Enhanced X-Ray Radiation Dose SR IOD Templates 

The Templates that comprise the Enhanced X-Ray Radiation Dose SR are interconnected as in Figure A-18b.

Enhanced X-Ray Radiation Dose Structured Report IOD Template Structure

Figure A-18b. Enhanced X-Ray Radiation Dose Structured Report IOD Template Structure


TID 10040 Enhanced X-Ray Radiation Dose

This template defines a container (the root) with subsidiary content items for documenting the radiation output for an ionizing radiation imaging device. This template supports cone-beam CT, in addition to other modalities.

Type:

Extensible

Order:

Non-Significant

Root:

Yes

Table TID 10040. Enhanced X-Ray Radiation Dose

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CONTAINER

EV (113701, DCM, "X-Ray Radiation Dose Report")

1

M

Root Node

2

>

HAS CONCEPT MOD

INCLUDE

DTID 1204 “Language of Content Item and Descendants”

1

M

3

>

HAS CONCEPT MOD

CODE

EV (121058, DCM, "Procedure reported")

1

M

4

>>

HAS CONCEPT MOD

CODE

EV (363703001, SCT, "Has Intent")

1

M

DCID 3629 “Procedure Intent”

5

>

HAS OBS CONTEXT

INCLUDE

DTID 1002 “Observer Context”

1-n

M

6

>

HAS OBS CONTEXT

CODE

EV (113705, DCM, "Scope of Accumulation")

1

M

DCID 10000 “Scope of Accumulation”

7

>

CONTAINS

INCLUDE

DTID 10041 “Accumulated Dose Data”

1-n

U

8

>

CONTAINS

INCLUDE

DTID 10042 “Irradiation Event Summary Data”

1-n

M

9

>

CONTAINS

INCLUDE

DTID 10043 “Irradiation Details”

1

M

10

>

CONTAINS

TEXT

EV (121106, DCM, "Comment")

1

U

11

>

CONTAINS

INCLUDE

DTID 1020 “Person Participant”

1

U

$PersonProcedureRole = EV (113850, DCM, "Irradiation Authorizing")

12

>

CONTAINS

CODE

EV (113854, DCM, "Source of Dose Information")

1-n

M

DCID 10020 “Source of Projection X-Ray Dose Information”


Content Item Descriptions

Row 7

TID 10041 may be included once for each source. Summary-level values that may include multiple sources, e.g., Dose Area Product Total from planes A and B of a biplane system,could be described by an instance of TID 10041 where the Identification of X-Ray Source indicates multiple sources, e.g., "A and B".

Row 8

TID 10042 shall be included once for each irradiation event. Irradiation event values that may include multiple sources, e.g., CTDI for a dual-source CT scanner,could be described by an instance of TID 10042 where the Identification of X-Ray Source indicates multiple sources, e.g., "1 and 2".

DICOM PS3.16 2022a - Content Mapping Resource