DICOM PS3.15 2022a - Security and System Management Profiles

A.5.3.6 DICOM Instances Accessed

This message describes the event of DICOM SOP Instances being viewed, utilized, updated, or deleted. This message shall only include information about a single patient and can be used to summarize all activity for several studies for that patient. This message records the studies to which the instances belong, not the individual instances.

If all instances within a study are deleted, then the EV(110105, DCM, "DICOM Study Deleted") event shall be used, see Section A.5.3.8.

Table A.5.3.6-1. Audit Message for DICOM Instances Accessed

Real World Entities

Field Name

Opt.

Value Constraints

Event

EventID

M

EV (110103, DCM, "DICOM Instances Accessed")

EventActionCode

M

Enumerated value:

C = create

R = read

U = update

D = delete

EventDateTime

M

not specialized

EventOutcomeIndicator

M

not specialized

EventTypeCode

U

not specialized

Active Participant:

Person and or Process manipulating the data

(1..2)

UserID

M

not specialized

AlternativeUserID

U

not specialized

UserName

U

not specialized

UserIsRequestor

M

not specialized

RoleIDCode

U

not specialized

NetworkAccessPointTypeCode

U

not specialized

NetworkAccessPointID

U

not specialized

Participating Object:

Studies (1..N)

ParticipantObjectTypeCode

M

Shall be: 2 = system

ParticipantObjectTypeCodeRole

M

Shall be: 3 = report

ParticipantObjectDataLifeCycle

U

not specialized

ParticipantObjectIDTypeCode

M

EV (110180, DCM, "Study Instance UID")

ParticipantObjectSensitivity

U

not specialized

ParticipantObjectID

M

The Study Instance UID

ParticipantObjectName

U

not specialized

ParticipantObjectQuery

U

not specialized

ParticipantObjectDetail

U

Not specialized

ParticipantObjectDescription

U

Not specialized

SOPClass

MC

See Table A.5.2-1

Accession

U

not specialized

NumberOfInstances

U

not specialized

Instances

U

not specialized

Encrypted

U

not specialized

Anonymized

U

not specialized

Participating Object:

Patient (1)

ParticipantObjectTypeCode

M

Shall be: 1 = person

ParticipantObjectTypeCodeRole

M

Shall be: 1 = patient

ParticipantObjectDataLifeCycle

U

not specialized

ParticipantObjectIDTypeCode

M

Shall be: EV (2, RFC-3881, "Patient Number")

ParticipantObjectSensitivity

U

not specialized

ParticipantObjectID

M

The patient ID

ParticipantObjectName

U

The patient name

ParticipantObjectQuery

U

not specialized

ParticipantObjectDetail

U

not specialized

ParticipantObjectDescription

U

not specialized


DICOM PS3.15 2022a - Security and System Management Profiles