DICOM PS3.16 2022a - Content Mapping Resource

TID 3507 Ventricular Pressure Measurement

The Ventricular Pressure Measurement Template consists of a CONTAINER, with an acquisition context of the measurement anatomic site, and containing systolic and end-diastolic measurements. This implies that the name of the measurement is effectively post-coordinated from the measurement name, the Hemodynamic Measurement Group container (procedure phase) name, and the acquisition context (finding site) value.

Type:

Extensible

Order:

Significant

Root:

No

Table TID 3507. Ventricular Pressure Measurement

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CONTAINER

EV (122122, DCM, "Ventricular pressure measurements")

1

M

2

>

INCLUDE

DTID 3530 “Hemodynamic Acquisition Context”

1

M

$LocationName = EV (363698007, SCT, "Finding Site")

$LocationValue = DCID 3609 “Ventricular Source Locations”

3

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

MC

IFF $LocationValue selected in Row 2 = (87878005, SCT, "Left Ventricle") or subsite thereof

$Measurement = EV (276780008, SCT, "Left Ventricular Systolic blood pressure")

$Units = DCID 3500 “Pressure Units”

4

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

MC

IFF $LocationValue selected in Row 2 = (87878005, SCT, "Left Ventricle") or subsite thereof

$Measurement = EV (276781007, SCT, "Left Ventricular End Diastolic pressure")

$Units = DCID 3500 “Pressure Units”

5

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

MC

IFF $LocationValue selected in Row 2 = (53085002, SCT, "Right Ventricle") or subsite thereof

$Measurement = EV (276772001, SCT, "Right Ventricular Systolic blood pressure")

$Units = DCID 3500 “Pressure Units”

6

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

MC

IFF $LocationValue selected in Row 2 = (53085002, SCT, "Right Ventricle") or subsite thereof

$Measurement = EV (276774000, SCT, "Right Ventricular End Diastolic pressure")

$Units = DCID 3500 “Pressure Units”

7

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

MC

IFF $LocationValue selected in Row 2 = (45503006, SCT, "Common Ventricle")

$Measurement = EV (122194, DCM, "Ventricular Systolic blood pressure")

$Units = DCID 3500 “Pressure Units”

8

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

MC

IFF $LocationValue selected in Row 2 = (45503006, SCT, "Common Ventricle")

$Measurement = EV (122191, DCM, "Ventricular End Diastolic pressure")

$Units = DCID 3500 “Pressure Units”

9

>

CONTAINS

INCLUDE

DTID 3550 “Pressure Waveform Measurements”

1-n

U


DICOM PS3.16 2022a - Content Mapping Resource