DICOM PS3.17 2023e - Explanatory Information
XXX.3.6.3.2 Volumetric Presentation State Cropping Module Recommendations

Table XXX.3.6-2. Volumetric Presentation State Cropping Module Recommendations

Attribute Name

Tag

Comment

Volume Cropping Sequence

(0070,1301)

The sequence contains three items, one for each segmented nodule type.

For brevity only the first item of the sequence is shown in this table.

>Volume Cropping Method

(0070,1302)

Set to "INCLUDE_SEG"

> Referenced Image Sequence

(0008,1140)

Set references to segments depicting the areas that make up the nodules marked by the segmentation.

>>Referenced SOP Class UID

(0008,1150)

Set to "1.2.840.10008.​5.​1.​4.​1.​1.​66.​4" (Segmentation Storage SOP Class UID)

>>Referenced SOP Instance UID

(0008,1155)

Set to the SOP Instance UID of the instance that contains the Segmentation.

>>Referenced Segment Number

(0062,000B)

Set to the identifier of the relevant segment within the Segmentation instance (e.g., "Large Nodules").

…


DICOM PS3.17 2023e - Explanatory Information