This Module is a library of target luminance characteristics to be achieved in the calibration of specific configurations of specific Display Subsystems.
Table C.32.2-1. Target Luminance Characteristics Module Attributes
|
Target luminance characteristics to which the Display Subsystem can be calibrated. |
||
|
Number that identifies this Luminance Characteristics Item. The value shall be unique within the Target Luminance Characteristics Sequence (0028,7008). |
||
|
The type of Display Function used to define the target luminance curve. See Section C.32.2.1 for Enumerated Values. |
||
|
The minimum luminance value of the target luminance curve, in candelas per square meter (cd/m2). |
||
|
The maximum luminance value of the target luminance curve, in candelas per square meter (cd/m2). |
||
|
The Gamma value for generating the target luminance curve. Required if the value of Display Function Type (0028,7019) is GAMMA. |
||
|
The number of luminance points in the user defined target luminance curve. Required if the value of Display Function Type (0028,7019) is USER_DEFINED. |
||
|
Luminance Response Values for the user defined target luminance curve. The number of Items shall match the value of Number of Luminance Points (0028,701B). Required if the value of Display Function Type (0028,7019) is USER_DEFINED. |
||
|
The integer Digital Driving Level (DDL) for this luminance value. The DDL Value for the first Item in the Sequence shall be 0. The DDL Value of the last Item in the Sequence shall be the maximum DDL value. Two Items in the Sequence shall not have the same DDL value. The DDL Value of the Items in the Sequence shall increase monotonically. |
||
|
Luminance in candelas per square meter (cd/m2) corresponding to the DDL Value. |
||
|
User defined text to describe the user defined target luminance curve. |
||
|
A single measured White Point recorded for the Display Subsystem. The white point is recorded as a CIE xy chromaticity with two dimensionless floating point values where the first is the x value and the second is the y value. |
||
|
Reflected ambient light in candelas per square meter (cd/m2) when the measurements were taken. |
||
|
Source of the reflected ambient light value. Enumerated Values: |
Enumerated Values for Display Function Type (0028,7019):
Grayscale Display Function; see PS3.14.
See CIE 1976 (L*,a*,b*) and CIELab Color Space by Gernot Huffman 1948.
See "A Standard Default Color Space for the Internet - sRGB" - http://www.w3.org/Graphics/Color/sRGB.