DICOM PS3.16 2022a - Content Mapping Resource

TID 5103 Vascular Ultrasound Section

Sections of a vascular ultrasound report are section containers of an anatomic region consisting of measurement group containers that contain the measurements.

Table TID 5103. Parameters

Parameter Name

Parameter Usage

$SectionScope

The concept name of the section heading modifier

$SectionLaterality

The laterality (if any) of the anatomy in this section heading

$Anatomy

The concept name of the vascular anatomy

$AnatomyRatio

The concept name of anatomy-coordinated ratio concepts


Type:

Extensible

Order:

Significant

Root:

No

Table TID 5103. Vascular Ultrasound Section

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CONTAINER

DT (59776-5, LN, "Findings")

1

M

2

>

HAS CONCEPT MOD

CODE

EV (363698007, SCT, "Finding Site")

1

M

$SectionScope

3

>

HAS CONCEPT MOD

CODE

EV (272741003, SCT, "Laterality")

1

M

$SectionLaterality

4

>

CONTAINS

INCLUDE

DTID 5104 “Vascular Ultrasound Measurement Group”

1-n

M

$AnatomyGroup = $Anatomy

5

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1-n

U

$Measurement = $AnatomyRatio


DICOM PS3.16 2022a - Content Mapping Resource