| DICOM PS3.3 2023e - Information Object Definitions |
|---|
Table C.8.2.2-1 specifies the Attributes of the Multi-energy CT Image Module, which describe a Multi-energy CT image.
Table C.8.2.2-1. Multi-energy CT Image Module Attributes
|
Human readable summary of the Multi-Energy technique applied during the acquisition. |
|||
|
>Include Table C.8.2.2-2 “Multi-energy CT X-Ray Source Macro Attributes” |
|||
|
>Include Table C.8.2.2-3 “Multi-energy CT X-Ray Detector Macro Attributes” |
|||
|
>Include Table C.8.2.2-4 “Multi-energy CT Path Macro Attributes” |
|||
|
>Include Table C.8-119 “CT Acquisition Details Macro Attributes” |
|||
|
Include Table C.8.15.3.13-1 “Multi-energy CT Processing Macro Attributes” |
|||
|
Include Table C.8.15.3.12-1 “Multi-energy CT Characteristics Macro Attributes” |
|||
Table C.8.2.2-2 specifies the Attributes of the Multi-energy CT X-Ray Source Macro, which describe the CT X-Ray Source(s).
Table C.8.2.2-2. Multi-energy CT X-Ray Source Macro Attributes
|
X-Ray Source (see Section C.8.2.2.1.1) information. |
|||
|
Identification number of this item in the Multi-energy CT X-Ray Source Sequence. The number shall be 1 for the first Item and increase by 1 for each subsequent Item. |
|||
|
Identifier of the physical X-Ray source. This might be the serial number. The X-Ray Source ID (0018,9367) will have the same value for different values of X-Ray Source Index (0018,9366) if a single source generates different nominal energies. |
|||
|
The date and time this X-Ray source (see Section C.8.2.2.1.1) was first used in this Multi-energy acquisition. |
|||
|
The date and time this X-Ray source (see Section C.8.2.2.1.1) was last used in this Multi-energy acquisition. |
|||
|
A number, unique within the sequence, to identify the switching phase. Required if Multi-energy Source Technique (0018,9368) is "SWITCHING_SOURCE". |
|||
|
Duration, in microseconds, that the energy is nominally in the target KV for this switching phase. I.e., the Switching Phase Nominal Duration does not include the Switching Phase Transition Duration (0018,936D). The target KV is the value of KVP (0018,0060) for the item in the CT X-Ray Details Sequence (0018,9325) that identifies the Multi-energy CT Path Index (0018,937A) that corresponds to this X-Ray Source. |
|||
|
Duration, in microseconds, that the energy has left the target KV for this switching phase, but has not yet reached the target KV for the next phase. The target KV is the value of KVP (0018,0060) for the item in the CT X-Ray Details Sequence (0018,9325) that identifies the Multi-energy CT Path Index (0018,937A) that corresponds to this X-Ray Source. |
|||
Each item in the Multi-energy CT X-Ray Source Sequence (0018,9365) might describe either a constant source, a constant source with energy selective filter or one output corresponding to a specific energy of a KV switching source. The Attributes will refer to a source, meaning one item in the Multi-energy CT X-Ray Source Sequence (0018,9365). The Attributes use the phrase "physical X-Ray Source" when it is necessary to refer to the actual device (tube) rather than the current item.
| DICOM PS3.3 2023e - Information Object Definitions |
|---|