The Templates that comprise the Implantation Plan SR Document IOD are interconnected as in Figure A-11.
This Template contains all the necessary information to position an Implant Assembly and its Components in a patient. Therefore, all the Components that comprise an Implant Assembly are listed. If the Implant Assembly consists of more than one Component, the relation between the Components will be described as well. It is also possible to describe the registration between the Components and the patient and between the Components themselves.
To reference the Components within this document the Implantation Plan Component ID is used.
The Component Connection links two Implantation Plan Components in a commutative way. This means that for each link between A and B only one Component Connection has to be defined and not two for A-B and B-A.
The terminology used is defined by illustration using the example in Figure A-12.
Table TID 7000. Implantation Plan
|
IFF related Implantation Plan Documents exist that are not referenced by row 4 |
||||||||
|
Only one Component Connection per Mating Feature Set is allowed |
||||||||
|
EV (112376, DCM, "Degree of Freedom Exact Translational Value") |
||||||||
|
EV (112377, DCM, "Degree of Freedom Minimum Translational Value") |
||||||||
|
EV (112378, DCM, "Degree of Freedom Maximum Translational Value") |
||||||||
|
EV (112379, DCM, "Degree of Freedom Exact Rotational Value") |
||||||||
|
EV (112380, DCM, "Degree of Freedom Minimum Rotational Value") |
||||||||
|
EV (112381, DCM, "Degree of Freedom Maximum Rotational Value") |
||||||||
|
EV (112367, DCM, "Planning Information for Intraoperative Usage") |
||||||||
|
References Spatial Registration SOP Instances and Deformable Spatial Registration SOP Instances |
||||||||
|
References SOP Instances except Images and Spatial Registrations |
||||||||
|
EV (112364, DCM, "Related Patient Data Not Used During Planning") |
|
If an Implant Assembly Template was used for the planning, it should be referenced here. |
|
|
ID given to this Implant Component. Used to reference this specific Component within the Implantation Plan. |
|
|
See description of Component Type Code Sequence (0076,0034) Attribute in Section C.29.2.1 “Implant Assembly Template Module” in PS3.3 |
|
|
Reference to the Template that describes that component. May be the same Implant Template as referenced in row 13. The target of the reference may not be needed or available during implantation. e.g., if the plan is opened in another hospital where those implant templates are not used. No purpose of reference is specified. |
|
|
This Frame of Reference is the Frame of Reference of the Implant Component (Frame of Reference UID (0020,0052) Attribute in the “Generic Implant Template Description Module” in PS3.3 ). This may help to find the right registration information (row 43). |
|
|
References the Original Template that was the basis for the Derived Template. May be the same Implant Template as referenced in row 11. The target of the reference may not be needed or available during implantation, e.g., if the plan is opened in another hospital where those implant templates are not used. |
|
|
If there is no Component Connection between sets of Implant Components, one Assembly must be used for each set. |
|
|
The ID of a planned Component that is defined in this document and that is part of this Relation. |
|
|
See description of Mating Feature Set ID (0068,63C0) Attribute in the “Generic Implant Template Mating Features Module” in PS3.3 |
|
|
See description of Mating Feature ID (0068,63F0) Attribute in the “Generic Implant Template Mating Features Module” in PS3.3 |
|
|
See description of Degree of Freedom ID (0068,6410) Attribute in the “Generic Implant Template Mating Features Module” in PS3.3 |
|
|
Defines the range or exact value that was selected or calculated by the planning application. |
|
|
Defines the calibrated Horizontal Pixel Spacing that was used by the planning application, which may be different from the spacing encoded in the referenced Image SOP Instance. |
|
|
Defines the calibrated Vertical Pixel Spacing that was used by the planning application, which may be different from the spacing encoded in the referenced Image SOP Instance. |
|
|
Any patient data other than Image IEs used for the planning, e.g., Surface Segmentations. |
|
|
Fiducials selected by the user for registration of implant components referenced in the parent Content Item. |
|
|
User comment about the Fiducial. This may be the reason it was selected, the intended use, the anatomical or non-anatomical structure that the Fiducial represents, or any other intent. |
|
|
All kinds of information in PDF form that are created by a planning application may be referenced here, e.g., drawings. |
|
|
All kinds of images that are created by a planning application should be referenced here, e.g., images that show patient images overlaid with contour information of the Implant Component, or images that show how several implant components may be composed, or merged patient images. |
|
|
References registration objects that contain registration data that is relevant for this Implantation Plan, e.g., registration of Implant Components. |
|
|
Identifies one or more items within the sequence of referenced Frames of Reference (Registration Sequence (0070,0308) in the Spatial Registration Module or Deformable Registration Sequence (0064,0002) in the “Deformable Spatial Registration Module” in PS3.3 ) that are relevant for this Implantation Plan. See Figure A-13. |
|
|
Any patient data created during the planning process that is not referenced in row 39 and 40, e.g., Surface Segmentation Instances created by the planning application. |
|
|
These Fiducials are derived from the Fiducials identified in Row 34. |
|
|
References to any relevant patient data containing IOD instances that were not used in planning or derived from it but belong to the patient model. Might be reports, images, surface segmentations, or other. |