DICOM PS3.17 2022a - Explanatory Information

QQQ Parametric Maps (Informative)

QQQ.1 

This Annex contains examples of the use of the Parametric Map IOD.

QQQ.1.1 

This Section contains an example of the use of the Parametric Map IOD to encode Ktrans for a Dynamic Contrast Enhanced (DCE) MR.

The frames comprise a single traversal of a regularly sampled 3D volume, described as a single stack and a single quantity, with dimensions of Stack ID, In-Stack Position Number and Quantity. A reference is also provided to the (single entire multi-frame) MR image from which the parametric map was derived. Only the Frame Content Sequence and Plane Position Sequence vary per-frame; all other functional groups are shared in this example.

Nesting

Attribute

Tag

VR

VL (hex)

Value

Specific Character Set

(0008,0005)

CS

000a

ISO_IR 100

Image Type

(0008,0008)

CS

0022

DERIVED\SECONDARY\PERFUSION\KTRANS

Instance Creation Date

(0008,0012)

DA

0008

20140312

Instance Creation Time

(0008,0013)

TM

000a

141900.944

Instance Creator UID

(0008,0014)

UI

0016

1.3.6.1.4.1.5962.99.3

SOP Class UID

(0008,0016)

UI

0016

1.3.6.1.4.1.5962.301.9

SOP Instance UID

(0008,0018)

UI

003e

1.3.6.1.4.1.5962.99.1.3078904268.1788845519.1394648340940.2.0

Study Date

(0008,0020)

DA

0008

20140312

Series Date

(0008,0021)

DA

0008

20140312

Content Date

(0008,0023)

DA

0008

20140312

Study Time

(0008,0030)

TM

000a

141900.944

Series Time

(0008,0031)

TM

000a

141900.944

Content Time

(0008,0033)

TM

000a

141900.944

Accession Number

(0008,0050)

SH

0000

Modality

(0008,0060)

CS

0002

MR

Manufacturer

(0008,0070)

LO

0000

Referring Physician's Name

(0008,0090)

PN

0008

Doe^John

Study Description

(0008,1030)

LO

002C

Dynamic magnetic resonance imaging of pelvis

Procedure Code Sequence

(0008,1032)

SQ

ffffffff

%item

Code Value

(0008,0100)

SH

000A

446315002

Coding Scheme Designator

(0008,0102)

SH

0004

SCT

Code Meaning

(0008,0104)

LO

002C

Dynamic magnetic resonance imaging of pelvis

%enditem

%endseq

Series Description

(0008,103E)

LO

0010

PK Model Results

Patient's Name

(0010,0010)

PN

0008

Doe^Jane

Patient ID

(0010,0020)

LO

0004

1234

Patient's Birth Date

(0010,0030)

DA

0000

Patient's Sex

(0010,0040)

CS

0000

Study Instance UID

(0020,000d)

UI

003e

1.3.6.1.4.1.5962.99.1.3078904268.1788845519.1394648340940.4.0

Series Instance UID

(0020,000e)

UI

003e

1.3.6.1.4.1.5962.99.1.3078904268.1788845519.1394648340940.3.0

Study ID

(0020,0010)

SH

0004

5678

Series Number

(0020,0011)

IS

0004

100

Instance Number

(0020,0013)

IS

0002

1

Patient Orientation

(0020,0020)

CS

0000

Frame of Reference UID

(0020,0052)

UI

003e

1.3.6.1.4.1.5962.99.1.3078904268.1788845519.1394648340940.5.0

Position Reference Indicator

(0020,1040)

LO

0000

Dimension Organization Sequence

(0020,9221)

SQ

ffffffff

%item

Dimension Organization UID

(0020,9164)

UI

003e

1.3.6.1.4.1.5962.99.1.3078904268.1788845519.1394648340940.6.0

%enditem

%endseq

Dimension Index Sequence

(0020,9222)

SQ

ffffffff

%item

Dimension Organization UID

(0020,9164)

UI

003e

1.3.6.1.4.1.5962.99.1.3078904268.1788845519.1394648340940.6.0

Dimension Index Pointer

(0020,9165)

AT

0004

(0020,9056)

Functional Group Pointer

(0020,9167)

AT

0004

(0040,9096)

Dimension Description Label

(0020,9421)

LO

0008

Stack ID

%enditem

%item

Dimension Organization UID

(0020,9164)

UI

003e

1.3.6.1.4.1.5962.99.1.3078904268.1788845519.1394648340940.6.0

Dimension Index Pointer

(0020,9165)

AT

0004

(0020,9057)

Functional Group Pointer

(0020,9167)

AT

0004

(0020,9111)

Dimension Description Label

(0020,9421)

LO

0018

In-Stack Position Number

%enditem

%item

Dimension Organization UID

(0020,9164)

UI

003e

1.3.6.1.4.1.5962.99.1.3078904268.1788845519.1394648340940.6.0

Dimension Index Pointer

(0020,9165)

AT

0004

(0040,9220) (the Quantity Definition Code Sequence)

Functional Group Pointer

(0020,9167)

AT

0004

(0020,9111)

Dimension Description Label

(0020,9421)

LO

0008

Quantity

%enditem

%endseq

Samples per Pixel

(0028,0002)

US

0002

0001

Photometric Interpretation

(0028,0004)

CS

000c

MONOCHROME2

Number of Frames

(0028,0008)

IS

0002

14

Rows

(0028,0010)

US

0002

256 dec

Columns

(0028,0011)

US

0002

256 dec

Bits Allocated

(0028,0100)

US

0002

0020 hex

Burned In Annotation

(0028,0301)

CS

0002

NO

Recognizable Visual Features

(0028,0302)

CS

0002

NO

Lossy Image Compression

(0028,2110)

CS

0002

00

Presentation LUT Shape

(2050,0020)

CS

0008

IDENTITY

Shared Functional Groups Sequence

(5200,9229)

SQ

ffffffff

%item

Derivation Image Sequence

(0008,9124)

SQ

ffffffff

%item

Source Image Sequence

(0008,2112)

SQ

ffffffff

%item

Referenced SOP Class UID

(0008,1150)

UI

001C

1.2.840.10008.5.1.4.1.1.4.1

Referenced SOP Instance UID

(0008,1155)

UI

002E

1.3.6.1.4.1.5962.1.1.0.0.0.1410021852.13877.0

Spatial Locations Preserved

(0028,135A)

CS

0004

YES

Purpose of Reference Code Sequence

(0040,A170)

SQ

ffffffff

%item

Code Value

(0008,0100)

SH

0006

121322

Coding Scheme Designator

(0008,0102)

SH

0004

DCM

Code Meaning

(0008,0104)

LO

002C

Source image for image processing operation

%enditem

%endseq

%enditem

%endseq

Derivation Code Sequence

(0008,9215)

SQ

ffffffff

%item

Code Value

(0008,0100)

SH

0006

113066

Coding Scheme Designator

(0008,0102)

SH

0004

DCM

Code Meaning

(0008,0104)

LO

000C

Time Course of Signal

%enditem

%endseq

%enditem

%endseq

Frame Anatomy Sequence

(0020,9071)

SQ

ffffffff

%item

Anatomic Region Sequence

(0008,2218)

SQ

ffffffff

%item

Code Value

(0008,0100)

SH

0008

41216001

Coding Scheme Designator

(0008,0102)

SH

0004

SCT

Code Meaning

(0008,0104)

LO

0008

Prostate

%enditem

%endseq

Frame Laterality

(0020,9072)

CS

0002

U

%enditem

%endseq

Plane Orientation Sequence

(0020,9116)

SQ

ffffffff

%item

Image Orientation (Patient)

(0020,0037)

DS

0064

0.99979773312597\​-.0160528955995\​.012115996823878\​.012116000683426\​0.96149705857037\​.274548008348208

%enditem

%endseq

Pixel Measures Sequence

(0028,9110)

SQ

ffffffff

%item

Slice Thickness

(0018,0050)

DS

0010

5.9999942779541

Pixel Spacing

(0028,0030)

DS

0022

1.01559996604919\1.01560020446777

%enditem

%endseq

Frame VOI LUT Sequence

(0028,9132)

SQ

ffffffff

%item

Window Center

(0028,1050)

DS

0004

2.5

Window Width

(0028,1051)

DS

0002

5

VOI LUT Function

(0028,1056)

CS

000c

LINEAR_EXACT

%enditem

%endseq

Pixel Value Transformation Sequence

(0028,9145)

SQ

ffffffff

%item

Rescale Intercept

(0028,1052)

DS

0002

0

Rescale Slope

(0028,1053)

DS

0002

1

Rescale Type

(0028,1054)

LO

0002

US

%enditem

%endseq

Real World Value Mapping Sequence

(0040,9096)

SQ

ffffffff

%item

LUT Explanation

(0028,3003)

LO

0006

Ktrans

Measurement Units Code Sequence

(0040,08ea)

SQ

ffffffff

%item

Code Value

(0008,0100)

SH

0004

/min

Coding Scheme Designator

(0008,0102)

SH

0004

UCUM

Code Meaning

(0008,0104)

LO

0004

/min

%enditem

%endseq

LUT Label

(0040,9210)

SH

0006

Ktrans

Real World Value Last Value Mapped

(0040,9211)

XS

0002

0005

Real World Value First Value Mapped

(0040,9216)

XS

0002

0000

Real World Value Intercept

(0040,9224)

FD

0008

0

Real World Value Slope

(0040,9225)

FD

0008

1

Quantity Definition Code Sequence

(0040,9220)

SQ

ffffffff

%item

Value Type

(0040,a040)

CS

0004

CODE

Concept Name Code Sequence

(0040,a043)

SQ

ffffffff

%item

Code Value

(0008,0100)

SH

000A

246205007

Coding Scheme Designator

(0008,0102)

SH

0004

SCT

Code Meaning

(0008,0104)

LO

0008

Quantity

%enditem

%endseq

Concept Code Sequence

(0040,a168)

SQ

ffffffff

%item

Code Value

(0008,0100)

SH

0006

126312

Coding Scheme Designator

(0008,0102)

SH

0004

DCM

Code Meaning

(0008,0104)

LO

0006

Ktrans

%enditem

%endseq

%enditem

%item

Value Type

(0040,a040)

CS

0004

CODE

Concept Name Code Sequence

(0040,a043)

SQ

ffffffff

%item

Code Value

(0008,0100)

SH

000A

370129005

Coding Scheme Designator

(0008,0102)

SH

0004

SCT

Code Meaning

(0008,0104)

LO

0012

Measurement Method

%enditem

%endseq

Concept Code Sequence

(0040,a168)

SQ

ffffffff

%item

Code Value

(0008,0100)

SH

0006

126340

Coding Scheme Designator

(0008,0102)

SH

0004

DCM

Code Meaning

(0008,0104)

LO

0014

Standard Tofts Model

%enditem

%endseq

%enditem

%endseq

%enditem

%endseq

Parametric Map Frame Type Sequence

(0040,9092)

SQ

ffffffff

%item

Frame Type

(0008,9007)

CS

0022

DERIVED\SECONDARY\PERFUSION\KTRANS

%enditem

%endseq

%enditem

%endseq

Per-frame Functional Groups Sequence

(5200,9230)

SQ

ffffffff

%item

Frame Content Sequence

(0020,9111)

SQ

ffffffff

%item

Stack ID

(0020,9056)

SH

0002

1

In-Stack Position Number

(0020,9057)

UL

0004

0001

Dimension Index Values

(0020,9157)

UL

000C

00000001,00000001,00000001

%enditem

%endseq

Plane Position Sequence

(0020,9113)

SQ

ffffffff

%item

Image Position (Patient)

(0020,0032)

DS

0032

-153.28300476074\-111.93399810791\-54.366100311279

%enditem

%endseq

%enditem

%

...

%endseq

Float Pixel Data

(7FE0,0008)

OF

380000

[]

DICOM PS3.17 2022a - Explanatory Information