C.8.27.2.4 Bits Allocated, Bits Stored and Samples Per Pixel

For IVOCT images, Bits Allocated (0028,0100), Bits Stored (0028,0101) Samples Per Pixel (0028,0002) are specified to use the following values for Presentation Intent Type (0008,0068) Attribute:

Table C.8.27.2-5. Bits Allocated/Stored

Presentation Intent Type

Bits Allocated / Bits Stored

Samples Per Pixel

FOR PRESENTATION

8/8, 16/12 or 16/16

1

FOR PROCESSING

8/8, 16/12 or 16/16

1