DICOM PS3.15 2022a - Security and System Management Profiles

E.3.10 Retain Safe Private Option

By definition, Private Attributes contain proprietary information, in many cases the nature of which is known only to the vendor and not publicly documented.

However, some Private Attributes may be necessary for the desired application. For example, specific technique information such as CT helical span pitch, or pixel value transformation, such as PET SUV rescale factors, may only be available in Private Attributes since the information is either not defined in Standard Attributes, or was added to the DICOM Standard after the acquisition device was manufactured.

When this Option is specified in addition to the Basic Application Level Confidentiality Profile, Private Attributes that are known by the de-identifier to be safe from identity leakage shall be retained, together with the Private Creator IDs that are required to fully define the retained Private Attributes; all other Private Attributes shall be removed or processed in the element-specific manner recommended by Deidentification Action (0008,0307), if present within Private Data Element Characteristics Sequence (0008,0300) (see PS3.3 Section C.12.1).

Whether or not an Attribute is known to be safe may be determined by:

When this Option is not specified, all Private Attributes shall be removed, as described in Table E.1-1.

Note

  1. A sample list of Private Attributes thought to be safe is provided here. Vendors do not guarantee them to be safe, and do not commit to sending them in any particular software version (including future products).

    Table E.3.10-1. Safe Private Attributes

    Data Element

    Private Creator

    VR

    VM

    Meaning

    (7053,xx00)

    Philips PET Private Group

    DS

    1

    SUV Factor - Multiplying stored pixel values by Rescale Slope then this factor results in SUVbw in g/l

    (7053,xx09)

    Philips PET Private Group

    DS

    1

    Activity Concentration Factor - Multiplying stored pixel values by Rescale Slope then this factor results in MBq/ml.

    (00E1,xx21)

    ELSCINT1

    DS

    1

    DLP

    (00E1,xx50)

    ELSCINT1

    DS

    1

    Acquisition Duration

    (01E1,xx26)

    ELSCINT1

    CS

    1

    Phantom Type

    (01F1,xx01)

    ELSCINT1

    CS

    1

    Acquisition Type

    (01F1,xx07)

    ELSCINT1

    DS

    1

    Table Velocity

    (01F1,xx26)

    ELSCINT1

    DS

    1

    Pitch

    (01F1,xx27)

    ELSCINT1

    DS

    1

    Rotation Time

    (0019,xx23)

    GEMS_ACQU_01

    DS

    1

    Table Speed [mm/rotation]

    (0019,xx24)

    GEMS_ACQU_01

    DS

    1

    Mid Scan Time [sec]

    (0019,xx27)

    GEMS_ACQU_01

    DS

    1

    Rotation Speed (Gantry Period)

    (0019,xx9E)

    GEMS_ACQU_01

    LO

    1

    Internal Pulse Sequence Name

    (0043,xx27)

    GEMS_PARM_01

    SH

    1

    Scan Pitch Ratio in the form "n.nnn:1"

    (0045,xx01)

    GEMS_HELIOS_01

    SS

    1

    Number of Macro Rows in Detector

    (0045,xx02)

    GEMS_HELIOS_01

    FL

    1

    Macro width at ISO Center

    (0903,xx10)

    GEIIS PACS

    US

    1

    Reject Image Flag

    (0903,xx11)

    GEIIS PACS

    US

    1

    Significant Flag

    (0903,xx12)

    GEIIS PACS

    US

    1

    Confidential Flag

    (2001,xx01)

    Philips Imaging DD 001

    FL

    1

    MR Image Chemical Shift

    (2001,xx02)

    Philips Imaging DD 001

    IS

    1

    MR Image Chemical Shift Number

    (2001,xx03)

    Philips Imaging DD 001

    FL

    1

    MR Image Diffusion B-Factor

    (2001,xx04)

    Philips Imaging DD 001

    CS

    1

    MR Image Diffusion Direction

    (2001,xx05)

    Philips Imaging DD 001

    SS

    1

    Graphic Annotation Parent ID

    (2001,xx06)

    Philips Imaging DD 001

    CS

    1

    MR Image Enhanced

    (2001,xx07)

    Philips Imaging DD 001

    CS

    1

    MR Image Type Edes

    (2001,xx08)

    Philips Imaging DD 001

    IS

    1

    MR Image Phase Number

    (2001,xx09)

    Philips Imaging DD 001

    FL

    1

    MR Image Prepulse Delay

    (2001,xx0a)

    Philips Imaging DD 001

    IS

    1

    Image Plane Number

    (2001,xx0b)

    Philips Imaging DD 001

    CS

    1

    Image Plane Orientation

    (2001,xx0c)

    Philips Imaging DD 001

    CS

    1

    MR Series Arrhythmia Rejection

    (2001,xx0e)

    Philips Imaging DD 001

    CS

    1

    MR Series Cardiac Cycled

    (2001,xx0f)

    Philips Imaging DD 001

    SS

    1

    MR Series Cardiac Gate Width

    (2001,xx10)

    Philips Imaging DD 001

    CS

    1

    MR Series Cardiac Sync

    (2001,xx11)

    Philips Imaging DD 001

    FL

    1

    MR Series Diffusion Echo Time

    (2001,xx12)

    Philips Imaging DD 001

    CS

    1

    MR Series Dynamic Series

    (2001,xx13)

    Philips Imaging DD 001

    SL

    1

    MR Series Epi Factor

    (2001,xx14)

    Philips Imaging DD 001

    SL

    1

    MR Series Nr Of Echoes

    (2001,xx15)

    Philips Imaging DD 001

    SS

    1

    MR Series Nr Of Locations

    (2001,xx16)

    Philips Imaging DD 001

    SS

    1

    MR Series Nr Of Phase Contrast Directions

    (2001,xx17)

    Philips Imaging DD 001

    SL

    1

    MR Series Nr Of Phases

    (2001,xx18)

    Philips Imaging DD 001

    SL

    1

    MR Series Nr Of Slices

    (2001,xx19)

    Philips Imaging DD 001

    CS

    1

    MR Series Partial Matrix Scanned

    (2001,xx1a)

    Philips Imaging DD 001

    FL

    3

    MR Series Pc Velocity

    (2001,xx1b)

    Philips Imaging DD 001

    FL

    1

    MR Series Prepulse Delay

    (2001,xx1c)

    Philips Imaging DD 001

    CS

    1

    MR Series Prepulse Type

    (2001,xx1d)

    Philips Imaging DD 001

    IS

    1

    MR Series Reconstruction Number

    (2001,xx1e)

    Philips Imaging DD 001

    CS

    1

    MR Series Reformat Accuracy

    (2001,xx1f)

    Philips Imaging DD 001

    CS

    1

    MR Series Respiration Sync

    (2001,xx21)

    Philips Imaging DD 001

    CS

    1

    MR Series Sel Part Inversion Recovery

    (2001,xx22)

    Philips Imaging DD 001

    FL

    1

    MR Series Water Fat Shift

    (2001,xx23)

    Philips Imaging DD 001

    DS

    1

    MR Series Flip Angle

    (2001,xx24)

    Philips Imaging DD 001

    CS

    1

    MR Series Is Interactive

    (2001,xx25)

    Philips Imaging DD 001

    SH

    1

    MR Series Echo Time Display

    (2001,xx26)

    Philips Imaging DD 001

    CS

    1

    Presentation State Subtraction Active

    (2001,xx27)

    Philips Imaging DD 001

    CS

    1

    Edge Enhancement Kernel Size

    (2001,xx28)

    Philips Imaging DD 001

    FL

    1

    Edge Enhancement Gain Factor Sub

    (2001,xx29)

    Philips Imaging DD 001

    FL

    1

    Edge Enhancement Gain Factor

    (2001,xx2a)

    Philips Imaging DD 001

    CS

    1

    Edge Enhancement Taste Adapt Sub

    (2001,xx2b)

    Philips Imaging DD 001

    CS

    1

    Edge Enhancement Taste Adapt

    (2001,xx2c)

    Philips Imaging DD 001

    FL

    1

    Harmonization Factor

    (2001,xx2d)

    Philips Imaging DD 001

    SS

    1

    Stack Number Of Slices

    (2001,xx2e)

    Philips Imaging DD 001

    CS

    1

    Harmonization Kernel Size

    (2001,xx2f)

    Philips Imaging DD 001

    FL

    1

    Harmonization Gain

    (2001,xx30)

    Philips Imaging DD 001

    UL

    1

    Log Subtraction Gain Step Taste

    (2001,xx31)

    Philips Imaging DD 001

    US

    1

    Mixing Nr Of Mask Image Numbers

    (2001,xx32)

    Philips Imaging DD 001

    FL

    1

    Stack Radial Angle

    (2001,xx33)

    Philips Imaging DD 001

    CS

    1

    Stack Radial Axis

    (2001,xx34)

    Philips Imaging DD 001

    CS

    1

    Mixing Mask Operation Type

    (2001,xx35)

    Philips Imaging DD 001

    SS

    1

    Stack Slice Number

    (2001,xx36)

    Philips Imaging DD 001

    CS

    1

    Stack Type

    (2001,xx37)

    Philips Imaging DD 001

    CS

    1

    Mixing Operation Type

    (2001,xx38)

    Philips Imaging DD 001

    FL

    1

    Overlay Mix Factor

    (2001,xx39)

    Philips Imaging DD 001

    FL

    1

    Overscan Factor

    (2001,xx3a)

    Philips Imaging DD 001

    CS

    1

    Pixel Shift

    (2001,xx3b)

    Philips Imaging DD 001

    FL

    4

    Pixel Shift Split Line Coordinates

    (2001,xx3c)

    Philips Imaging DD 001

    FL

    2

    Pixel Shift Shift Vectora

    (2001,xx3d)

    Philips Imaging DD 001

    UL

    1

    Contour Fill Color

    (2001,xx3e)

    Philips Imaging DD 001

    FL

    2

    Pixel Shift Shift Vectorb

    (2001,xx3f)

    Philips Imaging DD 001

    CS

    1

    Displayed Area Zoom Interpolation Meth

    (2001,xx40)

    Philips Imaging DD 001

    CS

    1

    Pixel Shift Split Screen

    (2001,xx41)

    Philips Imaging DD 001

    FL

    1

    Subtraction Land Marking Factor

    (2001,xx42)

    Philips Imaging DD 001

    CS

    1

    Subtraction Land Marking Active

    (2001,xx43)

    Philips Imaging DD 001

    IS

    2

    Ellips Displ Shut Major Ax Frst End Pnt

    (2001,xx44)

    Philips Imaging DD 001

    IS

    2

    Ellips Displ Shut Major Ax Scnd End Pnt

    (2001,xx45)

    Philips Imaging DD 001

    IS

    2

    Ellips Displ Shut Other Ax Frst End Pnt

    (2001,xx46)

    Philips Imaging DD 001

    CS

    1

    Graphic Line Style

    (2001,xx47)

    Philips Imaging DD 001

    FL

    1

    Graphic Line Width

    (2001,xx48)

    Philips Imaging DD 001

    SS

    1

    Graphic Annotation Id

    (2001,xx49)

    Philips Imaging DD 001

    IS

    1

    Trace First Image

    (2001,xx4a)

    Philips Imaging DD 001

    CS

    1

    Trace Taste

    (2001,xx4b)

    Philips Imaging DD 001

    CS

    1

    Interpolation Method

    (2001,xx4c)

    Philips Imaging DD 001

    CS

    1

    Poly Line Begin Point Style

    (2001,xx4d)

    Philips Imaging DD 001

    CS

    1

    Poly Line End Point Style

    (2001,xx4e)

    Philips Imaging DD 001

    CS

    1

    Window Smoothing Taste

    (2001,xx4f)

    Philips Imaging DD 001

    FD

    1

    Harmonization Offset

    (2001,xx50)

    Philips Imaging DD 001

    LO

    1

    Graphic Marker Type

    (2001,xx51)

    Philips Imaging DD 001

    IS

    1

    Overlay Plane Id

    (2001,xx53)

    Philips Imaging DD 001

    CS

    1

    Presentation Gl Trafo Invert

    (2001,xx54)

    Philips Imaging DD 001

    FL

    1

    Contour Fill Transparency

    (2001,xx55)

    Philips Imaging DD 001

    UL

    1

    Graphic Line Color

    (2001,xx56)

    Philips Imaging DD 001

    CS

    1

    Graphic Type

    (2001,xx57)

    Philips Imaging DD 001

    CS

    1

    Log Subtraction Taste

    (2001,xx58)

    Philips Imaging DD 001

    UL

    1

    Series Xray Contrast Transfer Taste

    (2001,xx59)

    Philips Imaging DD 001

    IS

    1

    Curve Id

    (2001,xx5a)

    Philips Imaging DD 001

    ST

    1

    Graphic Annotation Model

    (2001,xx5d)

    Philips Imaging DD 001

    ST

    1

    Measurement Text Units

    (2001,xx5e)

    Philips Imaging DD 001

    ST

    1

    Measurement Text Type

    (2001,xx5f)

    Philips Imaging DD 001

    SQ

    1-n

    Stack Sequence

    (2001,xx60)

    Philips Imaging DD 001

    SL

    1

    MR Series Nr Of Stacks

    (2001,xx61)

    Philips Imaging DD 001

    CS

    1

    Series Transmitted

    (2001,xx62)

    Philips Imaging DD 001

    CS

    1

    Series Committed

    (2001,xx63)

    Philips Imaging DD 001

    CS

    1

    Examination Source

    (2001,xx64)

    Philips Imaging DD 001

    SH

    1

    Text Type

    (2001,xx65)

    Philips Imaging DD 001

    SQ

    1-n

    Overlay Plane Sequence

    (2001,xx66)

    Philips Imaging DD 001

    SQ

    1-n

    Image Curve

    (2001,xx67)

    Philips Imaging DD 001

    CS

    1

    Linear Presentation Gl Trafo Shape Sub

    (2001,xx68)

    Philips Imaging DD 001

    SQ

    1-n

    Modality Gl Trafo Sequence

    (2001,xx69)

    Philips Imaging DD 001

    SQ

    1-n

    Display Shutter Sequence

    (2001,xx6a)

    Philips Imaging DD 001

    SQ

    1-n

    Spatial Transformation Sequence

    (2001,xx6b)

    Philips Imaging DD 001

    SQ

    1-n

    Edge Enhancement Sequence

    (2001,xx6d)

    Philips Imaging DD 001

    LO

    1

    Text Font

    (2001,xx6e)

    Philips Imaging DD 001

    SH

    1

    Series Type

    (2001,xx6f)

    Philips Imaging DD 001

    SQ

    1-n

    Mixing Sequence

    (2001,xx71)

    Philips Imaging DD 001

    CS

    1

    Graphic Constraint

    (2001,xx72)

    Philips Imaging DD 001

    IS

    2

    Ellips Displ Shut Other Ax Scnd End Pnt

    (2001,xx73)

    Philips Imaging DD 001

    SQ

    1-n

    Referenced Mask Image Sequence

    (2001,xx74)

    Philips Imaging DD 001

    DS

    1-n

    Window Center Sub

    (2001,xx75)

    Philips Imaging DD 001

    DS

    1-n

    Window Width Sub

    (2001,xx76)

    Philips Imaging DD 001

    UL

    1

    Presentation State Xray Contrast Transfer Taste

    (2001,xx77)

    Philips Imaging DD 001

    CS

    1

    Gl Trafo Type

    (2001,xx79)

    Philips Imaging DD 001

    SQ

    1-n

    Harmonisation Sequence

    (2001,xx7a)

    Philips Imaging DD 001

    FL

    1

    Window Rounding Factor

    (2001,xx7b)

    Philips Imaging DD 001

    IS

    1

    MR Series Acquisition Number

    (2001,xx7c)

    Philips Imaging DD 001

    UL

    1

    Frame Number

    (2001,xx7d)

    Philips Imaging DD 001

    OW/OB

    1

    Frame Pixel Data

    (2001,xx7e)

    Philips Imaging DD 001

    US

    1

    Edge Enhancement Gain Taste

    (2001,xx7f)

    Philips Imaging DD 001

    US

    1

    Edge Enhancement Gain Taste Sub

    (2001,xx80)

    Philips Imaging DD 001

    LO

    1

    Text Anchor Point Alignment

    (2001,xx81)

    Philips Imaging DD 001

    IS

    1

    MR Series Nr Of Dynamic Scans

    (2001,xx82)

    Philips Imaging DD 001

    IS

    1

    MR Series Echo Train Length

    (2001,xx83)

    Philips Imaging DD 001

    DS

    1

    MR Series Imaging Frequency

    (2001,xx84)

    Philips Imaging DD 001

    DS

    1

    MR Series Inversion Time

    (2001,xx85)

    Philips Imaging DD 001

    DS

    1

    MR Series Magnetic Field Strength

    (2001,xx86)

    Philips Imaging DD 001

    IS

    1

    MR Series Nr Of Phase Encoding Steps

    (2001,xx87)

    Philips Imaging DD 001

    SH

    1

    MR Series Nucleus

    (2001,xx88)

    Philips Imaging DD 001

    DS

    1

    MR Series Number Of Averages

    (2001,xx89)

    Philips Imaging DD 001

    DS

    1

    MR Series Percent Phase Field Of View

    (2001,xx8a)

    Philips Imaging DD 001

    DS

    1

    MR Series Percent Sampling

    (2001,xx8b)

    Philips Imaging DD 001

    SH

    1

    MR Series Transmitting Coil

    (2001,xx8c)

    Philips Imaging DD 001

    CS

    0-n

    Needs Processing

    (2001,xx90)

    Philips Imaging DD 001

    LO

    1

    Text Foreground Color

    (2001,xx91)

    Philips Imaging DD 001

    LO

    1

    Text Background Color

    (2001,xx92)

    Philips Imaging DD 001

    LO

    1

    Text Shadow Color

    (2001,xx93)

    Philips Imaging DD 001

    LO

    1

    Text Style

    (2001,xx94)

    Philips Imaging DD 001

    LO

    1

    Processing Order Specialization

    (2001,xx9a)

    Philips Imaging DD 001

    SQ

    1-n

    Graphic Number Sequence

    (2001,xx9b)

    Philips Imaging DD 001

    UL

    1

    Graphic Number

    (2001,xx9d)

    Philips Imaging DD 001

    LO

    1

    Subtraction Type

    (2001,xx9f)

    Philips Imaging DD 001

    US

    2

    Pixel Processing Kernel Size

    (2001,xxa1)

    Philips Imaging DD 001

    CS

    1

    Is Raw Image

    (2001,xxa2)

    Philips Imaging DD 001

    US

    1

    Log Subtraction Curve Taste

    (2001,xxa3)

    Philips Imaging DD 001

    UL

    1

    Text Color Foreground

    (2001,xxa4)

    Philips Imaging DD 001

    UL

    1

    Text Color Background

    (2001,xxa5)

    Philips Imaging DD 001

    UL

    1

    Text Color Shadow

    (2001,xxc0)

    Philips Imaging DD 001

    UL

    1

    Content Item Identifier

    (2001,xxc1)

    Philips Imaging DD 001

    LO

    1

    Nested Object Type Name

    (2001,xxc2)

    Philips Imaging DD 001

    FD

    1

    Subtraction Rescale Factor

    (2001,xxc3)

    Philips Imaging DD 001

    UL

    1

    Subtraction Offset

    (2001,xxc5)

    Philips Imaging DD 001

    SQ

    0-1

    Mask Image Lut Sequence

    (2001,xxc6)

    Philips Imaging DD 001

    SQ

    0-1

    Gain Lut Sequence

    (2001,xxc7)

    Philips Imaging DD 001

    SQ

    0-1

    Contrast Image Lut Sequence

    (2001,xxca)

    Philips Imaging DD 001

    SQ

    1

    Reversed Modality Lut

    (2001,xxcb)

    Philips Imaging DD 001

    IS

    1

    Redundant Overlay Plane Id

    (2001,xxd0)

    Philips Imaging DD 001

    AT

    1-n

    Frame Index Pointer

    (2001,xxd1)

    Philips Imaging DD 001

    US

    1-n

    Index Number Vector

    (2001,xxd2)

    Philips Imaging DD 001

    US

    1

    Frame Detector Number

    (2001,xxd3)

    Philips Imaging DD 001

    US

    1

    Frame Phase Number

    (2001,xxd4)

    Philips Imaging DD 001

    US

    1

    Frame Rotation Number

    (2001,xxd5)

    Philips Imaging DD 001

    US

    1

    Frame Rr Interval Number

    (2001,xxd6)

    Philips Imaging DD 001

    US

    1

    Frame Time Slot Number

    (2001,xxd7)

    Philips Imaging DD 001

    US

    1

    Frame Slice Number

    (2001,xxd8)

    Philips Imaging DD 001

    US

    1

    Frame Angular View Number

    (2001,xxd9)

    Philips Imaging DD 001

    US

    1

    Frame Time Slice Number

    (2001,xxda)

    Philips Imaging DD 001

    CS

    1

    Is Arrowhead

    (2001,xxdb)

    Philips Imaging DD 001

    US

    1

    PET Rr Interval Index

    (2001,xxdc)

    Philips Imaging DD 001

    US

    1

    PET Time Slot Index

    (2001,xxdd)

    Philips Imaging DD 001

    US

    1

    PET Time Slice Index

    (2001,xxde)

    Philips Imaging DD 001

    US

    1

    PET Slice Index

    (2001,xxdf)

    Philips Imaging DD 001

    UL

    1

    Voxel Number

    (2001,xxe9)

    Philips Imaging DD 001

    SQ

    0-n

    Per Frame Voxels Functional Group

    (2001,xxf1)

    Philips Imaging DD 001

    FL

    6

    MR Image Prospective Motion Correction

    (2001,xxf2)

    Philips Imaging DD 001

    FL

    1-n

    MR Image Retrospective Motion Correction

    (2001,xxf3)

    Philips Imaging DD 001

    CS

    1

    Dynamic Linear Voigl Trafo

    (2001,xxf4)

    Philips Imaging DD 001

    UL

    1

    Color Key

    (2001,xxf5)

    Philips Imaging DD 001

    CS

    1

    Rotated Text Allowed

    (2001,xxf6)

    Philips Imaging DD 001

    UL

    1

    Number Of Represented Images

    (2001,xxf7)

    Philips Imaging DD 001

    LO

    0-n

    Data Type Icons

    (2001,xxf9)

    Philips Imaging DD 001

    SQ

    0-n

    Flagging Sequence

    (2001,xxfb)

    Philips Imaging DD 001

    SQ

    0-1

    Bookmark Sequence

    (2001,xxfc)

    Philips Imaging DD 001

    SQ

    0-1

    Ris Code Sequence

    (2001,xxfd)

    Philips Imaging DD 001

    SQ

    0-n

    Workflow Step Sequence

    (2001,xxff)

    Philips Imaging DD 001

    OW/OB

    1

    Volume Pixel

    (2001,xx01)

    Philips Imaging DD 002

    US

    1

    Edr Setgp

    (2001,xx02)

    Philips Imaging DD 002

    FD

    1

    Edr Setsk

    (2001,xx03)

    Philips Imaging DD 002

    CS

    1

    Um Rank Type

    (2001,xx04)

    Philips Imaging DD 002

    SS

    1

    Um Rank Number

    (2001,xx05)

    Philips Imaging DD 002

    FD

    1

    Um Rank Enhancement

    (2001,xx06)

    Philips Imaging DD 002

    US

    1

    Um Kernel Size

    (2001,xx07)

    Philips Imaging DD 002

    CS

    1

    Um Gamma Type

    (2001,xx08)

    Philips Imaging DD 002

    FD

    1

    Um Gamma Angle

    (2001,xx09)

    Philips Imaging DD 002

    FD

    1

    Um Gamma Shift

    (2001,xx0a)

    Philips Imaging DD 002

    FD

    1

    Um Gamma Center

    (2001,xx0b)

    Philips Imaging DD 002

    FD

    1

    Drr Contrast Equalization

    (2001,xx0c)

    Philips Imaging DD 002

    US

    1

    Drr Contrast Equalization Kernel Size

    (2001,xx0d)

    Philips Imaging DD 002

    CS

    1

    Drr Sharp Lut Type

    (2001,xx0e)

    Philips Imaging DD 002

    FD

    1

    Drr Sharpening

    (2001,xx0f)

    Philips Imaging DD 002

    US

    1

    Drr Sharp Kernel Size

    (2001,xx10)

    Philips Imaging DD 002

    SS

    1

    Drr Window Width

    (2001,xx11)

    Philips Imaging DD 002

    SS

    1

    Drr Window Level

    (2001,xx12)

    Philips Imaging DD 002

    CS

    1

    Drr Lut Type

    (2001,xx13)

    Philips Imaging DD 002

    SS

    1

    Unique Processing Mode

    (2001,xx14)

    Philips Imaging DD 002

    FD

    1

    Unique Contrast Balance

    (2001,xx15)

    Philips Imaging DD 002

    FD

    1

    Unique Center Density

    (2001,xx16)

    Philips Imaging DD 002

    FD

    1

    Unique Bright Density

    (2001,xx17)

    Philips Imaging DD 002

    FD

    1

    Unique Detail Contrast

    (2001,xx18)

    Philips Imaging DD 002

    CS

    1

    Unique Density Curve

    (2001,xx19)

    Philips Imaging DD 002

    FD

    1

    Unique Density Min

    (2001,xx1a)

    Philips Imaging DD 002

    FD

    1

    Unique Density Max

    (2001,xx1b)

    Philips Imaging DD 002

    FD

    1

    Unique Gamma

    (2001,xx1c)

    Philips Imaging DD 002

    FD

    1

    Unique Gamma Min

    (2001,xx1d)

    Philips Imaging DD 002

    FD

    1

    Unique Gamma Max

    (2001,xx1e)

    Philips Imaging DD 002

    FD

    1

    Unique Structure Preference

    (2001,xx1f)

    Philips Imaging DD 002

    FD

    1

    Unique Noise Limit

    (2001,xx20)

    Philips Imaging DD 002

    FD

    1

    Unique Noise Band

    (2001,xx21)

    Philips Imaging DD 002

    FD

    1

    Unique Noise Step

    (2001,xx22)

    Philips Imaging DD 002

    FD

    1

    Unique Noise Compensation

    (2001,xx23)

    Philips Imaging DD 002

    FD

    1

    Unique Structure Boost

    (2001,xx24)

    Philips Imaging DD 002

    FD

    1

    Unique Strong Contrast Limit

    (2001,xx25)

    Philips Imaging DD 002

    FD

    1

    Unique Strong Contrast Factor

    (2001,xx26)

    Philips Imaging DD 002

    FD

    1

    Unique Structure Boost Offset

    (2001,xx27)

    Philips Imaging DD 002

    FD

    1

    Unique Weak Contrast Limit

    (2001,xx28)

    Philips Imaging DD 002

    US

    1

    Unique Smooth Gain

    (2001,xx29)

    Philips Imaging DD 002

    US

    1

    Unique Level

    (2001,xx2a)

    Philips Imaging DD 002

    US

    1

    Unique Gain Up Single

    (2001,xx2b)

    Philips Imaging DD 002

    SS

    1

    Unique Smh Limit One

    (2001,xx2c)

    Philips Imaging DD 002

    FD

    1

    Unique Film Density Min

    (2001,xx2d)

    Philips Imaging DD 002

    FD

    1

    Unique Film Density Max

    (2001,xx2e)

    Philips Imaging DD 002

    SS

    1

    Unique Version

    (2001,xx2f)

    Philips Imaging DD 002

    SS

    1

    Ranger Version

    (2001,xx30)

    Philips Imaging DD 002

    SS

    1

    Ranger Mode

    (2001,xx31)

    Philips Imaging DD 002

    SS

    1

    Ranger Field1

    (2001,xx32)

    Philips Imaging DD 002

    SS

    1

    Ranger Field2

    (2001,xx33)

    Philips Imaging DD 002

    SS

    1

    Ranger Percentile Key1

    (2001,xx34)

    Philips Imaging DD 002

    SS

    1

    Ranger Percentile Key2

    (2001,xx35)

    Philips Imaging DD 002

    FD

    1

    Ranger Dose1

    (2001,xx36)

    Philips Imaging DD 002

    FD

    1

    Ranger Dose2

    (2001,xx37)

    Philips Imaging DD 002

    FD

    1

    Ranger Manual Dose1

    (2001,xx38)

    Philips Imaging DD 002

    FD

    1

    Ranger Manual Dose2

    (2001,xx39)

    Philips Imaging DD 002

    CS

    1

    Unique Rox Shape

    (2001,xx3a)

    Philips Imaging DD 002

    SQ

    1

    Ranger Set Sequence

    (2001,xx3b)

    Philips Imaging DD 002

    SQ

    0-n

    Rox Sequence

    (2001,xx3c)

    Philips Imaging DD 002

    SQ

    1-n

    Xray Edge Enhancement Sequence

    (2001,xx3d)

    Philips Imaging DD 002

    SQ

    0-1

    Edr Lut Sequence

    (2001,xx3e)

    Philips Imaging DD 002

    SS

    1

    Unique Dose Decades

    (2001,xx3f)

    Philips Imaging DD 002

    SS

    1

    Unique Dose Unit

    (2001,xx40)

    Philips Imaging DD 002

    SS

    1

    Unique Density Unit

    (2001,xx50)

    Philips Imaging DD 002

    CS

    1

    Workflow Status

    (2001,xx52)

    Philips Imaging DD 002

    LO

    1

    Workflow Step Id

    (2001,xx53)

    Philips Imaging DD 002

    CS

    1

    Workflow Step Status

    (2001,xx57)

    Philips Imaging DD 002

    SQ

    0-1

    Workflow Step Input Sequence

    (2001,xx58)

    Philips Imaging DD 002

    SQ

    0-1

    Workflow Step Output Sequence

    (2001,xx5a)

    Philips Imaging DD 002

    LO

    1

    Workflow Step Type

    (2001,xx5c)

    Philips Imaging DD 002

    LO

    1

    Workflow Id

    (2001,xx5d)

    Philips Imaging DD 002

    UL

    1

    Pixel Data Representation Rows

    (2001,xx5e)

    Philips Imaging DD 002

    UL

    1

    Pixel Data Representation Columns

    (2001,xx5f)

    Philips Imaging DD 002

    SQ

    0-1

    Private Dicom Extension Sequence

    (2001,xx63)

    Philips Imaging DD 002

    ST

    1

    Isyntax Reference

    (2001,xx64)

    Philips Imaging DD 002

    SQ

    0-1

    Workflow Step Job Params Sequence

    (2001,xx65)

    Philips Imaging DD 002

    CS

    1

    Preserve Aspect

    (2001,xx66)

    Philips Imaging DD 002

    CS

    1

    Interpolated

    (2001,xx67)

    Philips Imaging DD 002

    LO

    1

    Error Flag

    (2001,xx68)

    Philips Imaging DD 002

    LO

    1

    Error Message Id

    (2001,xx6b)

    Philips Imaging DD 002

    LO

    0-n

    Suitable For

    (2001,xx71)

    Philips Imaging DD 002

    CS

    1

    Submit Mpps Job

    (2001,xx72)

    Philips Imaging DD 002

    FL

    2

    Displayed Area Bottom Right Hand Corner Fraction

    (2001,xx73)

    Philips Imaging DD 002

    FL

    2

    Displayed Area Top Left Hand Corner Fraction

    (2001,xx74)

    Philips Imaging DD 002

    LO

    1

    Interpolation Mode

    (2001,xx75)

    Philips Imaging DD 002

    SS

    0-n

    Graphic Annotation Id Reference

    (2001,xx01)

    Philips Imaging DD 097

    SQ

    1

    View Geometry

    (2001,xx02)

    Philips Imaging DD 097

    FD

    3

    Frame Geometry Origin

    (2001,xx03)

    Philips Imaging DD 097

    FD

    2

    Frame Geometry Extent

    (2001,xx04)

    Philips Imaging DD 097

    FD

    6

    Frame Geometry Orientation

    (2001,xx05)

    Philips Imaging DD 097

    SQ

    1-n

    Visual Sequence

    (2001,xx06)

    Philips Imaging DD 097

    SQ

    0-n

    Cut Sequence

    (2001,xx07)

    Philips Imaging DD 097

    FD

    1

    Visual Opacity

    (2001,xx08)

    Philips Imaging DD 097

    SQ

    1

    Opacity Map Sequence

    (2001,xx0a)

    Philips Imaging DD 097

    SQ

    1-n

    Light Sequence

    (2001,xx0b)

    Philips Imaging DD 097

    SQ

    1

    Color Map Sequence

    (2001,xx0d)

    Philips Imaging DD 097

    FD

    1

    Visual Threshold

    (2001,xx0e)

    Philips Imaging DD 097

    SQ

    0-n

    Scene Sequence

    (2001,xx0f)

    Philips Imaging DD 097

    FD

    2

    Frame Geometry Slab

    (2001,xx12)

    Philips Imaging DD 097

    SQ

    1

    Gradient Map Sequence

    (2001,xx17)

    Philips Imaging DD 097

    US

    1

    Display Rows

    (2001,xx18)

    Philips Imaging DD 097

    US

    1

    Display Columns

    (2001,xx19)

    Philips Imaging DD 097

    LO

    1

    Frame Geometry Type

    (2001,xx1a)

    Philips Imaging DD 097

    FD

    1

    Light Intensity

    (2001,xx1b)

    Philips Imaging DD 097

    UL

    1

    Light Color

    (2001,xx1c)

    Philips Imaging DD 097

    FD

    3

    Light Origin

    (2001,xx1d)

    Philips Imaging DD 097

    FD

    3

    Light Direction

    (2001,xx1e)

    Philips Imaging DD 097

    FD

    4-n

    Color Map Samples

    (2001,xx1f)

    Philips Imaging DD 097

    FD

    2-n

    Opacity Map Samples

    (2001,xx21)

    Philips Imaging DD 097

    UL

    1

    Visual Color

    (2001,xx22)

    Philips Imaging DD 097

    FD

    1

    Illumination Ambient

    (2001,xx23)

    Philips Imaging DD 097

    FD

    1

    Illumination Diffuse

    (2001,xx24)

    Philips Imaging DD 097

    FD

    1

    Illumination Specular

    (2001,xx25)

    Philips Imaging DD 097

    FD

    1

    Illumination Specular Power

    (2001,xx26)

    Philips Imaging DD 097

    FD

    1

    Depth Cue Begin

    (2001,xx27)

    Philips Imaging DD 097

    FD

    1

    Depth Cue Lambda

    (2001,xx28)

    Philips Imaging DD 097

    LO

    1

    Depth Cue Function

    (2001,xx2a)

    Philips Imaging DD 097

    LO

    1

    Cut Set Type

    (2001,xx2b)

    Philips Imaging DD 097

    FD

    3

    Cut Plane Origin

    (2001,xx2c)

    Philips Imaging DD 097

    FD

    3

    Cut Plane Normal

    (2001,xx2d)

    Philips Imaging DD 097

    UL

    1

    Volume Definition Number

    (2001,xx2e)

    Philips Imaging DD 097

    SQ

    0-1

    Volume Mask Sequence

    (2001,xx2f)

    Philips Imaging DD 097

    OB

    1

    Volume Mask Data

    (2001,xx30)

    Philips Imaging DD 097

    SL

    3

    Volume Mask Offset

    (2001,xx31)

    Philips Imaging DD 097

    SL

    3

    Volume Mask Size

    (2001,xx32)

    Philips Imaging DD 097

    SL

    2

    Volume Mask Alignment

    (2001,xx36)

    Philips Imaging DD 097

    LO

    1

    Light Anchor

    (2001,xx37)

    Philips Imaging DD 097

    FD

    3

    Cone Geometry Origin

    (2001,xx38)

    Philips Imaging DD 097

    FD

    9

    Cone Geometry Orientation

    (2001,xx39)

    Philips Imaging DD 097

    FD

    1

    Cone Geometry View Distance

    (2001,xx3a)

    Philips Imaging DD 097

    FD

    2

    Cone Geometry Slab Extent

    (2001,xx3b)

    Philips Imaging DD 097

    FD

    2

    Cone Geometry Field Of View

    (2001,xx3c)

    Philips Imaging DD 097

    SQ

    1

    Intensity Map Sequence

    (2001,xx3d)

    Philips Imaging DD 097

    FD

    1-n

    Intensity Map Samples

    (2001,xx3e)

    Philips Imaging DD 097

    SQ

    0-n

    Mesh Sequence

    (2001,xx40)

    Philips Imaging DD 097

    UL

    1

    Mesh Color

    (2001,xx41)

    Philips Imaging DD 097

    LO

    1

    Mesh Draw Style

    (2001,xx42)

    Philips Imaging DD 097

    SQ

    0-n

    Mesh Section Sequence

    (2001,xx44)

    Philips Imaging DD 097

    LO

    1

    Mesh Section Vertex Connectivity

    (2001,xx45)

    Philips Imaging DD 097

    SL

    1

    Mesh Section Vertex Size

    (2001,xx46)

    Philips Imaging DD 097

    LO

    1

    Mesh Section Vertex Format

    (2001,xx47)

    Philips Imaging DD 097

    OB

    1-n

    Mesh Section Vertex Data

    (2001,xx49)

    Philips Imaging DD 097

    SQ

    1

    Referenced Volume Definition Sequence

    (2001,xx4a)

    Philips Imaging DD 097

    OB

    1-n

    Mesh Section Index Data

    (2001,xx4b)

    Philips Imaging DD 097

    SL

    1

    Mesh Section Index Size

    (2001,xx4c)

    Philips Imaging DD 097

    LO

    1

    Shutter Interaction Type

    (2001,xxa1)

    Philips Imaging DD 097

    CS

    1-2

    Volume Type

    (2001,xxa2)

    Philips Imaging DD 097

    FD

    3

    Volume Origin

    (2001,xxa3)

    Philips Imaging DD 097

    FD

    9

    Volume Axis

    (2001,xxa4)

    Philips Imaging DD 097

    FD

    2

    Volume Pixel Spacing

    (2001,xxa5)

    Philips Imaging DD 097

    FD

    1-n

    Volume Slice Spacing

    (2001,xxa6)

    Philips Imaging DD 097

    US

    1

    Volume Slices

    (2001,xxa8)

    Philips Imaging DD 097

    IS

    1-n

    Referenced Volume Definition Number

    (2001,xxa9)

    Philips Imaging DD 097

    SQ

    0-n

    Linear Modality Gl Trafo Sequence

    (2001,xxaa)

    Philips Imaging DD 097

    FD

    1-n

    Volume Slice Offsets

    (2001,xxab)

    Philips Imaging DD 097

    IS

    1

    Number Of Volume Definitions

    (2001,xx00)

    Philips Imaging DD 129

    SQ

    1

    Presentation State Sequence

    (2001,xx01)

    Philips Imaging DD 129

    SQ

    1

    Embedded Original Presentation State Sequence

    (2001,xx02)

    Philips Imaging DD 129

    SQ

    0-1

    Planar Intersection Sequence

    (2001,xx03)

    Philips Imaging DD 129

    UL

    1

    Plane Separator Line Color

    (2001,xx04)

    Philips Imaging DD 129

    SQ

    1-n

    Plane Sequence

    (2001,xx05)

    Philips Imaging DD 129

    FD

    4

    Plane Equation

    (2001,xx06)

    Philips Imaging DD 129

    FD

    1

    Plane Opacity

    (2001,xx07)

    Philips Imaging DD 129

    CS

    1

    Plane Enabled

    (2001,xx08)

    Philips Imaging DD 129

    CS

    1

    Plane Selected

    (2001,xx09)

    Philips Imaging DD 129

    UL

    1

    Plane Highlight Color

    (0025,xx01)

    Philips ST80i

    OW

    1

    ST80i Stress Study file data

    (0019,xx0C)

    SIEMENS MR HEADER

    IS

    1

    B Value

    (0019,xx0D)

    SIEMENS MR HEADER

    CS

    1

    Diffusion Directionality

    (0019,xx0E)

    SIEMENS MR HEADER

    FD

    3

    Diffusion Gradient Direction

    (0019,xx27)

    SIEMENS MR HEADER

    FD

    6

    B Matrix

    (0043,xx39)

    GEMS_PARM_01

    IS

    4

    1stvalue is B Value

    (0043,xx6F)

    GEMS_PARM_01

    DS

    3-4

    Scanner Table Entry + Gradient Coil Selected

    (0025,xx07)

    GEMS_SERS_01

    SL

    1

    Images in Series

    (7E01,xx01)

    HOLOGIC, Inc.

    LO

    1

    Codec Version

    (7E01,xx02)

    HOLOGIC, Inc.

    SH

    1

    Codec Content Type

    (7E01,xx10)

    HOLOGIC, Inc.

    SQ

    1

    High Resolution Data Sequence

    (7E01,xx11)

    HOLOGIC, Inc.

    SQ

    1

    Low Resolution Data Sequence

    (7E01,xx12)

    HOLOGIC, Inc.

    OB

    1

    Codec Content

    (0099,xx01)

    NQHeader

    UI

    1

    Version

    (0099,xx02)

    NQHeader

    UI

    1

    Analyzed Series UID

    (0099,xx04)

    NQHeader

    SS

    1

    Return Code

    (0099,xx05)

    NQHeader

    LT

    1

    Return Message

    (0099,xx10)

    NQHeader

    FL

    1

    MI

    (0099,xx20)

    NQHeader

    SH

    1

    Units

    (0099,xx21)

    NQHeader

    FL

    1

    ICV

    (0199,xx01)

    NQLeft

    FL

    1

    Left Cortical White Matter

    (0199,xx02)

    NQLeft

    FL

    1

    Left Cortical Gray Matter

    (0199,xx03)

    NQLeft

    FL

    1

    Left 3rd Ventricle

    (0199,xx04)

    NQLeft

    FL

    1

    Left 4th Ventricle

    (0199,xx05)

    NQLeft

    FL

    1

    Left 5th Ventricle

    (0199,xx06)

    NQLeft

    FL

    1

    Left Lateral Ventricle

    (0199,xx07)

    NQLeft

    FL

    1

    Left Inferior Lateral Ventricle

    (0199,xx08)

    NQLeft

    FL

    1

    Left Inferior CSF

    (0199,xx09)

    NQLeft

    FL

    1

    Left Cerebellar White Matter

    (0199,xx0a)

    NQLeft

    FL

    1

    Left Cerebellar Gray Matter

    (0199,xx0b)

    NQLeft

    FL

    1

    Left Hippocampus

    (0199,xx0c)

    NQLeft

    FL

    1

    Left Amygdala

    (0199,xx0d)

    NQLeft

    FL

    1

    Left Thalamus

    (0199,xx0e)

    NQLeft

    FL

    1

    Left Caudate

    (0199,xx0f)

    NQLeft

    FL

    1

    Left Putamen

    (0199,xx10)

    NQLeft

    FL

    1

    Left Pallidum

    (0199,xx11)

    NQLeft

    FL

    1

    Left Ventral Diencephalon

    (0199,xx12)

    NQLeft

    FL

    1

    Left Nucleus Accumbens

    (0199,xx13)

    NQLeft

    FL

    1

    Left Brain Stem

    (0199,xx14)

    NQLeft

    FL

    1

    Left Exterior CSF

    (0199,xx15)

    NQLeft

    FL

    1

    Left WM Hypo

    (0199,xx16)

    NQLeft

    FL

    1

    Left Other

    (0299,xx01)

    NQRight

    FL

    1

    Right Cortical White Matter

    (0299,xx02)

    NQRight

    FL

    1

    Right Cortical Gray Matter

    (0299,xx03)

    NQRight

    FL

    1

    Right 3rd Ventricle

    (0299,xx04)

    NQRight

    FL

    1

    Right 4th Ventricle

    (0299,xx05)

    NQRight

    FL

    1

    Right 5th Ventricle

    (0299,xx06)

    NQRight

    FL

    1

    Right Lateral Ventricle

    (0299,xx07)

    NQRight

    FL

    1

    Right Inferior Lateral Ventricle

    (0299,xx08)

    NQRight

    FL

    1

    Right Inferior CSF

    (0299,xx09)

    NQRight

    FL

    1

    Right Cerebellar White Matter

    (0299,xx0a)

    NQRight

    FL

    1

    Right Cerebellar Gray Matter

    (0299,xx0b)

    NQRight

    FL

    1

    Right Hippocampus

    (0299,xx0c)

    NQRight

    FL

    1

    Right Amygdala

    (0299,xx0d)

    NQRight

    FL

    1

    Right Thalamus

    (0299,xx0e)

    NQRight

    FL

    1

    Right Caudate

    (0299,xx0f)

    NQRight

    FL

    1

    Right Putamen

    (0299,xx10)

    NQRight

    FL

    1

    Right Pallidum

    (0299,xx11)

    NQRight

    FL

    1

    Right Ventral Diencephalon

    (0299,xx12)

    NQRight

    FL

    1

    Right Nucleus Accumbens

    (0299,xx13)

    NQRight

    FL

    1

    Right Brain Stem

    (0299,xx14)

    NQRight

    FL

    1

    Right Exterior CSF

    (0299,xx15)

    NQRight

    FL

    1

    Right WM Hypo

    (0299,xx16)

    NQRight

    FL

    1

    Right Other

    (2005,xx0D)

    Philips MR Imaging DD 001

    FL

    1

    Scale Intercept

    (2005,xx0E)

    Philips MR Imaging DD 001

    FL

    1

    Scale Slope

    (0119,xx00)

    SIEMENS Ultrasound SC2000

    LO

    1

    Acoustic Meta Information Version

    (0119,xx01)

    SIEMENS Ultrasound SC2000

    OB

    1

    Common Acoustic Meta Information

    (0119,xx02)

    SIEMENS Ultrasound SC2000

    SQ

    1

    Multi Stream Sequence

    (0119,xx03)

    SIEMENS Ultrasound SC2000

    SQ

    1

    Acoustic Data Sequence

    (0119,xx04)

    SIEMENS Ultrasound SC2000

    OB

    1

    Per Transaction Acoustic Control Information

    (0119,xx05)

    SIEMENS Ultrasound SC2000

    UL

    1

    Acoustic Data Offset

    (0119,xx06)

    SIEMENS Ultrasound SC2000

    UL

    1

    Acoustic Data Length

    (0119,xx07)

    SIEMENS Ultrasound SC2000

    UL

    1

    Footer Offset

    (0119,xx08)

    SIEMENS Ultrasound SC2000

    UL

    1

    Footer Length

    (0119,xx09)

    SIEMENS Ultrasound SC2000

    SS

    1

    Acoustic Stream Number

    (0119,xx10)

    SIEMENS Ultrasound SC2000

    SH

    1

    Acoustic Stream Type

    (0119,xx11)

    SIEMENS Ultrasound SC2000

    1

    Stage Timer Time

    (0119,xx12)

    SIEMENS Ultrasound SC2000

    1

    Stop Watch Time

    (0119,xx13)

    SIEMENS Ultrasound SC2000

    IS

    1

    Volume Rate

    (0119,xx21)

    SIEMENS Ultrasound SC2000

    SH

    1

    (0129,xx00)

    SIEMENS Ultrasound SC2000

    SQ

    1

    MPR View Sequence

    (0129,xx02)

    SIEMENS Ultrasound SC2000

    UI

    1

    Bookmark UID

    (0129,xx03)

    SIEMENS Ultrasound SC2000

    1

    Plane Origin Vector

    (0129,xx04)

    SIEMENS Ultrasound SC2000

    1

    Row Vector

    (0129,xx05)

    SIEMENS Ultrasound SC2000

    1

    Column Vector

    (0129,xx06)

    SIEMENS Ultrasound SC2000

    SQ

    1

    Visualization Sequence

    (0129,xx07)

    SIEMENS Ultrasound SC2000

    UI

    1

    Bookmark UID

    (0129,xx08)

    SIEMENS Ultrasound SC2000

    OB

    1

    Visualization Information

    (0129,xx09)

    SIEMENS Ultrasound SC2000

    SQ

    1

    Application State Sequence

    (0129,xx10)

    SIEMENS Ultrasound SC2000

    OB

    1

    Application State Information

    (0129,xx11)

    SIEMENS Ultrasound SC2000

    SQ

    1

    Referenced Bookmark Sequence

    (0129,xx12)

    SIEMENS Ultrasound SC2000

    UI

    1

    Referenced Bookmark UID

    (0129,xx20)

    SIEMENS Ultrasound SC2000

    SQ

    1

    Cine Parameters Sequence

    (0129,xx21)

    SIEMENS Ultrasound SC2000

    OB

    1

    Cine Parameters Schema

    (0129,xx22)

    SIEMENS Ultrasound SC2000

    OB

    1

    Values of Cine Parameters

    (0129,xx29)

    SIEMENS Ultrasound SC2000

    OB

    1

    (0129,xx30)

    SIEMENS Ultrasound SC2000

    CS

    1

    Raw Data Object Type

    (0139,xx01)

    SIEMENS Ultrasound SC2000

    SL

    1

    Physio Capture ROI

    (0149,xx01)

    SIEMENS Ultrasound SC2000

    FD

    1-n

    Vector of BROI Points

    (0149,xx02)

    SIEMENS Ultrasound SC2000

    FD

    1-n

    Start/End Timestamps of Strip Stream

    (0149,xx03)

    SIEMENS Ultrasound SC2000

    FD

    1-n

    Timestamps of Visible R-waves

    (7FD1,xx01)

    SIEMENS Ultrasound SC2000

    OB

    1

    Acoustic Image and Footer Data

    (7FD1,xx09)

    SIEMENS Ultrasound SC2000

    UI

    1

    Volume Version ID

    (7FD1,xx10)

    SIEMENS Ultrasound SC2000

    OB

    1

    Volume Payload

    (7FD1,xx11)

    SIEMENS Ultrasound SC2000

    OB

    1

    After Payload

    (7FD1,xx01)

    SIEMENS SYNGO ULTRA-SOUND TOYON DATA STREAMING

    OB

    1

    Padding

    (7FD1,xx09)

    SIEMENS SYNGO ULTRA-SOUND TOYON DATA STREAMING

    UI

    1

    Version ID

    (7FD1,xx10)

    SIEMENS SYNGO ULTRA-SOUND TOYON DATA STREAMING

    OB

    1

    Volume Payload

    (7FD1,xx11)

    SIEMENS SYNGO ULTRA-SOUND TOYON DATA STREAMING

    OB

    1

    After Payload


  2. One approach to retaining Private Attributes safely, either when the VR is encoded explicitly or known from a data dictionary (such as may be derived from published DICOM Conformance Statements or previously encountered instances, perhaps by adaptively extending the data dictionary as new explicit VR instances are received), is to retain those Attributes that are numeric only. For example, one might retain US, SS, UL, SS, FL and FD binary values, and IS and DS string values that contain only valid numeric characters. One might assume that other string Value Representations are unsafe in the absence of definite confirmation from the vendor to the contrary; code strings (CS) may be an exception. Bulk binary data in OB Value representations is particularly unsafe, and may often contain entire proprietary format headers in binary or text or XML form that includes the patient's name and other identifying information.

The safe private attributes that are retained shall be described in the Conformance Statement.

DICOM PS3.15 2022a - Security and System Management Profiles