| DICOM PS3.3 2023e - Information Object Definitions |
|---|
Table C.8.25.16-5 specifies the Attributes of the Calculated IOL Macro.
Table C.8.25.16-5. Calculated IOL Macro Attributes
|
Constants used in calculation of intraocular lens power. These constants are a characteristic of the model of intraocular lens being considered for use in cataract surgery. |
|||
|
Constant type used in calculation of intraocular lens power. |
|||
|
Information needed to select the intraocular lens power for cataract surgery. |
|||
|
The intraocular lens power, in diopters. If Type of Optical Correction (0022,1046) is TORIC, this value represents the spherical equivalent of the toric intraocular lens power. |
|||
|
The toric intraocular lens power. Only a single Item shall be included in this Sequence. Required if Type of Optical Correction (0022,1046) is TORIC. |
|||
|
>>Include Table C.8.25.16-7 “Calculated Toric Power Macro Attributes” |
|||
|
The predicted postoperative refractive error (i.e., amount of near or far sightedness), in diopters. |
|||
|
The predicted postoperative toric error. Only a single Item shall be included in this Sequence. Required if Type of Optical Correction (0022,1046) is TORIC. |
|||
|
>>Include Table C.8.25.16-7 “Calculated Toric Power Macro Attributes” |
|||
|
Indicates whether the intraocular lens specified by this Sequence Item has been pre-selected for implantation or not. Only one Item in IOL Power Sequence (0022,1090) shall contain the value YES. |
|||
|
The IOL power that would be required to achieve exact emmetropia, or no need for glasses at distance after surgery, in diopters. If Type of Optical Correction (0022,1046) is TORIC, this value represents the spherical equivalent of the toric intraocular lens power for exact emmetropia. |
|||
|
The toric IOL power that would be required to achieve exact emmetropia. Zero or one Item shall be included in this Sequence. Required if Type of Optical Correction (0022,1046) is TORIC. |
|||
|
>Include Table C.8.25.16-7 “Calculated Toric Power Macro Attributes” |
|||
|
The IOL power that would be required to exactly achieve Target Refraction (0022,1037), in diopters. If Type of Optical Correction (0022,1046) is TORIC, this value represents the spherical equivalent of the toric intraocular lens power for exact target refraction. |
|||
|
The toric IOL power that would be required to exactly achieve Target Refraction (0022,1037). Zero or one Item shall be included in this Sequence. Required if Type of Optical Correction (0022,1046) is TORIC. |
|||
|
>Include Table C.8.25.16-7 “Calculated Toric Power Macro Attributes” |
|||
|
Comment, hints or warnings related to the intraocular lens calculation(s). |
|||
Specific behaviors associated with Calculation Comment Type (0022,112B) are left to implementations to define. For example, the user might be expected to be notified about a comment with a type of WARNING whereas they might have to seek out comments with a type of INFORMATIVE. However, since this is an optional Attribute, and one that was added to this Module after its original definition, and there are no SOP Class specific constraints on display behavior, the sender cannot be guaranteed that a receiving system or user will not ignore or discard the comments.
| DICOM PS3.3 2023e - Information Object Definitions |
|---|