| DICOM PS3.16 2022a - Content Mapping Resource |
|---|
This template describes the positions of objects and locations that are defined in the X-Ray source reference coordinate system (e.g., output measurement point, filter position, etc.).
Table TID 10051. Beam Position
|
The DateTime of the beginning of the time period over which the content items are applicable. |
|
|
The DateTime of the end of the time period over which the content items are applicable. |
|
|
Identification the X-Ray source. This designation shall not change for a given source throughout the entire RDSR. For systems with multiple X-Ray sources, each source shall be described with a separate instance of this template. |
|
|
The Output Measurement Point Position describes the position where the measure or machine estimate of radiation output occurs. It is described in the X-Ray source reference coordinate system. It may be the same position as the reference point position (the reference point may be defined in TID 10041). |
|
|
If a reference point is defined in TID 10041, its position in the X-Ray source reference coordinate system shall be described here. |
|
|
Provide the model and position of each X-Ray filter described in TID 10046, if known. |
|
|
Each X-Ray filter model and position description includes the identification of the X-Ray filter, which shall match the identification used in TID 10046. |
|
|
A 4-by-4 matrix of dimensionless numbers of the form defined in Section C.20.2.1.1 “Frame of Reference Transformation Matrix” in PS3.3 . The matrix describes the rigid transformation matrix (including translation and rotation) that transforms the reference coordinate system for the X-Ray filter to the X-Ray source reference coordinate system. The translation described by this matrix indicates the position and orientation of the X-Ray filter within the X-Ray source reference coordinate system. The RCS for the X-Ray filter shall be a right-handed Cartesian coordinate system. If the model data in Row 11, 12, or 13 was not originally in a right-handed Cartesian coordinate system, it must transformed prior to inclusion in this template. |