Table C.8.21.2.3-1 specifies the Attributes of the Breast Tomosynthesis Contributing Sources Module, which describe Breast Tomosynthesis Contributing Sources. The Attributes in this Module specify the overall characteristics of one or more source images that were used to create a Breast Tomosynthesis Image SOP Instance.
The intention of this Module is to supply acquisition information without the need to have access to all the source projection images.
Table C.8.21.2.3-1. Breast Tomosynthesis Contributing Sources Module Attributes
|
A Sequence that describes characteristics of the sources that are used to create a derived SOP Instance. |
|||
|
>Include Table 10-13 “General Contributing Sources Macro Attributes” |
|||
|
>Include Table 10-14 “Contributing Image Sources Macro Attributes” |
|||
|
The ID or serial number of the detector used to acquire this image. |
|||
|
The date on which the detector used to acquire this image as identified in Detector ID (0018,700A) was last calibrated. |
|||
|
The time at which the detector used to acquire this image as identified in Detector ID (0018,700A) was last calibrated. |
|||
|
Physical distance between the center of each detector element, specified by a numeric pair: row spacing value (delimiter) column spacing value in mm. See Section 10.7.1.3 for further explanation of the value order. |
|||