| DICOM PS3.17 2023e - Explanatory Information |
|---|
This Section provides examples of the relationship between the Ophthalmic Tomography Image SOP Instance(s) and Ophthalmic Optical Coherence Tomography B-scan Volume Analysis SOP Instance(s).
Ophthalmic Tomography Image SOP Instance UID is "1.2.3.4.5" and contains five frames.
Ophthalmic Optical Coherence Tomography B-scan Volume Analysis SOP Instance encodes five frames (e.g., one frame for each ophthalmic tomography frame).
References are encoded via the Per-Frame Functional Groups Sequence (5200,9230) using Attributes Derivation Image Sequence (0008,9124) and Source Image Sequence (0008,2112).
Figure UUU.2-1. Ophthalmic Tomography Image and Ophthalmic Optical Coherence Tomography B-scan Volume Analysis IOD Relationship - Simple Example
Below is a more complex example.
Ophthalmic TomographyImage SOP Instance UID is "2.3.4.5" and contains 3 frames.
Ophthalmic TomographyImage SOP Instance UID is "1.6.7.8.9" and contains 2 frames.
Ophthalmic Optical Coherence Tomography B-scan Volume Analysis SOP Instance encodes five frames (e.g., one frame for each Ophthalmic TomographyFrame from the two Ophthalmic Tomography Image SOP Instances).
Figure UUU.2-2. Ophthalmic Tomography Image and Ophthalmic Optical Coherence Tomography B-scan Volume Analysis IOD Relationship - Complex Example
| DICOM PS3.17 2023e - Explanatory Information |
|---|