| DICOM PS3.16 2022a - Content Mapping Resource |
|---|
This Template records details related to the use of the NEMA Dose Check Standard [NEMA XR 25-2019].
Table TID 10015. CT Dose Check Details
|
IFF value of Row 2 = (373066001, SCT, "Yes") |
||||||||
|
IFF value of Row 3 = (373066001, SCT, "Yes") |
||||||||
|
IF Accumulated DLP Forward Estimate (Row 6) exceeds DLP Alert Value (Row 4) |
||||||||
|
IF Accumulated CTDIvol Forward Estimate (Row 7) exceeds CTDIvol Alert Value (Row 5) |
||||||||
|
IFF Accumulated DLP Forward Estimate (Row 6) exceeds DLP Alert Value (Row 4) or Accumulated CTDIvol Forward Estimate (Row 7) exceeds CTDIvol Alert Value (Row 5) |
||||||||
|
IF Accumulated DLP Forward Estimate (Row 6) exceeds DLP Alert Value (Row 4) or Accumulated CTDIvol Forward Estimate (Row 7) exceeds CTDIvol Alert Value (Row 5) |
$PersonProcedureRole = EV (113850, DCM, "Irradiation Authorizing") |
|||||||
|
IFF value of Row 11 = (373066001, SCT, "Yes") |
||||||||
|
IFF value of Row 12 = (373066001, SCT, "Yes") |
||||||||
|
IF DLP Forward Estimate (Row 15) exceeds DLP Notification Value (Row 13) |
||||||||
|
IF CTDIvol Forward Estimate (Row 16) exceeds CTDIvol Notification Value (Row 14) |
||||||||
|
IFF DLP Forward Estimate (Row 15) exceeds DLP Notification Value (Row 13) or CTDIvol Forward Estimate (Row 16) exceeds CTDIvol Notification Value (Row 14) |
||||||||
|
IFF DLP Forward Estimate (Row 15) exceeds DLP Notification Value (Row 13) or CTDIvol Forward Estimate (Row 16) exceeds CTDIvol Notification Value (Row 14) |
$PersonProcedureRole = EV (113850, DCM, "Irradiation Authorizing") |
|
Indicates whether a DLP Alert Value was configured (e.g., by the institution) for the exam to which this irradiation event belongs. |
|
|
Indicates whether a CTDIvol Alert Value was configured (e.g., by the institution) for the exam to which this irradiation event belongs. |
|
|
The configured value applicable to the current exam that would trigger an alert if the accumulated DLP were projected to exceed it. |
|
|
The configured value applicable to the current exam that would trigger an alert if the Accumulated CTDIvol at any given location were projected to exceed it. |
|
|
The value estimated prior to performing this irradiation event of the projected DLP accumulated during this exam, including this irradiation event. The estimate may include assumptions such as those described in the NEMA Dose Check Standard [NEMA XR 25-2019]. |
|
|
The value estimated prior to performing this irradiation event of the projected CTDIvol accumulated during this exam, including this irradiation event. The value is for the location with the highest estimated accumulation. The actual location is not recorded. The estimate may include assumptions such as those described in the NEMA Dose Check Standard [NEMA XR 25-2019]. |
|
|
The reason provided by the operator for proceeding with an irradiation event projected to exceed an alert value. |
|
|
Person responsible for authorizing irradiation projected to exceed an alert value. |
|
|
Indicates whether a DLP Notification Value was configured (e.g., by the institution) for the Protocol Element Group to which this irradiation event corresponds. |
|
|
Indicates whether a CTDIvol Notification Value was configured (e.g., by the institution) for the Protocol Element Group to which this irradiation event corresponds. |
|
|
The configured value applicable to the current irradiation event that would trigger a notification if the DLP were projected to exceed it. |
|
|
The configured value applicable to the current irradiation event that would trigger a notification if the CTDIvol were projected to exceed it. |
|
|
The value estimated prior to performing this irradiation event of the DLP for this irradiation event. The estimate may include assumptions such as those described in the NEMA Dose Check Standard [NEMA XR 25-2019]. |
|
|
The value estimated prior to performing this irradiation event of the CTDIvol for this irradiation event. The value is for the location with the highest estimated value. The actual location is not recorded. The estimate may include assumptions such as those described in the NEMA Dose Check Standard [NEMA XR 25-2019]. |
|
|
The reason provided by the operator for proceeding with an irradiation event projected to exceed a notification value. |
|
|
Person responsible for authorizing irradiation projected to exceed a notification value. |
| DICOM PS3.16 2022a - Content Mapping Resource |
|---|