DICOM PS3.17 2022a - Explanatory Information

XXX.3.8.3 Encoding Details

XXX.3.8.3.1 Volumetric Presentation State Relationship Module Recommendations

Table XXX.3.8-1. Volumetric Presentation State Relationship Module Recommendations

Attribute Name

Tag

Comment

Volumetric Presentation State Input Sequence

(0070,1201)

Contains two items: one for the CT volume input and one for the PET volume input.

>ITEM 1

>Presentation Input Type

(0070,1202)

Set to "VOLUME".

>Referenced Image Sequence

(0008,1140)

Set reference(s) to the image(s) that make(s) up the CT input volume.

>Referenced Spatial Registration Sequence

(0070,0404)

Set to reference the Spatial Registration instance which registers the CT input to the VPS RCS.

>Window Center

(0028,1050)

Set either Window Center and Window Width or VOI LUT Sequence (0028,3010).

>Window Width

(0028,1051)

>Crop

(0070,1204)

Set to "NO".

>ITEM 2

>Presentation Input Type

(0070,1202)

Set to "VOLUME".

>Referenced Image Sequence

(0008,1140)

Set reference(s) to the image(s) that make(s) up the PET input volume.

>Referenced Spatial Registration Sequence

(0070,0404)

Set reference to the Spatial Registration instance which registers the PET input to the VPS RCS.

>Window Center

(0028,1050)

Set either Window Center and Window Width or VOI LUT Sequence (0028,3010).

>Window Width

(0028,1051)

>Crop

(0070,1204)

Set to "NO".


DICOM PS3.17 2022a - Explanatory Information