DICOM PS3.3 2023e - Information Object Definitions

A.74.4 Breast Projection X-Ray Image Functional Group Macros

Table A.74-2 specifies the use of the Functional Group Macros used in the Multi-frame Functional Groups Module for the Breast Projection X-Ray Image IOD.

Table A.74-2. Breast Projection X-Ray Image Functional Group Macros

Functional Group Macro

Section

Usage

Frame Content

C.7.6.16.2.2

M - May not be used as a Shared Functional Group.

Referenced Image

C.7.6.16.2.5

U

Derivation Image

C.7.6.16.2.6

C - Required if the Image Type (0008,0008) Value 1 equals DERIVED or Value 1 is ORIGINAL and Presentation Intent Type equals FOR PRESENTATION. May be present otherwise.

Frame Anatomy

C.7.6.16.2.8

M

Identity Pixel Value Transformation

C.7.6.16.2.9b

M

Frame VOI LUT With LUT

C.7.6.16.2.10b

M

Contrast/Bolus Usage

C.7.6.16.2.12

C - Required if the Enhanced Contrast/Bolus Module is present.

Frame Display Shutter

C.7.6.16.2.16

U

Irradiation Event Identification

C.7.6.16.2.18

M

X-Ray Frame Characteristics

C.8.19.6.1

U

X-Ray Field of View

C.8.19.6.2

M

X-Ray Frame Pixel Data Properties

C.8.19.6.4

M

X-Ray Frame Detector Parameters

C.8.19.6.5

U

X-Ray Calibration Device Usage

C.8.19.6.6

U

X-Ray Frame Acquisition

C.8.19.6.8

U

X-Ray Collimator

C.8.19.6.12

M

Breast X-Ray Positioner

C.8.31.2

C - Required if X-Ray Source moves relative to the Patient. May be present otherwise.

Breast X-Ray Detector

C.8.31.3

C - Required if X-Ray Detector plane is not normal to the X-Ray beam vector. May be present otherwise.

Breast X-Ray Geometry

C.8.31.4

M

Breast X-Ray Acquisition Dose

C.8.31.5

M

Breast X-Ray Isocenter Reference System

C.8.31.6

M

X-Ray Grid

C.8.31.7

U

X-Ray Filter

C.8.31.8

U


A.74.4.1 Breast Projection X-Ray Image Functional Group Macros Content Constraints

A.74.4.1.1 Frame Anatomy Functional Group Macro

For Anatomic Region Sequence (0008,2218) DCID 4013 “Anatomic Region for Mammography” shall be used.