Table C.8-69 specifies the Attributes of the DX Anatomy Imaged Module, which describe the anatomy contained in a DX IOD.
Table C.8-69. DX Anatomy Imaged Module Attributes
|
Laterality of (possibly paired) body part (as described in Anatomic Region Sequence (0008,2218)) examined. NoteThis Attribute is mandatory, in order to ensure that images may be positioned correctly relative to one another for display. Shall be consistent with any laterality information contained in Primary Anatomic Structure Modifier Sequence (0008,2230), if present. |
|||
|
Include Table 10-6 “General Anatomy Required Macro Attributes” |
Anatomic Region Sequence DCID 4009 “DX Anatomy Imaged” for humans, and CID 7482 “DX Anatomy Imaged for Animal” for non-human organisms. This is the anatomic region that is placed on the table or bucky for examination. |
||
The Attributes in this Module extend the function of Body Part Examined (0018,0015) as used in other IODs, and are intended to be used to facilitate the management of images and Series in terms of routing, storage and display, as well as to dictate certain Conditions on Attributes and Modules in the DX IOD.