| DICOM PS3.3 2023e - Information Object Definitions |
|---|
Table A.70-1 specifies the Modules of the Legacy Converted Enhanced CT Image IOD.
Table A.70-1. Legacy Converted Enhanced CT Image IOD Modules
The Photometric Interpretation (0028,0004) defined in Section C.8.15.2 is MONOCHROME2.
If the value of Photometric Interpretation (0028,0004) in the source single frame images is MONOCHROME1, which is permitted by Section C.8.2.1, lossless conversion of the pixel data to MONOCHROME2 and updating of any related Attributes is necessary.
The General Image Module, Overlay Plane Module, VOI LUT Module, Supplemental Palette Color Lookup Table Module and Graphic Annotation Module shall not be used in a Standard Extended SOP Class of the Legacy Converted Enhanced CT Image.
In order to annotate images, whether during acquisition or subsequently, SOP Instances of the Grayscale Softcopy Presentation State Storage or the Structured Report Storage SOP Classes that reference the image SOP Instance, may be used.
No standard mechanism is provided for inclusion of annotations within the image SOP Instance itself, and implementers are discouraged from using private extensions to circumvent this restriction.
Grayscale Softcopy Presentation State Storage Instances that are generated during conversion shall be referenced from the Image SOP Instance by using the Referenced Grayscale Presentation State Sequence in the Enhanced CT Image Module. See Section C.8.15.2.
The Curve Module (Retired) has been retired from DICOM. It is still not permitted to be present. See PS3.3-2004.
| DICOM PS3.3 2023e - Information Object Definitions |
|---|