DICOM PS3.3 2023e - Information Object Definitions

C.17.6 Key Object Selection Modules

C.17.6.1 Key Object Document Series Module

Table C.17.6-1 defines the Attributes of the Key Object Document Series.

Note

Series of Key Object Selection Documents are separate from Series of Images or other Composite SOP Instances. Key Object Documents do not reside in a Series of Images or other Composite SOP Instances.

Table C.17.6-1. Key Object Document Series Module Attributes

Attribute Name

Tag

Type

Attribute Description

Modality

(0008,0060)

1

Type of device, process or method that created the Instances in this Series.

Enumerated Values:

KO

Key Object Selection

Series Instance UID

(0020,000E)

1

Unique identifier of the Series.

Note

No specific semantics are specified.

Series Number

(0020,0011)

1

A number that identifies the Series.

Note

No specific semantics are specified.

Series Date

(0008,0021)

3

Date the Series started.

Series Time

(0008,0031)

3

Time the Series started.

Protocol Name

(0018,1030)

3

Description of the conditions under which the Series was performed.

Note

This may represent the type of analysis used in creation of the KO SOP Instances.

Series Description

(0008,103E)

3

Description of the Series.

Series Description Code Sequence

(0008,103F)

3

A coded description of the Series.

Only a single Item is permitted in this Sequence.

>Include Table 8.8-1 “Code Sequence Macro Attributes”

No Baseline CID is defined.

Referenced Performed Procedure Step Sequence

(0008,1111)

2

Uniquely identifies the Performed Procedure Step SOP Instance for which the Series is created.

Zero or one Item shall be included in this Sequence.

Note

See notes on this Attribute in Section C.17.1

>Include Table 10-11 “SOP Instance Reference Macro Attributes”

Treatment Session UID

(300A,0700)

3

A unique identifier of the RT Treatment Session to which Instances in this Series belong.


DICOM PS3.3 2023e - Information Object Definitions