DICOM PS3.3 2023e - Information Object Definitions

C.26.2 Substance Approval Module

Table C.26-2 specifies the Attributes used to approve the administration of a contrast agent, drug, or device to a patient.

Table C.26-2. Substance Approval Module Attributes

Attribute Name

Tag

Attribute Description

Substance Administration Approval

(0044,0002)

Status of request for substance administration.

Enumerated Values:

APPROVED

Use of the substance for the patient is approved, with related notes (e.g., appropriate dose for age/weight) in Approval Status Further Description (0044,0003)

WARNING

The substance may be used for the patient subject to warnings described in Approval Status Further Description (0044,0003)

CONTRA_INDICATED

The substance should not be used for the patient for the reasons described in Approval Status Further Description (0044,0003)

Approval Status Further Description

(0044,0003)

Description of warning or contra-indication, or notes on approval.

Approval Status DateTime

(0044,0004)

Timestamp for the Substance Administration Approval response.


DICOM PS3.3 2023e - Information Object Definitions