f9a2b9-eb16-45be-9dfd-178094b03bb7">%item</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_8104f71a-bf0b-431a-9587-b00f08723863">1.4</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_e5eb555e-b2a8-4eb5-9b4f-abe101a7efb5">&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_14f324a6-5461-443c-8240-c2303c6d047d">Code Value</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_d8e6f485-24ec-4294-8cd6-b60676666fe2">(0008,0100)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_113fe39c-578f-4d37-8c7f-c2ac6f43485f">SH</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_0c7f1307-91c9-42c7-8ef9-46b42948a9b6">0006</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_aea70058-7d36-49f2-9a3f-75238d9e6b79">121058</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_96c94be0-df0e-4dd6-b6ca-e09212a77440">1.4</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_7bff31c7-aa94-4661-996b-85043a577c2f">&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_f95810eb-7e7a-47bc-9641-b9f5924ef7a5">Coding Scheme Designator</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_1874b586-386f-4533-8065-c1dc7f151069">(0008,0102)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_f510ab50-8cae-4c0e-a0a8-a0a6c72744dc">SH</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_7563f8b9-a58e-4c8c-b5e5-d4d0db7e7ce8">0004</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_59a20c41-e230-48de-bf62-eb3207e2e65e">DCM</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_e0a2db5d-7758-4769-9dd0-2fbcb7b56136">1.4</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_781d6ff5-5623-49ec-aa16-f52fa9a1e334">&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_fbb43e24-47e9-46d0-9a97-ca8fe5c70fc2">Code Meaning</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_c68da1b9-272e-491a-95ec-aeddce7fe596">(0008,0104)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_d763dea0-6bb3-4609-938f-796e0135f6be">LO</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_4e1aa534-ba1e-4419-bc3a-157db9b4ea61">0012</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_1a45561d-30fa-4053-9594-982af48b3030">Procedure reported</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_ba82d360-9bec-4bdd-a046-b3213b89c0e0">%enditem</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_b64c8d34-c7b6-465a-bd38-c32df69442b0">%endseq</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_22d5f16f-d9f3-4cae-a942-c4e66f3cad07">1.4</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_3030f517-2768-4e80-bed9-52175bb269f3">&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_2e9fcf6f-f31a-4913-ac33-1daf9cbdf68e">Concept Code Sequence</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_41644a97-1342-4702-8f9c-80b9407d9073">(0040,a168)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_3fb573b9-7ff2-426f-ab13-e2b4bf0269f8">SQ</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_37ef2fa8-e754-4fb2-aae1-3acea37ec0c2">ffffffff</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_2e88343b-a7b2-4eb0-b300-a47fb0a3e449"/>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_edae67fe-34bc-482b-8a20-79d8d7e7cb6d">%item</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_37d5e566-1c2c-4de5-8b79-3c8fd39656fa">1.4</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_c638d6c3-cb68-4271-ba3a-37ce3d2cca8a">&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_f30411ec-edc3-44f1-ae78-b569a80ca2f4">Code Value</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_55814160-b714-4bca-85e6-442843232a20">(0008,0100)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_cccb6603-0746-4ab1-900e-cc4af6f14b7f">SH</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_3db57d4e-d351-45a8-b6d2-4f49b592efe6">0008</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_28e140c3-a520-4ba3-b305-62f1d8fe61c6">44136-0</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_8829d4bf-6e9c-4f56-a4cc-b35d6edcf82e">1.4</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_af39217b-0d83-48f9-ad2a-0eb1cb0ec2b0">&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_f0427048-8c98-4738-8f82-0061292ac37d">Coding Scheme Designator</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_e9f772e0-1c5f-4590-9eb7-0887aaf3b0c2">(0008,0102)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_161d9b9a-f234-4cd7-bb34-79291cc50e3e">SH</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_96fbdc8b-5b9d-45e3-970e-1971db175199">0002</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_04902697-8b42-49a9-a77b-0258b342fb8c">LN</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_45d0ee7f-ce59-46e2-8e5f-15c497e0174b">1.4</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_7ad8bbdc-5bbf-4271-aa8e-bac1ef462898">&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_6d981f33-7571-4e26-878e-d2fa2ed02679">Code Meaning</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_550faf4d-943d-4d8e-92ea-42178083ea21">(0008,0104)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_2f719f3b-0c0a-4ce5-8b5f-f4a5732902fe">LO</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_c98a021f-459a-479e-a0e0-5d570d282706">001c</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_07676da9-8fab-48f5-94ca-95c1ceba1209">PET unspecified body region</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_735f023b-7ddf-4e5a-ae37-62df81931028">%enditem</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_f387d6ed-333f-4095-9d05-01e35b00611f">%endseq</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_6b1e7b8e-dde4-40a9-a419-1271e136b2b6">%enditem</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_f4089c62-b99e-4a51-93ce-f0eb268c5ba2">%item</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_10677863-f711-4396-a278-20ce43036487">1.5</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_54c8fea9-74b2-4cc8-9444-cdca1dd1a4b0">&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_00da28d3-92b6-4936-9b2b-86c464b4b138">Relationship Type</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_1e4ad544-77a3-4242-9bc5-7fa6255dd860">(0040,a010)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_4865cc5d-1759-404f-892f-ef6ca8f8e0b6">CS</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_7c01cc41-fdfc-4e34-a750-0dd49bfa57df">0008</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_55118bea-3e53-477c-84d5-44084610f47a">CONTAINS</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_2695d6af-0854-4a99-906d-9420549a64d8">1.5</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_89254fb8-70f8-407b-8132-85955ab81485">&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_6de5a575-86af-4f82-8fc8-b88f172a3881">Value Type</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_f0ae756e-2fdd-4714-9516-7f2cb641f134">(0040,a040)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_e47acc23-ff8e-4a28-8d72-a867ca1e381f">CS</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_6bca782e-61e1-41b0-8ebc-7277ce0d6e25">000a</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_92203937-c195-4e21-80e6-08ae93baa987">CONTAINER</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_36bf7488-7a19-4307-9cd1-7821e2369f54">1.5</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_0ebad18d-ac3b-40ed-ba15-1f14516e1704">&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_72c745c5-507d-4b8b-86a9-e82434dbef9d">Concept Name Code Sequence</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_1a9d705f-2ab2-407e-bb27-54592ef209eb">(0040,a043)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_b678f4b4-948d-4464-bfc7-c86f1739252f">SQ</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_28bff4f7-cefd-4a1f-afaa-09eabb3522c1">ffffffff</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_67b8d0d9-8e91-428a-bb8d-78ac1fad63fd"/>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_92598d09-d84b-4f9c-b5db-a17a429bd3f0">%item</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_16049575-2aa0-40fa-b15e-85909ebeac5e">1.5</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_9845b181-9898-4b4c-bd96-8d1f0c09936e">&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_aea9bd63-3ba8-42c6-bab3-c7490756b23d">Code Value</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_581bc1ae-aff6-42b2-b755-a678a9c37664">(0008,0100)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_27e446ff-f17c-441c-a100-69dfce5e5f9d">SH</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_59feb823-c4e2-452a-aa7d-6cbf38b5571f">0006</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_8be6aaeb-3bf1-41ee-80c0-cc2b4c1bfc7b">111028</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_e487940f-66af-4b6d-94ce-5456f86ec2c6">1.5</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_cdefd4e8-5c1b-4d73-8e1b-6fdeb6e473a0">&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_fdec47ae-3078-46c6-8e9d-f3f4538c1c90">Coding Scheme Designator</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_01920273-5559-4b11-9a1d-befc627e4891">(0008,0102)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_7dd51a3d-6eb9-4639-b6d8-deca8484d638">SH</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_6913af15-d1f6-48e8-a6b1-04953c7b0acb">0004</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_2a709983-4cb5-4d69-ac36-a4f6e9389caa">DCM</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_31cbf44c-8139-4954-a42d-3f55d345c166">1.5</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_3a6ae78b-7ea1-408b-a4aa-63bc8a480f23">&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_58aeda74-f65c-46fe-88bd-a9ccfbafd802">Code Meaning</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_c34800ee-7df2-4e95-a5bb-8220396685ac">(0008,0104)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_fab46a36-ec53-4e48-b89b-c5560e5253ab">LO</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_65ae08f2-c552-488a-9a79-5cec2f88efbc">000e</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_8af42fe4-9085-46b7-bb34-bf279dd60530">Image Library</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_965cd206-2e1f-4dc3-8bb7-b691cd31f769">%enditem</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_91c43cbe-576e-4858-b2cc-53807c7af593">%endseq</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_f67a3f45-fe54-4618-9886-7adf2e3cc143">1.5</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_c1fa0397-d8f3-46a3-a1da-18ce92028f61">&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_92fae534-8c67-4079-bffb-51a05740b24d">Continuity Of Content</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_7e431883-cc75-48c0-8ce3-3c30e3973ef8">(0040,a050)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_ead49334-a938-4367-a769-5673f907388a">CS</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_73600b2f-cda6-4cf6-a44e-1ac75cfee542">0008</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_4287b138-2f59-474f-b0be-d6dbe69a815f">SEPARATE</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_2760afc4-f2a7-4071-8664-b749116a83cf">1.5</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_685cd840-ae96-4999-b25b-aad7bd077806">&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_1387e9f9-eeaf-468f-91d4-e611ac541e79">Content Sequence</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_7f7ec0bf-19ca-48ca-b37d-f82e78652f71">(0040,a730)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_71455f50-11a4-41a5-92a5-ded6e9ff87a5">SQ</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_7918e041-687a-46f9-a710-ceff56a0613d">ffffffff</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_32a87cc8-f08f-4f01-bdd9-f4777d76becf"/>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_f9356aa7-b4ca-4740-a0d9-0c7b6c33fae2">%item</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_a2302112-5705-4056-90c1-6aaa45294aa1">1.5.1</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_5de682e5-bef9-40e0-89cb-9e11b8299da8">&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_e56a0325-f5a0-4908-aa5c-887539887fbd">Relationship Type</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_93e1973a-f04e-48dc-bf5c-01452e320cf2">(0040,a010)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_3b6b04b3-5041-4894-9ae1-4f72731312bf">CS</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_2e233e5e-9f0b-4abc-b5f2-e121ff19fda9">0008</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_9e799f6a-49e3-46af-9fe1-8b6e861c1504">CONTAINS</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_e7c1cce6-a109-42d6-b5eb-b2b3c7e498b9">1.5.1</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_8d2305c2-1583-47a9-bd83-c57803bb1af6">&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_505dbabc-2559-47a6-9513-d67a8e488c98">Value Type</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_fb7e7f00-8ff2-450f-930f-d6e6ebaeedd1">(0040,a040)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_22931139-7eac-4b5e-a367-0177485d9489">CS</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_dac8d723-9059-4dc4-b8ed-70ec89cde038">000a</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_88abbff3-8571-4884-9621-ed8c79ad83f2">CONTAINER</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_8acf50b4-357e-4c5b-989b-753018b6c656">1.5.1</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_fa74dd80-f9c7-416a-b942-74ef43dac219">&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_a9b515fa-8663-45a7-9667-66f92bd36470">Concept Name Code Sequence</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_2ebc9749-706c-49dc-bf07-6f78c84a2490">(0040,a043)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_20612426-89b9-43c0-a4c8-517f8c4c2c89">SQ</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_25328fc6-d8a7-46df-8951-d083d2fb9780">ffffffff</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_dfe18f8d-654e-401b-a7d0-1bcaaa1d9928"/>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_37d211c0-7535-48dd-ae96-811404996977">%item</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_f5e7b799-1304-4581-9542-a118a1d3056b">1.5.1</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_8f21246b-f156-4cc3-9b68-aae6862e139a">&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_e647961e-2c75-4211-b566-a4a961eea135">Code Value</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_adfad3fd-46ce-4f6d-97b4-79440512e357">(0008,0100)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_2a6a1249-effd-4da4-838a-5b87c6a7ad4b">SH</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_a5befb4e-d5c9-484d-8583-92c00c9bbfad">0006</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_318c6780-9a13-4355-8094-69eb38529226">126200</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_eb360143-5ebd-4b9f-84dc-3e3b0e746b8e">1.5.1</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_fd4e4282-efa1-4365-a980-a13a6a2feee9">&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_5ab6990b-4f1d-41a9-87d1-8ee2a6fbd27f">Coding Scheme Designator</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_1d4963b6-e9f5-4427-b261-4e5569b80c3a">(0008,0102)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_10270158-2881-4b09-ba38-c5e7615ba794">SH</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_ed1c2177-784d-4eb1-9c8c-f1df893ea02f">0004</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_48b10fef-ac07-46f2-8bba-94e2c8a24ef2">DCM</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_ae62e44b-ef47-417a-bbd5-d97ce38b2c56">1.5.1</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_40125570-ae8e-4840-ae13-5c7346b0efd2">&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_feae60ad-0cd1-4336-a3ab-6fcf968740ff">Code Meaning</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_55e65456-c2df-4ae7-ae8a-20286c76c813">(0008,0104)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_28741d9f-9b2f-428d-9b05-d3decadc414d">LO</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_67ba9b74-d0d7-4cf4-9a8c-349a836f5799">0014</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_bfb3eaa3-25db-4998-8f59-47b11b01ab58">Image Library Group</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_376ad0ae-0e29-4652-a16a-4e692ba67d8a">%enditem</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_992fd932-57b8-4b95-a61e-215b711cab5c">%endseq</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_e944125c-4564-4542-bec3-33906ace19ec">1.5.1</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_9bde0230-62f6-4594-86e7-892049ff168b">&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_f499b038-49be-4f6c-abf5-f68cefc6eef6">Continuity Of Content</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_d7b02d7f-f386-4289-b740-8405fb1e5ca9">(0040,a050)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_d0e9a77a-caf2-460a-a79f-1dd79a9c3e23">CS</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_89a46614-061d-4186-8218-7ea400e602ec">0008</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_375a3809-492e-45a1-83c1-328143cbfbfd">SEPARATE</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_25a8e5cb-31ab-4f67-9c46-3f09dbeeb16d">1.5.1</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_e7173efe-d148-480c-b08d-e4c355ee2d35">&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_960276c3-f026-4e55-bb9c-824ff2ef99b4">Observation UID</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_10835cd3-3ef5-496c-a4c6-a74fe045eeae">(0040,a171)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_3a6ccfd5-8ff8-4825-b85b-924eb484a8f5">UI</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_d63d4330-b515-4185-ab2b-426b907b3197">002c</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_6dceeac0-7303-4af0-a217-3e789a631536">
            2.25.239108061065263370785162033783811931375</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_db71a08b-df29-4769-b41e-7a2533494054">1.5.1</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_5cbbfc9a-1b0d-446d-b350-74b120452848">&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_904db7ea-a6de-4a91-942c-ecb8f200deb8">Content Sequence</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_38deafbd-7a65-4ecb-9e29-ab615c20fad3">(0040,a730)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_00fab628-29b4-4dcf-b381-e5c1af9fc87e">SQ</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_9c02094e-2cc8-4cd9-9237-562435991830">ffffffff</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_fc2b6ad5-0c91-46d8-842e-de4edb728644"/>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_0bd3408a-6a46-4732-a90d-41f3b0df9e6a">%item</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_c2f9560d-9161-479d-9d7c-07271b232ea5">1.5.1.1</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_249d8bd0-9bc9-41e5-b2ca-aa3349da61ba">&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_cd8cb14b-a7bb-4a72-8154-8912dc9c8b03">Referenced SOP Sequence</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_f72b7627-551d-4a69-b88b-9cb0e4694bd8">(0008,1199)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_b0bfb3d9-6397-4df2-8b19-1f190c1d6c54">SQ</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_4ed17476-4823-4b17-a85a-8d08a064b2e4">ffffffff</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_caa60664-b8bd-488a-8dd8-0769ca1d481f"/>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_6676e61c-92c1-4db6-869e-22edad731501">%item</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_96f3942d-a999-44fb-8e94-f1d02c3445f3">1.5.1.1</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_b67c45b8-f51b-4595-8653-54b2d61e763a">&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_28dfa26c-9245-4604-bd1c-4f1313f57c00">Referenced SOP Class UID</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_c41dbc7b-6aa5-45fe-a62a-e258f3a80a08">(0008,1150)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_0ee9e16a-b4ba-45a0-bf92-c5ba0d5695ee">UI</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_17f88eb3-3628-45c6-9845-5b843d4b41db">001c</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_1c05c600-d91e-4fa7-a0f8-020699d04100">1.2.840.10008.5.1.4.1.1.128</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_263b0c9d-d85a-4d53-a78b-54a77f0a7642">1.5.1.1</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_9d149f7f-2eff-48c5-85cc-47d35197f6f3">&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_8179a817-90d3-432f-9370-f0ede5346c69">Referenced SOP Instance UID</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_f545385f-f339-448e-9da9-32563c2c7e9e">(0008,1155)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_4d3269fd-f486-4ad6-ab6f-67daac8be376">UI</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_5143e46f-daf6-4bcb-9e22-b66ad38ae3b6">002c</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_296f6a2d-5953-48ac-af8e-0fcea8c44e12">
            2.25.319214308104243787945491694789635628411</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_fd90f90e-cec7-45c8-90bb-f096311f6526">%enditem</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_aa4b47f8-a8d0-4deb-884e-67f05d466e25">%endseq</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_5985713c-1644-4a1c-b759-3ad0831fed1d">1.5.1.1</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_9b4c8ea8-b64c-4989-9625-3f167368c87b">&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_f68249cc-46cf-46de-8eac-228b351dcc86">Relationship Type</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_0fdea705-2f77-4025-8dce-5eff189832d3">(0040,a010)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_bc2ab9a5-8086-4a2d-a94a-67020228d110">CS</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_38de39ae-f4fd-4765-9e6f-d62535836571">0008</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_df14a622-c7ac-439c-a5d6-c30d4fae71cb">CONTAINS</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_3296dbac-0e44-41a2-a6e7-d574cfe150b8">1.5.1.1</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_3b2b2bc5-1eb6-4372-8134-328ed6d35769">&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_de6a8b0d-6efe-453e-96ee-2e2c509c881a">Value Type</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_1d1c1091-8f8b-4a1d-b5d2-1c8ce4028ddb">(0040,a040)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_f37368b6-278b-4deb-b9c9-9de1aefeeef9">CS</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_44077411-b361-46d1-ae4e-ca8eb508cee3">0006</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_b119eb87-f02f-484a-afab-08a50de9aaf1">IMAGE</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_1035ef05-3f18-47e9-bcdd-2ee38b9c1606">1.5.1.1</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_2e22ac90-1435-44e4-ac92-30bf5204fc7a">&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_4e1329aa-2f9d-482d-8283-22082387a58b">Content Sequence</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_1dffbee5-a3ac-44be-b243-9a90c097b375">(0040,a730)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_e9a0cb4c-3331-423c-89de-f4e2a81f71a9">SQ</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_b6ac37ab-bf91-4e19-88bb-04eaae24549a">ffffffff</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_352ee97e-d826-47d0-9799-6978ecda4589"/>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_20b48538-2a37-4bc6-9cf2-76d3516187d9">%item</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_03922284-156f-484c-872e-f995bf556b77">1.5.1.1.1</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_aaca7858-8ffe-4efa-b739-1f7f93da74d0">&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_65e4fa78-7e32-4d19-a49b-ecd28f3deedf">Relationship Type</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_29a41d97-1cfc-4705-aee6-5d42f79c4375">(0040,a010)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_138d9deb-fa72-439e-abe9-e54d068541d8">CS</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_92921b35-4403-43d1-9d44-abae7a417719">0010</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_b1709441-c647-42c4-bf3a-a8e6255728f8">HAS ACQ CONTEXT</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_74aa270f-5d9b-4ecc-8b5f-90e94e7dca48">1.5.1.1.1</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_24f19d5b-fc64-4056-9083-f5d360ed0de4">&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_65c62b0a-bd56-4ba0-96d2-89aca25068ab">Value Type</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_abda1071-001d-4a5c-bf99-36c66f10a780">(0040,a040)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_d948b106-8b45-4e0c-9fc1-92cf5b596c79">CS</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_56fa5c1f-95f3-4a77-bd0c-be410eca7afc">0004</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_1b71e176-656d-426a-82ff-256f7231d967">CODE</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_c16b0fe0-98e9-455e-81e3-a911fb21d3fd">1.5.1.1.1</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_c733f64b-083d-438d-ab12-e1718067e74e">&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_616eeacb-3a93-4d77-9060-c3c0b51c61c6">Concept Name Code Sequence</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_139c3f07-81da-42b0-b4ce-619e9ee3d75f">(0040,a043)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_9b688a2f-5475-46f2-bf5f-398b31004913">SQ</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_04106e99-47ff-41f0-a597-37f0f3819092">ffffffff</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_ad5c29c7-f270-4dbc-81e1-a70310aa13e1"/>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_bdc1e62f-5ec2-4493-b3bf-410223f04821">%item</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_72fdbc48-55de-4429-895b-5e623fc8eeb0">1.5.1.1.1</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_ec2d6d1c-ce5e-41fd-916b-85fbf9e8759c">&gt;&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_dc947536-4689-44ee-b960-5239eafad857">Code Value</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_3866c5cf-cf51-4091-994f-568d446259fa">(0008,0100)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_036f2919-6b2f-42f4-9340-5269ab792c40">SH</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_e15445b1-e4ee-4473-9f70-6bc81e682dbe">0006</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_1f550372-a759-46d7-971b-cb91d85503e7">121139</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_efefc8c1-40b4-412e-8be4-126d94f60d53">1.5.1.1.1</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_6270a498-120d-488c-840d-0b0892e1514a">&gt;&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_d726d39c-3033-4b18-8c82-41a0cb2e9e86">Coding Scheme Designator</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_5d25df5b-54f6-4f81-a861-4a85812f5083">(0008,0102)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_d25b4e5c-9943-40fe-bae2-426d4fad30b6">SH</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_c3f63fa5-abf8-46ee-b2fd-925077aa365d">0004</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_5a24779a-d0aa-4b16-be8e-4cf2d95e2767">DCM</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_1e04462b-2ae9-4be4-ab5e-94bf3b45fa35">1.5.1.1.1</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_5cff6e57-60c3-4d6b-92ea-06f37363779a">&gt;&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_bd1824e7-38c4-455d-9cb6-9a8c569a7f3d">Code Meaning</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_70860551-05db-4574-92a3-9adf9a409a88">(0008,0104)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_cd256e9e-6817-4d87-9eac-3571655a155f">LO</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_9ad147bd-105b-418d-9599-6e214a4071a9">0008</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_dff53f1a-243f-485a-bd28-c3b8e7e4d797">Modality</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_605179f7-c991-4355-add1-4d0d1df686e3">%enditem</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_2ddd37e1-242d-47b2-bb8c-cc4e8e661c2d">%endseq</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_88458e4d-b4c1-4f51-b299-384f443533f9">1.5.1.1.1</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_21ec6590-93c1-4ca2-9cfb-e367d4c52b29">&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_1efad2e6-6c90-4b8a-a7a9-7bb1ba3200d2">Concept Code Sequence</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_b424dc0d-c612-4c97-8bda-37e9475c6ea9">(0040,a168)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_88484c99-4ab6-41ee-bce2-a48634a3d9a0">SQ</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_129c23f3-32c0-47ba-9d33-f491e5942a87">ffffffff</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_4c049d43-a861-4525-aee1-af2a4c270483"/>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_9f93f631-f122-4d2d-a163-a8f239dba5ac">%item</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_7aafce63-5a64-4b9b-9bfb-5f30aff631f2">1.5.1.1.1</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_1afa2ee8-80e6-4d03-93d6-d8e574c57e47">&gt;&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_3154dfed-f667-4d74-9fd9-f09b459454a5">Code Value</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_bf1051da-23c6-41c8-a9bd-f13144df0122">(0008,0100)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_2e144729-18b7-419d-bb60-5f7fffefc660">SH</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_538538b6-3bb4-48a4-8f88-7cda56b5839b">0002</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_49651476-09dd-45c2-a114-127defcc2d36">PT</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_cca62703-4999-4710-b27c-ccff0aaf39e3">1.5.1.1.1</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_b37160b6-31ae-4c3c-89df-1a07f71b7023">&gt;&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_2b20b850-b4e2-4000-9e8a-5e9f674a9e33">Coding Scheme Designator</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_93833d4b-2e71-4d54-9645-bdc9da50aa1e">(0008,0102)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_bbe85377-d35c-410c-b422-b036e3c9d05f">SH</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_d951fa88-b4ee-429d-a79e-a922cee1bfde">0004</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_e066fd9e-6713-455a-a8bb-81b862565c29">DCM</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_99ae594d-a2b6-431b-9ff1-3918cf9af1bc">1.5.1.1.1</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_b5a071a0-e491-4ee6-b778-f216a42c7221">&gt;&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_11af4a9e-8ae4-4afd-af46-5b8f6416651c">Code Meaning</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_3b98378f-4623-4d9e-8db4-ebd872ddc48b">(0008,0104)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_8bf77bbc-bb23-47a9-9d05-91c94a11f4fb">LO</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_1f4e355c-e925-44ab-9e94-395013130573">001c</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_03b9e32a-1a65-44b5-a10e-5df080d36cfc">Positron emission tomography</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_315bfead-b7ff-4efa-93df-33f0514646c6">%enditem</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_6e672340-5f83-46ef-8e85-b5488f81af5e">%endseq</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_5f516b6a-37ac-4b7c-880f-e847c1280354">%enditem</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_af942b98-ef7f-4bdb-8762-11570ab5adf0">%item</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_6ab15903-6e1d-43a2-bef6-c2e6b6b21033">1.5.1.1.2</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_b579133a-f61e-4dae-a7a5-fe81734a9e22">&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_c4d28a14-29f3-4b63-8518-77962b19511e">Relationship Type</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_f9782090-76e0-4636-b44a-ca34f8c46837">(0040,a010)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_c815f635-2fd6-4136-b81a-9e95a4708a87">CS</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_9245d30c-f3a0-4138-a46e-ea14b331d7c7">0010</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_3322c5cf-9028-450e-b39a-4a910625c1e5">HAS ACQ CONTEXT</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_848b9c4e-cd2c-408b-a21d-3c77c010a1f2">1.5.1.1.2</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_3017d70a-2ef5-44cb-a4d2-a151fdeed6d7">&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_b1a1e23c-32f5-4ff7-a5e0-e99af9c063d5">Value Type</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_ea7214b0-6434-43db-9f50-082a45abe97a">(0040,a040)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_ea50e893-0c42-4747-b00c-eac1f1f884fa">CS</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_b84e42ba-494d-48e3-a85c-8d2494e0b817">0004</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_6c281558-3603-4cee-900e-748e28e22aeb">TEXT</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_fd552a5f-deac-42be-a4e0-53cd8b652ddb">1.5.1.1.2</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_0fe1adc3-0611-4f95-8e0e-869048f94d8b">&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_85afc426-3669-48f0-a06f-82f65b9f9b7b">Concept Name Code Sequence</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_aa635348-39b4-44a4-b160-1fd0dc30a7cb">(0040,a043)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_47d18d28-6c59-4c59-b815-1bfad0d9c9e6">SQ</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_5f0d348c-6849-4981-a086-e2be6eeeb4a6">ffffffff</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_002f8350-28dd-44ea-85a8-109b9b237053"/>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_9cb28ec5-263c-4573-aa33-c5e43391cb66">%item</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_39b481df-6551-42dc-8b76-70cfd2865cf9">1.5.1.1.2</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_3d76cfc7-a3dd-41b1-a9a9-e42f276a47d4">&gt;&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_464f7906-b413-43d9-802a-698eb6551db9">Code Value</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_697b8cb8-fbd0-4f97-90eb-594068c27189">(0008,0100)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_777ba75a-26f6-4308-bfe9-d176212483e5">SH</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_cadd98c6-e462-4bef-845a-8a40d6c96c95">0006</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_9e16f0a9-856e-4e04-9fc6-ef031386ddc5">121022</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_52b96f60-52b2-46f2-8acd-00f338bff152">1.5.1.1.2</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_047814e2-f899-4420-bc71-2eb082f8ef1b">&gt;&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_9d4c2b6d-502c-45b6-86ef-b42f0354c40b">Coding Scheme Designator</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_502d2a56-0ffa-438c-a03e-b27f6dbd6a02">(0008,0102)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_f7cef73e-6633-4bec-9e60-0bcbebb77905">SH</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_7c4140e1-f6c4-4da8-a9df-da6522683ddc">0004</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_139e1502-cfce-4aaf-843d-d978c99956e7">DCM</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_8f3adb50-ea3b-405f-b6c3-ceb3466f8d13">1.5.1.1.2</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_4493e6a0-dc4a-41a9-8702-ddaa7e7b4e4b">&gt;&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_a46ba16c-373e-4983-889c-9e56386436a6">Code Meaning</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_28eb8e17-cf67-4ae7-bdbd-8871a4082fd8">(0008,0104)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_d2fd6a4e-2666-452a-b4e9-57208e11080f">LO</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_504c5324-13bf-40d1-8437-304871fa0b3f">0010</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_1193432e-91da-436b-a5e1-a0faf1e67306">Accession Number</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_67cc3d85-465e-462f-99af-ecd835e0a93c">%enditem</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_1fec1ed2-edd5-4013-8b63-9af2aaa8ed2b">%endseq</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_7c5ff48c-4a9c-4e2f-b5d5-4b045815b5a5">1.5.1.1.2</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_dadb8343-a6c9-4409-a931-7a504d849847">&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_9275fcc9-0f53-4fa6-8581-5e274e360ac9">Text Value</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_ee88a3c6-ba05-4cf8-96c7-b16fd1bdadee">(0040,a160)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_d8d12dd4-5e4f-4815-9c07-cba7b491ae05">UT</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_199231bb-04db-459d-837b-088a30812c59">000a</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_f78b53ef-43c0-41c7-9eeb-05db121b4223">AN1234IMG</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_6eb7f304-d71f-4186-a713-36fdfc4ac004">%enditem</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_d854f2ca-1b71-492a-81b0-b249c1cf478c">%item</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_31c785b5-fd62-42e8-b093-a853a81e57b6">1.5.1.1.3</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_a31bdf3c-14bc-4f00-a9df-81bd7646baef">&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_bcd3a224-e7da-4153-827b-91a312cf6e09">Relationship Type</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_9c850576-d3ef-4188-93e6-aa58767239cf">(0040,a010)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_aa8ea210-dd22-44fc-a04e-ae844a5ae2d8">CS</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_c78f2d5a-3f93-4564-b091-a5d652b213b1">0010</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_c1ac75d3-dfa1-417d-b456-3e89c69b7727">HAS ACQ CONTEXT</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_773039b6-2659-4f9e-9b08-73182feb4230">1.5.1.1.3</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_a7bee178-cccf-45f5-ae86-5f05e02d23b4">&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_b2e33a61-c8fd-4571-a4ab-07921e103261">Value Type</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_0a564afd-336d-4e0f-b3dc-a33fe44fbc7d">(0040,a040)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_890cede8-8fd1-4a15-b126-e561935df1da">CS</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_f51be236-db63-4308-9f82-78d7b6b27a52">0004</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_0d8b2916-1edb-4895-a27e-04c391b3abb3">DATE</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_b342a7b3-8e91-44ae-9cee-8e9e66d7aa82">1.5.1.1.3</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_b0d8d2f8-2fc8-42f0-8e75-9778fec313f4">&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_5ecc0bd3-e81a-45c2-b50d-7c316117c538">Concept Name Code Sequence</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_fccf1087-9781-4916-a999-82c2dfba2883">(0040,a043)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_2fb66bf9-32ff-43be-bb3e-8714eef1023d">SQ</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_3de353bc-c115-443f-983a-bb24795ea480">ffffffff</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_92f29fe7-7171-4f96-82db-9061ae3670a7"/>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_7a433355-fc87-40ca-80a0-5efa620d42eb">%item</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_06ec5187-75f4-4ba1-a8d1-480d7e6d8f1b">1.5.1.1.3</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_f077a3f2-1a7d-47a1-bce7-92d6d9141b91">&gt;&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_1a9e0539-b88b-4fe9-a0ba-0d5746e7cbdc">Code Value</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_cc11ac52-528c-402f-97b5-e203601becb7">(0008,0100)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_345cfdb9-c831-4487-abe5-8c23f2c72314">SH</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_f1d52679-ef36-4216-9a07-1bc40d9bd096">0006</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_223d428e-739d-410d-a27e-111f59f40d72">111060</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_c2181dff-9c10-4a78-aca3-47e524504886">1.5.1.1.3</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_b7d29944-59f3-46a3-9082-b93921e270cc">&gt;&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_3d522b64-dfe3-4c88-9a26-fc29f0a3ee6d">Coding Scheme Designator</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_fa0d195b-7369-43a9-9de4-1debd68c25fa">(0008,0102)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_0b6d9de0-4f21-453c-92df-973d73ceb85d">SH</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_db0d37a5-1864-4682-a1d8-e6d557f3065e">0004</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_b04e5164-20ca-4f14-aa8f-560f0ec8b577">DCM</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_ad6e3536-dcbb-4cd1-8b13-f82759c274a2">1.5.1.1.3</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_42fe0890-9c85-4ce7-8698-655a2f300a17">&gt;&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_e3225ee6-6007-4105-ae4a-89b4e03a304d">Code Meaning</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_c5d62999-0846-4d8e-bead-f163b6750913">(0008,0104)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_28b6a95a-f3ea-4c26-bbc2-eeb822ed3c11">LO</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_f1c9e879-61f2-4090-bc47-3f062777451c">000a</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_65f17445-bdb5-47fc-a0ac-792796712992">Study Date</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_53432b11-86a3-40e9-831e-53712462a262">%enditem</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_eddd9b12-1847-4fde-9a4b-49b33aa12543">%endseq</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_e9ff2792-b897-4f9c-bdc5-b2eb297307d9">1.5.1.1.3</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_6f107f37-3e52-42f9-8f0c-87faafcaeca0">&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_aa9dc732-dbf3-4639-8897-6c3db2d18ece">Date</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_b697323a-9fa8-44f0-bf3b-4043703e2741">(0040,a121)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_c8bc944f-8449-491b-8ff1-c7c2150ea99a">DA</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_f6436e38-0159-4a48-a8e9-6dc70d352882">0008</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_6b62d1cc-41f6-4f42-bc61-574bf54007fd">20170113</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_c1e00776-59f6-43db-b654-b0f15a100803">%enditem</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_fe2bba99-320e-4545-9816-14930477bf7a">%item</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_1ce2f166-edc2-4df2-9f59-0f147ae8c12c">1.5.1.1.4</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_1a1534e4-1bb8-4aef-88ff-653bcdb2dd40">&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_6d2f3681-9104-48fb-a725-c0424e47620a">Relationship Type</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_f3902c14-48a2-4bd8-9750-3eda8f5758cc">(0040,a010)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_c272d5be-81de-40c4-9082-0d84e4f3c3d5">CS</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_adacd85d-ee4b-461c-b455-d5ff7de4260b">0010</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_3ba03bc0-8e1f-4d1b-8c67-9cb7dd07feb2">HAS ACQ CONTEXT</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_86c90312-a713-4b8c-9735-bbaf4f830f38">1.5.1.1.4</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_994d340f-cd88-453c-ad9c-fe423ee4bffd">&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_d5dd16fd-fc62-4137-a3e9-aeddc4ee632e">Value Type</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_ee6370f5-f1f5-43af-b40b-36f94b6501cb">(0040,a040)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_9add07ee-54e5-4ce3-9518-b50d2076344f">CS</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_070a4de6-7dcb-4672-930d-9eac6ffcd9db">0004</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_de566c66-558b-453f-8982-3be76d2fc907">TIME</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_84d0fbaa-ec6f-4808-b98b-6463b337f1eb">1.5.1.1.4</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_5fb5f910-f422-4026-9ca5-ac903458809b">&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_2afeb023-ec13-47b3-b72e-1bcb212faa06">Concept Name Code Sequence</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_93d7b4d8-740b-4a3e-a255-4bcd381edc18">(0040,a043)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_54e17f86-4e6a-4b86-8998-ea7d8451f508">SQ</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_f87f1a9d-de81-4934-bab5-a6766a66cbe4">ffffffff</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_ad4b2f8e-a5cb-4b4d-aed9-0b34c04824cc"/>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_28d2a4a2-6c6f-4fae-a79e-86897ff39e9c">%item</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_d7460c07-7900-4cfa-b8c8-22671e00dd07">1.5.1.1.4</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_36da45cd-836c-4f2c-ac9d-c6f33c01e8d6">&gt;&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_6fff8c2a-0c04-4b3d-a0a0-f2f2552e5d9c">Code Value</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_6610a2a4-245d-4cdd-b579-c171d0816ed0">(0008,0100)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_f6cd59fc-b603-41b3-92b5-d64dc23a4b51">SH</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_31643ec5-9396-440a-be6d-bf83f2a1aba3">0006</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_046cfed1-62fd-4339-a32e-4d322656b366">111061</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_6a2db5ea-e94d-4fc6-bd4a-29ec0d4c713c">1.5.1.1.4</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_de9faa72-d9ef-4c8d-a588-364f5bfe9c02">&gt;&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_eb0d87bb-b09a-4637-85e5-e201f4e9c43c">Coding Scheme Designator</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_bcd5e2f9-ebb4-452a-a132-21e396f52558">(0008,0102)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_d67af828-1bf4-4345-a5eb-592e8e921eca">SH</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_1974199f-bead-4f44-bd9f-20fa314fe63a">0004</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_8befeabb-b634-4d4d-976d-02db22edc6a7">DCM</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_072587a1-45e2-4403-9bea-5ca0c9ab777a">1.5.1.1.4</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_cf30f5ea-4709-431a-8523-ba0304b15869">&gt;&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_fcb35a6d-f1da-419b-b249-2d3235f1ab78">Code Meaning</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_09b90969-83b1-408b-a69e-d7ab5390a364">(0008,0104)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_b3872ab9-6a92-4f21-ab74-e363aa011507">LO</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_f5dbd855-9caf-4f57-8a6a-0acc1c44feaf">000a</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_778f9205-d73f-435a-b3c7-b7564dc60fef">Study Time</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_2f36819a-2db7-4583-ac29-fed2c96c8302">%enditem</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_80cd8d41-266c-4069-b824-8687227d2eee">%endseq</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_c2511f2c-4a2a-4e4e-bdb9-abc36a4f549a">1.5.1.1.4</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_f1d05c5b-4be9-44a2-b9ed-7d26f336e44d">&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_e0a295b4-d468-4dff-b2ad-085833b9e80b">Time</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_1580ea4c-caab-4cd9-94ea-0ad66da0bc0d">(0040,a122)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_593ce3e9-cb53-4af3-a1ca-1d6ae9207618">TM</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_9782a872-c0b7-4e1f-b4bb-19787aedb967">0006</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_44efd284-19a8-4806-9c69-ba03809a3c9f">070844</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_88719164-52ca-4500-a22b-fa1ec6521c48">%enditem</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_21cb13e9-4ee7-42f2-a446-4b552664beb0">%endseq</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_b0b96788-cb76-4c8f-9994-bf35c3ef24bc">%enditem</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_2a95bb69-474b-40b2-b8f0-9435bfa9c19d">%endseq</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_696dd2f1-6760-40d7-9e44-d8166142e3e7">%enditem</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_cc6f5203-6f46-4ba1-aadd-b2fd3b68e248">%endseq</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_7191bddd-2561-480a-ab72-0f9c21f7f326">%enditem</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_d0ec95ac-5d50-4505-ba01-417410de73a1">%item</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_33a071e4-eacb-4d29-917f-9e25d850cec1">1.6</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_be505135-9670-4dbc-ab7a-ec4c9a3adb73">&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_a87a2ba4-b2db-4c5e-8cf0-b8d7dbeb37ad">Relationship Type</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_568319a2-da7b-416a-8fb3-a20edf8bf162">(0040,a010)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_5788cc95-9a61-46e5-b4b2-265869232f98">CS</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_684acb33-19df-49e9-89e3-91a7f484da60">0008</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_afcf7802-6f9e-42c6-9e00-547328a9faf7">CONTAINS</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_a4ddbdb2-c1e5-42d1-9f1c-ce0998ff5ca6">1.6</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_0769b0f5-8a7f-4cf5-bc65-f0f1533bed55">&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_980010a0-1010-4ff2-a7fd-1e625930473d">Value Type</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_36a8dbcd-49e9-4bef-ad86-bea8e5a24b47">(0040,a040)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_f434b8a3-9949-4c7d-823b-98539be252e2">CS</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_852f9e22-07f5-4977-b642-e38d9f924c0c">000a</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_308ee8a8-c688-457a-b8a9-8fa9e65d0f82">CONTAINER</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_baf453b7-ce72-4240-80ab-55d5f9bbdc0a">1.6</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_96a1a997-0fd3-45f3-bc1a-1f2614be92cd">&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_d469d2b8-d965-4a97-85a1-bb852abbcc1e">Concept Name Code Sequence</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_9d83304b-f62e-4af3-a76c-374b9d8a7edd">(0040,a043)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_73c83fc7-8ee5-49fc-b31a-d1fc3703f1c5">SQ</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_cacc5af0-1699-4c5f-85ca-b28758465cb4">ffffffff</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_319df901-4e49-4cc6-81f2-f1d419f1e959"/>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_5d234b35-cd73-470b-a21e-197f34f8d64a">%item</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_0c4a9415-ad88-4cc4-bdad-e3498bf91c22">1.6</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_e3eb5225-bd3f-4509-8b63-80609c32fe02">&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_ffeb6e0a-9efb-4fc3-9f0a-5906ab91aa78">Code Value</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_0606e431-99b4-4f1f-ba45-49d63c96dd99">(0008,0100)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_4d369223-ab7a-48c9-b3a8-112165ed6a4e">SH</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_23fb5189-fb6e-466c-9c51-fde648e64836">0006</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_4419dc0e-7c49-411c-87aa-a2aa025d6495">126010</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_777dbabe-e916-45b0-b071-9f2a762eaf1b">1.6</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_3e8cce7e-f142-4705-8d85-8de5072e6311">&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_34b21b5e-a8e3-4b44-801e-658093264586">Coding Scheme Designator</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_b1cbf6b1-af44-4176-801e-586643cab049">(0008,0102)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_b4a65abe-0a5b-4e3c-bcba-bd1e76cb934b">SH</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_bdfab28b-f717-47e1-81bb-76d2be6dd1c1">0004</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_c9a7058c-c87e-483c-9164-63e60781574b">DCM</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_80788fe6-0bd0-4003-9317-22e16e1c6842">1.6</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_d16ba37e-5fff-460f-9bfd-4ea50ef67848">&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_2c087d71-28a5-4c4c-86cf-e10e37a0dcc2">Code Meaning</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_9ee0b13b-94a1-4278-b764-6299ab950898">(0008,0104)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_acb8aaae-a587-4abb-9875-b1cbb30bfb06">LO</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_685b0d41-3468-40ff-a557-63c70dfaf3ee">0014</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_8213f45d-ce82-48af-a934-50f81cad6ed2">Imaging Measurements</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_6cf4a3fb-45c4-454e-977e-d8cea93cc032">%enditem</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_e08f1f03-8faf-458a-b3c3-9e8b960fa14a">%endseq</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_eb580c86-0354-4d8e-9f94-4229502acdd0">1.6</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_618a2779-f86e-4e4a-b52a-d093a78bea1b">&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_600bfa7e-9966-4e24-987b-dd91c7a3d2db">Continuity Of Content</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_ee0efd16-56fa-4155-9a24-7abac33c6fea">(0040,a050)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_e9329ed9-ab7a-4c7c-9f46-85f45f4b476a">CS</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_dfebc509-7dae-4b11-a088-b70a38295227">0008</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_046e1322-e296-401c-ae00-01ad9a76474c">SEPARATE</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_b0739c48-3190-46cd-80dd-12e0c5ebf507">1.6</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_d68bec23-21f1-4338-9a3d-809e7d176552">&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_d372b214-8302-4273-985a-4955febdbe3a">Content Sequence</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_25314ba5-ee25-40bd-a11d-a7b449ea856b">(0040,a730)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_6f5cb295-7bb0-4fef-b11b-af7b3934677a">SQ</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_15fe9fd1-d408-4d92-9304-a50ffcf98e3b">ffffffff</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_481b605d-da95-4bd7-8ede-c5fbb7d26577"/>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_1093c356-bd89-4c83-b181-74563e897323">%item</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_e02804fe-5bc6-46a6-98dc-fb4ced22e44a">1.6.1</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_5227c460-326d-4c51-9e38-06b7c721eedf">&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_65323509-951b-4e4c-a56b-07fb2d5e1113">Relationship Type</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_f30c1f9c-f876-41f7-822b-fdf1809a8a25">(0040,a010)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_79819be1-0085-4a65-9ba2-b38aef6adb06">CS</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_f708cfbc-0f23-4d71-bc18-9d06961b9dfe">0008</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_96cf10a3-da5d-4e1e-833c-b93cd3015a2c">CONTAINS</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_4cff7b24-9780-4061-8f8a-76e99c87630b">1.6.1</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_00a067f5-bd70-44c1-8cb0-5d622e08b28c">&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_b04b2b7a-58ca-4ad6-955d-54725a20b27c">Observation DateTime</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_c79e0d1f-4522-4df8-8765-f5d7471983c0">(0040,a032)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_086cc7f5-7810-4175-9b02-61274bbb3cc2">DT</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_081e1cd4-db3c-43a3-ad2e-d914b0f90024">000e</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_ea7743b9-f4af-4a19-b46e-21103e002e0b">20170201180043</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_651e8383-ea72-49dc-bafa-e091061af535">1.6.1</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_7a95563e-c0a6-43dc-aac0-700d2768a79a">&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_6fcf0835-1b57-4621-9522-f0f701393ac2">Value Type</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_f2108bda-5761-46ea-be4d-51e82373ed7a">(0040,a040)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_10b32cce-a535-4f4f-ba4d-ef974f89e2a6">CS</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_a2de968b-164d-4081-ad23-8a27e44e94a7">000a</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_ddc71a75-d0fd-4488-aacb-fb61640fb469">CONTAINER</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_7f8e7d44-c382-460b-a70b-06cd121c9f4c">1.6.1</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_8b146c6d-5ba6-4ea4-bc6e-4084cb729390">&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_5e8d0b63-f068-4194-b913-586aa99789db">Concept Name Code Sequence</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_25c67e50-5b9e-4606-813e-6c2733dbe47e">(0040,a043)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_0d07cf72-ff67-4681-8f54-46b2cb857faf">SQ</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_b03e1f86-88cb-432d-9547-4ae5ac3f0750">ffffffff</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_e98d9ae9-3ad2-491b-b505-571b21e8d291"/>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_2796b4ff-94ca-4b6a-9ea7-ecce34e433f4">%item</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_323dd11b-a78a-4fcf-92e5-00a8fceec043">1.6.1</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_e1a380be-8e1d-4543-b0b9-76c6d46c68af">&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_b4cc468c-cd50-4016-97c9-de6e7a4d1f4a">Code Value</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_dd2eec45-30b3-47c9-9469-c1e71b3bc0ff">(0008,0100)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_0715e9d9-821e-4970-bffa-fa9ffb566124">SH</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_5d428ec6-2710-42dd-bedf-986f85cf23f0">0006</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_c8bb1293-b414-4523-ae9f-b0a25cea3f01">125007</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_b5ef6d1a-c65f-41d2-8c92-5c1d3c0e2edf">1.6.1</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_d5b291d1-0118-4c66-b02c-408ee2ba0156">&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_5f52de3e-82b0-41bf-b6af-888286abc72f">Coding Scheme Designator</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_4106a3f3-a985-4ef9-aa42-528841f9389a">(0008,0102)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_2b0cba08-eb08-4f14-a7f3-ae5883000d45">SH</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_2db0a5ac-0222-40f8-9698-9fc2ed161238">0004</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_aea99c07-7307-448d-9b27-2eeb150fa0a0">DCM</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_4617057a-7b0c-4c89-87f2-cdbb65c50853">1.6.1</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_3e0146fc-7faf-4d2c-a442-461b8968448f">&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_dd7e369b-eac3-43e2-b36a-86842522c38d">Code Meaning</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_8e7b03a4-31ea-4efb-9d19-2ea9238ab327">(0008,0104)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_59a5270d-7af4-4c80-a905-ac73afa5eeef">LO</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_d0885908-d066-4e25-9bbe-07178471971e">0012</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_3b6bcbc7-28cf-4303-9578-05eb9067259c">Measurement Group</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_965a6ea9-e2fe-476f-b588-85c0a2c72be3">%enditem</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_5d67787f-3dfd-4c30-b07d-85365781721f">%endseq</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_409bd792-731f-40d9-8817-6f80abdaf60a">1.6.1</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_f18b1078-f20f-433d-bf67-92eb795cfab5">&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_1e69302b-ec8b-48f0-b7f7-bcf37eec253d">Continuity Of Content</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_1cd09b6b-f720-435d-9525-c2356b4c1216">(0040,a050)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_92f135af-e8dd-48f4-80f1-0049e1e08359">CS</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_37c38db0-705d-4fed-809b-28d55f1d5249">0008</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_db3e4ace-28a1-4bf6-9039-d2d2c4339337">SEPARATE</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_7b09fa7f-2fb2-4907-a607-fd7206926ec3">1.6.1</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_e6b546b5-58d8-44ce-8649-81a49de7b279">&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_66004c77-a089-49d4-819c-afa09cbb4c1c">Observation UID</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_ed36d1eb-728a-4c11-84c9-796f845f708f">(0040,a171)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_4ef1a552-aae2-4932-b130-3530e1ca10d3">UI</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_7a54e8b0-4aef-4c9f-8833-7565b4a7c6cb">002c</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_b4f5eb29-3ab8-4ce0-a865-47c4a9328cad">
            2.25.56002466128627498886935079903172938041</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_fdfceddc-42ef-4689-921e-39bd4481aed2">1.6.1</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_ae275e9f-f79a-41aa-82cf-effaa32bb126">&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_e6919f09-73e0-4a88-8e19-b2ac5f9051ed">Content Sequence</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_78a20ff0-2a24-4800-b0c7-ab5429380dc6">(0040,a730)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_19ffa2cb-9417-43a2-9f58-a0d068d01d47">SQ</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_45eaed07-7bed-4700-b870-3539530dfe9c">ffffffff</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_9166107d-5924-40b4-b589-3d4ed8f96ad2"/>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_d09de6e2-3014-4726-b648-d94fbd7cc594">%item</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_b1f416ff-7557-4ad3-9b7e-153ca987b585">1.6.1.1</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_e9a28ad3-6dd0-44ae-91ea-1603bf4bd446">&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_cef884b7-d0b5-4e3b-8a7b-f96e2da5a6e6">Relationship Type</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_37619ae9-d011-413e-a0c6-58dc34333f0a">(0040,a010)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_e2f2421a-ae72-4263-956b-d86ab29ff903">CS</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_38753d2c-ff17-49be-9168-d345e8c29b9e">0010</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_8825115d-8622-4bf8-8dba-3b3152c55f1f">HAS OBS CONTEXT</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_f13535e3-bddd-40fa-b96a-5428476d26d2">1.6.1.1</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_7ed0184f-0906-4866-974f-8130335fce6d">&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_c7dff675-16f5-4952-b8b7-3a0d05c41e91">Value Type</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_f45dd817-0a41-4b15-a4e5-dd61b5f6e996">(0040,a040)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_6df07562-be15-4dd1-af6b-07ba4ca2626a">CS</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_244492bf-3bda-4312-8d99-2bf80f7a1f08">0004</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_c35d4c7a-1747-496e-91b6-3a83362c4d36">TEXT</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_d69eb840-bbc0-4ace-9ce3-48a20bceb9cd">1.6.1.1</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_9456a148-738d-4099-b7fc-3bc48e381ff5">&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_61ce8d26-ef7a-45d1-b187-f240ff24ed93">Concept Name Code Sequence</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_de903cf2-3371-4945-abba-97941f7cf7e7">(0040,a043)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_0c6c9499-b29d-4ebb-9d50-96d5f320a4b2">SQ</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_a90986a1-e103-4d40-8817-d932ca3f942e">ffffffff</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_2dc4dfae-0656-478f-9b66-22c24d147a7f"/>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_8fcb1394-f541-4ea7-8f8a-573600ed8faf">%item</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_1055e161-de2a-4a91-a3c9-e7d8f987983a">1.6.1.1</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_3e13bc78-d711-4722-aa27-a0c2b3fe8807">&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_b9fca7cf-eb6c-4b6b-9303-63746d3737c8">Code Value</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_0ecfe154-320c-40d4-afda-300f975e1610">(0008,0100)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_087cc964-8864-476e-be43-545bf1aaba32">SH</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_723e33f4-a58e-4480-854d-7886e6c9366f">0006</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_5780c7fe-b8c1-4d81-812d-221c6ffb1485">112039</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_0090c29a-03ee-4b31-9257-9792f3ef1ead">1.6.1.1</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_319be73a-6e0a-4775-8c05-8fb0da9d30c6">&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_a678c5be-448a-4776-95a4-0f7757e91f66">Coding Scheme Designator</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_619a62f0-bd87-4163-be6c-02330447e2a7">(0008,0102)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_956b9132-b5f9-4860-9efd-7eccf0f734e7">SH</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_ca4ade54-28bb-46da-89d7-464e0094dee0">0004</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_84bbed04-c82b-4b1d-b383-0aaee64f9528">DCM</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_f11d454d-0320-4b2d-abae-bf993eac6a21">1.6.1.1</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_3dd7a1f0-dbe2-4bcc-8d8e-5da42182faa8">&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_0c5aac67-6858-4bd2-9ec3-d2e3d91b6e77">Code Meaning</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_5e0703e5-cf8d-4cac-8881-e4ef3842b44b">(0008,0104)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_c442b06b-65c2-4843-b565-a6d2c74d9523">LO</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_12fe666f-727c-47b1-90be-04c97d7d660b">0014</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_e6a5f1b6-aeb8-4cc9-b7f4-716bcce9ec4b">Tracking Identifier</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_504dfe1a-c27e-4124-896a-a4de5ede75ce">%enditem</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_4ee142c3-1f2a-4072-a61e-02aeb0c001d2">%endseq</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_7ee440e3-cc28-4e89-b6fc-c55d7ec5a667">1.6.1.1</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_3d8a64ea-0321-4337-b38d-208923732491">&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_3889628b-cb76-446c-bfa0-2ddedb4d8514">Text Value</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_3cc4fd0a-dc37-4121-b5c4-370ee182e3d7">(0040,a160)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_9d92a015-7840-43a4-9971-7293636bbddc">UT</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_e13aceb8-5f47-439c-8020-aec1f90e5143">0008</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_82200568-5de2-40ca-91a7-a592e41cf67f">Lesion1</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_adb60ac1-3e52-4e73-896b-34123dc680ad">%enditem</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_6521575d-527d-4e6c-a0af-872f7cfca1a4">%item</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_7f808b4a-33f8-4802-8d16-7d9ed0aae625">1.6.1.2</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_22087f31-0aa5-4ed8-b7f6-3b02d76c4eca">&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_e3d38d5d-4d28-4085-8f91-e263f8bc0c37">Relationship Type</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_98122172-5e88-449f-bad5-a405284e8191">(0040,a010)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_2651d5af-758f-4c72-a183-b126002f32f0">CS</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_7bef8255-7cc6-466e-ab0b-95e57783563f">0010</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_0255bca5-4b77-4431-a843-49e52fb6f744">HAS OBS CONTEXT</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_ac1a6ec7-9412-4d86-b9d0-6c0662d1ed10">1.6.1.2</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_65ab5c8d-faa1-4e76-8e6e-b3d8578fc28d">&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_e44a06a5-35a0-449c-bfb6-a0cec45c1b50">Value Type</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_6d449a92-f25d-40e3-8676-6b797904315d">(0040,a040)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_ee634faf-b9bd-4cda-b9ba-eefc397f5e3a">CS</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_a7afcff6-f20e-45df-9911-fe9ae5715976">0006</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_94be3937-1021-45c7-ad09-976d8b813ccd">UIDREF</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_ea996fc1-654a-4428-9818-8385aa80fee3">1.6.1.2</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_74db584b-3aa8-4bfd-9cd4-f7f058979143">&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_730af3c4-687b-4c3a-b046-c567db8b359f">Concept Name Code Sequence</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_7d0a33e4-697f-4d01-8870-8a68037bb525">(0040,a043)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_6fe3099b-4dfb-410c-b4c2-b6cd7be90b15">SQ</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_6db83b37-e08e-4a50-b4e7-00edbf40963a">ffffffff</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_19922abb-2dc1-4bff-a825-e27c1651d713"/>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_c3b66b90-0ee8-4d89-b42c-70fb38435906">%item</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_0eaf286a-6914-4450-a96f-cd2733e53aad">1.6.1.2</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_dffab26a-0d9b-4866-93a0-3a1e60caebbd">&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_576336a3-dc89-4aa9-9924-8f1a528d5552">Code Value</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_feb64ed6-b862-4576-ae20-681621908936">(0008,0100)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_ffea925e-390a-41b7-99ae-fcb4b75bcb75">SH</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_ef69bbed-e56b-4fef-bfb7-227ca398550d">0006</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_7fc88330-6c30-4e75-99f3-926c1b33d077">112040</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_b81d2094-8b24-4da5-bc35-e2fd66a95397">1.6.1.2</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_2168e36c-c013-448a-9baf-e9af1f49e19f">&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_5f75d096-373c-4b53-ba63-acbe77c7f19d">Coding Scheme Designator</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_4276eb77-43d8-498f-808f-6917a89725fd">(0008,0102)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_6c2ff8ac-cf3b-481b-9039-5dbd6e4ed459">SH</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_f5cc489f-a6b1-45b2-84db-df396b1148bd">0004</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_b57ae540-9331-4ab5-81b5-cf784e02bd37">DCM</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_2a83d41a-93bc-408a-b8c6-fffc27b383e3">1.6.1.2</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_f46fd113-010f-491f-a271-0351897bcff6">&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_a1e00b0f-10a8-42b6-a0f9-93a0112c7829">Code Meaning</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_58810924-dcdf-42bf-b373-e06630e90a7e">(0008,0104)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_60b901ba-adf9-48be-8c28-95a66503d00d">LO</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_94ece67f-24a9-45bc-8073-d9d2efa1d4c1">001a</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_64075085-eb6e-4edf-9d2f-cf2ea6d2a2a0">Tracking Unique Identifier</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_9eac6e07-6a1a-4870-b58d-ea4e6a0d239c">%enditem</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_4f547226-addb-4898-897a-d6ab27ca87b7">%endseq</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_2d317936-199c-425c-90c9-431081190275">1.6.1.2</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_0eb5aeaa-82ab-4865-be73-b4916b84cca6">&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_b4d64766-84c2-49cd-b231-1f0ee869b39a">UID</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_df843ce1-4459-4a3c-93e1-7dc3302e202a">(0040,a124)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_559e8726-c115-4f3a-ae40-458ea4bf0fbf">UI</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_2d40543e-5edc-4ae8-9f71-458e9e2bd1d6">002c</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_9310d72f-7f51-4e5f-a755-4cc2bbd7b1ba">
            2.25.165294254063588909770717555738008800301</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_f4d1a9d7-040a-45db-b69a-b3065e6cd61a">%enditem</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_77bacb53-07a7-465f-a60b-54993e929ced">%item</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_348e132e-f41e-41d8-84e2-f270eb3af0cc">1.6.1.3</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_10127b23-ef82-487e-bda1-8533e171761e">&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_838530db-14fa-4cb2-b65d-ef14250665ec">Relationship Type</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_fc82cc17-7deb-4f37-98a7-60bc412bff27">(0040,a010)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_4ae2fc47-a629-4314-86d1-dbecd6882927">CS</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_effd84d4-ef5c-4ba4-ba43-59febbc08e5f">0008</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_e4ef144d-d32f-43a9-b7c4-68394ddb8c92">CONTAINS</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_4d0db742-e881-4740-b60f-0d13f5e44e0a">1.6.1.3</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_89bf94f8-1b31-41db-82d1-025e2bf31bf0">&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_28f93f4d-2fbc-4929-b258-58e3282640ae">Value Type</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_f4b3cf47-8e03-4006-ae7b-652a8bfacfdb">(0040,a040)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_2cb2555d-6aa5-48d9-8656-18217eb41801">CS</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_ae1cc81d-dd4a-4a8f-94ab-7a37c94b1f58">0004</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_a79ffce7-6059-47d5-acd9-5725c73faff9">CODE</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_2947d38a-ce18-48c1-b6ec-f400d08e6779">1.6.1.3</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_ea1ee5f0-dfda-466d-83a6-b95ef4ff1088">&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_1b4c2a9f-02ff-4020-b0e0-058faa121a82">Concept Name Code Sequence</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_0b169ee6-410e-4891-bc69-c55e67a2b81c">(0040,a043)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_106a6a50-d77b-4f18-b482-9bb1ad8d5a02">SQ</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_d8ad895a-3e6f-4d2c-b37e-630a57e42666">ffffffff</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_fddef0e7-7cec-4ce4-9cfb-268a74466b8e"/>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_6cb33b2a-dca3-49ef-9d33-108935f3cb1a">%item</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_e83103d9-3058-404e-b822-90b3319e078e">1.6.1.3</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_75f466a7-0df9-4052-886f-3514c34b575e">&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_2c94454a-725f-413b-9486-fae62780562d">Code Value</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_35bc85a8-a359-4161-856f-4f81df39d2b5">(0008,0100)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_c550d0c5-a6a5-4c65-aa86-c11d7303066e">SH</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_bbd4c1b8-abdb-4e9b-b0ae-d89fee9aee09">0006</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_5867fc5d-6d21-4c29-acd3-3734984f51aa">121071</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_860b52c4-731d-4f04-a49d-bac977c89c7e">1.6.1.3</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_68d9e500-40ad-4f8e-a3c7-9a2a62ed16fc">&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_c6cde463-5ff3-410f-aeaa-d26c2a009f76">Coding Scheme Designator</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_8e65cd43-04ea-4e69-8d4e-cee5824a8d7a">(0008,0102)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_05aef364-dbaa-4dd0-81b5-f1739e4d1c47">SH</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_e356d070-3cf8-444d-94b7-65f130381873">0004</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_2d5103a8-1230-4655-a2e8-a3c23a457416">DCM</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_52659b7a-a853-4741-b727-6909add20249">1.6.1.3</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_8346a010-c7a0-48ea-b2e4-595dc830cf3d">&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_387f887c-6c71-4679-9090-99ba7a91822f">Code Meaning</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_e493052b-faf4-40b5-8026-3a6c78b971aa">(0008,0104)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_5b7c5309-94ab-4452-b348-dccbbb41c06d">LO</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_fe0d13bd-71c6-4e4e-8112-dda8f6b622b6">0008</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_41d99ada-1735-426b-bc3e-ff6d163cc46b">Finding</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_f433a4ed-723f-4613-bd37-6631d737f19c">%enditem</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_d0ff87d4-5a1a-4516-96dd-d602277c15f1">%endseq</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_9981550c-7ba8-43e2-8cbf-0dce750b6196">1.6.1.3</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_707624ed-7872-4664-b387-17402feed598">&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_f0aec7c5-1f46-4b51-9e7b-416f1ca9c0f3">Concept Code Sequence</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_4089fdf7-bb31-48d9-aecc-13052d4a5699">(0040,a168)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_e0c741d9-2aa0-4de8-bbd4-9278c80cbeef">SQ</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_56fd94e4-9774-4be8-995b-2ff8af83fb26">ffffffff</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_29aa10d4-12c2-45ca-987b-f6a7766ac5ee"/>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_0c5ca9b9-39b0-482c-9d08-00259e763844">%item</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_9d33929c-4f66-493b-b406-b16883484b35">1.6.1.3</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_6f76d85d-f267-4570-89a9-85121fce653a">&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_b82d05b7-8194-4635-9d67-f364f5ebed92">Code Value</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_7ec1c3bb-e687-4e87-941f-35cc2c41e719">(0008,0100)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_53355611-49a6-4046-93ee-ed9259245abf">SH</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_482bd008-75e6-451e-9435-930b4e715d24">0008</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_5acc40eb-2be7-4f89-84a3-28ab2faffd99">52988006</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_712805eb-0861-42b6-a513-2e867ae7388f">1.6.1.3</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_5ec9c5f9-467f-43ba-a66c-e42edfa4a279">&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_ec1fd8c2-f789-4b23-aa10-3b9622a3ac31">Coding Scheme Designator</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_7adc581f-15eb-4d4d-8685-9f34894166fa">(0008,0102)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_8c506625-afab-475c-9231-51398a15f672">SH</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_6c8653dd-5d6f-4ecc-9648-e444c223ad60">0004</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_8e96ba47-f0a3-4403-8304-1670c27f1bbc">SCT</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_ca12c2b7-bf95-4884-b56e-5ef35ded40cc">1.6.1.3</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_9c4fe676-a4a5-4298-92a3-7bd2793b549a">&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_97ceeb36-aaa4-4273-9e5a-10f99ac5943c">Code Meaning</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_5aaf8f2c-6cd5-469d-9fed-71f7705eb96b">(0008,0104)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_16f7895e-f514-465c-85dc-032fb7485bb7">LO</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_afca4e8d-eee3-4a01-b563-16a6e4aa117e">0006</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_9f6962fb-1d5c-4a52-9c28-e50dea970d1b">Lesion</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_1a6bc348-c5a8-433f-9290-6640e041ef65">%enditem</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_05fe2256-1429-4b75-a04e-775ef786e187">%endseq</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_249e0900-7342-48b7-ba45-2b8c93456155">%enditem</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_461d81d4-f15a-49bd-816a-6a8695a98582">%item</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_c1d5f481-3a45-41bc-a721-0ec2fa6e8f40">1.6.1.4</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_230ea986-01ef-4d31-89f2-23b7738f0887">&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_71322f3c-1a78-49c3-ad1d-5b87f4866f67">Referenced SOP Sequence</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_d3fc8ff7-7f4d-40b6-a8cf-723e8c63320b">(0008,1199)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_9d983829-2102-4724-bc6a-594bcde4ae2a">SQ</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_024379e4-e957-46f7-bd96-ee5db0938dc0">ffffffff</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_e1fa864c-76e2-4dd5-b5ab-805fb2ed1bb5"/>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_719814bb-2b78-4a24-afa9-2e3130913f5b">%item</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_ecb80286-329d-4632-942a-e47141bf3324">1.6.1.4</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_476ae530-778e-4bfe-896e-ae12e5e22ccb">&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_2e3c066f-cef7-4e04-b74f-f58f1e173b1e">Referenced SOP Class UID</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_d51c30a8-bdcc-4deb-a93e-8a9cd53ac0b3">(0008,1150)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_a20fac82-087b-4793-a7af-fd9b9b0ecbcb">UI</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_cc1031b1-690e-4b54-9074-e6d512d2e87e">001c</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_890fe7c3-d9d0-43da-84db-522831925db5">1.2.840.10008.5.1.4.1.1.66.4</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_0628713f-fbef-4b89-81c6-96bcd1262c28">1.6.1.4</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_56affbab-318c-47da-af92-5007f95a41a3">&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_10dd49f1-40a2-4c9a-b06d-4fe5eca121d4">Referenced SOP Instance UID</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_6f0dfd55-29bd-4a76-9687-3d20e492e669">(0008,1155)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_deec2e82-93a0-4c94-9f31-7d956b12535d">UI</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_f2a03e5b-3c16-47eb-aca6-b53cd0be04b9">002c</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_695c0492-ef32-4be0-a4cb-4eb1bbdbbd3a">
            2.25.134884066033959077306435705240550195701</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_2f0411c3-3c6c-41fa-95f7-047721b7a78d">1.6.1.4</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_2171c6ce-ba4b-46f2-96d6-27ce6c2f9978">&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_ded14508-ad0a-460b-a2a7-0e1b2f5142c1">Referenced Segment Number</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_53923790-51ad-43c2-8ce9-cd56927fdcce">(0062,000b)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_36d2ad7c-a276-4537-a9db-dd017aa99c27">US</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_232383ec-90f6-45a1-9f3d-14e983cf3cd8">0002</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_4275b4ee-8a32-4b31-8cdc-60a555d7ae1e">0x0001</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_36c7b1f0-1486-41b1-9648-ca54d4c0ff32">%enditem</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_157f3d96-c3f3-47cf-958c-af43de0cc703">%endseq</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_f82dd433-1342-446f-819e-d1774c6a56c9">1.6.1.4</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_bc45f4fa-1f95-45ac-aa64-945460de8d7c">&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_ec7cfd64-5eef-4308-bf74-7e821b789572">Relationship Type</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_1a4cfd8c-d405-4e93-a332-39cf2a3747eb">(0040,a010)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_3f36953e-3534-4240-b434-eedefaec3570">CS</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_12e731a9-76a9-4ea4-9c18-494afb2107e4">0008</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_f198d40d-03e7-45bf-8ecc-aab173e8a073">CONTAINS</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_cb54f916-1cc4-431c-85f6-c9ddd59ab6ce">1.6.1.4</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_538f8426-a654-4cb7-85eb-43fd28bf6908">&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_672122d9-1865-4fff-b214-6edcf6c84309">Value Type</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_14159f63-a11b-4070-9dd3-7264071f951f">(0040,a040)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_8dc04dc2-93ab-475b-84a3-296ea7b983cd">CS</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_8981ee4b-0d46-438a-9658-354032d12cca">0006</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_4567385c-2213-4dc9-92e2-876be0257de0">IMAGE</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_39a7978f-d10a-4122-beb2-14c0112c947b">1.6.1.4</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_d0b4cef9-e72c-42e1-bebb-39e618d0b8f4">&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_239738a1-4050-403e-bfcf-7d3ea4487761">Concept Name Code Sequence</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_6832b325-d560-47d8-8984-050b9f403fb5">(0040,a043)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_997ed393-3958-46fb-9080-ea0b718f2dc1">SQ</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_d050d53d-967f-4160-8788-093d96efa23e">ffffffff</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_c94e8124-b1b5-4cdb-b08b-981c12c0dc33"/>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_55c8db98-39ad-410d-8d66-7a29fba9f096">%item</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_d710ef13-0ecc-430e-8405-b245c4e1567c">1.6.1.4</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_c8b97fab-3420-4350-9731-22e8e2edffd8">&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_b5a1cde6-2d7b-4ecd-99f4-46b2bcf9e5b0">Code Value</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_d841c01c-4da4-44e9-a356-df9bd7c57e3e">(0008,0100)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_e268c0ad-3809-4e03-9093-4712e811f6f4">SH</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_0d39d8ff-e3c9-4e78-9ad9-133dbb1d6bc3">0006</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_30a31707-77de-4d88-9fd7-e002352f813b">121191</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_60d860ed-3bb6-4a32-897c-9cd53796d364">1.6.1.4</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_d9dd6899-48b0-491d-9018-5c14bb95149d">&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_599902e7-7eec-4e99-956a-5db59d31b918">Coding Scheme Designator</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_d592116b-1bc4-422c-91f2-560d23c19dd4">(0008,0102)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_a02a6f7d-610f-47da-ab1d-f1fdbf64923f">SH</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_eae5d645-82e2-475d-b793-54c188c67ebf">0004</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_75cbf965-78af-42da-ad36-925e48995d03">DCM</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_6a5d58d1-ad5f-4548-8081-7c54271b7ff1">1.6.1.4</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_95ba06db-8cee-4984-b873-eee054587aeb">&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_f6fe29c6-22ca-4bb0-990b-e745e0c6c614">Code Meaning</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_4d677c1f-bd2e-4e02-88ec-041b33dc7da9">(0008,0104)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_920d401c-7ef3-444c-a21e-afa54cbcb261">LO</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_0f046e7a-cc47-4b27-9870-6ce87aec08b0">0012</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_1d915bc8-d5c9-48b3-8c28-3b9010fb1ce9">Referenced Segment</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_80ffa596-142a-410f-9c6c-00a718de4c9f">%enditem</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_836402a3-5014-476c-85d9-2a3b750c444e">%endseq</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_a0b1d77c-87b8-4e2a-9199-af0d03dae7a8">1.6.1.4</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_fe185f87-3d1e-4165-954d-76700ce6dcce">&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_c79f400f-5811-4561-a5d4-fa93cfb88f95">Observation UID</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_b132addd-3560-4d87-8e5b-2986a14b5817">(0040,a171)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_47988198-8155-456b-a597-fa7689a08423">UI</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_8793051e-dd04-405e-94e9-8da093881144">002c</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_17d5fb2f-b76f-4b1a-8872-6664a488f7f1">
            2.25.318310842062810077214341266367812728264</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_74ff9a06-60e9-4e5e-83a6-19e0fa4b285a">%enditem</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_f69394f5-e645-4e3b-b552-ec37982942b5">%item</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_84cb8c97-f944-4aa7-8fee-fc13845f6d9f">1.6.1.5</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_be149266-cb33-4e1f-ac35-45884c11e730">&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_a4182758-79d9-4824-8765-5da21b98293f">Referenced SOP Sequence</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_8e6d3408-184d-44d6-bfff-b829d7aed01e">(0008,1199)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_7d623476-cd3e-489a-b923-a4d7b789b820">SQ</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_ee0265e2-df4c-4d10-99b6-3c53f1e340b4">ffffffff</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_6a31d3e0-5a61-43e2-9306-cb3dcab60d36"/>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_55150df1-31e6-45e7-907f-e16fc4d14c7f">%item</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_ec8fb64e-16e0-45d6-be90-b26e96fd1248">1.6.1.5</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_26a53992-1630-4885-923a-d67b499bc2fe">&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_cfd8431b-f05f-4949-8900-f22d3d96e4bd">Referenced SOP Class UID</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_e3d391b9-eb3a-4c55-98b9-4a2c851041f9">(0008,1150)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_6026b321-44ad-428b-a978-7f841f1ffdd1">UI</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_01c3c509-d826-485d-b58a-b8b4e84b756e">001c</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_6cdedb95-d70f-46d0-95d8-6b8ee11968d9">1.2.840.10008.5.1.4.1.1.128</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_f9969062-7763-469a-ab44-5baa81dbe3fe">1.6.1.5</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_8c9bd11e-bb97-46b1-a0a9-3ad038a6e0af">&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_e2cec4ba-ff99-418e-8612-7642a61be21f">Referenced SOP Instance UID</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_d91a6d4a-6342-4673-877f-7b8e82dbaed5">(0008,1155)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_56208936-d0b0-4cf5-bde9-59056400d4f9">UI</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_7689b314-5494-473f-98d0-a708d46368ed">002c</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_ce711813-082b-494c-b425-7e326baa3439">
            2.25.319214308104243787945491694789635628411</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_79f5d0a8-11a9-4bcb-9a97-7de9b0c93a26">%enditem</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_5613423f-669d-479b-ba41-4bc836e70a17">%endseq</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_782112f7-d376-404e-90cf-433cf81f7edb">1.6.1.5</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_cac9e9ba-af2b-489e-aa47-b0a83ffb8f43">&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_7e29b74e-4395-44d1-8f40-e5531e36fad6">Relationship Type</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_cfa7bfc7-ad77-4e50-95ea-be5a3984633e">(0040,a010)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_843d3ff0-b8a0-453d-a685-cd55564f58cc">CS</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_80f504a0-eb3c-4b47-a246-5e6f28f2bc00">0008</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_6e45a4a3-3275-4dcf-bce7-b8c862f5860c">CONTAINS</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_1875ca0d-22d0-4c3c-a59d-a6426b5569a7">1.6.1.5</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_98c175c5-74a8-4256-ac23-f487d4468bcd">&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_ace8a11a-4ad7-48ca-8417-ed25839ce362">Value Type</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_23a63b9f-f568-4dea-93cf-37bdbfddc56a">(0040,a040)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_23a0c271-2c9a-41e8-b511-b685a6b5a48b">CS</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_2735c8e9-efca-4e4f-a201-3da1a1c7f9e9">0006</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_7d70fbb6-bacd-4a0e-91eb-53afca76eee4">IMAGE</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_fd2ad195-7b76-41af-8961-840b3500cc36">1.6.1.5</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_157b0f98-2f49-412d-acbf-6ad2a07ac0ef">&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_daab4a3f-b5b2-49d2-a465-aeacc8eede1e">Concept Name Code Sequence</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_64bbded3-5b13-4998-a07f-2b8ca5426870">(0040,a043)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_b2cb4729-af22-4a21-8203-cbf8ac3d6110">SQ</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_7b349f02-867a-4e15-9779-f3de1d0fdd50">ffffffff</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_a4897b90-3930-455d-8d81-4a9d125caf52"/>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_c1669ec7-fff2-4ac6-9329-32d955fc67f3">%item</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_a7afc587-1209-45b5-ad0e-a694fc52fd22">1.6.1.5</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_f6ec2db3-5034-4796-91ad-c0b20e7e9dcd">&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_49a32679-4e74-4530-89cc-912337295924">Code Value</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_66a1c3f1-97df-497f-9593-221b73c9b8ae">(0008,0100)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_eb57967e-f667-4949-a599-dfd28af68cc5">SH</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_a3ceb1bd-5637-42cd-b2fd-cb7d9487658e">0006</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_75d11be1-840d-4f76-94d7-f1bcb1afd259">121233</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_c6afcb41-e87c-4066-92d0-63d86367f636">1.6.1.5</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_4ce7a74b-2ac4-4a3d-b819-d5ad4014d675">&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_bf957f24-e92a-4fd2-82ad-0351331801ae">Coding Scheme Designator</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_b288c7db-6f5b-4aa7-9e3b-dcba844f830f">(0008,0102)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_5c7fc2a0-62ac-4c66-87f2-b8f03ae9df53">SH</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_b5cecb8c-9677-41bd-8e71-6f2bd292c638">0004</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_1dc95e4f-72c9-4beb-8d97-2e50e98fc040">DCM</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_23ae3343-2a8e-4f8a-b167-9a9e94294a46">1.6.1.5</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_abea5148-0e7c-45b0-8485-4c7932339f56">&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_4f3c089b-3f33-436a-9f6e-529ac164016e">Code Meaning</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_cb311c75-7fbb-4fcc-b82f-1df1536323e5">(0008,0104)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_1732bfcf-a831-4664-9076-df143313d077">LO</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_da64015d-609e-43d8-860b-2fde52e46a12">001e</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_5ccf9867-3532-4d59-9b81-3c48fdb91d4d">Source image for segmentation</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_4b5c3b41-f77c-40ef-82fc-b8e2d1bdc5bc">%enditem</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_deb83757-cc3f-4d72-9d4f-4d2e9b401b76">%endseq</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_56d091b0-2c33-4c2a-8e63-5e6cddab0756">%enditem</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_a06b05e7-2fbb-4d4b-92fc-ef8815268ced">%item</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_922af6f7-95be-4692-bf4f-3ceeb667e8ea">1.6.1.6</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_e910cd8d-3638-4764-b233-c675827cd35f">&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_fac2fe70-c3b6-4645-a498-93a112d8b424">Relationship Type</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_e72155f6-2e41-4b10-a184-f473a0793880">(0040,a010)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_6760cc10-78bc-4de8-baa6-580f75a2083e">CS</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_5e0eaed8-5c79-4b36-8712-2d7919727929">0008</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_e6ba45aa-d15b-48f3-a124-fd96d0d715a5">CONTAINS</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_34326790-dbe6-4180-b82d-2d3e146d2a5f">1.6.1.6</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_9de5e302-c62d-4b20-887c-8abdb8be27ed">&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_c406875a-bc75-45ab-8a04-9c1309f26d8a">Value Type</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_37a8c484-30d3-4360-ba84-2f2e5de9f23b">(0040,a040)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_3f51f5e4-db22-4a7b-a28f-108f9a07456f">CS</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_c7653af4-f881-494a-b03c-13646f9bfb66">0004</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_5e1590f4-e736-4797-b16d-cc00c0194b8e">NUM</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_da49f12d-9a34-4d81-b897-3464d53012cd">1.6.1.6</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_947600ac-bd3b-491a-898d-50d770e7ccd9">&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_d374da03-1b93-4e80-86c2-9a07b28922a3">Concept Name Code Sequence</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_ab6c1a30-1a04-4f92-b90b-0555c42819e2">(0040,a043)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_525cad1c-3ed1-46a4-aa85-f305752591e9">SQ</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_dc107472-b1e2-4cf6-9d9e-695db2415774">ffffffff</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_c4a4845c-3c9b-4c9f-aa3e-42e6410105c4"/>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_7f52d675-b353-4d45-ad33-43efdca74414">%item</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_094d7ccc-bbf6-4761-a410-6da56733b301">1.6.1.6</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_275d210b-c01c-431d-ac00-a0b185857e5f">&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_6ae7f0bf-1619-4f36-889c-aa6972341328">Code Value</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_2c67417e-421e-4ca1-83e0-d3d0c780e496">(0008,0100)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_aa9b29fb-39bf-48af-b2c6-77c5193b4f48">SH</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_3a00cae0-2790-400d-8456-60afe154f7dc">0006</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_19351149-3ea0-4208-b283-8efdfef567c7">126401</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_49b604fb-6309-410e-922a-f143e26c6138">1.6.1.6</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_0717b028-fdf8-4806-9241-9f9c6477e586">&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_7da6c859-06b2-45ce-9d65-9cb3837170ac">Coding Scheme Designator</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_f6ce934c-c8c1-466b-bb24-4f8fd941e940">(0008,0102)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_a0f416c9-4b19-490c-8092-53c2ed264424">SH</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_1dece95c-9551-421c-8095-a8d8ff367ddb">0004</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_6daa55c7-055b-4b18-b4e8-690d7e519e72">DCM</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_9c27de30-d6d4-47be-97ab-15700113d916">1.6.1.6</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_21253f5c-db6e-4c0d-b872-9659ed15fae7">&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_6a6a3491-dde0-4bdf-9ba0-e14be4c86f63">Code Meaning</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_024f614f-b796-4c2f-b091-a20ea1877f11">(0008,0104)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_4cd85f4f-d428-474e-930f-43bfe41a1245">LO</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_174ea449-fc54-4de5-abdb-361a8b725385">0006</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_0d2dbf03-db33-402e-9922-f022f4d79b5f">SUVbw</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_d91ce0d5-e095-48b7-a76c-64fafd67d208">%enditem</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_59e61ce7-45af-4ccc-bc33-073397161ec1">%endseq</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_a21470a8-14bc-430f-8a4c-80ec68bc32d6">1.6.1.6</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_e38c6a03-deef-4f68-a80d-62429bd23baa">&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_6571fe33-fdf0-4a44-bf32-d45b7544622e">Observation UID</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_40e0e6b4-84ff-443c-8e72-3ada3175e25a">(0040,a171)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_e7a3e4f5-7cf3-4160-8bf9-78a6edc9488a">UI</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_57f1182a-2645-409a-9c1d-cef7fdebbbb8">002c</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_760d9fed-0cdf-4a7f-9c7f-31d8acefef28">
            2.25.51420968257530981243824658943871973198</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_7000a726-814f-43e6-a803-c6ba2d8ae1bf">1.6.1.6</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_67b7755d-03b0-4c99-ad94-4880a70666ad">&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_6a11ca30-6237-4b4c-9f0c-fa3b130477bd">Measured Value Sequence</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_299186f4-dd55-48ee-bf10-55b3df71980d">(0040,a300)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_17ea4808-7048-4410-9e57-9e270a69098f">SQ</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_3a9a17cd-3862-4e2e-a1b6-010e5817a023">ffffffff</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_5538a5fd-33e0-4521-90c2-b04467b67d42"/>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_b4f7400a-7f8d-4790-ac70-95a48d892998">%item</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_d31b9755-7df5-48cc-b0c0-7f48b9c66b79">1.6.1.6</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_31dd0be8-be89-4fe6-a370-dc3f478691fa">&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_0c003b3f-2bc1-4d26-8168-05672fe3273c">Measurement Units Code Sequence</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_1c711f72-0363-4002-845c-b897410ec7db">(0040,08ea)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_94e746c7-c91f-48ef-91d6-d91b2c2664e1">SQ</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_dd8c98ea-c45f-4918-ad57-f8347d1fb857">ffffffff</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_0fcc983b-4666-47ac-b878-89149c783ac3"/>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_84694d1d-a31a-480e-bbb5-d249ae650ed2">%item</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_73c3e74b-bd62-42c0-84db-83efe4127366">1.6.1.6</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_273b2049-5d92-4581-8946-5f12773570e7">&gt;&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_47bd427e-a6ff-487e-a46d-4c61dd2f4320">Code Value</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_ea086c3e-8a26-41c8-9e68-5a847ed7f8ac">(0008,0100)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_f1ae3d88-6231-4edb-8f3d-a8a294fd43ba">SH</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_95904fa5-aabb-4702-a946-e9cd449333bf">000c</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_ad30f8a1-65a0-44c1-881d-7c25f5a340d2">g/ml{SUVbw}</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_a65341a8-c20f-4cf8-9065-ec7b154102e6">1.6.1.6</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_c2a2c0c2-5b10-4e9b-b794-dc69228a943a">&gt;&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_d3b4bdb5-4c00-4409-b954-9540ad2478ab">Coding Scheme Designator</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_148ae4e8-acb4-4453-9787-e1c98d3028f8">(0008,0102)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_54a76a50-d3cb-4b73-aaa2-734fd431fa34">SH</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_fb1f6eba-60e7-4375-981a-c7cc079ec53e">0004</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_81adc6b7-6c1c-4ba0-ad75-b6e3d1809be1">UCUM</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_b20035d7-a00d-45d2-94f0-2e105530d1d0">1.6.1.6</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_63a86ddc-a613-4c26-a6f5-8bfe1966f8f4">&gt;&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_5e2dcc8d-29d0-49bb-a609-fb7442bba995">Code Meaning</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_f407d0e8-66f7-4031-b3fe-22812f2d9442">(0008,0104)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_facb3d3c-b692-4511-bbc5-a09d96d97877">LO</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_558e79a4-c2d4-4e59-9a34-797133bd9d90">000c</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_57f58ac4-da59-4c2f-b69d-e1df0a9417e1">g/ml{SUVbw}</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_14a7e1dd-f7d3-45bb-97f3-f5103de784a2">%enditem</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_b71d1f20-1d83-4514-abb0-b2fa462761b6">%endseq</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_e525d126-5afa-40d2-989d-9b056f83aa93">1.6.1.6</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_aa53982b-cda2-4854-8296-5261ffc338b0">&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_d9f16996-c200-47f2-be68-2f6dbc5c3146">Numeric Value</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_0a65d508-6322-47d6-97d9-0248b5b36208">(0040,a30a)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_86139340-3d00-4d47-a70d-82bbfeea2f66">DS</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_b7640931-a46d-4938-bd47-cf17abe1df20">0008</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_67f6714d-1cfb-4a61-b6ad-1ff9cbcfd3f4">1.98024</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_182ff4fe-4c20-4a09-ad65-30c5b16f95de">%enditem</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_36943bf2-d9e9-4083-9984-1a5477c31134">%endseq</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_abe37834-b84d-4395-aecd-70de400a5a6d">1.6.1.6</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_684b8006-9518-40a5-8fec-4658364b287f">&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_e9812088-0a44-457e-a004-bbe7798b7923">Content Sequence</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_d092ff7a-a729-4196-b538-8ed5b5b50ca6">(0040,a730)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_0f8b8a00-5f06-4c91-a8b5-26316cf30944">SQ</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_5c716b5a-a0af-47ed-8412-1f93b4f9da6c">ffffffff</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_4091be64-9aaa-4471-9d2d-b3ff2ff17eb2"/>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_429626e9-a65a-4a70-a612-f577d4e3b004">%item</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_56bf1dd7-e9fb-400a-9536-df36b9fcfaef">1.6.1.6.1</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_07ffeeb7-0fbc-47ec-b08e-fc543748535c">&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_81ca2baa-a16f-4185-85b9-535548054738">Relationship Type</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_ab83a9ca-bbc2-4bea-84d2-cba87180dc41">(0040,a010)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_3a1a74d2-52cb-4867-a81e-5a57af5d248a">CS</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_5b1d0fe1-cfc6-4851-ac73-5e44bd9dca9b">0010</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_9c811dca-e837-4bd5-8f46-5f30893766fd">HAS CONCEPT MOD</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_21e37b7b-6331-4c42-bc02-519f222efe4d">1.6.1.6.1</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_79403907-9cd1-4f86-8959-a67790cb5c37">&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_4132ea0d-c4bf-40aa-bd2f-00a2049e5b5a">Value Type</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_57159184-4d5c-4a6b-bc00-0fe891d22cb8">(0040,a040)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_4bef357c-cec7-4c52-8ff3-55f032c7e598">CS</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_60051687-bb22-4e40-a21d-6a7093d8190e">0004</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_62e3bfc8-6472-4979-bc42-a12bd0b45a84">CODE</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_ec597abe-f989-4f31-9a92-452d9f3ca2bd">1.6.1.6.1</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_a3691cf0-c5b9-4ec5-946e-6956930722d9">&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_20178887-465e-466a-a933-6bf911e32967">Concept Name Code Sequence</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_4aea99cb-db21-4735-8358-f05c98bf7bf8">(0040,a043)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_bc031c83-1a0c-46f3-9e3a-943bb74352d4">SQ</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_f94aa7e5-158f-4576-ab34-1858496f360c">ffffffff</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_78a8ac16-2f96-40df-bafe-c50d477d0976"/>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_e48d33a4-3d18-4444-889c-928e35e3cb60">%item</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_19fbbd35-efaa-4fd7-b4ae-ef301d27107d">1.6.1.6.1</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_5fac9d35-7a9c-4fd6-ab34-7efdb1ebbe0d">&gt;&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_82796730-a5ce-4a85-b176-1cf3f5dfdced">Code Value</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_fc1e5971-6d32-4757-b536-8abf4d0f6a8b">(0008,0100)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_507e4339-aee7-40ae-9df2-9d76b0c508e5">SH</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_92d35461-2c85-4944-90f9-1bd21982fd57">0006</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_6be31491-0227-4c85-876e-136a80156ec5">121401</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_36830e2f-6854-4055-80a7-ee42341ed2bc">1.6.1.6.1</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_ef0e6aa9-51a0-49e6-90a0-9e77216ab274">&gt;&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_192c81eb-9f3b-466f-93aa-a7f9322604f9">Coding Scheme Designator</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_2831046c-6604-4d0d-9424-fb9c8b256175">(0008,0102)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_17f0cace-2db4-485f-919d-0b41235b690a">SH</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_a7d03211-4434-4fbc-92c2-2d0f0be85b55">0004</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_de96eabe-dd97-4d99-bc2c-031d4cb07c89">DCM</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_e2e7df89-ff53-4511-8662-20769e55e5be">1.6.1.6.1</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_0c8d89dd-cbd8-4915-817f-29d536bb4c7f">&gt;&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_0ac80d73-cf79-47d0-9828-ce6cb0889669">Code Meaning</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_8f783058-fced-4e8f-a275-bcc2a6bf0a11">(0008,0104)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_0616405b-80d4-421b-93e4-b8fa456f1918">LO</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_fd7ecd54-127f-4faf-b23d-a963378d6293">000a</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_6760a7f2-9fdf-4634-bd68-e5483ce34523">Derivation</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_88e6b97e-a9fb-4b85-9a91-cd8a20ff0806">%enditem</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_a5f0709c-7f24-4234-8d61-dc6c46b05619">%endseq</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_4419f449-e85d-4ca0-920f-50f78e353bca">1.6.1.6.1</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_f259d9ae-3c0d-4fef-bb9a-e8b78bcf4e99">&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_49937a35-01ba-43e2-8d9a-6f1247bdce19">Concept Code Sequence</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_3b29638c-9de3-476e-9758-4b9b8c1ab2c7">(0040,a168)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_52f54cb1-3076-42d5-9c13-7d33e22b3922">SQ</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_dbf0f414-7ef0-4f4d-a257-56076c9b7f8d">ffffffff</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_916e1eaf-3e42-447b-895e-1ff21b8596e9"/>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_c6d26326-6d1d-4f1c-8ef5-09caad6aeb53">%item</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_24bf525d-89b1-45fb-8eed-d74bd4c176e4">1.6.1.6.1</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_93eff5a5-8c46-4561-9513-e5c030936ea6">&gt;&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_d51b4771-58d8-4736-a77b-b7185f20a5a2">Code Value</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_037cca23-6680-4869-a968-2b79074ab162">(0008,0100)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_26e19bd1-6d21-4a23-95ab-ce869430d2ca">SH</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_8425b430-c57c-4743-8396-79e3c5dab8a7">000a</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_7d3cd608-e518-4df1-8895-e1e411a590cc">255605001</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_d948d0b3-7cb3-4cda-bbcd-26e7289851dd">1.6.1.6.1</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_a494d87c-721f-4762-98f5-6a2761c6e93f">&gt;&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_88d636c6-de39-4140-9fa0-3d771e90fddb">Coding Scheme Designator</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_c5e03dfb-e3a1-4bcc-94e2-abed6f870a55">(0008,0102)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_d99c4a05-cb37-44bb-a3d2-cca71176a348">SH</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_2181c405-7f60-4d58-85f4-493801ab2c83">0004</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_5da026e3-e04c-4e0a-8da0-c6d54430232d">SCT</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_ce4c502c-0f68-4705-9aed-984b0abbacdb">1.6.1.6.1</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_8cd53a10-86df-4557-8d45-b3e4958e740c">&gt;&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_8c6882e9-2fe5-41cc-b259-f4c1be2173a4">Code Meaning</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_e4ffc70b-a62b-4678-b67d-30bfdd0face8">(0008,0104)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_a8951b4c-e0ec-47cc-b58d-5a414932a949">LO</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_7d5c1e21-a2d5-43d9-8d3d-ca33f651ed31">0008</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_733b062c-f659-4a9f-9840-f5c7a49f8bea">Minimum</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_e475a8d0-43f0-4654-b491-4570a001d4a1">%enditem</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_db8297e3-de3f-4c10-b1f6-c37561df1c85">%endseq</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_5bbb99fa-1dd0-4109-8871-d19b1320e124">%enditem</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_90dfa229-8aa2-49ac-b800-edbdf4870f96">%endseq</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_e64d7109-4268-47d2-a407-76258f3e4598">%enditem</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_6184732b-af25-426e-af00-a6e13e6a039c">%item</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_8a93c79c-9a92-41bb-bbfa-eb06c588dee6">1.6.1.7</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_6282fd19-0052-47b9-865d-eeba3bc0f883">&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_86075728-6fa0-4acc-9b90-24548bdf508a">Relationship Type</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_d6a24c2d-06a7-4151-bcdd-2bf0cf3818b7">(0040,a010)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_7eff282c-962d-4e59-b3dd-dde2ef1c0d2d">CS</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_c19e0a5d-51c5-4bc3-95bd-b5a6f7375feb">0008</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_9184ac79-4e73-4534-83ec-54605ec7c168">CONTAINS</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_5ce8c3de-c782-447e-82e6-ba35bd5c58a3">1.6.1.7</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_10e3e5a1-ac90-4b04-a380-eee692f8c5e3">&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_cb0e696d-5d06-4068-8326-5850177c53bb">Value Type</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_a67763be-c6c1-4ad9-9692-c45ba5f77739">(0040,a040)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_4def5bf7-2abd-423d-a307-8dfe81586379">CS</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_a68e1743-cf98-4f27-a200-5ba3c3a2a724">0004</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_0702a57f-e234-4881-963c-4daeb3024d64">NUM</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_19fd3035-1f0c-4a94-8c98-bd244bdbd4e4">1.6.1.7</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_4117aa32-440b-4efc-bcb0-37c87eecda8a">&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_3160c877-dea5-4e8c-8c3c-390a416201f8">Concept Name Code Sequence</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_ddb31221-25dd-4fee-95c9-5c2683945f6e">(0040,a043)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_2dd5aaf6-3783-47dc-8668-95ca79bb200e">SQ</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_47e609c0-7e99-467b-9ac6-c2752687158e">ffffffff</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_3927dbd6-5506-4149-ae93-9e2863936fe1"/>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_77e2bb35-d5b1-4e46-95d6-bfa31afaf6f2">%item</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_09ec4e6a-b4e4-4f8a-b1af-3fc2d270143a">1.6.1.7</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_04fe1181-f211-4ea5-9f23-58335baa33b9">&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_f9de71a3-2f41-4480-9140-c4da4ba6b7f3">Code Value</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_a33b67a0-4f06-4317-bc70-33fe5d693965">(0008,0100)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_7ae71a06-9873-4a9c-9027-496d71e90c88">SH</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_f3242fcc-776c-42f9-b748-5b38c978157b">0006</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_218e8022-40d6-4f99-9c9d-3f2caaf6ec5e">126401</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_535b9d9c-2bf6-4008-aa70-fd9ef8713bff">1.6.1.7</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_00f1f917-e239-4fac-91e4-d7166bd55fec">&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_b8fe2d1d-7cc8-4f51-8e95-74e8e5ef50a0">Coding Scheme Designator</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_d433eb87-dbb1-4303-ba82-1bb87ef002c6">(0008,0102)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_4b1f9f98-fe23-4e31-af65-8d26bfd57344">SH</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_9eb14276-6373-4001-ac22-8924c8b20941">0004</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_b9b97ee3-b843-49c5-a0e1-65369d76f31c">DCM</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_85ccb0ea-29b5-44cf-b2e6-43f7d777cdc4">1.6.1.7</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_5be17db4-431f-4ad4-a561-d92000c3188f">&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_534ac663-f8f5-41ac-881c-0b2139263c78">Code Meaning</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_f1c02ad7-4592-4dfb-8456-a02153c62f1d">(0008,0104)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_8b4b24b1-81cc-4924-acb8-7554851d81d6">LO</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_7be5a792-8b49-4a74-8917-9722f7094da9">0006</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_79d54cf2-433f-44f0-852f-fd83b886c6e7">SUVbw</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_5a14194a-1931-48ca-92b1-ec7e5b64b14d">%enditem</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_4a47eb98-ea24-4bd9-bb23-1f0be6230aaf">%endseq</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_5794a204-22b7-47d6-8a7c-7217df4e7b24">1.6.1.7</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_80ac52fc-e402-42f2-9de5-8fc0633ce6c0">&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_c1c8b6c2-f3ef-495e-88b5-3aeef60af051">Observation UID</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_ffb30845-31ee-4930-bfe2-ba6afdd4febb">(0040,a171)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_b5be067b-2695-458b-87ca-9cea76291bab">UI</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_5b5f02c2-c066-43b5-a00d-2ab2fe0ec5da">002c</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_7ad3c4fe-bed7-4bfa-84da-771a724e165f">
            2.25.205292243885258032428819330909580896146</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_393713d0-4262-44b6-96de-2268b23029ef">1.6.1.7</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_a342c456-5e28-499b-87d4-e39f871749bf">&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_8cc9d748-f787-40c4-b272-2f0e6685ea51">Measured Value Sequence</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_64b958f4-327c-4894-b842-55245dc30270">(0040,a300)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_28e918dd-1d0f-4c67-979f-f5bfa963a465">SQ</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_8a7ab4ab-7801-4035-8d18-ba8ab16f24a9">ffffffff</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_bbab05d2-4870-45fa-aa94-38bb6819e2c9"/>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_e51e7657-3165-451f-bda5-3a051b5c341d">%item</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_73d4db96-2dbb-425c-a5e2-07cb9e519fb1">1.6.1.7</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_1fd10971-1205-4f54-9de0-a8cd0b44b12d">&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_d6528283-b7a0-4f4d-86e1-bf4e10a71559">Measurement Units Code Sequence</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_e054e88e-737c-48c8-ab5c-b436f007f54c">(0040,08ea)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_999c135c-0ff3-45e6-ae32-4f70edb27e72">SQ</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_e5771b37-b04c-48cf-81eb-9cc3eff14f0c">ffffffff</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_295fdfdf-674e-45ef-a48e-6808083a955c"/>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_0928a257-931b-41c4-9610-6ea4d03b500b">%item</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_4b75264b-46f0-4d5d-bf0d-df5ddc87ebc8">1.6.1.7</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_3f41ac0d-602c-4651-8260-b014d6fd31ec">&gt;&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_c326f372-dcbf-4031-8bd3-a0f824188dfe">Code Value</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_f200c16e-5ef1-4d26-8a70-f783ded6ae02">(0008,0100)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_fa97cb41-3c01-4179-b9dc-701d763bb9a4">SH</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_2ed8bd7c-a536-4812-8d95-b1ca523ab492">000c</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_92fcac14-22de-43b6-929b-864b288b0924">g/ml{SUVbw}</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_5c4b12b6-3711-422c-a21d-134bb666ec44">1.6.1.7</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_3c503267-9189-4e27-895a-e868a6ea668b">&gt;&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_b28db403-9545-4e7d-980e-7ce623afa653">Coding Scheme Designator</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_aa9f7929-43a6-4d97-9b84-40d5a42f5941">(0008,0102)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_8d4bfde1-f901-4f52-b22e-baba9fe912a2">SH</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_76133f60-2500-47dd-ac80-406d610d2211">0004</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_2cecb627-b532-4c57-bdc6-a12737d92272">UCUM</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_f16e944c-f547-4cba-8be7-e0cf96a35bb0">1.6.1.7</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_c96e7dfe-6e6c-4f10-b2ad-cf902f2c686e">&gt;&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_55344aa2-234f-4648-8c7a-0ff7a62d2459">Code Meaning</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_74e467df-52c0-429f-ac03-52c86734ae31">(0008,0104)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_a23131f9-16d7-435d-b103-85fbeb631ae9">LO</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_48bce22d-ae47-4d9d-b2b6-ce8d1ebe3310">000c</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_13940a3b-3841-4049-98b6-6fb02bee1578">g/ml{SUVbw}</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_5c68a3a1-f55a-4de2-9055-ae153d761627">%enditem</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_5117e67d-b61c-4c98-a022-7041e8288b8a">%endseq</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_4534256d-75c9-4310-9b9c-ad986385b268">1.6.1.7</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_7dbab596-a6b5-425c-967a-cd271c80bd43">&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_78d6f4f0-3d7c-41fd-91ab-44aab26dc31a">Numeric Value</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_9b408d87-2b3a-467c-a344-f043eb1611b5">(0040,a30a)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_a14b4a71-38c9-438b-8a96-5a88cc4b3870">DS</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_2c3a0d48-8b87-49c8-ba07-b66c3ea10918">0008</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_3923f9d4-3ebe-44f7-bf48-9e7be6366b54">5.68816</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_769b969e-8e74-495b-848c-f4bccde5dcfc">%enditem</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_5b7f17ec-b72a-4e76-8254-c75929166879">%endseq</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_51da1b35-cda1-4cdb-9382-facf54d4e4cb">1.6.1.7</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_64056a7d-1208-4c8c-95cb-c91a6e135505">&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_44c1e264-f1aa-4a88-bbe9-02a344ed2cfd">Content Sequence</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_49974e81-0684-4500-b576-337d3d174822">(0040,a730)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_8a5873aa-96a0-4230-bcb5-aa08fb04be9b">SQ</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_5a33b546-fc89-4fa2-9de1-1c562a441613">ffffffff</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_9dc50d15-3a58-4ff8-a3ee-af50d93ab87f"/>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_341c55d6-fb9b-4b4e-be57-8ae5719b429f">%item</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_26166b06-ede7-45c9-861e-b56184e71d89">1.6.1.7.1</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_5a86fcda-2bbc-490f-b5e5-73d293c4f198">&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_ece1c28b-1948-4004-8b56-f09faaef50a4">Relationship Type</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_e0646d8d-d21a-4df8-b8bd-5481af99aca3">(0040,a010)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_b967fafb-30bf-4c01-b360-8631e30de4ee">CS</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_2c65d022-e7eb-4c93-83f7-00f2057105c3">0010</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_ee88466c-2b5b-420e-b2ae-9dc8cc3f9c1f">HAS CONCEPT MOD</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_3f43c56c-5848-4398-9b27-e28ba895c38c">1.6.1.7.1</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_b222c589-082f-424a-8f65-32f5017aa2c4">&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_04b50aaa-2541-4bfd-b0a8-745344ac9e2f">Value Type</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_8f91cd86-97fb-4d8e-90f8-71a5bd0c4008">(0040,a040)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_e0ef1179-3781-434c-b8a3-c90bf71b2db6">CS</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_62ed1748-5a82-4e9c-a7d0-806671e81d49">0004</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_1b4b3bcd-280e-4b4f-89dc-02228f858e12">CODE</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_8eee3b97-c7c2-4dbd-8371-4b03051ee7b4">1.6.1.7.1</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_c764b9f3-bb02-4086-8d6d-eaf125134b1a">&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_309be5c5-a5dd-49bb-a455-c56f8a327285">Concept Name Code Sequence</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_b0811f87-50a0-4f96-890d-87dfd9118b80">(0040,a043)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_a3ac6014-e627-4ada-ac1d-5eff291fd06c">SQ</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_40c55e31-b62f-4c07-b5ee-aad674704545">ffffffff</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_9301eee6-bbff-4329-8cd7-2824253f7b1a"/>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_903ee2f2-3e03-499f-892b-fdcf737340e4">%item</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_b6d4a099-6f0f-4b7c-a206-0107e97b4830">1.6.1.7.1</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_0fa41a77-5b35-4901-8784-4fcd30476800">&gt;&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_6ebd54db-4455-43e1-bf85-2b9785790677">Code Value</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_d7ba6a2e-6d03-497a-9a57-7fc4487c9484">(0008,0100)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_ae47b524-a1ce-4553-b182-24f1025edfb8">SH</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_c73420dc-504a-4870-9a10-6a79fbacdf36">0006</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_cc1a5f60-120a-4413-a657-03ac1945324e">121401</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_77864095-a0cd-44d6-b020-2c6ca780ed67">1.6.1.7.1</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_4778724b-5e08-44d8-91e3-3144c01052a7">&gt;&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_3464e8f7-bae5-4ce6-8da3-4a9bc4626cae">Coding Scheme Designator</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_76732e28-23da-4cc5-90de-fc93b4d1bde4">(0008,0102)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_8d0b7778-508d-465f-abe5-1773f0360adc">SH</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_8f547a29-05f6-4e87-8977-008697c0544c">0004</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_aa4f4f5b-9bb5-4df2-9b86-0f16db852832">DCM</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_8808851e-74bf-4010-bfd5-4514015d5fcc">1.6.1.7.1</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_727637d2-f898-48f1-b14c-98afd515e4dd">&gt;&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_6cb6dd1b-658b-429e-a163-68b50141fbee">Code Meaning</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_64587bee-0d85-4d69-adaa-b04eec62fee3">(0008,0104)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_091f27b3-523d-4d50-9b89-d42d42682531">LO</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_3823baff-3108-414b-b400-4cd0c0b79b3c">000a</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_f9c103e8-728c-4c5e-8e32-6f5603c5d9d3">Derivation</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_6b5ccb6f-26bd-4507-8e44-9239d9815d75">%enditem</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_879230ba-1d0a-417f-9daa-ea59f3963f4b">%endseq</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_df9c241e-ace1-4b9c-8dd1-13ac80481a08">1.6.1.7.1</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_5cdfe02d-48c9-4cdd-9729-710daf9bd68a">&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_b50033bf-a0dc-46a7-ba34-2b345aa07737">Concept Code Sequence</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_64fae7d4-f908-4f71-8394-a1cc44884f58">(0040,a168)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_6654cdf1-adc9-45b3-a611-86e4477edd8e">SQ</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_b83451ac-598b-42d6-9cd4-96f187aeac40">ffffffff</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_953b737b-eecd-43d0-b56e-57b2a2b9dcb4"/>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_f272ed61-e10f-4659-889a-fe0d9a028971">%item</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_b6c252d7-189c-46c8-a2f6-8b3bf8a8e8bf">1.6.1.7.1</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_30b94de5-6947-400e-a6b1-926b0348afa7">&gt;&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_1c6a2f8c-dc60-4f96-bf62-798d4b2b38a3">Code Value</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_abf7ebca-391f-43d2-bf5b-8b701a14ccbd">(0008,0100)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_57a814aa-6f45-4c28-9f8b-f5da61817698">SH</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_7a88f53a-02d2-44f7-9bb6-0ca8d7d1a705">0008</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_3f516496-3eb2-4d3a-b29e-9410d14f288a">56851009</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_cbacf1e1-ee2f-438e-a4d8-952083293b3c">1.6.1.7.1</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_43505ff9-6fc6-4502-b765-0f1bd05765e9">&gt;&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_5864af8f-b18b-4c81-afb1-a26a869e6445">Coding Scheme Designator</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_f60cad5c-033e-4ff6-9fd9-0e3e53cc370d">(0008,0102)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_6033ad8c-dc08-41ef-8b7f-61a6dddcbcae">SH</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_845249de-ff57-4897-8cd5-eb0896f79d65">0004</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_01d0ac98-3c94-44b2-96a2-70e0bd195132">SCT</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_befa9aee-449a-47dc-8014-d28d8c4746c0">1.6.1.7.1</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_3debaaac-a02d-4da8-9d91-116c992fb84e">&gt;&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_3788ffb2-868c-4c74-8d00-ac9412f80b61">Code Meaning</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_dba8a485-2370-41df-8885-137bb8518950">(0008,0104)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_ee8485a5-3e1e-4a79-bbaa-8ad953d40641">LO</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_09ba2938-c4d3-49d5-9a37-01a4f52a8b59">0008</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_53e29cf7-eec7-46ec-94ed-3554875cab0d">Maximum</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_7f1a2285-412f-4799-904f-18c1b5bcbd9e">%enditem</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_4fa6f437-a791-4f00-973e-b9fb1bfba38d">%endseq</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_8d9144c3-cc44-4069-a6a8-b0458d397827">%enditem</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_a1da930e-adf8-46e0-9da0-fa2e24292d50">%endseq</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_a5e02a18-7706-4504-8ca4-cdcf1e80620e">%enditem</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_557a961f-31ab-4295-bb3a-860a140ac1ec">%item</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_d4812660-07c8-4112-9edd-eb5f2ac51f46">1.6.1.8</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_372448fd-14e0-495a-a43d-ce984565acda">&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_a43e1cd5-116e-4cfd-a88e-32fba6d31519">Relationship Type</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_105e8c6a-922a-4284-83d8-85ecbcaff369">(0040,a010)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_b5fc5e81-aa84-4bc6-ad08-4b0a42f87791">CS</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_783041db-b46d-450b-8273-08624d7c085a">0008</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_59b9daa4-bdae-4f86-b2eb-07ec57865854">CONTAINS</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_46a9337d-bfe3-491f-a654-37f8f549f578">1.6.1.8</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_594d6988-7e61-4a09-9624-02e1ed968c24">&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_0317875f-a656-4313-b6a3-f845e90094f3">Value Type</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_f1686878-34db-456f-84e2-174eccc9fb35">(0040,a040)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_dcde63de-7701-4c8d-a187-7e154706604f">CS</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_3a1f9df9-3b1c-432f-9c8d-7b2610f5043c">0004</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_d933c30d-ada2-4940-8701-149b2bb4158c">NUM</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_62f400a9-720d-4a72-a33b-423cad132427">1.6.1.8</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_8a28a5f5-d576-41ff-b9e0-edcd78363cf1">&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_718e41cd-67a2-4f03-b170-2ae01179d2fe">Concept Name Code Sequence</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_e56abc15-4b49-499d-a5f0-b1b11737c687">(0040,a043)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_6435252e-c174-4c31-9652-69bf680e1c1c">SQ</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_3d32b16b-5d36-48c1-b260-5488ad74d550">ffffffff</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_7c826396-7769-4cbd-8881-cde3affca403"/>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_9013202f-203b-4733-b12a-1236971a0b89">%item</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_8412c6a2-e28e-4f7d-865d-b63d3fcb0af1">1.6.1.8</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_941f07bd-1e07-4b0b-8414-c7945936c90b">&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_ec768228-b73d-413a-b973-c8635bf35cab">Code Value</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_765084bc-d586-4334-a4a8-44ea5577164b">(0008,0100)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_e9f0fbb2-1915-404d-8066-890512d4d03f">SH</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_0dcaf8ae-9f08-4dc6-960e-d8100ee8df79">0006</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_2f43f71d-6ab7-49c1-a567-8362bdc856b3">126401</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_89dea629-b9a8-42b6-89a8-ee689facfad3">1.6.1.8</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_96e5e963-b096-4563-8694-b14356758f70">&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_63bd0ae9-3950-43a4-8364-aa5e511305c4">Coding Scheme Designator</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_59774388-66b1-4714-95a0-310a6f15a437">(0008,0102)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_affd2088-74d4-4573-bfab-5da09ca3c395">SH</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_6cadaab5-bebc-4eb0-bbb5-d4462db4b381">0004</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_483537e3-74b4-4c41-a058-75c66b5f4933">DCM</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_144b14ea-7034-4909-82fd-137dcb73d1a1">1.6.1.8</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_5e03f63c-d60b-44f2-a995-d52d82fce73c">&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_b7bcb9f6-6a99-4152-bf1e-145e70fd48b5">Code Meaning</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_ceca654a-60b5-4d3a-90e5-72628b734dac">(0008,0104)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_dfa7edf3-8710-4447-b5a9-b90a534733d8">LO</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_f368c487-3592-404e-9381-b14009555437">0006</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_85e70cb9-f6cd-4984-afa2-39e8c3b17126">SUVbw</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_5438b0c3-d7ad-4902-8a51-034201802fcf">%enditem</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_94872c97-361e-4e63-b6ce-7436c2e0106d">%endseq</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_bab378cd-ed31-4390-8066-605d91bb55eb">1.6.1.8</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_cbd6a42a-64b0-4c9f-a789-8c99333ea007">&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_6e6baa43-f330-4a31-a8b6-386970eedc86">Observation UID</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_f0ce08c8-29d3-47a9-a825-0e72afa32c86">(0040,a171)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_cf76dd64-2ac4-42a7-be86-854f8fa185c5">UI</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_cf4d3a99-c857-49c1-b261-526bbad3e6c4">002c</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_6543f1d9-6a19-4059-97e6-fbd2cabea17f">
            2.25.70160252080234577167847509948368893276</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_e3880707-851a-40a2-bd00-ff0569bf527a">1.6.1.8</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_44535bf2-46b6-4fc3-a329-d13b3a0cfa23">&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_5f5f0f58-1c4f-4a3e-87fe-191660afdd4f">Measured Value Sequence</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_0e4486cc-6844-4dbe-8237-46561167493a">(0040,a300)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_08983676-a347-4964-9abb-aac7c3ead834">SQ</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_2c945a0c-52cc-47b8-b9bd-48808b6e8615">ffffffff</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_3c563232-9285-4002-bbd3-02042adf0c58"/>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_5de8747e-abf8-486d-9043-e6cc20b73934">%item</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_7921435e-e98d-4fe9-a656-0b3c646111af">1.6.1.8</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_7559ece4-f178-4404-a9f2-812c0f7729d4">&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_93749ddb-14d3-4ee4-a7c8-0e99b833b137">Measurement Units Code Sequence</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_85ec091b-7536-4c84-973d-5f468fa6f4f6">(0040,08ea)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_87f684b7-d0d8-4f51-8f39-4449f0ba9d2e">SQ</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_3766e700-ae88-483b-bc63-2d572f3a8e55">ffffffff</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_cfe10e5d-03ef-4633-bd2c-2549966e3770"/>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_afb729c1-fa11-4b06-81e6-9f1dcc5523b0">%item</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_3486ca1c-70c7-4aa6-99aa-3671fd8c4bd5">1.6.1.8</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_3d1ee1d8-9bc3-445a-9b25-1cf08bfbd24e">&gt;&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_5d252e00-8705-4463-9ae6-e6e072fc36a2">Code Value</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_c2d1222d-da6f-47c9-8956-e025aa134ffa">(0008,0100)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_5ce7c998-7ae1-49c6-a619-d6be96a81abf">SH</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_a7cd6d83-6b66-4c7d-add9-300f25739d60">000c</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_6c7fcd49-69e2-4f2a-b034-c0b5fbe75b9a">g/ml{SUVbw}</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_ea3e3151-1636-4715-8797-f8b53a3f0aaf">1.6.1.8</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_ef2c82c1-bb50-4b45-95d2-8028408c1379">&gt;&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_51d97363-9393-4ee8-b4e3-ac24878c85a2">Coding Scheme Designator</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_2f19fdf8-0e6e-45fe-8b8f-85e9dbbe3661">(0008,0102)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_5b496519-2f75-4532-a69e-27d75520418e">SH</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_0a534027-ac43-4707-8c11-325669d31b06">0004</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_d391a97b-13a6-4017-ade7-0d0ffa963beb">UCUM</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_7020bab1-0e91-4fce-b4d7-0cfa151cb3dd">1.6.1.8</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_01bbd7da-e54e-40fb-88fd-5d7385b52bf9">&gt;&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_0b1bc5d7-4174-4f47-9c7c-b360e74298f3">Code Meaning</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_ee2a3053-f2da-4730-9019-7e6a92a2f04e">(0008,0104)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_cd2703ef-ac50-4aad-ae72-81918b1bfd60">LO</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_2158683d-aea5-4caa-9da6-4be19e18fd3d">000c</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_a7587faf-9383-4f80-9cc2-7a0efaed4011">g/ml{SUVbw}</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_fd35849b-ecdb-4280-aad2-e6cba480edac">%enditem</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_fd598db4-7dc0-4f52-b814-74ac08ba03d4">%endseq</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_005de05a-2018-4773-a344-c567ead923d5">1.6.1.8</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_0263e47a-8890-49cc-8895-1bd53d291e6a">&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_5062c2d2-6a40-4673-bb8a-65aa83e775ac">Numeric Value</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_25e81926-7db6-470a-ad7f-f29a34af8113">(0040,a30a)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_2aab07ab-2a3c-4c8e-9345-846fd53f3e99">DS</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_a79bf87e-ce01-4d94-b26b-5229cf4c7c85">000e</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_851e439d-71da-4518-86ab-5399f9149a38">2.329186593407</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_12abf09e-6a19-4e22-a62f-99bd306f6ab2">%enditem</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_5133fdbc-535e-4640-99d9-fb4af6b8c3fa">%endseq</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_258052ae-4af8-4dcf-9c4f-cbeb40dcbdbb">1.6.1.8</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_caa5d416-1f31-4c16-9d88-3473f8173463">&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_d23fb9c4-0658-4a2a-9a97-ac35cff85cea">Content Sequence</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_4d7f006e-9bbc-4f0f-992a-29640c782b12">(0040,a730)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_fca25a45-68ad-48ec-8a7d-7ff2856983b0">SQ</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_c7247f95-fb1f-41f5-b112-cc258a3e640d">ffffffff</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_379fba04-7b22-4e8c-8de5-5a1eaba424c4"/>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_3b4c1aa9-07f7-4cd9-837a-beb604349980">%item</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_eba65cbe-2b80-425c-bf01-e0334c3dbb42">1.6.1.8.1</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_65e868b3-bb94-4ace-a346-9e4751413b48">&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_68730835-8bcc-423a-88ca-eefec8ecd56d">Relationship Type</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_fddb9e8f-22af-46fd-891d-31b1c2cc7234">(0040,a010)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_2ffb87de-94b5-45ec-9372-175fcc45ab15">CS</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_45b0048e-44ab-4617-9ed5-3e11cc50e33b">0010</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_d8c4336b-91ea-4916-9249-7488c02d2d78">HAS CONCEPT MOD</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_cae9571c-efc8-4c6d-a1fe-3c98c12362bb">1.6.1.8.1</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_88d9b833-7d3b-46e8-badf-9a7bd08e5428">&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_bd6c25fe-7f3d-4399-bdab-d5d96767fa54">Value Type</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_f9ab7309-bc3e-441e-9117-3d694504c6f0">(0040,a040)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_0d1336df-8777-488d-b5dc-588e2c5675fa">CS</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_8b6d4013-4a67-4b6d-9379-ec21b68a0cb4">0004</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_61bfc0b9-8a03-49ed-b9b8-bfb528502f42">CODE</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_25a553dd-a133-4609-adff-4cac1ebdf2b0">1.6.1.8.1</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_ca4ae64b-129a-4761-b93c-f780bf87c4ea">&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_e1dd1e92-5f4d-493b-88b1-37a69005181f">Concept Name Code Sequence</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_8f5741f6-af1e-493c-8a1e-af442b417802">(0040,a043)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_94799c01-8e9c-498c-b30e-78ea0dd4d428">SQ</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_61e61122-51a9-4f3f-abaf-1d164665d778">ffffffff</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_97af910f-da10-4e5b-b1d8-5edcd2626f15"/>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_7e861fc7-071b-41b0-a16b-3b22928595d6">%item</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_1804ff03-ea68-42b9-93ad-c425cf5b73b4">1.6.1.8.1</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_117906de-3973-42bd-9cf1-d40413e76a2a">&gt;&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_492de656-f7ef-4539-9fae-97268755f8bf">Code Value</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_9db7a980-c8b3-4484-952f-760737adac9f">(0008,0100)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_b999941a-c68b-40a1-9568-79fdd2be9f6f">SH</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_10481fce-df6e-4e76-bb53-44d1bdc695c2">0006</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_7f6c25b2-e34f-4223-8261-7a9086cdb283">121401</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_2a8ad029-875d-4b8a-bb33-0e74981534d5">1.6.1.8.1</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_3920c701-ff83-425e-b98d-c426d8303a5e">&gt;&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_25f6c844-2d5a-4db2-b243-3d4d5bb8873c">Coding Scheme Designator</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_cb340cfc-995e-4746-ad81-94b232ec38c9">(0008,0102)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_971e5a7a-962b-4ba5-9742-aebc03f07c14">SH</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_0387284b-5f83-4726-a360-747cbf76edd0">0004</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_3e5c18c0-d960-456a-b4c9-5bea468de346">DCM</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_eabd9347-5b1f-44c9-939a-a14ac702b58a">1.6.1.8.1</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_c3531e7f-c54d-4010-9c9b-66e8b6413f9c">&gt;&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_7bb22f13-cb4f-417f-92df-1b8f5a6bfac2">Code Meaning</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_bf199545-d279-49f3-be2f-2942ad80ed96">(0008,0104)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_83a6c486-582d-40a8-bb8f-f042df2e3ed6">LO</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_753b6c8a-c76e-4def-9455-e0b53d8255e6">000a</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_e0dc875f-f195-4145-8bde-b4df58f8656d">Derivation</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_9b1cd3c6-90c9-42de-9ebd-31c0050fe5ec">%enditem</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_8174a937-0971-46ad-b639-4d134a2c9aa4">%endseq</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_4dfcbf03-907e-4c43-adf2-2a3285d8f247">1.6.1.8.1</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_cef5974c-c168-41eb-ac14-b9d27bcead32">&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_0e1b589b-4049-4cc6-ac11-bfe1dfa7fd52">Concept Code Sequence</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_041fd634-07d5-4010-9273-5ffc9080890f">(0040,a168)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_6f171019-d357-4af4-90ae-8d8339b95298">SQ</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_fe600d05-ba43-4258-b47a-6f1e0d42e268">ffffffff</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_fbfe691c-5740-4b81-800f-147aaa4a3871"/>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_53cff9ce-602c-4216-a023-898343b4a558">%item</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_0c87c41a-a168-40e5-851a-4d66b56372d3">1.6.1.8.1</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_38014502-8c03-44c0-baa7-8a774a634421">&gt;&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_b2394eff-8d8e-41b8-9794-51210487c05d">Code Value</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_ee3c17db-b8ff-43d8-b5a3-6293017f163c">(0008,0100)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_df165060-5400-41f9-8fa9-5c08ca2cc47a">SH</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_4fb8307e-2881-48fa-a0a6-450d2d52f242">000a</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_ef28cba0-3529-4346-82b9-ae7ac5c48e8f">373098007</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_29a9f9a0-223d-4035-b79c-e7ad38939184">1.6.1.8.1</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_69b9f380-2e73-4448-9510-1ebb5695aca8">&gt;&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_b474a0eb-1bad-4501-9ff3-c7fd5d9dfe97">Coding Scheme Designator</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_07c14631-b4fe-473f-bc61-f55af4d3e2fc">(0008,0102)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_6b69dcbf-d24b-4dca-8b25-2987ed011cb5">SH</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_4962b433-cb80-4e87-bc83-3b17f6a4eebd">0004</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_bf703d7b-78e9-4971-9079-33800d668245">SCT</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_1d3e8bdd-b586-4ac1-aadf-dbf06cb162c9">1.6.1.8.1</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_81fe54d4-f909-4c27-84b8-3c436159db33">&gt;&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_03e88562-fa14-4be0-bf3d-6f17f0de1558">Code Meaning</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_c2cc3946-188c-4efd-a2b2-2ddff59578d5">(0008,0104)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_3e235690-6b6e-42dc-a698-32fba683f19d">LO</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_09d6a966-bd4a-4c01-905f-6b37e7b1135c">0004</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_877df436-178b-4378-af52-e666aacc155a">Mean</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_9cca10a2-316a-4947-8d6a-8774710fcc78">%enditem</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_ef43ccdd-28e0-4d45-a903-cf9ca7ed5618">%endseq</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_6a268e39-577e-4346-9f28-036b957f7278">%enditem</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_c10137f9-7226-4966-978c-bf5f1643f32a">%endseq</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_98e943f4-a68d-486b-a150-b963e18d4f39">%enditem</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_70cb2da9-9272-4aa7-9911-2575a30baac8">%item</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_ffec225a-17d0-4e19-b0c9-6f557092b393">1.6.1.9</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_81364d19-855b-437f-991c-5e50953202a4">&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_ecaba1d8-623d-49f0-9a6a-0e679ff6798c">Relationship Type</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_5d40fb74-a273-46e5-8642-7c264db7e2fb">(0040,a010)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_2c18379b-969a-4385-81f9-626360df1b93">CS</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_96f12d47-1eb9-48b1-b645-efd87df34c7b">0008</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_8cd05293-b32b-44dc-bb86-2852d836db4f">CONTAINS</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_fecdcbcd-d277-41a3-b6c2-8d657ebf6206">1.6.1.9</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_35dd5a6b-cd71-4efe-884a-ae1b8ea93fb5">&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_a48a45bd-b2e1-4fd8-875c-21d786990701">Value Type</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_bfde43b6-aa3e-47f2-b8e9-80b5f6b6ad31">(0040,a040)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_05bc390a-9bd6-416e-9c8a-6c6f04055e2f">CS</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_ebaaa12b-84e3-4987-bf78-0247e7de28e4">0004</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_0062c47b-88e0-498d-a670-53cbba34df99">NUM</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_e2dc4bf8-cf57-4a38-aa98-246644d0bb0d">1.6.1.9</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_7850d76f-6f64-485e-a1ef-5289d2d63d75">&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_656b1182-32b5-46da-9e97-16be85d2213b">Concept Name Code Sequence</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_5470dbab-13fe-4de0-a829-6c0c319197a4">(0040,a043)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_6ad407b2-c5b6-4a33-bce1-2ed6d39bfb62">SQ</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_4ebd90fc-9576-4cbd-a4c1-33ec093cc4ed">ffffffff</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_c99db7e9-5f58-4aae-aa3d-a6656ccaa85d"/>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_fc3d2acb-e70f-4949-9e05-2c19ca774d08">%item</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_780c46fa-fd24-40fa-b63a-60e4aee546b7">1.6.1.9</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_239d7c5d-2a56-495b-a565-f1cd278c78c1">&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_885b7403-e797-4923-aed2-368222afb9ed">Code Value</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_24ae0a7f-2dc6-4d5b-8c3e-09164ac3428e">(0008,0100)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_7851cab4-312d-4105-9224-43f0c7e3b74f">SH</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_740165b1-298a-4b8c-9ad2-70634b805df3">0006</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_f58ebbcf-ff5f-4d0b-8aa1-f5097a2da780">126401</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_46c6228c-2bc3-408d-b630-17ccb99f04f8">1.6.1.9</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_46135be3-0d1e-4f4b-8f18-4c1279db9371">&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_6095191e-91f1-4d7b-b6a6-da60e8503823">Coding Scheme Designator</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_be88f991-e325-4ce2-af73-b7bb6b056a59">(0008,0102)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_e58dd842-06e1-4170-bd11-7db92da9e772">SH</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_da0c018a-f1b9-46d9-8799-16691877e0f5">0004</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_621be77c-49d2-4812-8113-32b6ebb09af2">DCM</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_cb117ab9-7ce4-40c2-9a63-be9370cc42bf">1.6.1.9</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_71d7da19-ed0f-453a-a423-23a28619156d">&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_8586b713-b777-495e-bc92-35fd8b9edc9f">Code Meaning</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_74f55fb2-1e34-4b55-b6bb-bc851d96f223">(0008,0104)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_0ff41060-618b-4f94-aa18-51b8405ab667">LO</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_dd554890-6a20-4a69-b2c5-24a13e653432">0006</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_cbea9d79-a7d7-4b2f-b58e-7b1f0955d851">SUVbw</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_55f660f6-23c9-49cd-ae65-719775e6d124">%enditem</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_3f7c73ea-3ab7-43c5-a3b5-5ad03d55ec0c">%endseq</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_dcd0238b-12f4-4996-88d2-bd2aa9717dd3">1.6.1.9</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_dc97e17e-d51c-46ef-b2b6-a1ee77c61330">&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_822a50e1-3ec6-412e-ae93-d9a8e008d3a9">Observation UID</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_e8f2bc1a-1769-4bbd-8785-737fe5592c8b">(0040,a171)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_405d29ee-f947-416a-aeae-b6d9ce76243f">UI</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_2f8d2b15-4101-4011-b529-c87f85964185">002c</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_28e87f10-6ae9-4803-96af-7195966163eb">
            2.25.140657026119469861895824082767088344984</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_805f4fd2-193c-47e2-aae7-5f7bd8578f3f">1.6.1.9</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_223902c4-363a-43fe-bcbe-63d41757ce3d">&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_a07c8f5f-9c00-46e0-9561-f4e8ee2fccaa">Measured Value Sequence</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_98203d6a-9c77-406d-8ac4-729a06635ffc">(0040,a300)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_7d2b48ec-888f-49aa-8638-ee34d5c47a5a">SQ</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_cec6f17a-1ae8-425a-8dc9-75ea5344c214">ffffffff</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_b07983b5-1de1-4170-9f2d-a4da8b960afa"/>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_97842315-d7df-4645-851e-94c6264a5345">%item</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_99a5f8cd-01db-4aed-8f82-4486e3d31b1e">1.6.1.9</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_5e7559ea-60b6-45cc-885a-55a528061e6b">&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_9516a0c8-5a32-4e9d-980e-66a245e84591">Measurement Units Code Sequence</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_95ccda80-fac9-4ae9-9151-65d728d8343b">(0040,08ea)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_f8a93f59-7a37-48f7-a439-018d4a999163">SQ</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_3568bedb-b1e0-4036-857b-e75ea021ef88">ffffffff</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_1bd97edc-53b3-4a2b-9f56-a3f5a170e21b"/>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_aacc8967-37cb-4cd7-ad9f-951589e254bc">%item</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_a8c4dbbd-d574-4c37-b174-b63985954180">1.6.1.9</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_d857537d-0ed1-4e5d-a6b0-cb5ac8efdfb8">&gt;&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_cf3a8a51-8297-4549-b693-cbcce9643ba5">Code Value</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_26608eee-64b1-4a54-a6df-826f595f5381">(0008,0100)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_82cf29c9-e19c-4ee2-86c0-8179fd0cd824">SH</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_1e82971f-6bec-42fd-b521-f66240c81f06">000c</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_810858f1-99df-467b-9d5c-30430bc060bd">g/ml{SUVbw}</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_1bc953b2-de6c-4b5f-93f0-7f545d10d4d0">1.6.1.9</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_2c396736-a146-475a-826c-5cc8008ce1fc">&gt;&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_70717827-122a-43ec-b4ec-bda91f3ea3e7">Coding Scheme Designator</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_46eaa63d-c404-4a00-ae25-6137b35254b0">(0008,0102)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_dc7ee881-ac8f-4d7e-b7fe-76cd34f35797">SH</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_09ee646b-83b5-42e0-acd7-5273fb84972a">0004</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_b209fa45-4312-4b27-a84b-5099950df9f9">UCUM</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_2a514726-c9c8-4ab5-827d-ce29dd3b9fbe">1.6.1.9</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_2707909f-276c-489c-a5b1-a6eccb4f5683">&gt;&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_10f25b5b-dd71-4b4d-a4a1-cd8e3bc6b646">Code Meaning</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_5d5a77c5-2afa-4f8a-acd3-b6b618c9f46b">(0008,0104)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_c5f38ac4-0bdf-452a-9e84-ba8229d773f8">LO</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_f188e1ae-189d-4c5e-afef-2ad17a1b353f">000c</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_410f8306-5b66-4f8d-bcb2-489f20f7c6fc">g/ml{SUVbw}</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_a92bb1fe-832c-4673-939f-28b2190a95b7">%enditem</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_aac0d49d-6ce4-4b8f-8d66-8e9403f7c8ef">%endseq</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_0973e95e-b305-4b19-ae89-4a2cf4c35085">1.6.1.9</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_f01bec76-8cb6-4922-bc1b-59384b7e2fcd">&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_914a9e4a-000b-4fa7-ae62-f23f914be6f8">Numeric Value</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_ee02225a-dfbc-4590-9175-51fe819958cf">(0040,a30a)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_b874d8e6-65c5-43be-8973-c53ac5e00409">DS</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_3914aad1-49fd-4e0f-9ff5-2175800f4a26">0010</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_2bfd5c61-530f-4c18-85f7-0d5f984ce27f">1.8828952323684</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_17c7f59d-79e4-4c3a-988a-061a190bee8a">%enditem</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_25a60881-3bea-4413-a44c-b46cf6daaa76">%endseq</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_3d5c560e-d1cd-4f37-90f1-cc2c89c3ea99">1.6.1.9</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_113ad1bd-9971-415a-adb7-66088ed4f820">&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_a781ffd1-9eca-4678-a8ea-048719cf9f8d">Content Sequence</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_75418542-c1b9-48ad-9bfe-384517e4a48f">(0040,a730)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_3072db05-1a8b-46f6-b5b5-0f6da53e43ae">SQ</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_f74bd577-c371-42b4-846c-447fc680cf1e">ffffffff</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_d5c33b11-c7f8-4916-a568-721c90964747"/>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_8ccbe0c3-972a-47ad-aab6-298979439235">%item</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_4e475af1-610e-4cdb-923b-ead4155f45a2">1.6.1.9.1</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_840e5859-6e5f-478f-b632-ebc9fb3401c1">&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_468d98de-3b2d-4cb9-9275-155bfab8451c">Relationship Type</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_c2ccc440-a6f1-4277-848c-63e49358a5ff">(0040,a010)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_fdee6c30-8799-4fe2-aec6-65945b13a7ff">CS</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_350d2050-46ee-436d-b931-c0b400688019">0010</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_42292484-0366-4f80-b594-224faf55b070">HAS CONCEPT MOD</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_40d009b8-1c27-4c38-8393-8453699b86b4">1.6.1.9.1</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_2ea4e588-500d-4acd-9915-eb52344bf391">&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_a1d25dac-1914-42e4-8345-a71529eabd14">Value Type</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_c53bf14a-1c7f-4bfb-b516-c6bb8e76759b">(0040,a040)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_2b6bdd03-35c9-4812-aa12-89f8e3746eb8">CS</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_a878d553-8874-462d-a3e5-bb96a2bea811">0004</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_4b2cfde6-5d84-443a-83d2-fd720c79a30c">CODE</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_ef48255e-ac77-4e71-b1ff-792b0eb1cfe0">1.6.1.9.1</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_5fba8064-5758-47fa-90ae-3c4bf000901c">&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_f807a8be-2b27-4d42-9b5f-31d07b0f6a5c">Concept Name Code Sequence</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_8a727670-0b8f-4171-8b99-36df325771e7">(0040,a043)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_bbff6812-e4b7-40d5-96b4-5c21ac5e192e">SQ</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_421832b7-6cd0-4937-a648-4655ce118b6e">ffffffff</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_83478b66-3a7e-4ec0-9031-da8d1fb17588"/>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_30044043-c47a-4113-9537-62500f4699f9">%item</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_add39718-5dea-4a2f-8553-651f482ca91c">1.6.1.9.1</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_9d1d0e4b-4a3d-465f-8297-d2949350eb70">&gt;&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_8fa70ffc-0796-4f67-869c-d5c8acdda5ba">Code Value</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_1662654b-2bc0-4425-b197-8245e5fe82bc">(0008,0100)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_8823d9c0-2e31-45dd-a7e7-45c042568eaf">SH</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_b8be5ee2-2408-4869-86bf-555ab5b672c8">0006</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_ddaaaef5-61f5-45f1-99f9-72c383811652">121401</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_5f814111-687b-4ca7-95b7-5ea846620004">1.6.1.9.1</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_7a373f3b-8c4a-4227-9ce7-77882b92998c">&gt;&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_ee8d52a5-172d-463f-aa46-8e2ed4eddd5c">Coding Scheme Designator</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_350847ac-ff97-4d30-8315-b1be68bdae8c">(0008,0102)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_767fcb5c-13bf-40e0-92b7-e4bc8ae331ae">SH</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_80005a31-b4df-42bc-8c10-0c18fabf6e74">0004</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_f610aeb9-a8de-4f2e-bdab-f79184a894e8">DCM</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_b2d5bc93-64e6-4d2d-a7b2-5701322beaa3">1.6.1.9.1</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_336403b8-e2d1-455c-bbc4-ea22f009b17a">&gt;&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_1b726d9c-da27-45fb-b748-086ab22c5b4e">Code Meaning</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_11345570-4ec9-463b-8f88-fe9050859303">(0008,0104)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_6264b32c-5fc9-4336-8333-96a5de869cb8">LO</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_44324499-2f0c-4300-85bf-2a8e9296dba8">000a</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_760a8985-1078-4415-aaf1-e3920fa86aad">Derivation</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_68c4feab-0261-4e9f-ac31-03626ca32a18">%enditem</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_eeec6fa4-e40a-4c24-a270-227c313f221c">%endseq</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_ea92ac69-c90b-4ad9-a9de-097ee7423b5b">1.6.1.9.1</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_04b178d4-c5a9-40f0-aaec-b765f057209f">&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_b3dbc050-0911-433f-8154-80ec2f00aa51">Concept Code Sequence</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_8a758753-4cc3-41ae-ac19-5b5345bad7cf">(0040,a168)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_a9b242ff-e88e-4c8f-b704-59a5d2faac76">SQ</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_90ee779e-1abf-474b-adad-d2706a060ec1">ffffffff</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_7f008b33-eb45-40e8-bb6c-fc284f85e47b"/>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_9ed085d7-d3d1-4fb2-ba3b-302a23ddb552">%item</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_5081ad66-bfb4-4771-80fb-efba9517946b">1.6.1.9.1</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_e3e02f71-066d-426f-9a45-3e6efce4ac6c">&gt;&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_bc5408bc-4db3-4197-b1f9-f09c540e6d4f">Code Value</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_f1cb9dbc-23de-42d5-8934-6971d109e8ec">(0008,0100)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_571abb51-f69b-46cc-9850-45b147e78b5e">SH</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_5668bd76-2b76-461c-9912-321a1cf576ce">000a</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_630338c0-cf00-4a3e-95b8-f0f549014a69">386136009</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_e9357712-02ea-4fc8-a815-2ebec15250db">1.6.1.9.1</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_0d524b4a-0960-4c13-a5ec-6ed24da762bc">&gt;&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_0f01d2aa-f926-48cb-9e9a-4218e0d97d09">Coding Scheme Designator</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_b5045240-6a0a-44f4-a190-04d501d1d0db">(0008,0102)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_1f1bf4da-abfc-4e81-b494-5120f73dba83">SH</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_31ae3660-fc09-47fb-a7bb-e951d7b1c93f">0004</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_d3a9203a-3581-45c2-9caa-1b20fdabdb7d">SCT</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_812424d0-0b73-43cf-b1e5-57269ddca8e9">1.6.1.9.1</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_b3983906-6c3c-4e38-9907-d3055992ec50">&gt;&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_2e0b4900-e8e0-4e75-bfa3-416aabc972c6">Code Meaning</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_ebed6834-c0d0-4b05-a6a4-133c4c5509de">(0008,0104)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_b7b82769-586d-4034-83ec-06312c5ae4c1">LO</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_adb26967-7f9e-4f67-95db-237e79951a7d">0012</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_690359ab-0236-43a8-bc2f-c9272bc01265">Standard Deviation</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_b8aa68ac-7eb9-47cd-844c-dd478c307bbb">%enditem</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_bda3ee54-70ca-4965-aa63-e0ebc59500dd">%endseq</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_ebda83b5-87cf-4c58-8b1d-e510fb87b157">%enditem</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_afe51983-c07c-47ef-8d78-91b5d407d6cd">%endseq</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_166c74f8-15a4-4156-9c3c-3c7014d053b8">%enditem</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_8851243f-c90a-421c-85c5-3741406d1e63">%item</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_c2c840ab-483a-4295-804d-34bf6adfe1de">1.6.1.10</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_19eb6a13-aebc-448e-bf44-2e8804a8ce16">&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_d5674f9a-fd34-4d15-bb3c-7320166b5bdc">Relationship Type</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_a5f5cbf4-f455-4af6-b681-0b065af85c0a">(0040,a010)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_03d6c215-58f1-4ca2-94c5-8fc45dcfc0ab">CS</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_689889db-a03c-4bde-8de3-4f4073678ef4">0008</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_308949ff-a335-4434-b2df-b10bab095916">CONTAINS</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_9ec20151-d81f-4dcd-aa6a-3ca28e3b9bbc">1.6.1.10</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_f011dd30-b6aa-43e6-9a03-0ccfa5f75119">&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_63e593f2-4a0d-436f-8dc9-36e2f043e9d0">Value Type</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_b1e8ff5b-d1de-4add-995a-562b4f0734a6">(0040,a040)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_f0f965a7-d18c-45b6-8ee6-376f6c3fdd1e">CS</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_3c3b0875-cfea-4a2a-a342-ece33bd22e89">0004</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_2b12c3d0-49a2-4337-9c2f-f701c7b18012">TEXT</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_3811bf8c-ee78-4b2d-8e7f-ed23e0373687">1.6.1.10</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_96567340-6d9f-45d0-8492-cbaabe2ee3e8">&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_ff97bd53-b22c-43ec-8ee5-ff6bdd793873">Concept Name Code Sequence</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_25a030b7-959f-46f4-82ce-3719eab2ca19">(0040,a043)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_5be1cc97-5a8e-4a4f-b1af-46db5b2c044b">SQ</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_fc5d7cab-a0cf-4859-b277-43af309b494f">ffffffff</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_9988e712-d9d1-4b7f-a1f1-44c1aa19ca8d"/>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_31bc7f83-9f85-4f30-9c39-f50139588130">%item</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_3af850e1-3bb3-4751-8da7-ab2f6a4f76df">1.6.1.10</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_29190be7-1f3f-4a04-8c24-58e786165710">&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_716ae389-2bce-4588-8245-d4b6a21597f0">Code Value</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_0af0fb5c-6116-4ac5-8953-5c4d4d75292e">(0008,0100)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_612ce731-0bf9-4170-8075-b813a54e348b">SH</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_54ba7db1-91fe-43c4-9d8e-2d70234192b9">0006</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_be3d668a-41a7-4949-9e57-e292262df379">121106</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_cd292bc5-58ab-49e4-a753-2deec96aa154">1.6.1.10</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_49e11147-0ec4-4bbc-97de-b9f6cc2fc784">&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_5697549a-a583-4b1a-a82c-ed6e2effd8e1">Coding Scheme Designator</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_f629c9ed-ff03-4a9b-bfab-ef2bf0221018">(0008,0102)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_4a347a91-4262-40c9-a30c-f950549b2858">SH</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_eabb4a75-f74b-445f-af18-4f733c4e244c">0004</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_eeddcb0e-8dd4-4715-89f6-a71e95f8c0f7">DCM</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_11c71996-075b-451e-9d56-6dce4f129920">1.6.1.10</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_e4334bfa-9c96-4e1f-a0e3-afa9e537c1c8">&gt;&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_393ddd5b-37f0-4b07-9c2d-44083394eabf">Code Meaning</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_41ac85d3-81b4-477a-8db5-740dbf935705">(0008,0104)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_23c39fe0-2785-4909-a3f9-469cc8e169cd">LO</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_d8bd640e-28e2-4c52-b45e-3409b52bc20c">0008</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_2b864d9b-9db5-465b-863b-3c8e2760bc33">Comment</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_9eb3a5c2-60df-44d1-88b4-6cb721d98b86">%enditem</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_ca170e22-7ca6-4b7c-bfd8-36f80b5021a2">%endseq</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_4c9a4d29-99c8-4561-9db2-594ea32c1765">1.6.1.10</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_e015c753-6d4e-4b6a-95e1-aa3bbc8224c3">&gt;&gt;&gt;</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_f24c6703-f8b5-4c34-a228-899c2a4d8089">Text Value</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_384a66f4-7ca5-4ee3-bc7f-9ecb20e8ab3c">(0040,a160)</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_e314565b-c7bd-4ead-a5a3-17a463ae4031">UT</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_0c2b3380-6238-48f6-ae96-768294cbe4a9">0014</para>
          </td>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_069962dc-007d-4d4e-a016-956a53e6c89d">PT / WB NAC P600 / 0</para>
          </td>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_a2c9f4bc-7fbf-47db-bc08-e42b863f8a13">%enditem</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_769bbdc8-4dee-4810-b7a4-98386488e7fa">%endseq</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_f126272a-5e13-4d4b-9d29-8565c534e11d">%enditem</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_ca35c246-b9ac-4bf2-ae82-7ff712fd013f">%endseq</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_58930e45-82a5-4fad-8696-8b0b1a445498">%enditem</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
        <tr valign="top">
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1">
            <para xml:id="para_e5194a3f-7f64-4eb6-90cc-2d17a0267b1a">%endseq</para>
          </td>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
          <td align="left" colspan="1" rowspan="1"/>
        </tr>
      </tbody>
    </table>

      </section>
      
    </section>
    
    <section label="A.8" status="2" xml:id="sect_A.8">
      <title>Overview of Data Types</title>
      <para xml:id="para_48f6d8a7-8476-4e58-8e59-0c50d37d7c23">DICOM data types are specified in PS3.5 of the Standard (<olink targetdoc="PS3.5" targetptr="PS3.5" xrefstyle="select: labelnumber"/>).</para>
      <para xml:id="para_507e25fb-bc91-4855-ad62-34bbdb838df5">The AIM V4 data types are a subset of <xref linkend="biblio_ISO_21090"/>, which are in turn based on <xref linkend="biblio_HL7V3DTR1"/>. The AIM V4 data types used are documented in <xref linkend="biblio_AIMV3V4Changes"/> and <xref linkend="biblio_AIM_Model_Extending"/>.</para>
      <para xml:id="para_d8623ebb-74da-4498-9ae2-beb3dbcd1548">While a complete comparison of DICOM and <xref linkend="biblio_ISO_21090"/> data types, cardinality and optionality is beyond the scope of this mapping guide, some hints are given on topics that are relevant for transforming AIM instances and DICOM SR Measurement Reports.</para>
      <para xml:id="para_53a145b4-586f-4ad4-b659-d069878e7d76">The AIM V4 model uses the data types as specified in <xref linkend="table_A.8-2" xrefstyle="select: label"/> from <xref linkend="biblio_ISO_21090"/>, of which only a subset are encountered in use cases described by this mapping. In XML encoded AIM instances, the data type is not explicitly encoded, though it is defined in the UML model.</para>
      <table frame="box" label="A.8-1" rules="all" xml:id="table_A.8-1">
        <caption>ISO 21090 Data Types used in AIM V4</caption>
        <thead>
          <tr valign="top">
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_35fb66c6-97ea-4e2b-a98a-fa933828b6ea">ISO 21090 Data Types</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_9d8bbeed-2830-466e-ad21-613995bd6426">Description</para>
            </th>
          </tr>
        </thead>
        <tbody>
          <tr valign="top">
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_c9b8fe00-2916-4a15-90ed-955ca8e6a23e">BL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d51d2be2-e64b-47ec-a74f-315964dad25d">Boolean (two-valued logic). A BL value can be either true or false, or may have a nullFlavor.</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_1a5c6420-1c40-4fe0-a9d0-5766d0a53bb5">CD</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_de5df6d8-bcad-4ed2-bd98-d34427ca1eb5">Concept Descriptor. A reference to a concept defined in an external code system, terminology or ontology, or an expression in some syntax defined by the referenced code system.</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_bb623c56-3126-46d6-b2a4-4b9ed39ed12d">II</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_8c4f8447-8b9a-4760-b468-343b6a493fad">Instance Identifier. An identifier that uniquely identifies a thing or object. Instance identifiers are usually defined based on ISO object identifiers.</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_6476708c-2413-421b-b858-17058059ca15">INT</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9f082121-bc48-4cd3-a547-04bfb0ea7ab7">Integer. No arbitrary limit is imposed on the range of integer numbers.</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_8552da0e-4c92-41de-b140-aa838433891c">REAL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5cfbd59b-3627-467a-b50e-23b8197c31a4">Fractional numbers. The typical representation is decimal.</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_6b425b8b-cf7e-4681-9fdd-22ab1bc3e036">ST</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3933925e-1e21-4295-a44d-21e61bdd321d">Character string. Shall have at least one character or else have a nullFlavor.</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="center" colspan="1" rowspan="1">
              <para xml:id="para_b220b2ee-24e0-48ed-b1eb-207fb996e90f">TS</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_22a68169-a7ca-4a3a-ad44-40635a10b53d">Point in time. A quantity specifying a point on the axis of natural time. Most often represented as a calendar expression.</para>
            </td>
          </tr>
        </tbody>
      </table>
      <para xml:id="para_9c06b578-bd3f-4324-bc9e-0b9bf975452a">Additional data type mapping considerations include:</para>
      <orderedlist numeration="loweralpha">
        <listitem>
          <para xml:id="para_4da7a3c6-e2de-4863-b667-572e0185d583">Optionality and nullFlavor</para>
          <para xml:id="para_36b258c8-a3ae-4a9c-aac5-50f3581a85a7">If the original AIM instance does not include values that are required or mandatory in DICOM SR TID 1500, fixed values are specified since empty values are not permitted in DICOM SR TEXT and CODE entries and omitting the Content Item would violate the template constraints. All nullFlavor values are treated as empty, except for numeric values. </para>
        </listitem>
        <listitem>
          <para xml:id="para_c5f98aef-acd6-450e-bf9b-69bf7c03b556">Character Sets</para>
          <para xml:id="para_4238feb2-2ff5-4e6f-ab8e-fecd9d8df7e0">DICOM provides information on the interpretation of text data types by specifying a default character set (ISO-IR 6) and "Specific Character Set" (0008,0005) values that are used if the Basic Graphic Set is expanded or replaced. For AIM, the XML declaration attribute "encoding" (overall document) and the attribute "charset" (for ST data type values) may be used to provide information on character sets. See the description of Specific Character Set in <xref linkend="sect_A.6.1.1.11" xrefstyle="select: label quotedtitle"/>.</para>
        </listitem>
        <listitem>
          <para xml:id="para_eb4013fa-e798-46bd-9623-53c6ade63383">Character strings</para>
          <para xml:id="para_c2ed2f0f-6fe9-4949-b1f0-70c0a8996b94">In general, ST text value attributes in AIM XML elements are mapped to DICOM Text Value (0040,A160) of value type TEXT (with a VR of Unlimited Text (UT)) in the SR Content Tree. No maximum length is specified for AIM elements and attributes.</para>
          <para xml:id="para_f93a0529-230a-4f73-8b1e-bd4eb2530c10">Some text value attributes in AIM XML elements are mapped to Attributes in the DICOM header, and DICOM length limits may apply to character strings such as Long String (LO), e.g., Patient ID.</para>
        </listitem>
        <listitem>
          <para xml:id="para_8983729c-4199-4452-ab39-4730cce83ab0">Identifiers</para>
          <para xml:id="para_e7c56dde-ddfb-4a23-aa28-686e27ac18be">Unique identifiers in AIM V4 are encoded as the root attribute of an XML element (aim:uniqueIdentifier/​@root), which has an II data type, and are mapped to the DICOM UI VR, which is limited to 64 bytes.</para>
        </listitem>
        <listitem>
          <para xml:id="para_db4bab48-c3df-4033-bce9-d8ba0387e524">Codes</para>
          <para xml:id="para_dced8fda-6d5b-4ad5-a351-d01d7205d74a">Codes in AIM V4 are encoded as attributes of the aim:typeCode XML element and are mapped as specified in <xref linkend="table_A.8-2" xrefstyle="select: label"/> below for the <xref linkend="biblio_ISO_21090"/> code data type (CD). The code and codeSystemName attributes are encountered as attributes of the aim:typeCode XML element, but the displayName is the value attribute of a child element, aim:typeCode/​iso:displayName/​@value. Note that codeSystem (Coding Scheme UID) is usually not sent, even though it is required by <xref linkend="biblio_ISO_21090"/>. DICOM also supports other Attributes for encoding code values that exceed 16 characters in length.</para>
          <table frame="box" label="A.8-2" rules="all" xml:id="table_A.8-2">
            <caption>Mapping between DICOM Basic Code Attributes and AIM ISO 21090 Code Data Types (CD)</caption>
            <thead>
              <tr valign="top">
                <th align="center" colspan="2" rowspan="1">
                  <para xml:id="para_9db5430f-df94-46c4-a64f-86caddfc1a0d">DICOM Attribute and VR</para>
                </th>
                <th align="center" colspan="2" rowspan="1">
                  <para xml:id="para_cb0de65d-7aa6-4547-9f3d-49c88eb863ee">AIM Element and Attribute and ISO 21090 Data Type</para>
                </th>
              </tr>
            </thead>
            <tbody>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_696ffc4a-6844-4e71-bd01-0d6a118c3d24">Code Value (0008,0100)</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_7cebad70-2304-4f35-ac78-1c00f51a9945">SH</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_13e957f8-9cdb-4607-9594-95230d15d0fd">aim:typeCode/​@code</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_00287a54-56c1-442b-b988-77a1d5f63708">CD.characterstring</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_efdab805-facf-470e-a627-7c8c1b69f5a1">Coding Scheme UID (0008,0x010C)</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_57f5848a-8be4-4b08-904c-5c7b174c936d">UI</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_13992445-25a4-4790-8d9c-7a3916c16c02">aim:typeCode/​@codeSystem</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_b25e7b6b-4580-4638-89f9-3a44f080afd2">CD.characterstring</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_cc0ae466-5ea7-4f82-bb4b-408f87dcd404">Coding Scheme Designator (0008,0102)</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_316b8c46-4736-4a6e-9f70-2b69b1f10e94">SH</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_05029024-5dd2-42ae-87eb-167200eeac05">aim:typeCode/​@codeSystemName</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_c72fb1cd-2105-421d-b57e-1cb50dfe3380">CD.characterstring</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_fc248382-c9ae-4883-ab88-0b4e4c6cd2dc">Coding Scheme Version (0008,0103)</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_1774b314-a45e-4931-a4e0-47ffb8d5245e">SH</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_8e3aeb0c-8a9d-42bf-bdf2-7258ef7598b0">aim:typeCode/​@codeSystemVersion</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_41a38884-ea26-4f9c-8d02-88fdcb48130e">CD.characterstring</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_7caac131-f8ab-4508-bbcc-0989676a459c">Code Meaning (0008,0104)</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_5b2c4c42-c673-4611-86f0-1cf6189936bd">LO</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_a5710642-5aca-4c35-916a-9d3609fc55af">aim:typeCode/​iso:displayName/​@value</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_6196c909-f68b-44c4-bf29-83819a3f04c3">CD.ST</para>
                </td>
              </tr>
            </tbody>
          </table>
        </listitem>
        <listitem>
          <para xml:id="para_ae193aac-5851-4505-baa2-3672bb2a5f23">Date and Time</para>
          <itemizedlist>
            <listitem>
              <para xml:id="para_d05901bb-5da9-499d-a53b-867b8a644b4d">The AIM V4 XML element aim:dateTime/​@value attribute corresponds to the ISO 21090 TS data type, and is mapped to the DICOM DateTime (DT) VR, or the combination of separate Date (DA) and Time (TM) Attributes.</para>
            </listitem>
            <listitem>
              <para xml:id="para_43158009-00e7-4ec9-9471-a7c6145d8e9f">DICOM DT matches TS except for the number of decimal places of fractional seconds (6 for DT versus 4 for TS).</para>
            </listitem>
            <listitem>
              <para xml:id="para_cfaa81af-8d5a-4ca2-9c94-2ad72fd13931">DICOM DA matches the TS part YYYYMMDD (Y=Year, M=Month, D=Day), except that TS may be missing DD or MMDD.</para>
            </listitem>
            <listitem>
              <para xml:id="para_04858f57-afde-4841-ae16-d21c605ea277">DICOM TM matches the TS part HHMMSS.UUUUUU (H=Hour, M=Minute, S=Second, U=Fractional Second) except for the number of decimal places of fractional seconds (6 for DT versus 4 for TS).</para>
            </listitem>
            <listitem>
              <para xml:id="para_2013d00e-532b-4121-be44-29dbb5a69899">If available, the DICOM Timezone Offset From UTC (0008,0201) used for DA or TM data types may be populated using time zone offset values from the ISO 21090 TS value.</para>
            </listitem>
            <listitem>
              <para xml:id="para_cb6de56c-7d08-4350-b4c6-4f61c015fca0">ISO TS allows for separators; these need to be removed for conversion to DT, DA and TM.</para>
            </listitem>
          </itemizedlist>
        </listitem>
        <listitem>
          <para xml:id="para_1364fa5a-e7a5-4389-97fe-ff6bc340b06b">Person Names</para>
          <itemizedlist>
            <listitem>
              <para xml:id="para_00d25ea2-80cd-4864-b36a-7ee2512a36d1">DICOM Person Name (PN) shall be mapped from <xref linkend="biblio_ISO_21090"/> data type Person Name (PN) as described in <xref linkend="table_A.8-3" xrefstyle="select: label"/>.</para>
              <table frame="box" label="A.8-3" rules="all" xml:id="table_A.8-3">
                <caption>Mapping between DICOM Person Name (PN) ans ISO 21090 Data Type Person Name (PN)</caption>
                <thead>
                  <tr valign="top">
                    <th align="center" colspan="1" rowspan="1">
                      <para xml:id="para_537fd510-73a4-4a83-b92a-38f5b0d8880f">DICOM Person Name (PN)</para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                      <para xml:id="para_0183cf65-fb78-4fd9-aed7-427f32822ce9">ISO 21090 Data Type: Person Name (PN)</para>
                    </th>
                  </tr>
                </thead>
                <tbody>
                  <tr valign="top">
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_56fdfd46-3fb3-4cbc-9108-70bf9b95c199">&lt;family_name_complex&gt;</para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_0de6e3f2-0710-48d6-8c60-546b4b8bba1e">Family Part type</para>
                    </td>
                  </tr>
                  <tr valign="top">
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_d79eadbe-51c1-4059-91fa-0aaaa8d049c4">&lt;given_name_complex&gt;</para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_4f4a0417-6654-4d41-a9d8-d8f091c1ec58">Given Part type</para>
                    </td>
                  </tr>
                  <tr valign="top">
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_0752239a-0f09-45a5-a6cb-cfd3faaa87db">&lt;middle_name&gt;</para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_46796ac3-47e1-48d9-8d81-a591d1b6d8a5">Given Part type - order of parts matters</para>
                    </td>
                  </tr>
                  <tr valign="top">
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_ce4d4bf3-85c0-4f9a-9af5-dffbdd562d85">&lt;name_suffix&gt;</para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_5c657628-e776-4742-a2a4-abcd0fe5a290">Suffix Part type</para>
                    </td>
                  </tr>
                  <tr valign="top">
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_e95639b4-cf4a-4ef0-a521-dbf0e7961a56">&lt;name_prefix&gt;</para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                      <para xml:id="para_d8b6e3dd-c23b-4ad9-bb2b-0eb238f3175e">Prefix Part type</para>
                    </td>
                  </tr>
                </tbody>
              </table>
            </listitem>
            <listitem>
              <para xml:id="para_0b6b6460-639b-4f2b-a472-a48ab7d92c4e">
<xref linkend="biblio_ISO_21090"/> PN may contain multiple given names. DICOM PN Middle Name shall be mapped to <xref linkend="biblio_ISO_21090"/> PN Given Name Part type.</para>
              <example label="A.8-1" xml:id="example_A.8-1">
                <title>Person Name Example</title>
                <para xml:id="para_b18d64ec-f1b7-4b32-984b-d8abc3c0d27b">John Robert Morrison, Ph.D. "Morrison^John Robert^^^Ph.D." [One family name; two given names; no middle name; no prefix; one suffix] can be represented as a <xref linkend="biblio_ISO_21090"/> Person Name (PN) in the following way:</para>
                <programlisting><![CDATA[<name>
   		<given>John</given>
   		<given>Robert</given>
		<family>Morrison</family>
 		<suffix>Ph.D.</suffix>
	</name>
]]></programlisting>
              </example>
            </listitem>
            <listitem>
              <para xml:id="para_413d0071-cc8c-47f7-b38d-4d45bf862bc6">The following <xref linkend="biblio_ISO_21090"/> PN use codes may be used to represent multi-part DICOM person names: ABC (Alphabetic), IDE (Ideographic), SYL (Phonetic).</para>
              <example label="A.8-2" xml:id="example_A.8-2">
                <title>HL7 V3 Multi-Part Person Name Example</title>
                <programlisting><![CDATA[<name use="ABC">
   		<family>KIMURA</family>
   		<given>MICHIO</given>
	</name>
	<name use='IDE'>
   		<family>木村</family>
   		<given>道男</given>
	</name>
	<name use="SYL">
   		<family>きむら</family>
   		<given>みちお</given>
	</name>
]]></programlisting>
              </example>
            </listitem>
          </itemizedlist>
        </listitem>
        <listitem>
          <para xml:id="para_57e1f0f3-6c47-4807-80a3-e12fb28b781f">Numeric Measurements</para>
          <para xml:id="para_3ed6843c-c06a-4254-9a70-462b7aebee55">DICOM Numeric Measurement value types shall be mapped from the <xref linkend="biblio_ISO_21090"/> data types as specified in <xref linkend="table_A.8-4" xrefstyle="select: label"/>.</para>
          <table frame="box" label="A.8-4" rules="all" xml:id="table_A.8-4">
            <caption>Mapping between DICOM Numeric Measurement Value Types and ISO 21090 Data Types</caption>
            <thead>
              <tr valign="top">
                <th align="center" colspan="2" rowspan="1">
                  <para xml:id="para_13d90cce-ad82-45a4-a010-02d799638a2b">DICOM <olink targetdoc="PS3.3" targetptr="PS3.3" xrefstyle="select: labelnumber"/>, <olink targetdoc="PS3.5" targetptr="PS3.5" xrefstyle="select: labelnumber"/> and <olink targetdoc="PS3.16" targetptr="PS3.16" xrefstyle="select: labelnumber"/>: Numeric Measurement (NUM) Value Type</para>
                </th>
                <th align="center" colspan="2" rowspan="1">
                  <para xml:id="para_2c634e3f-d27a-4779-a349-f45c25bcfb49">AIM Path and ISO 21090 Data Type</para>
                </th>
              </tr>
            </thead>
            <tbody>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_f7c12c2e-a54a-4849-8f50-2a185e3bbd46">Measured Value Sequence (0040,A300) &gt; Concept Name Code Sequence (0040,A043)</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_16d57463-ed7d-4e75-879c-28a2262b56ad">Code Sequence Macro</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_62e93a8f-5eea-4a79-aae3-b52f83bf284f">CalculationEntity/​typeCode[1]</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_3f06afa0-ab09-45c7-8ff1-547259d5cfd0">CD</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_99bd176e-e361-40a3-89f3-f69fe99096fd">Measured Value Sequence (0040,A300) &gt; Numeric Value (0040,A30A)</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_cf5f10df-f297-430f-941e-990f85ae4d71">DS</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_7f65d216-ebbf-442d-b180-6f81cf8df488">CalculationEntity/​calculationResultCollection/​CalculationResult/​​@value</para>
                  <para xml:id="para_78c4423d-f517-43cc-8a0f-e910aed6b50e">CalculationEntity/​calculationResultCollection/​CalculationResult/​​calculationDataCollection/​CalculationData/​@value</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_b959c226-553d-4b08-b565-4f0ed4cabac3">ST</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_df163ffd-5592-47e5-bb54-3dceb7f49f58">Measured Value Sequence (0040,A300) &gt; Measurement Units Code Sequence (0040,08EA)</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_e9ebe78f-f38d-42d1-b7c5-5f16abae1cbf">Code Sequence Macro</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_e5639dae-1dd3-4b4d-87b9-c2014014fd22">CalculationEntity/​calculationResultCollection/​CalculationResult/​unitOfMeasure</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_594c760c-8777-429e-972d-03058cd494b8">ST</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_769a78b3-606d-4aeb-9471-62af957fa815">Numeric Value Qualifier Code Sequence (0040,A301)</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_4a5bb8b4-5cef-447e-93ab-4175f7d313b3">Code Sequence Macro</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_ee3dacf4-dc84-441e-90ec-72d10c3305d3">CalculationEntity/​calculationResultCollection/​CalculationResult/​calculationDataCollection/​CalculationData/​@value</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_5bca4224-0564-4116-be80-fd8401d647c7">ST</para>
                </td>
              </tr>
            </tbody>
          </table>
          <para xml:id="para_ce8e42f3-5629-4b08-9f17-109416cb3c9e">The <xref linkend="biblio_ISO_21090"/> PQ data type is not used in AIM.</para>
          <para xml:id="para_d6d34ef3-b9dd-42f1-af4d-7ed0a3c544d4">The Concept Name of the measurement is usually pre-coordinated in a single CalculationEntity/​typeCode entry. If there is more than one CalculationEntity/​typeCode, the first is assumed to be the primary concept and the others may be modifiers that, if recognized as such, may be mapped to method and derivation, or if otherwise recognized and name-value pair of concepts can be constructed can be encoded as generic modifiers, but otherwise have to be ignored.</para>
          <para xml:id="para_1d61f48e-3e93-4f7a-b516-6d8b848e8e90">The Numeric Value may be found as the single value of a CompactCalculationResult (i.e., value child of CalculationResult) or the first value of an ExtendedCalculationResult (i.e., nested within calculationResultCollection). This can give rise to a difference in representation in a round trip conversion.</para>
          <para xml:id="para_8c79d718-a9ff-4aea-b52a-ad9a92e98345">Units of measurement shall be converted from a text string (ST) to a Coded Sequence entry using the UCUM Code Values and "UCUM" as the Coding Scheme Designator (in AIM, CalculationResult/​unitOfMeasure is defined as "A string representation of UCUM unit for the value of the calculation").</para>
          <para xml:id="para_683000ee-7bbe-4949-b7f4-754ac53c1f41">The AIM CalculationData/​@value shall be assumed to be in the US English locale (i.e., periods are used as the decimal point, not commas, etc.).</para>
          <para xml:id="para_3eea8356-ef7b-401b-bc00-9116f896290e">The length of the AIM CalculationData/​@value ST is not limited, but the DICOM DS value representation is limited to 16 characters. Values of CalculationData/​@value that are too long shall be truncated or rounded to fit in an implementation-dependent manner.</para>
          <para xml:id="para_6f6ac400-ee61-434f-b70f-0a0d03d30644">The CalculationResult/​dataType (e.g., Double, Integer) is not encoded in the DICOM mapping, since all DICOM SR numeric values are encoded as a Decimal String (DS), so in a round trip from AIM to DICOM and back to AIM will not be recovered (i.e., will always be encoded as Double). For the use cases for this mapping, it is likely that all measurements will be Double anyway.</para>
          <para xml:id="para_0be77456-db43-4acb-9217-fded9d3126ae">DICOM allows the Measured Value Sequence (0040,A300) to be sent zero length (empty) if there is no value. In such cases the Numeric Value Qualifier Code Sequence (0040,A301) may be used in DICOM to send a code indicating why, either because of an invalid floating point result (e.g., <olink targetdoc="PS3.16" targetptr="DCM_114000">(114000, DCM, "Not a number")</olink> corresponding to <xref linkend="biblio_IEEE754"/> NaN), or for more general reasons (e.g., <olink targetdoc="PS3.16" targetptr="DCM_114006">(114006, DCM, "Measurement failure")</olink>). See <olink targetdoc="PS3.16" targetptr="sect_CID_42" xrefstyle="select: labelnumber quotedtitle"/>. <xref linkend="table_A.8-4" xrefstyle="select: label"/> indicates that a non-numeric CalculationData/@value may be mapped to Numeric Value Qualifier Code Sequence (0040,A301). Various possible mappings of AIM string values to a subset of DICOM codes corresponding to <xref linkend="biblio_IEEE754"/> are defined in <xref linkend="table_A.8-5" xrefstyle="select: label"/>. These are based
            on the:</para>
          <itemizedlist>
            <listitem>
              <para xml:id="para_b3cc4d54-d949-4e70-a8ee-f209839ee264">Java Double.toString(double) definition (see <link xl:href="https://docs.oracle.com/javase/8/docs/api/java/lang/Double.html#toString-double-"/>)</para>
            </listitem>
            <listitem>
              <para xml:id="para_09e95f78-fc4b-4dd4-bae2-634c92770c2e">
<xref linkend="biblio_XSD_DataTypes"/>
</para>
            </listitem>
            <listitem>
              <para xml:id="para_f3609879-1e3a-4087-b046-b0b09ab1e808">
<xref linkend="biblio_ISO_21090"/>
</para>
            </listitem>
          </itemizedlist>
          <para xml:id="para_39cd7bab-6a9c-49d7-8bc9-5248801308a4">No similar standard C or C++ mapping is known to exist (e.g., for libc dtostr() or sprintf()). Other languages offer some flexibility (e.g., Python supports case insensitive variants of "NaN" and "Infinity", the latter with or without a sign; see <link xl:href="http://docs.python.org/3/library/functions.html#float"/>). For JavaScript, see <link xl:href="https://tc39.github.io/ecma262/#sec-tostring-applied-to-the-number-type"/>, <link xl:href="https://tc39.github.io/ecma262/#sec-parsefloat-string"/> and <link xl:href="https://tc39.github.io/ecma262/#sec-number.parsefloat"/>. The table describes a subset of possible values, the mapping may not be exact (e.g., the definitions of NaN may differ), the mapping is ambiguous (since AIM does not define which string source to use), and the mapping of other values is undefined.</para>
          <table frame="box" label="A.8-5" rules="all" xml:id="table_A.8-5">
            <caption>Mapping between DICOM Numeric Value Qualifier Code Sequence and AIM ST</caption>
            <thead>
              <tr valign="top">
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_29293882-c619-4023-8be7-ab5f4ad7a6f7">DICOM Code</para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_72948da2-bbf3-4981-af85-9089bc188158">Java String</para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_2bf28d24-cfb0-4628-b96a-0e6cc9d8585c">XML Schema</para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_0b063ca6-de37-49ab-b293-a15c3fd29803">ISO 21090 Null Flavor</para>
                </th>
              </tr>
            </thead>
            <tbody>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_725a20a9-b809-4832-8418-5494c2f11f35">
<olink targetdoc="PS3.16" targetptr="DCM_114000">(114000, DCM, "Not a number")</olink>
</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_5eeb62f3-10a5-4c47-91e3-4860a809f23b">NaN</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_f8ea391c-547b-4bdb-9d01-6234e3b3caec">NaN</para>
                </td>
                <td/>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_2d2be3c4-b844-4ce5-8d28-f863e7545e5f">
<olink targetdoc="PS3.16" targetptr="DCM_114001">(114001, DCM, "Negative Infinity")</olink>
</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_6c8bcec9-64aa-47a3-911a-fabe67198271">-Infinity</para>
                </td>
                <td>
                  <para xml:id="para_df1e636b-891e-4c49-9f3c-152ff00f3a6b">-INF</para>
                </td>
                <td>
                  <para xml:id="para_92619187-219f-4b39-aff5-c16d051b247d">NINF</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_d3458423-7b01-4739-88a6-f57f0d36c8d0">
<olink targetdoc="PS3.16" targetptr="DCM_114002">(114002, DCM, "Positive Infinity")</olink>
</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_6b4f83ec-5023-42d5-a897-5659513b2b49">Infinity</para>
                </td>
                <td>
                  <para xml:id="para_7dcccf47-c579-49b9-b36a-e6a7ea3daf4c">INF</para>
                </td>
                <td>
                  <para xml:id="para_4a30013c-259a-4246-8a3c-4b51b62efed6">PINF</para>
                </td>
              </tr>
            </tbody>
          </table>
        </listitem>
        <listitem>
          <para xml:id="para_c9dfb59a-72c4-41ca-b243-e1d40cab18e7">Image and segmentation references</para>
          <para xml:id="para_a656ba8d-f4bb-4e76-8ce4-1ade9fb41278">DICOM image references may be mapped as specified in <xref linkend="table_A.8-6" xrefstyle="select: label"/>.</para>
          <table frame="box" label="A.8-6" rules="all" xml:id="table_A.8-6">
            <caption>DICOM Image references to AIM Path</caption>
            <thead>
              <tr valign="top">
                <th align="center" colspan="2" rowspan="1">
                  <para xml:id="para_a6a40482-85cc-479f-a29d-f20dce17572a">DICOM <olink targetdoc="PS3.3" targetptr="PS3.3" xrefstyle="select: labelnumber"/>, <olink targetdoc="PS3.5" targetptr="PS3.5" xrefstyle="select: labelnumber"/> and <olink targetdoc="PS3.16" targetptr="PS3.16" xrefstyle="select: labelnumber"/>: Image Reference (IMAGE) Value Type</para>
                </th>
                <th align="center" colspan="2" rowspan="1">
                  <para xml:id="para_1696caeb-6742-4adb-8675-8c9b90a734e9">AIM Path and ISO 21090 Data Type</para>
                </th>
              </tr>
            </thead>
            <tbody>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_67cd09cd-2d44-47b3-b466-60f999b8aa4a">Referenced SOP Sequence &gt; Referenced SOP Class UID</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_f62ccd49-830c-4870-8af5-e80961f24d58">UI</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_701fecdc-2a8b-47d4-8b52-02ada8ed1d9a">/Image​Annotation​Collection/​image​Annotations/​Image​Annotation/​imageReferenceEntityCollection/​ImageReferenceEntity/​imageStudy/​imageSeries/​imageCollection/​Image[sopInstanceUid/​@root=​imageReferenceUid/​@root]/​sopClassUid/​@root</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_bb4d641a-7d59-47e9-9d92-37d5123dd98c">II</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_01601a3b-c634-44e1-981e-d4f93b62916f">Referenced SOP Sequence &gt; Referenced SOP Instance UID</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_d6fdfa26-703d-4fef-9e45-dd7633169b2a">UI</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_cbbacd9e-cc9f-42d2-8c99-d3b45d5ba736">imageReferenceUid/​@root</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_8597471f-6777-4235-a8e5-779d5c78e854">II</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_3d1fe3ef-01cf-4d07-b337-28ec3ce35414">Referenced SOP Sequence &gt; Referenced Frame Number</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_2c5f6a96-2b97-480d-94fb-26616834e305">IS</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_abc17cac-b963-4394-91f3-d943b7e04afb">referencedFrameNumber/​@value</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_db2edbf9-27f2-40ca-bd76-44c90e8a06c9">INT</para>
                </td>
              </tr>
            </tbody>
          </table>
          <para xml:id="para_99c05498-1a1e-48c8-940a-0b436e8b04dc">An image reference in the AIM tree locally consists of the SOP Instance UID only, without SOP Class, which is described elsewhere in the tree in the imageReferenceEntityCollection (which, similar to the DICOM Current Requested Procedure Evidence Sequence or Pertinent Other Evidence Sequence, also contains the Study and Series level information). Hence the use of the predicate "sopInstanceUid/​@root=$sopInstanceUID" in the path in the table.</para>
            <para xml:id="para_4605072d-31ab-4635-981b-6cf14d8eced6">The AIM version 4.2 model includes an optional AccessionNumber in the imageStudy class used in the imageReferenceEntityCollection. This may be preserved in a DICOM SR instance in the ImageLibrary.</para>
            <para xml:id="para_f47b35af-1e76-48bf-b9a8-8d95b34168a5">DICOM segmentation references may be mapped as specified in <xref linkend="table_A.8-7" xrefstyle="select: label"/>.</para>
          <table frame="box" label="A.8-7" rules="all" xml:id="table_A.8-7">
            <caption>DICOM Segmentation references to AIM Path</caption>
            <thead>
              <tr valign="top">
                <th align="center" colspan="2" rowspan="1">
                  <para xml:id="para_4e9395bc-e5da-4207-9b25-aa6069656309">DICOM <olink targetdoc="PS3.3" targetptr="PS3.3" xrefstyle="select: labelnumber"/>, <olink targetdoc="PS3.5" targetptr="PS3.5" xrefstyle="select: labelnumber"/> and <olink targetdoc="PS3.16" targetptr="PS3.16" xrefstyle="select: labelnumber"/>: Image Reference (IMAGE) Value Type</para>
                </th>
                <th align="center" colspan="2" rowspan="1">
                  <para xml:id="para_ee663b02-c7de-40b6-8e89-f1a8b1e54a57">AIM Path and ISO 21090 Data Type</para>
                </th>
              </tr>
            </thead>
            <tbody>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_8808a374-08a0-4e07-b384-841b6e50ad67">Referenced SOP Sequence &gt; Referenced SOP Class UID</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_60d9d4fb-e068-49f5-b0aa-e87e37aa435b">UI</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_e1fcb82d-83fe-49bf-b3c1-f2d683a532ab">SegmentationEntity/​sopClassUid/​@root</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_89f4ded5-b8ca-4b39-94e2-2e0d381bd2f1">II</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_30e9ddcb-2922-49f4-a637-97634aba32f9">Referenced SOP Sequence &gt; Referenced SOP Instance UID</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_50ec909c-9461-42c5-88af-feae78186996">UI</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_c20b2378-8516-4a1c-8e63-e54a48bcb5b9">​SegmentationEntity/​sopInstanceUid/​@root</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_dfce09d3-2c05-4a3b-ab7c-a162c6e8a327">II</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_b4d7f987-49ce-4dc7-8993-31379fa5cab1">Referenced SOP Sequence &gt; Referenced Segment Number</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_e135f5d9-3c99-4f1f-9074-01f0a0a7de0b">US</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_703f6bdc-a059-451b-b025-1b5ea096feba">SegmentationEntity/​segmentNumber/​@value</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_2ec1264a-075d-421c-9063-fb6adc5d9fea">INT</para>
                </td>
              </tr>
            </tbody>
          </table>
          <para xml:id="para_0e1bf246-ccfd-4a35-bc6d-8f3b67c29f42">The SOP Class UID is included locally in the AIM tree with the reference, rather than being factored out into the imageReferenceEntityCollection, in which it is not present.</para>
          <para xml:id="para_56daa39b-d85b-4727-b44f-24a939ec7402">Ideally, all segmentation references would be included in either Current Requested Procedure Evidence Sequence or Pertinent Other Evidence Sequence as appropriate.
              There is optional information in the AIM 4.2 model to identify the Study and Series, but if the Study and Series Instance UIDs are absent,
              they cannot be assumed to be those of any related images.</para>
            <para xml:id="para_1ea039d9-8f12-46d3-8de6-160a1e5767fc">The reference to the original image that was segmented or a representative image on which the segmentation may be displayed,
                which may be present in SegmentationEntity/​referencedSopInstanceUid/​@root, may be encoded in a separate Content Item if supported by the template (e.g., TID 1410, TID 1411) in <olink targetdoc="PS3.16" targetptr="DCM_121233">(121233, DCM, "Source image for segmentation")</olink>.</para>
        </listitem>
      </orderedlist>
    </section>
  </chapter>
  
</book>