| DICOM PS3.3 2023e - Information Object Definitions |
|---|
Table C.8.31.3-1 specifies the Attributes of the Breast X-Ray Detector Macro.
Table C.8.31.3-1. Breast X-Ray Detector Macro Attributes
|
A Sequence that describes the geometrical position of the detector. |
|||
|
Angle of the X-Ray beam in the row direction in degrees relative to the normal to the detector plane. Positive values indicate that the X-Ray beam is tilted toward higher numbered columns. Negative values indicate that the X-Ray beam is tilted toward lower numbered columns. See C.8.11.7.1.2. Required if the detector can be moved in the row direction. May be present otherwise. |
|||
|
Angle of the X-Ray beam in the column direction in degrees relative to the normal to the detector plane. Positive values indicate that the X-Ray beam is tilted toward lower numbered rows. Negative values indicate that the X-Ray beam is tilted toward higher numbered rows. See C.8.11.7.1.2. Required if the detector can be moved in the column direction. May be present otherwise. |
| DICOM PS3.3 2023e - Information Object Definitions |
|---|