DICOM PS3.16 2022a - Content Mapping Resource

ECG Report Templates 

TID 3700 ECG Report

The ECG Report Template is the root structure for the representation of analysis of an ECG waveform, potentially in comparison to a prior ECG waveform analysis. The analyzed waveform may or may not be stored as a DICOM SOP Instance.

Type:

Extensible

Order:

Significant

Root:

Yes

Table TID 3700. ECG Report

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CONTAINER

EV (28010-7, LN, "ECG Report")

1

M

Root node

2

>

HAS CONCEPT MOD

CODE

EV (121058, DCM, "Procedure reported")

1

U

BCID 3670 “ECG Procedure Types”

3

>

HAS CONCEPT MOD

INCLUDE

DTID 1204 “Language of Content Item and Descendants”

1

M

4

>

CONTAINS

INCLUDE

DTID 1002 “Observer Context”

1-n

M

5

>

CONTAINS

CONTAINER

EV (18785-6, LN, "Indications for Procedure")

1

U

6

>>

CONTAINS

CODE

EV (121071, DCM, "Finding")

1-n

U

DCID 3671 “Reason for ECG Exam”

7

>>

CONTAINS

TEXT

EV (121071, DCM, "Finding")

1

U

8

>

CONTAINS

INCLUDE

DTID 3802 “Cardiovascular Patient History”

1

U

9

>

CONTAINS

INCLUDE

DTID 3704 “Patient Characteristics for ECG”

1

U

10

>

CONTAINS

INCLUDE

DTID 3702 “Prior ECG Exam”

1

U

11

>

CONTAINS

INCLUDE

DTID 3708 “ECG Waveform Information”

1

M

12

>

CONTAINS

CONTAINER

EV (122144, DCM, "Quantitative Analysis")

1

M

13

>>

CONTAINS

INCLUDE

DTID 3713 “ECG Global Measurements”

1

U

14

>>

CONTAINS

INCLUDE

DTID 3714 “ECG Lead Measurements”

1-n

U

One instantiation per reported lead

15

>

CONTAINS

INCLUDE

DTID 3717 “ECG Qualitative Analysis”

1

U

16

>

CONTAINS

INCLUDE

DTID 3719 “Summary, ECG”

1

U


DICOM PS3.16 2022a - Content Mapping Resource