A.90 Confocal Microscopy IODs
The Confocal Microscopy IODs specify images that are acquired by means of a confocal microscope.
The confocal microscopy may be performed in-vivo or ex-vivo in reflectance or fluorescence mode.
Confocal images may be simple (non-tiled) or tiled.
Separate IODs have been defined for simple confocal microscopy images, tiled images, and tiled pyramidal images.
Simple confocal images may be encoded according the to the Confocal Microscopy Image IOD.
A SOP Instance may contain one or more frames (multi-frame).
A movie acquisition may be encoded as a multi-frame cine image.
A.90.1 Confocal Microscopy Image IOD
A.90.1.1 Confocal Microscopy Image IOD Description
The Confocal Microscopy Image IOD specifies the Attributes of a simple (non-tiled) Confocal Microscopy Image.
A.90.1.2 Confocal Microscopy Image IOD Description Entity-Relationship Model
The Confocal Microscopy Image IOD uses the DICOM Composite Instance IOD Entity-Relationship Information Model defined in Section A.1.2, with only the Image IE below the Series IE.
A.90.1.3 Confocal Microscopy Image IOD Module Table
Table A.90.1.3-1 specifies the Modules of the Confocal Microscopy Image IOD.
Table A.90.1.3-1. Confocal Microscopy Image IOD Modules
|
IE
|
Module
|
Reference
|
Usage
|
|
Patient
|
Patient
|
C.7.1.1
|
M
|
|
Clinical Trial Subject
|
C.7.1.3
|
U
|
|
Study
|
General Study
|
C.7.2.1
|
M
|
|
Patient Study
|
C.7.2.2
|
U
|
|
Clinical Trial Study
|
C.7.2.3
|
U
|
|
Series
|
General Series
|
C.7.3.1
|
M
|
|
Clinical Trial Series
|
C.7.3.2
|
U
|
|
Frame of Reference
|
Frame of Reference
|
C.7.4.1
|
M
|
|
Synchronization
|
C.7.4.2
|
C - Required if time synchronization was applied
|
|
Equipment
|
General Equipment
|
C.7.5.1
|
M
|
|
Enhanced General Equipment
|
C.7.5.2
|
M
|
|
Acquisition
|
General Acquisition
|
C.7.10.1
|
M
|
|
Image
|
General Image
|
C.7.6.1
|
M
|
|
General Reference
|
C.12.4
|
U
|
|
Image Pixel
|
C.7.6.3
|
M
|
|
Multi-frame Functional Groups
|
C.7.6.16
|
M
|
|
Multi-frame Dimension
|
C.7.6.17
|
M
|
|
Specimen
|
C.7.6.22
|
C - Required if the Imaging Subject is a Specimen
|
|
Acquisition Context
|
C.7.6.14
|
M
|
|
Confocal Microscopy Image
|
C.8.35.1
|
M
|
|
Cutaneous Confocal Microscopy Image Acquisition Parameters
|
C.8.35.3
|
C - Required for cutaneous confocal microscopy
|
|
Optical Path
|
C.8.12.5
|
M
|
|
SOP Common
|
C.12.1
|
M
|
|
Common Instance Reference
|
C.12.2
|
U
|
|
Frame Extraction
|
C.12.3
|
C - Required if the SOP Instance was created in response to a Frame-Level retrieve request
|
A.90.1.4 Confocal Microscopy IOD Content Constraints
The value of Modality (0008,0060) shall be CFM.
A.90.1.4.2 Acquisition Context Module
The Defined TID for Acquisition Context Sequence (0040,0555) is TID 8300 “Skin Imaging Acquisition Context”.
Note
Any lesion level Attributes apply to the single lesion seen in the acquired image.
A.90.1.4.3 Referenced Image Sequence
In cutaneous confocal microscopy the Referenced Image Sequence (0008,1140) may be used to identify the SOP instance of a Dermoscopic, or Visible Light image correlated to the Confocal Microscopy acquisition
in which case the value of Purpose of Reference Code Sequence (0040,A170) shall be (121311, DCM, "Localizer").
A.90.1.4.4 Anatomic Region Sequence
For dermatology applications:
A.90.1.5 Confocal Microscopy Image Functional Group Macros
Table A.90.1.5-1. Confocal Microscopy Image Functional Group Macros
|
Functional Group Macro
|
Section
|
Usage
|
|
Pixel Measures
|
C.7.6.16.2.1
|
M - Shall be used as a shared functional group
|
|
Derivation Image
|
C.7.6.16.2.6
|
C - Required if the image or frame has been derived from another SOP Instance.
|
|
Optical Path Identification
|
C.8.12.6.2
|
C - Required if Dimension Organization Type (0020,9311) is not TILED_FULL; may be present otherwise.
|
|
Referenced Image
|
C.7.6.16.2.5
|
C - Required if the image or frame has been planned on another image or frame. May be present otherwise.
|
|
Frame Content
|
C.7.6.16.2.2
|
U - Shall not be used as a shared functional group
|
|
Real World Value Mapping
|
C.7.6.16.2.11
|
U - May be used only if Photometric Interpretation (0028,0004) is MONOCHROME2.
|
|
Plane Position (Slide)
|
C.8.12.6.1
|
C - Required if the Frame of Reference is defined in the Slide Coordinate System.
|
|
Confocal Microscopy Image Frame Type
|
C.8.35.4.1
|
M
|
|
Frame Anatomy
|
C.7.6.16.2.8
|
M
|