| DICOM PS3.17 2022a - Explanatory Information |
|---|
In this example, the slices of the two volumes are reconstructed in the axial plane of the patient; the row direction is aligned in the positive x-direction of the patient (right-left) and the column direction is aligned in the positive y-direction of the patient (anterior-posterior).
The full field of view reconstruction in encoded with the Instance UID "Z1" and consists of a 512 cube volume of 0.2 mm of voxel size. The sub-region reconstruction in encoded with the Instance UID "Z2" and consists of a 256 cube volume of the voxel size of 0.1 mm.
Both volumes share the same Frame of Reference UID.
Figure TTT.2.3-2. Attributes of 3D Reconstruction of the full field of view of the projection frames
| DICOM PS3.17 2022a - Explanatory Information |
|---|