DICOM PS3.16 2022a - Content Mapping Resource

TID 4004 Mammography CAD Composite Feature

This Template collects a composite feature for a lesion, non-lesion object, or correlation of related objects. The details of the composition are expressed in the Mammography CAD Composite Feature Body (see TID 4005 “Mammography CAD Composite Feature Body”). The data from which the details are inferred, are expressed in the Composite Features (see TID 4004 “Mammography CAD Composite Feature”) and/or Single Image Findings (see TID 4006 “Mammography CAD Single Image Finding”), of which there may be several.

A Composite Feature shall be INFERRED FROM any combination of two or more Composite Features or Single Image Findings or mixture thereof.

Type:

Non-Extensible

Order:

Significant

Root:

No

Table TID 4004. Mammography CAD Composite Feature

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CODE

EV (111015, DCM, "Composite Feature")

1

M

DCID 6016 “Mammography Composite Feature”

2

>

HAS CONCEPT MOD

CODE

EV (111056, DCM, "Rendering Intent")

1

M

DCID 6034 “Intended Use of CAD Output”

3

>

HAS OBS CONTEXT

INCLUDE

DTID 4108 “Tracking Identifier”

1

U

4

>

HAS PROPERTIES

INCLUDE

DTID 4005 “Mammography CAD Composite Feature Body”

1

M

5

>

INFERRED FROM

INCLUDE

DTID 4004 “Mammography CAD Composite Feature”

1-n

MC

At least two items shall be present: two of row 5, two of row 6, or one of each.

6

>

INFERRED FROM

INCLUDE

DTID 4006 “Mammography CAD Single Image Finding”

1-n

MC

At least two items shall be present: two of row 5, two of row 6, or one of each.

7

>

HAS OBS CONTEXT

INCLUDE

DTID 4022 “CAD Observation Context”

1

MC

Shall be present only if this feature is incorporated from a different report than its parent.


Content Item Descriptions

Rendering Intent

This Content Item constrains the SCP receiving the Mammography CAD SR IOD in its use of the contents of this Template and its Target Content Items. Mammography CAD devices may opt to use data marked "Not for Presentation" or "Presentation Optional" as input to subsequent mammography CAD processing steps. Refer to Section O.2 “Structured Reporting Storage SOP Class SCU and SCP Behavior” in PS3.4 .

DICOM PS3.16 2022a - Content Mapping Resource