DICOM PS3.16 2022a - Content Mapping Resource

TID 10052 Attenuator Position

This template describes the details of the attenuator positions.

Type:

Extensible

Order:

Non-Significant

Root:

No

Table TID 10052. Attenuator Position

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CONTAINER

EV (130528, DCM, "Attenuator Position")

1

M

2

>

CONTAINS

DATETIME

DT (111526, DCM, "DateTime Started")

1

M

3

>

CONTAINS

DATETIME

DT (111527, DCM, "DateTime Ended")

1

M

4

>

CONTAINS

CONTAINER

EV (128472, DCM, "X-Ray Beam Attenuator Model")

1-n

MC

IF an X-Ray Attenuator is described in TID 10047 AND the position is known

5

>>

CONTAINS

TEXT

EV (130527, DCM, "Identification of the Attenuator")

1

M

6

>>

CONTAINS

IMAGE

EV (128470, DCM, "X-Ray Attenuator Model Data")

1

MC

XOR Rows 7, 8

7

>>

CONTAINS

COMPOSITE

EV (128470, DCM, "X-Ray Attenuator Model Data")

1

MC

XOR Rows 6, 8

8

>>

CONTAINS

UIDREF

EV (128470, DCM, "X-Ray Attenuator Model Data")

1

MC

XOR Rows 6, 7

9

>>

CONTAINS

TABLE

EV (130520, DCM, "Transformation Matrix")

1

M

NCOLUMNS = 4

NROWS= 4

CELL VR = FD


Content Item Descriptions

Row 2

The DateTime of the beginning of the time period over which the content items are applicable.

Row 3

The DateTime of the end of the time period over which the content items are applicable.

Row 4

Provide the model and position of each X-Ray attenuator described in TID 10047, if known.

Row 5

Each X-Ray attenuator model and position description includes the identification of the X-Ray attenuator, which shall match the identification used in TID 10047.

Row 6, 7, 8

Reference to a three-dimensional model of the X-Ray attenuator

Row 9

A 4-by-4 matrix of dimensionless numbers of the form defined in Section C.20.2.1.1 “Frame of Reference Transformation Matrix” in PS3.3 . The matrix describes the rigid transformation matrix (including translation and rotation) that transforms the reference coordinate system for the X-Ray attenuator to the RDSR reference coordinate system. The translation described by this matrix indicates the position and orientation of the X-Ray attenuator within the RDSR reference coordinate system. The RCS for the X-Ray attenuator shall be a right-handed Cartesian coordinate system. If the model data in Row 6, 7, 8 was not originally in a right-handed Cartesian coordinate system, it must transformed prior to inclusion in this template.

DICOM PS3.16 2022a - Content Mapping Resource