DICOM PS3.3 2023e - Information Object Definitions

C.17.2.4 Identified Person or Device Macro

Table C.17-3b defines the Attributes that identify a person or a device. This Macro contains content equivalent to TID 1002 (see PS3.16).

Table C.17-3b. Identified Person or Device Macro Attributes

Attribute Name

Tag

Type

Attribute Description

Observer Type

(0040,A084)

1

The type of entity being identified.

Enumerated Values:

PSN

Person

DEV

Device

Person Name

(0040,A123)

1C

Name of the person observer identified.

Required if Observer Type value is PSN.

Person Identification Code Sequence

(0040,1101)

2C

Coded identifier of the person identified.

Zero or one Item shall be included in this Sequence.

Required if Observer Type value is PSN.

>Include Table 8.8-1 “Code Sequence Macro Attributes”

No Baseline CID is defined.

Organizational Role Code Sequence

(0044,010A)

3

The organizational capacity in which the person observer is participating.

>Include Table 8.8-1 “Code Sequence Macro Attributes”

No Baseline Context ID defined.

Station Name

(0008,1010)

2C

Name of the identified device.

Required if Observer Type value is DEV.

Device UID

(0018,1002)

1C

Unique identifier of the device identified.

Required if Observer Type value is DEV.

Manufacturer

(0008,0070)

1C

Manufacturer of the device identified.

Required if Observer Type value is DEV.

Manufacturer's Model Name

(0008,1090)

1C

Model Name of the device identified.

Required if Observer Type value is DEV.

Station AE Title

(0008,0055)

3

Application Entity Title of the device identified.

Device Serial Number

(0018,1000)

3

Manufacturer's serial number of the identified device.

Note

While the serial number will be unique within the scope of the Manufacturer and Model, it might not be universally unique.

Software Versions

(0018,1020)

3

Manufacturer's designation of software version of the identified device.

See Section C.7.5.1.1.3.

Date of Manufacture

(0018,1204)

3

The date the identified device was originally manufactured or re-manufactured (as opposed to refurbished).

Date of Installation

(0018,1205)

3

The date the identified device was installed in its current location. The device may or may not have been used prior to installation in its current location.

Institution Name

(0008,0080)

2

Institution or organization to which the identified person is responsible or accountable, or that manages the identified device.

Institution Code Sequence

(0008,0082)

2

Institution or organization to which the identified person is responsible or accountable, or that manages the identified device.

Zero or one Item shall be included in this Sequence.

>Include Table 8.8-1 “Code Sequence Macro Attributes”

No Baseline CID is defined.

Institutional Department Name

(0008,1040)

3

Department in the institution or organization to which the identified person is responsible or accountable, or that manages the identified device.

Institutional Department Type Code Sequence

(0008,1041)

3

A coded description of the type of Department or Service within the healthcare facility.

Note

This might be obtained from a corresponding HL7v2 message containing PV1:10 Hospital Service.

Only a single Item is permitted in this Sequence.

>Include Table 8.8-1 “Code Sequence Macro Attributes”

BCID 7030 “Institutional Department/Unit/Service”.


DICOM PS3.3 2023e - Information Object Definitions