DICOM PS3.17 2023e - Explanatory Information

AAA.4 Implantation Plan SR Document Dental Drilling Template Example

The following example shows the result of a planning activity for a dental implantation using a dental drilling template. The implant positioning is based on a CT-Scan during which the patient has been wearing a bite plate with 3 markers. In this example the markers (visible in the patient's CT images) are detected by the planning application. After the implants have been positioned, the bite plate, in combination with the registration information of the implants, can be used to produce the dental drilling template.

In the following example, two implants are inserted that are not assembled using Mating Points.

The markers of the bite plate are identified and stored as 3 Fiducials in one Fiducial Set. This Fiducial Set has its own Frame of Reference (1.2.3.4.100).

The Registration Object created by the planning application uses the patient's CT Frame of Reference as main Frame of Reference (see Figure AAA.4-1).

Spatial Relations of Implant, Implant Template, Bite Plate and Patient CT

Figure AAA.4-1. Spatial Relations of Implant, Implant Template, Bite Plate and Patient CT


Table AAA.3-2. Dental Drilling Template Example

Node

Code Meaning of Concept Name

Code Meaning or Example Value

TID

1

Implantation Plan

TID 7000

1.1

Language of Content Item and Descendants

English

TID 1204

1.2

Observation Context

TID 1001

1.2.1

Person Observer Name

Dr. Michael Mueller

TID 1003

1.2.2

Subject Name

John Smith

TID 1007

1.2.3

Subject ID

1.2.3.4.5.6.7.8.9

TID 1007

1.2.4

Subject Species

(337915000, SCT, "homo sapiens")

TID 1007

1.3

Implant Component List

TID 7000

1.3.1

Selected Implant Component

TID 7000

1.3.1.1

Component ID

1

TID 7000

1.3.1.2

Component Type

112305

TID 7000

1.3.1.3

Reference to Implant Template "DI1000" (derived)

TID 7000

1.3.1.4

Frame Of Reference UID

1.2.3.4.1

TID 7000

1.3.1.5

Manufacturer Implant Template

Reference to Implant Template "DI1000" (original)

TID 7000

1.3.2

Implant Component Selection

TID 7000

1.3.2.1

Component ID

2

TID 7000

1.3.2.2

Component Type

(112306, DCM, "Acetabular Cup Insert")

TID 7000

1.3.2.3

Reference to Implant Template "DI2000" (derived)

TID 7000

1.3.2.4

Frame Of Reference UID

1.2.3.4.2

TID 7000

1.3.2.5

Manufacturer Implant Template

Reference to Implant Template "DI2000" (original)

TID 7000

1.4

Information used for planning

TID 7000

1.4.1

Patient Image

Reference to CT Image01

TID 7000

1.4.1.2

Horizontal Pixel Spacing

0.3 mm/pixel

TID 7000

1.4.1.3

Vertical Pixel Spacing

0.3 mm/pixel

TID 7000

1.5

Planning Information for Intraoperative Usage

TID 7000

1.5.1

Derived Planning Images

Reference to Visualization01

TID 7000

1.5.2

Spatial Registration

Reference to Registration01

TID 7000

1.5.2.1

Frame Of Reference UID

1.2.3.4.1

TID 7000

1.5.2.2

Frame Of Reference UID

1.2.3.4.2

TID 7000

1.5.2.3

Frame Of Reference UID

1.2.3.4.3

TID 7000

1.5.2.4

Frame Of Reference UID

1.2.3.4.100

TID 7000

1.5.3

Derived Planning Data

Reference to Fiducial 01

TID 7000

1.5.3.1

Derived Fiducial

1.2.3.4.3

TID 7000

1.5.3.1.1

Fiducial Intent

Bite Plate Marker

TID 7000

1.5.3.2

Derived Fiducial

1.2.3.4.4

TID 7000

1.5.3.2.1

Fiducial Intent

Bite Plate Marker

TID 7000

1.5.3.3

Derived Fiducial

1.2.3.4.5

TID 7000

1.5.3.3.1

Fiducial Intent

Bite Plate Marker

TID 7000


DICOM PS3.17 2023e - Explanatory Information