This Section describes the Confocal Microscopy Image Module, the Confocal Microscopy Tiled Pyramidal Image Module, and the Cutaneous Confocal Microscopy Image Acquisition Parameters Module.
The Confocal Microscopy Image Module and the Confocal Microscopy Tiled Pyramidal Image Module contain Attributes specific to Confocal Microscopy images.
The Cutaneous Confocal Microscopy Image Acquisition Parameters Module contains Attributes that are specific to Cutaneous Confocal Microscopy images.
Table C.8.35.1-1 specifies the Attributes that describe confocal microscopy images.
Table C.8.35.1-1. Confocal Microscopy Image Module Attributes
|
Image identification characteristics. See Section C.8.35.1.1.1 for specialization. |
|||
|
Specifies the intended interpretation of the pixel data. See Section C.8.12.1.1.1 for specialization of this Attribute. |
|||
|
Number of bits allocated for each pixel sample. Each sample shall have the same number of bits allocated. See Section C.8.12.1.1.2 for specialization of this Attribute. |
|||
|
Number of bits stored for each pixel sample. Each sample shall have the same number of bits stored. See Section C.8.12.1.1.2 for specialization of this Attribute. |
|||
|
Most significant bit for pixel sample data. Each sample shall have the same high bit. See Section C.8.12.1.1.2 for specialization of this Attribute. |
|||
|
Data representation of the pixel samples. Each sample shall have the same pixel representation. See Section C.8.12.1.1.3 for specialization of this Attribute. |
|||
|
Number of samples (planes) per image. See Section C.8.12.1.1.4 for specialization of this Attribute. |
|||
|
Indicates whether the pixel data are encoded color-by-plane or color-by-pixel. Required if Samples per Pixel (0028,0002) has a value greater than 1. See Section C.8.12.1.1.5 for specialization of this Attribute. |
|||
|
Specifies whether an Image has undergone lossy compression (at a point in its lifetime). Enumerated Values: |
|||
|
Whether the images were acquired by the confocal microscope in reflectance or fluorescence mode. |
|||
|
Whether the tissue that is the subject of the image is in the body (i.e., in-vivo) or an excised tissue sample (i.e., ex-vivo). |
Image Type (0008,0008) is specified to be Type 1 with the following constraints:
Value 1 shall have a value of ORIGINAL or DERIVED.
Value 2 shall have a value of PRIMARY.
Value 3 (Image Flavor) shall have the Defined Terms specified in Table C.8.35.1.1.1-1.
Table C.8.35.1.1.1-1. Confocal Microscopy Image Flavors
Value 4 (Derived Pixel) shall have the Defined Terms specified in Table C.8.35.1.1.1-2.