DICOM PS3.17 2023e - Explanatory Information
XXX.3.12.3.5 Render Display Module Recommendations

Table XXX.3.12.3.5-1. Render Display Module Recommendations

Attribute Name

Tag

Comment

Pixel Presentation

(0008,9205)

Set to TRUE_COLOR

Volume Stream Sequence

(0070,1A08)

Item #1 of Volume Stream Sequence

>Volumetric Presentation Input Set UID

(0070,1209)

Set to UID1

>Presentation State Classification Component Sequence

(0070,1801)

>Item #1 of Presentation State Classification Component Sequence

>>RGBA Transfer Function Description

(0070,1A09)

Set to "Liver, semi-opaque"

>>Component Type

(0070,1802)

Set to ONE_TO_RGBA

>>Component Input Sequence

(0070,1803)

>>> Volumetric Presentation Input Index

(0070,1804)

Set to 1

>>RGB LUT Transfer Function

(0028,140F)

TABLE

>>Alpha LUT Transfer Function

(0028,1410)

TABLE

>>Red, Green, Blue Palette Color Lookup Table Descriptors and Data

Orange tint

>>Alpha Palette Color Lookup Table Descriptor and Data

Semi-transparent across liver H.U.range

>Item #2 of Presentation State Classification Component Sequence

>>RGBA Transfer Function Description

(0070,1A09)

Set to "Tumor mass, solid"

>>Component Type

(0070,1802)

Set to ONE_TO_RGBA

>>Component Input Sequence

(0070,1803)

>>> Volumetric Presentation Input Index

(0070,1804)

Set to 2

>>RGB LUT Transfer Function

(0028,140F)

TABLE

>>Alpha LUT Transfer Function

(0028,1410)

NONE

>>Red, Green, Blue Palette Color Lookup Table Descriptors and Data

Yellow tints

>Item #3 of Presentation State Classification Component Sequence

>>RGBA Transfer Function Description

(0070,1A09)

Set to "Contrast-enhanced vessels, red"

>>Component Type

(0070,1802)

Set to ONE_TO_RGBA

>>Component Input Sequence

(0070,1803)

>>> Volumetric Presentation Input Index

(0070,1804)

Set to 3

>>RGB LUT Transfer Function

(0028,140F)

TABLE

>>Alpha LUT Transfer Function

(0028,1410)

NONE

>>Red, Green, Blue Palette Color Lookup Table Descriptors and Data

Red tints

Item #2 of Volume Stream Sequence

>Volumetric Presentation Input Set UID

(0070,1209)

Set to UID2

>Presentation State Classification Component Sequence

(0070,1801)

>Item #1 of Presentation State Classification Component Sequence

>>RGBA Transfer Function Description

(0070,1A09)

Set to "Contrast-enhanced vessels, blue"

>>Component Type

(0070,1802)

Set to ONE_TO_RGBA

>>Component Input Sequence

(0070,1803)

>>>Volumetric Presentation Input Index

(0070,1804)

Set to 4

>>RGB LUT Transfer Function

(0028,140F)

TABLE

>>Alpha LUT Transfer Function

(0028,1410)

NONE

>>Red, Green, Blue Palette Color Lookup Table Descriptors and Data

Blue tints

>Item #2 of Presentation State Classification Component Sequence

>>RGBA Transfer Function Description

(0070,1A09)

Set to "Contrast-enhanced vessels, orange"

>>Component Type

(0070,1802)

Set to ONE_TO_RGBA

>>Component Input Sequence

(0070,1803)

>>>Volumetric Presentation Input Index

(0070,1804)

Set to 5

>>RGB LUT Transfer Function

(0028,140F)

TABLE

>>Alpha LUT Transfer Function

(0028,1410)

NONE

>>Red, Green, Blue Palette Color Lookup Table Descriptors and Data

Orange tints

Presentation State Compositor Component Sequence

(0070,1805)

Item #1

>Weighting Transfer Function Sequence

(0070,1806)

Specify tables that give equal weight to both volumes

ICC Profile

(0028,2000)

Value corresponding to IEC 61966-2-1:1999

Color Space

(0028,2002)

Set to SRGB


DICOM PS3.17 2023e - Explanatory Information