| DICOM PS3.3 2023e - Information Object Definitions |
|---|
Table C.36.25-1 specifies the Attributes of the RT Treatment Preparation Module, which contain information describing the procedures to prepare the patient for a radiotherapeutic treatment. These procedures may include aligning the patient position, setting up required accessories, such as fixation and shielding devices applied to the patient, administrating medications, etc.
Table C.36.25-1. RT Treatment Preparation Module Attributes
|
Include Table 10.32-1 “Entity Long Labeling Macro Attributes”. |
See Section C.36.25.1.1. |
||
|
The RT Radiation Set, RT Radiation Instances or the Treatment Position Groups to which the treatment preparation parameters apply. |
|||
|
>Include Table C.36.2.3.3-1 “RT Patient Position Scope With Legacy Support Macro Attributes”. |
|||
|
The position of the patient at the start of the treatment preparation. |
|||
|
>Include Table C.36.2.3.2-1 “RT Patient Position Macro Attributes”. |
|||
|
User-defined description of patient treatment preparation technique. |
|||
|
>Include Table 10-11 “SOP Instance Reference Macro Attributes” |
|||
|
The value of Patient Treatment Preparation Procedure Index (300A,0795) from Patient Treatment Preparation Procedure Sequence (300A,0790) corresponding to the Patient Treatment Preparation Procedure to which this Sequence Item refers. Required if this Patient Setup Photo is associated with a Patient Treatment Preparation Procedure. |
|||
|
Index of the Patient Treatment Preparation Procedure in this Sequence. |
|||
|
>>Include Table C.36.2.2.3-1 “RT Accessory Device Identification Macro Attributes” |
|||
|
>Patient Treatment Preparation Procedure Parameter Description |
User-defined description of patient treatment preparation procedure parameter(s). |
||
|
>>Include Table 10.2.1-1 “Content Item with Modifiers Macro Attributes” |
See C.36.25.1.2 for Baseline TID. |
||
A free-text description of the treatment preparation instruction, if present, shall be encoded in Attribute Entity Description (3010,0037) of the Entity Labeling Macro.
In Table C.36.25-2, if the Patient Treatment Preparation Procedure Code Sequence (300A,0791) has the code value specified in the left column, the Template for Patient Treatment Preparation Procedure Parameter Sequence (300A,0793) shall be the one specified in the right column.
Table C.36.25-2. Patient Treatment Preparation Procedure Parameter Codes