DICOM PS3.16 2022a - Content Mapping Resource

TID 3510 Vital Signs

The Vital Signs Template consists of a CONTAINER containing the various vital signs measurements. These measurements may be acquired automatically from patient monitoring equipment, or may be entered based on manual measurements.

Type:

Extensible

Order:

Significant

Root:

No

Table TID 3510. Vital Signs

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CONTAINER

EV (8716-3, LN, "Vital Signs")

1

M

2

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

U

$Measurement = EV (271649006, SCT, "Systolic blood pressure")

$Units = DCID 3500 “Pressure Units”

$Method = BCID 3560 “Blood Pressure Methods”

3

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

U

$Measurement = EV (271650006, SCT, "Diastolic blood pressure")

$Units = DCID 3500 “Pressure Units”

4

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

U

$Measurement = EV (8867-4, LN, "Heart rate")

$Units = EV ({H.B.}/min, UCUM, "BPM")

5

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

U

$Measurement = EV (8310-5, LN, "Body temperature")

$Units = EV (Cel, UCUM, "C")

6

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

U

$Measurement = DCID 3526 “Blood Gas Saturation”

$Units = EV (%, UCUM, "%")

7

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

U

$Measurement = EV (86290005, SCT, "Respiration rate")

$Units = EV (/min, UCUM, "breaths/min")

8

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

U

$Measurement = EV (122195, DCM, "Pulse Strength")

$Units = DT ({0:4}, UCUM, "range 0:4")

9

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

U

$Measurement = EV (225908003, SCT, "Pain Score")

$Units = DT ({1:10}, UCUM, "range 1:10")

10

>

CONTAINS

CODE

DT (8884-9, LN, "Cardiac Rhythm")

1

U

BCID 3415 “Cardiac Rhythms”

11

>

CONTAINS

CODE

DT (9304-7, LN, "Respiration Rhythm")

1

U

BCID 3416 “Respiration Rhythms”


DICOM PS3.16 2022a - Content Mapping Resource